CombiTemp TFRN

CombiTemp TFRN
TM
Industrial – RTD temperature sensor
Main features
„„
Pt100 sensor element,
„„
ATEX
„„
Programmable by touch screen
„„
Easy and full programmable
2- or 4-wire
„„
HART®, PA
„„
Built in graphical display,
­CombiViewTM DFON optional
„„
Head mounted 4...20 mA
t­ransmitter, FlexTop type 22xx
with FlexProgrammer 9701
Applications
„„
Oil and Gas
„„
Chemical
„„
Energy
„„
General Process Industrie
Sensor element and electrical specification
Accuracy
Pt100 output as to
DIN/EN/IEN 60751
Transmitter output
<±0.1 °C / <±0,25 °C
Connections
Threaded
Sensor type
RTD type Pt100 (acc. to DIN/EN/
IEC 60751)
Single or Double
2-wire or 4-wire
Accuracy
Class 1/1 B ±(0,3 + (0,005 x T)) °C
Class 1/3 B ±1/3 x
(0,3 + (0,005 x T)) °C
Class 1/6 B ±1/6 x
(0,3 + (0,005 x T)) °C
Class 1/1 A ±(0,15 + (0,002 x T)) °C
Analog output
4-20 mA, 4-20mA+HART®,
­Profibus®
See separate data sheet, series
22xx
Technical specification
Measuring principle
Resistance Temperature Detector
(RTD)
Measuring ranges
-50...400 °C
Immersion tube, diameter
Immersion tube, length
Immersion tube, tip
Ø6 mm, Ø8 mm
Min. 20 mm - Max. 3000 mm
Normal response - Ø 6/Ø 8 mm
Fast response - Ø 6/Ø 4 or
Ø 8/Ø 4 mm
Process connections
See page 4
Environment
Time constant, τ 0,5
Medium
Liquid
Air
Air
Velocity
0,4 m/sec.
0 m/sec.
3 m/sec.
Temperature, Ambient
-40…160 °C
ø 6 mm
<6,1
<138
<27,2
- w. transmitter
-40…85 °C
ø 6/4 mm
<1,5
<136
<21,4
- w. display
-30...80 °C
ø 8 mm
<7,6
<201
<47,7
Protection rating, IEC 529
IP67 / IP69K, depending on
­electrical connection
With BattTemp : IP54
ø 8/4 mm
<1,5
<181
<33,6
Humidity, IEC 68-2-38
98%, condensing
Vibrations
DNV high vibration strain, class B
1.6 mm, 2...25 Hz
IEC60068-2-6, test FC
25...100 Hz, 4.0 g
Material
SS 1.4404, AISI 316L
Housing
SS 1.4301, AISI 304
Sealing
See ordering table
Approvals
www.baumer.com
Process
°C
400
350
Standard, no cooling neck
With cooling neck
With cooling neck and distance piece
300
250
Process connection
Apply to
Temperature curve
200
150
100
0
EMC directive 2004/108/CE
in accordance with
EN61000-6-2, EN 61000-6-3
Pressure directive 97/23/CE
85°C
50
85°C
0
20
40
60
80
100
Ambient
°C
Data sheet TFRN
Design and speci ications subject to change without notice
-50 … 400 °C
EN/2016-04-21
Main characteristics (20 °C)
Standard process temperature
Page 1 / 5
CombiTemp TFRN
TM
Industrial – RTD temperature sensor
Technical Data
Transmitter, type FlexTop 2221 - Standard
Transmitter, type FlexTop 2202 - Standard
Input
Pt100
Input
Pt100 / Pt1000 (universal)
Output
Accuracy
4...20 mA
Output
Accuracy
4...20 mA / HART
Input
Output
Range
Minimum span
<0.25°C
<0.1% signal span (16mA)
-200...850°C
25°C
Supply
8...35 VDC
Programmability
By FlexProgrammer 9701
For further information please see data sheet for FlexTop 2202
<0.1°C
<0.1% signal span (16mA)
Range
Minimum span
-200...850°C
25°C
Supply
8...35 VDC
Programmability
By FlexProgrammer 9701
By HART terminal/modem
For further information please see data sheet for FlexTop 2221
Transmitter, type FlexTop 2211 - Performance
Input
Pt100 / Pt1000 (universal)
Output
Accuracy
4...20 mA
Input
Output
Input
Output
<0.1°C
<0.1% signal span (16mA)
Range
Minimum span
-200...850°C
25°C
Supply
8...35 VDC
Programmability
By FlexProgrammer 9701
For further information please see data sheet for FlexTop 2211
Display DFON
Transmitter, type FlexTop 2231 - Performance
Input
Pt100 / Pt1000 (universal)
Output
Accuracy
Profibus PA
Input
<0.1°C
Range
Minimum span
-200...850°C
25°C
Supply
9...32 VDC
Programmability
By FlexProgrammer 9701
For further information please see data sheet for FlexTop 2231
BattTemp
Type
Graphically LCD
Front glass
Polycarbonate
Display modes
8 modes, programmable, e.g.
value, bar graph, analogue, tank
illustration
Background
Measuring range
Type
LCD
Front glass
Polycarbonate
Temperature
sensor
Pt-100, 2-wire (DIN/EN/IEC
60751)
White, green, red - programmable
Power supply
Battery 1,5 V, type AA
-9999...99999
Range
-200…850°C
Digit height
Max. 22 mm
Measuring range
-9999…9999
Accuracy
0,1% @ ambient -10…70 °C
Unit
°C
Voltage drop
4V...6.5 V
Output
2 configurable relay output, 60 Vp,
75 mA
Digit height
Touch screen or FlexProgrammer
9701
Further information can be found in separate data sheet for DFON,
D21.09.
