63W Desk Top Type Medical Power Supplies MPU63 series Features: Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz IEC-320-C14 Input Inlet Single Output Optional Output Connector (See page appendix) Over Voltage and Over Load protection Output Voltage Protection (Crowbar Design) Class I Energy Star 2.0, Efficiency level V 3 year warranty Electrical Characteristics: Safety Approvals Input Voltage Range 100~240VAC Operate Voltage Range 90~264VAC fin Input Frequency 47~63Hz Po Output Power Range See rating chart Vo Output Voltage Range See rating chart Io Output Current Range Iil Input Current (Low Line) Io=Full load, Vin=100VAC 1.62A Iih Input Current (High Line) Io=Full load, Vin=240VAC 0.72A Low Line Inrush Current Io=Full load, 25°C, Cool start, Vin=115VAC 45A (max.) High Line Inrush Current Io=Full load, 25°C, Cool start, Vin=230VAC 90A (max.) Efficiency Io=Full Load, Vin=230VAC 77~88% Line Regulation Io=Full Load 0.5~1% Vin=230VAC 3~7% Vin Ir Eff REG-i REG-o Load Regulation Over Voltage Protection 112~132% Over Current Protection 110~150% Ttr Time of Transient Response Io=Full Load to Half Load, Vin=100VAC 4mS (max.) Th Hold-Up Time Io=Full Load, Vin=110VAC 16mS (min.) Ts Start Up Time Io=Full Load, Vin=100VAC 0.3~2S Ripple & Noise(Peak to Peak) Full Load, Vin=90VAC 1% (max.) Ilk Safety Ground Leakage Current Vin=240VAC/60Hz 0.1mA (max.) TC Temperature Coefficient All output ±0.04%/°C OA Operating Altitude Pno No-Load Power Consumption No load, Vin=230VAC See rating chart Vps Dielectric Withstanding Voltage Primary to secondary 5911VDC (min.) Vpg Dielectric Withstanding Voltage for Primary to PE Primary to PE 2249VDC (min.) Ris Isolation Resistance Test Voltage=500VDC 50M Up to 3000m (min.) Environmental To Operating Temperature See derating curve Ts Storage Temperature -40~85°C Ho Operating Humidity 0~95% Hr Storage Humidity 0~95% Operating Temperature at 25°C, Calculated per MIL-HDBK-217F 0.1M Hrs (min.) Pd Derate linearly from 100% load at 50°C to 50% load at 70°C Safety Approvals: S&E OVP MTBF - Medical Touch Panel PC - Patient Monitoring System - Ultrasound System See rating chart OCP Vp-p Application: RT PS E EN 60601-1 IEC 60601-1 RoHS 2002/95/EC COMPLIANT PS E 63W Desk Top Type Medical Power Supplies MPU63 series Output Voltage And Current Rating Chart ( Single Output ) : Model Number Output Voltage Output Current Total Regulation Max . Output Power Pno (max.) MPU63-102 5 VDC 9.00 A 7% 45W 0.5W MPU63-103 7 VDC 7.85 A 7% 55W 0.5W MPU63-104 9 VDC 6.44 A 5% 58W 0.5W MPU63-105 12 VDC 5.25 A 5% 63W 0.5W MPU63-106 15 VDC 4.20 A 5% 63W 0.5W MPU63-107 18 VDC 3.50 A 5% 63W 0.5W MPU63-108 24 VDC 2.62 A 3% 63W 0.5W MPU63-109 30 VDC 2.10 A 3% 63W 0.5W MPU63-110 36 VDC 1.75 A 3% 63W 0.5W MPU63-102~103 are required to use AWG#16 / 5C / 4FT output cable. MPU63-104~107 are required to use AWG#16 /2C/ 4FT output cable. MPU63-108~110 are required to use AWG#18 /2C/ 6FT output cable. The regulation and efficiency will be changed by modified output cable. Mechanical Specifications: Note: 1. Dimensions are shown in mm. 2. Weight: 510~560gs approx. 3. Optional output connector: See page Appendix. Derating Curve : 1. Operating Temperature: 0 to 70°C 2. Derate linearly from 100% load at 50°C to 50% load at 70°C 120% Output Watts (%) 100% 80% 60% 40% 20% 0% 0 10 20 30 40 50 Temperature (deg c) 60 70 80