Download Data Sheet

PSIM65 SERIES
40~65 Watts
Single Output
Medical AC/DC Power Supplies
Rev B
FEATURES
• RoHS Compliant
• Single Outputs Ranging from 5VDC to 56VDC
• High Power Density
• -20°C to +70°C Operating Temperature Range
• 2” x 3” Open Frame Footprint
• No load Power Consumption < 0.2W at 230VAC
• 90~264 VAC Input Voltage Range
• Over Voltage, Over Load, and Short Circuit
• 87% High Efficiency (Except 5V models
Protection
• Ultra Low Leakage Current < 189µA at 264VAC
which are 85%)
• Up to 65 Watts Output Power
• Both Medical 60601-1 3rd Ed. MOPP & ITE 60950-1
• Meets DOE Level VI
Approvals
DESCRIPTION
Size:
3.00 x 2.00 x 1.13 inches
76.2 x 50.8 x 28.7 mm
Weight:
4.23oz (120g)
The PSIM65 series of AC/DC medical power supplies provides up to 65 Watts of output power in a
compact 2 x 3 inch open frame package. This series consists of single output models ranging from
5VDC to 56VDC with a 90~264VAC input voltage range. These supplies also feature a low leakage
current of less than 189μA at 264VAC, no load power consumption less than 0.2W, and 88% typical
efficiency. These supplies are also protected against short circuit, over voltage, and over load
conditions. The PSIM65 series is RoHS compliant and has both medical 60601-1 3rd edition and ITE
60950-1 safety approvals.
MODEL SELECTION TABLE
PSIM040B-1Y05
5 VDC
Output Current
Min
Max.
0A
8A
PSIM065B-1Y12
12 VDC
0A
Model Number
Input Voltage
Range
Output Voltage
5.416A
Output Power
Ripple & Noise
(1)
Output Regulation
-20°C~0°C 0°C~40°C
±10%
±5%
40W
150mVp-p
65W
240mVp-p
±10%
±3%
PSIM065B-1Y15
15 VDC
0A
4.333A
65W
300mVp-p
±10%
±3%
PSIM065B-1Y19
19 VDC
0A
3.421A
65W
300mVp-p
±10%
±3%
20 VDC
0A
3.25A
65W
300mVp-p
±10%
±3%
PSIM065B-1Y24
24 VDC
0A
2.708A
65W
300mVp-p
±10%
±3%
PSIM065B-1Y28
28 VDC
0A
2.321A
65W
300mVp-p
±10%
±3%
PSIM065B-1Y48
48 VDC
0A
1.354A
65W
300mVp-p
±10%
±3%
PSIM065B-1Y56
56 VDC
0A
1.161A
65W
300mVp-p
±10%
±3%
PSIM065B-1Y20
09/28/2015
90 - 264 VAC
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-7789797
website: www.wallindustries.com • e-mail: [email protected]
Page 1 of 4
PSIM65 SERIES
40~65 Watts
Single Output
Medical AC/DC Power Supplies
Rev B
SPECIFICATIONS: PSIM65 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
264
63
1
0.6
65
130
VAC
Hz
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
90
47
115VAC and full load
230VAC and full load
115VAC, cold start, 25°C
230VAC, cold start, 25°C
Input Current (rms)
Inrush Current
A
A
OUTPUT SPECIFICATIONS
Output Voltage
Output Regulation
See Table
See Table
40
65
See Table
5VDC output model
Others
Output Power
Output Current
Minimum Load
0
Ripple & Noise (20MHz BW)
Hold-up Time
Measured at 20MHz limited bandwidth and with a 100µF
electrolytic capacitor and a 0.1µF ceramic capacitor in parallel
across the output
115VAC and 75% load
10
Shutdown and latch-off
Automatic recovery
105
%
See Table
ms
PROTECTION
Over Voltage Protection
Over Load Protection
Short Circuit Protection
AC recycle
170
Automatic recovery
%
GENERAL SPECIFICATIONS
Efficiency
5V output model
Others
115VAC and full load
85
87
%
ENVIRONMENTAL SPECIFICATIONS
Operating Ambient Temperature
Derating linearly 2.5% per °C from 41°C to +70°C
Storage Temperature Range
Humidity
MTBF
Non-condensing
Full load and 25°C ambient temperature
-20
+70
°C
-20
+85
°C
0
100,000
90
%
hours
PHYSICAL SPECIFICATIONS
Weight
Dimensions (L x W x H)
Input Connector (CN1)
Output Connector (CN2)
Mates with JST VHR-3N or equivalent
Mates with JST VHR-4N or equivalent
4.23oz (120g)
3.00 x 2.00 x 1.13 inch
(76.2 x 50.8 x 28.7 mm)
JST B3P-VH-B pitch: 7.92mm or equivalent
JST B4P-VH-B pitch: 3.96mm or equivalent
SAFETY & EMC
Safety Approvals
EMC Standards
ITE
Medical
UL60950-1, CSA-C22.2 NO.950-1, EN60950-1, CB IEC60950-1
UL60601-1 3rd ed.., EN60601-1 3rd ed., CSA-C22.2 No.60601-1 3rd ed., IEC EN60601-1 3rd ed.
ITE
EN55022 Class B, CISPR22 Class B, FCC Part 15 Class B, CE
Medical
EN60601-1-2, FCC Part 18 Class B, EN55011 Class B, CE
NOTES
1. Ripple & noise is measured at 20MHz limited bandwidth and with a 100µF electrolytic capacitor and a 0.1µF ceramic capacitor in parallel across
the output
09/28/2015
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-7789797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 4
Rev B
PSIM65 SERIES
40~65 Watts
Single Output
Medical AC/DC Power Supplies
DERATING CURVES
MECHANICAL DRAWING
09/28/2015
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-7789797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 4
Rev B
PSIM65 SERIES
40~65 Watts
Single Output
Medical AC/DC Power Supplies
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will
provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one example
of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its
customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing support for
your project above and beyond the design and production phases. Give us a call today to discuss your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
09/28/2015
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-7789797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 4