MP45-05 Dual Output SPEC_B0 140423.cdr

45W Dual Outputs Open Frame Medical Power Supply
MP45 series
Features:
Universal AC input / Full range
Medical safety approval: ANSI / AAMI ES 60601-1 /
TUV EN 60601-1: 2006; IEC 60601-1 / 2xMOPP
Protection: Short circuit, Over load, Over voltage
Brown-out (Low AC Input Voltage)
High operation temperature up to 70°C
Withstand 2G vibration test
3 years warranty
MEDICAL
Model
Output
Input
Output Number
DC Voltage Range
Rated Current
Current Range
Rated Power
Peak Load (10sec.)
Ripple & Noise (Max.)
Voltage Tolerance
Line Regulation
Load Regulation
Setup,Rise Time
Hold Up Time
Voltage Range
Frequency Range
Efficiency (Typ.) at 230VAC
AC Current (Typ.)
Inrush Current (Typ.)
Leakage Current
Over Load
Protection
Over Voltage
Environment
Safety & EMC
Others
Note
Working Temp.
Working Humidity
Storage Temp. & Humidity
Temp. Coefficient
Vibration
MP45-05F
E324816
MP45-05K
CH1
CH2
CH1
CH2
5V
5V
12V
24V
3.5A
1.6A
4.2A
1.1A
0.5 ~ 3.85A
0.5 ~ 4.7A
0.1 ~ 1.76A
0.1 ~ 1.21A
43.9W
40.2W
44.62W
48.29W
80mVp-p
80mVp-p
80mVp-p
100mVp-p
+3 / -2%
+3 / -2%
+8 / -4%
±6.0%
±0.5%
±1.0%
±0.5%
±1.0%
±1.5%
±2.0%
±1.5%
±2.0%
200ms,15ms / 230VAC, 200ms,15ms / 115VAC at full load
70ms / 230VAC, 14ms / 115VAC at full load
90 ~ 264VAC, 127 ~ 373VDC (Withstand 300VAC surge for 5sec. Without damage)
47 ~ 63Hz
80%
82%
1.3A / 115VAC, 0.7A / 230VAC
Cold Start 30A / 115VAC, 60A / 230VAC
For earth < 200uA/264VAC, For patient < 100uA / 264VAC
Above 110% rated output power
Protection type: Hiccup mode, recovers automatically after fault condition is removed
115 ~ 140% rated output voltage
Protection type: latch-off mode
-20 ~ 70°C (Refer to de-rating curve)
20 ~ 90% RH non-condensing
-40 ~ +85°C, 10 ~ 95% RH
±0.03% / °C (0 ~ 50°C)
10 ~ 500Hz, 2G 10min. / 1cycle, period for 60min. each along X, Y, Z axes
Certified UL / cUL (ANSI / AAMI ES 60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99
Safety Standards
Certified TUV (EN 60601-1; IEC 60601-1); MOPP approval
I/P-O/P: 4KVAC, I/P-FG: 1.8KVAC, O/P-FG: 1.5KVAC
Withstand Voltage
Isolation Resistance
I/P-O/P: 100M Ohms / 500VDC
EMI Conduction & Radiation
Certified EN 55011
Harmonic Current
Certified EN 61000-3-2; EN 61000-3-3
Certified EN 60601-1-2; IEC 61000-4-2, 3, 4, 5, 6, 8, 11
EMS Immunity
MTBF
215.48K HRS Certified MIL-HDBK-217F
Dimension (WxHxD)
50.8x29.0x101.6 mm / 2.000x1.142x4.000 inch
Packing
0.151kg; 80pcs / 13.08kg / 1.43CUFT
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
3. Tolerance: includes setup time tolerance, line regulation and load regulation.
4. Line regulation is measured from low line to high line at rated load.
5. Load regulation is measured from 0% to 100% rated load.
6. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed
that it still meets EMC directives.
7. Length of setup time is measured at first cold start. Turning ON/OFF the power supply continuously may increase the setup time.
8. Max. 33% duty cycle within every 30 sec. Average output power should not exceed the rated power.
REV. B0
14/04/23
1
45W Dual Outputs Open Frame Medical Power Supply
MP45 series
Mechanical Drawings:
Unit : mm[inch]
1
2
3
4
5
6
50.8 [2.000]
CN1
3
2
1
44.45 [1.75]
3.175 [0.125]
CN2
3.175 [0.125]
95.25 [3.75]
4-Φ3.3
29 [1.142]
101.6 [4.000]
3 [0.118] max.
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
Mating Housing
AC / N
No Pin
AC / L
JST VHR
or equivalent
DC Output Connector (CN2) : JST B4P-VH or equivalent
Terminal
Pin No.
1,2
3,4
5
6
JST SVH-21T-P1.1
or equivalent
Assignment
Mating Housing
Terminal
V1
COM
V2
NC
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
Block Diagram:
RECTIFIERS
&
FILTER
EMI
FILTER
I/P
POWER
SWITCHING
RECTIFIERS
&
FILTER
V2
RECTIFIERS
&
FILTER
V1
COM
DETECTION
CIRCUIT
O.L.P.
FG
CONTROL
O.V.P.
De-rating Curve:
Load V.S Temp.
Load V.S I/P Voltage
Ta=25°C
100
100
90
Load (%)
Load (%)
80
60
50
80
70
60
40
50
20
-20
40
0
10
20
30
40
50
Ambient Temperature (°C)
60
70
(HORIZONTAL)
90
95
105
115
120
140
160
180
264Vac
Input Voltage(VAC) 60Hz
REV. B0
14/04/23
2