Accuracy
-200...-51°C
+/- 0,6 +/- 1 digit
-50...300°C
+/- 0,3 +/- 1 digit
301...850°C
+/- 0,6 +/- 1 digit
For further information please see datasheet for BattTemp
Design and speci ications subject to change without notice
Programming
11,5 mm.
Electrical
Electricalconnections
connections
M12, 8-wire
Cable gland
11
Transmitter
1
+24VDC / - 4...20mA
2
- 24VDC / +4...20mA
1
Red clip (FlexProgrammer)
2
Black clip (FlexProgrammer)
22
4...20 mA
RTD
1+2
3+4
1
2
3
4
5
Single Double
Pt100-1
Pt100-1
Pt100 – 1
Pt100 – 1
Pt100 – 2
Pt100 – 2
N.C.
N.C.
11
22
33
44
55
supply,4...20
4...20mA
mA
+- supply,
Commonfor
forrelays
relays
Common
supply,
4...20
mA
- +supply,
4...20
mA
Relay11
Relay
Relay22
Relay
11
22
33
44
55
66
77
88
N.C.
N.C.
supply,4...20
4...20mA
mA
+ -supply,
Relay
Relay1 1
Relay1 1
Relay
Relay2 2
Relay
Relay2 2
Relay
+ supply,
4...20
- supply,
4...20
mAmA
N.C.
N.C.
RTD
Single
White White
Red
Red
Double
White
White Red
Red
Display
+ 4...20
1
N.C. mA
- 4...20
2
N.C. mA
3
Relay 1
4
Relay 1
5
Relay 2
6
Relay 2
Com 1
Red clip (FlexProgrammer)
Com 2
Black clip (FlexProgrammer)
EN/2016-04-21
M12, 5-wire
Dimensions (mm)
Front view
www.baumer.com
Bottom connection
Rear connection
Data sheet TFRN
Page 2 / 5
CombiTemp TFRN
TM
Industrial – RTD temperature sensor
ATEX data for temperature transmitters and displays
Transmitter, type FlexTop 2202 - ATEX
Transmitter, type FlexTop 2221 - ATEX
Approval
Ex ia IIC T5/T6, ATEX II 1G
Ex nA II T5, ATEX II 3G
Approval
Ex ia IIC T5/T6, ATEX II 1G
Ex nA II T5, ATEX II 3G
Supply
8...28 VDC
Supply
8...30 VDC (Ex nA : 12...30 VDC)
Internal inductivity
Internal capacity
Li ≤ 10 µH
Ci ≤ 10 nF
T1...T5: -40 <Tamb < 85°C
T6:
-40 <Tamb < 50°C
Internal inductivity
Internal capacity
Li ≤ 15 µH
Ci ≤ 5 nF
T1...T5: -40 <Tamb < 85°C
T6:
-40 <Tamb < 50°C
Barrier data
U: ≤ 28 VDC
I: ≤ 0.1A
P: ≤ 0.75 W
Approval
Ex ia IIC T5/T6, ATEX II 1G
Ex nA II T5, ATEX II 3G
Supply
6.5...30 VDC
Internal inductivity
Internal capacity
Li ≤ 1 5 µH
Ci ≤ 5 nF
T1...T5: -40 <Tamb <85°C
T6:
-40 <Tamb <50°C
Barrier data
U: ≤ 30 VDC
I: ≤ 0.1A
P: ≤ 0.75 W
Display DFON - ATEX
ATEX Gas ia and for ATEX Dust ia
Voltage drop
UDisp Temperature class
Li ≤ 1 5 µH
Ci ≤ 5 nF
Temperature class
4.5 ... 6.5 VDC
T1...T5 Zone 0 and 20
-20 ... 60 °C
Zone 1/2 and 21/22
-40 ... 65 °C
Internal inductivity
L i
<10 µH
Internal capacity
Ci <15 nF
Ui Ii
Pi
<30 VDC
<0.1 A
<0.75 W
Barrier data
U: ≤ 30 VDC
I: ≤ 0.1A
P: ≤ 0.75 W
Approval
Internal inductivity
Internal capacity
Temperature class
Coupler/link
FISCO approved
Barrier data
Ex ia IIC T5/T6, ATEX II 1G
Ex nA II T5, ATEX II 3G
Li ≤ 10 µH
Ci ≤ 2 nF
T1...T4: -40 <Tamb < 85°C
T5:
-40 <Tamb < 60°C
Ui ≤17.5 VDC
Pi ≤0.75 W
Li ≤10 µH
Ii ≤ 0.275 A
Ci ≤2nF
Ui ≤20 VDC
Pi ≤0.75 W
Li ≤10 µH
Ii ≤ 0.1 A
Ci ≤ 2nF
ATEX data for BattTemp
II 1 G, Ex ia IIC T5 Ga
II 1 D, Ex ia IIIC T100°C Da
Approval Gas Zone 0/1
Dust Zone 20/21
Barrier data
Transmitter, type FlexTop 2231 - ATEX
Transmitter, type FlexTop 2211 - ATEX
Temperature class
Temperature class
Approval
Ex ia IIC T5/T6, ATEX II 1G
Temperature class
