Download Data Sheet

DTAM200 Series
Rev. A
Active PFC
200 Watt Single Output
AC/DC Medical Desktop Power Supply
FEATURES
● High Reliability
● Universal Input
● 100% Hi-Pot Tested
● 100% Burn-In Tested
● IEC-320-C14 AC Inlet
● Safety Approvals, EMC Compliance
● 40 mm Cooling Fan with Fan Speed Control
● Meets CEC and Energy Star Level IV Requirements
● Short Circuit, Over Load, and Over Voltage Protected
SPECIFICATIONS: DTAM200 Series
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
Input Current (rms)
OUTPUT SPECIFICATIONS
Output Current
Output Voltage
Output Power
Voltage Regulation
Ripple & Noise (20MHz BW)
Hold Up Time
PROTECTION
Short Circuit Protection
Over Voltage Protection
Over Load Protection
GENERAL SPECIFICATIONS
90 - 264VAC (Full Range)
47 - 63Hz
3A @ 90VAC max; 1.5A @ 264VAC max
See Table
See Table
200W max.
±5%
See Table
>16ms typical at full load and 115VAC
Auto recover
Latch, AC Recycle
Auto recover
Efficiency
Power Factor
No Load Power Consumption (green functions)
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Humidity
Cooling
MTBF
PHYSICAL SPECIFICATIONS
Dimensions (L x W x H)
AC Input
SAFETY & EMI
Safety Standards
EMC Standards
> 87% @ full load and 115VAC
> 87% @ average load and 115/230VAC for CEC compliant models.
> 0.9 @ full load and 115 / 230VAC
< 500mW
0°C to +40°C
-10°C to +85°C
20% ~ 90%
40 mm cooling fan with fan speed control
>80,000 hours at full load and 25°C ambient temperature
7.42 x 4.11x 2.36 inches
188.5 x 104.5 x 60.0 mm
IEC 320-C14 Inlet
UL60601-1, CSA C 22.2-No. 60601-1, TUV EN60601-1, CB IEC60601-1
EN60601-1-2, FCC Part 18 Class B, EN55011 Class B, CE
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797
DTAM200 Series
Active PFC
200 Watt Single Output
AC/DC Medical Desktop Power Supply
Rev. A
OUTPUT VOLTAGE / CURRENT RATING CHART
Model
Output
Voltage
Output
Current
Output
Power
Output
Regulation
Ripple &
Noise
CEC (IV)
Energy
Star
DTAM-200A3-1Y12E
DTAM-200A3-1Y16E
DTAM-200A3-1Y19E
DTAM-200A3-1Y20E
DTAM-200A3-1Y24E
DTAM-200A3-1Y36E
DTAM-200A3-1Y48E
12 VDC
16 VDC
19 VDC
20 VDC
24 VDC
36 VDC
48 VDC
16.66 A
12.5 A
10.53 A
10 A
8.33 A
5.555 A
4.166 A
200W
200W
200W
200W
200W
200W
200W
±5%
±5%
±5%
±5%
±5%
±5%
±5%
240mVp-p
300mVp-p
300mVp-p
300mVp-p
300mVp-p
300mVp-p
300mVp-p
V
V
V
V
V
V
V
V (1.1)
V (2.0)
V (2.0)
V (2.0)
V (2.0)
V (2.0)
V (2.0)
NOTES
1. Ripple & Noise is measured at 20MHz bandwidth with a tantalum 10uF in parallel with a 0.1uF ceramic capacitor.
MECHANICAL DRAWING
Unit: inches [mm]
STANDARD OUTPUT CABLE
1. 12VDC: AWM 2464 #16AWG / 6C + SHIELDING, UL 80°C 300V VW-1, 1000mm ±30mm (MIN)
16VDC and up: AWM 2464 #16AWG / 4C + SHIELDING, UL 80°C 300V VW-1, 1800mm ±30mm (MIN)
2. Depends on customer’s requirements.
Wall Industries, Inc. 5 Watson Brook Road Exeter, NH 03833
603-778-2300 www.wallindustries.com Fax 603-778-9797