MPU63 series

63W Desk Top Type
Medical Power Supplies
MPU63 series
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single Output
Optional Output Connector (See page appendix)
Over Voltage and Over Load protection
Output Voltage Protection (Crowbar Design)
Class I
Energy Star 2.0, Efficiency level V
3 year warranty
Electrical Characteristics:
Safety Approvals Input Voltage Range
100~240VAC
Operate Voltage Range
90~264VAC
fin
Input Frequency
47~63Hz
Po
Output Power Range
See rating chart
Vo
Output Voltage Range
See rating chart
Io
Output Current Range
Iil
Input Current (Low Line)
Io=Full load, Vin=100VAC
1.62A
Iih
Input Current (High Line)
Io=Full load, Vin=240VAC
0.72A
Low Line Inrush Current
Io=Full load, 25°C, Cool start, Vin=115VAC
45A (max.)
High Line Inrush Current
Io=Full load, 25°C, Cool start, Vin=230VAC
90A (max.)
Efficiency
Io=Full Load, Vin=230VAC
77~88%
Line Regulation
Io=Full Load
0.5~1%
Vin=230VAC
3~7%
Vin
Ir
Eff
REG-i
REG-o Load Regulation
Over Voltage Protection
112~132%
Over Current Protection
110~150%
Ttr
Time of Transient Response
Io=Full Load to Half Load, Vin=100VAC
4mS (max.)
Th
Hold-Up Time
Io=Full Load, Vin=110VAC
16mS (min.)
Ts
Start Up Time
Io=Full Load, Vin=100VAC
0.3~2S
Ripple & Noise(Peak to Peak)
Full Load, Vin=90VAC
1% (max.)
Ilk
Safety Ground Leakage Current
Vin=240VAC/60Hz
0.1mA (max.)
TC
Temperature Coefficient
All output
±0.04%/°C
OA
Operating Altitude
Pno
No-Load Power Consumption
No load, Vin=230VAC
See rating chart
Vps
Dielectric Withstanding Voltage
Primary to secondary
5911VDC (min.)
Vpg
Dielectric Withstanding Voltage
for Primary to PE
Primary to PE
2249VDC (min.)
Ris
Isolation Resistance
Test Voltage=500VDC
50M
Up to 3000m
(min.)
Environmental
To
Operating Temperature
See derating curve
Ts
Storage Temperature
-40~85°C
Ho
Operating Humidity
0~95%
Hr
Storage Humidity
0~95%
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
0.1M Hrs (min.)
Pd
Derate linearly from 100% load at 50°C to 50% load at 70°C
Safety Approvals:
S&E
OVP
MTBF
- Medical Touch Panel PC
- Patient Monitoring System
- Ultrasound System
See rating chart
OCP
Vp-p
Application:
RT
PS
E
EN 60601-1
IEC 60601-1
RoHS
2002/95/EC
COMPLIANT
PS
E
63W Desk Top Type
Medical Power Supplies
MPU63 series
Output Voltage And Current Rating Chart ( Single Output ) :
Model Number
Output Voltage
Output Current
Total
Regulation
Max .
Output Power
Pno (max.)
MPU63-102
5 VDC
9.00 A
7%
45W
0.5W
MPU63-103
7 VDC
7.85 A
7%
55W
0.5W
MPU63-104
9 VDC
6.44 A
5%
58W
0.5W
MPU63-105
12 VDC
5.25 A
5%
63W
0.5W
MPU63-106
15 VDC
4.20 A
5%
63W
0.5W
MPU63-107
18 VDC
3.50 A
5%
63W
0.5W
MPU63-108
24 VDC
2.62 A
3%
63W
0.5W
MPU63-109
30 VDC
2.10 A
3%
63W
0.5W
MPU63-110
36 VDC
1.75 A
3%
63W
0.5W
MPU63-102~103 are required to use AWG#16 / 5C / 4FT output cable.
MPU63-104~107 are required to use AWG#16 /2C/ 4FT output cable.
MPU63-108~110 are required to use AWG#18 /2C/ 6FT output cable.
The regulation and efficiency will be changed by modified output cable.
Mechanical Specifications:
Note:
1. Dimensions are shown in mm.
2. Weight: 510~560gs approx.
3. Optional output connector:
See page Appendix.
Derating Curve :
1. Operating Temperature: 0 to 70°C
2. Derate linearly from 100% load at 50°C to 50%
load at 70°C
120%
Output Watts (%)
100%
80%
60%
40%
20%
0%
0
10
20
30
40
50
Temperature (deg c)
60
70
80