Battery : Energizer Lithium FR6
L91 AA
T1…T4: -10<Tamb<70°C
T5
: -10<Tamb<50°C
Battery: Duracell Alkaline
MN1500 LR6 AA
T1…T3: -10<Tamb<70°C
T4
: -10<Tamb<60°C
Design and speci ications subject to change without notice
Temperature class
ATEX Gas nA
II 3 G, Ex nA II T5
Approval Gas Zone 2
Voltage drop
UDisp Temperature class
T1...T5 -30 < Tamb< 65 °C
4.5 ... 6.5 VDC
Internal inductivity
L i
Internal capacity
Ci <15 nF
Maximum voltage
Umax
<35 VDC
Maximum current
Umax
<35 VDC
EN/2016-04-21
<10 µH
www.baumer.com
Data sheet TFRN
Page 3 / 5
CombiTemp TFRN
TM
Industrial – RTD temperature sensor
Dimensions (mm)
Ø80 mm housing front view
Ø80 mm housing bottom connection
Ø80 mm housing rear connection
68
91
95.5
Ø55
Ø91
66
60
82
60
54
Ø22
Ø22
www.baumer.com
1/2 - 14 NPT
22
L
Ø4 mm
EN/2016-04-21
Ø6 mm
Ø8 mm
Design and speci ications subject to change without notice
25
NV 24
R 1/2
17
17
L
14
L
L
Ø8 mm
NV 24
G1/2 DIN3852-2-A
G1/2 DIN 3852
Ø6 mm
25
NV 27
Ø22
L
28
NV 27
25
6
54
Ø22
17 mm
Data sheet TFRN
Page 4 / 5
CombiTemp TFRN
TM
Industrial – RTD temperature sensor
Ordering details
–
5
6
7
1
3
5
B
9
0
1
3
0
1
2
4
6
0
1
2
3
4
5
0
1
3
9
0
1
2
3
0
1
2
5
6
7
8
A
B
C
1
2
A
B
C
D
Design and speci ications subject to change without notice
0
4
5
6
10
11
12
13
30
0
1
2
3
0
1
2
0
1
2
0
x
x
x
x
If the product wanted is not available from above list please inquire.
www.baumer.com
Data sheet TFRN
Page 5 / 5
EN/2016-04-21
Model
TFRN
CombiTempTM
Housing material
Ø80 mm, Stainless steel, AISI 304 Bottom
Ø80 mm, Stainless steel, AISI 304 Rear
Field housing Ø55, stainless steel, AISI 304
Electrical connection
M12, 5 pins
M12, 8 pins
Cable gland, M16
Cable gland, M20
No electrical connection - for BattTemp only
Material el. connection
None - for BattTemp only
Plastic
AISI 304
Display
Without display, Ø55 housing
Without display
With display
No relays activated
With display
With activated relays
With BattTemp
Transmitter / socket
Flying leads
Ceramic socket
Pt100
Transmitter 2202 4-20 mA
±0,25 °C (Accuracy class)
Transmitter 2211 4-20 mA
±0,10 °C (Accuracy class)
®
Transmitter 2221 4-20 mA / HART ±0,10 °C (Accuracy class)
Transmitter 2231 Profibus®
±0,10 °C (Accuracy class)
Safety
Standard
Ex ia IIC T4/T5 1G (Gas)
Ex nA II T4/T5 3G (Gas)
Ex ia II 1 G Eex ia IIC, Zone 0, simple apparatus
Configuration
No configuration
Configuration of Range
Configuration of Range + Display
Configuration of Range + Display incl. 2 x relays
Sensor element (DIN/EN/IEC 60751)
None (for cable sensor)
1x Pt100 Class 1/1 B
2x Pt100 Class 1/1 B
1x Pt100 Class 1/3 B
2x Pt100 Class 1/3 B
1x Pt100 Class 1/6 B
2x Pt100 Class 1/6 B
1x Pt100 Class 1/1 A
2x Pt100 Class 1/1 A
1x Pt100 Class 1/1 B, < 600°C
Sensor insert type
Sensor tube with embedded sensor element 2-wire
Sensor tube with embedded sensor element 4-wire
Cable sensor Pt100 Class 1/1 B (1xPt100 only, for ø 8 mm only)
Cable sensor Pt100 Class 1/3 B (1xPt100 only, for ø 8 mm only)
Cable sensor Pt100 Class 1/6 B (1xPt100 only, for ø 8 mm only)
Cable sensor Pt100 Class 1/1 A (1xPt100 only, for ø 8 mm only)
Cooling neck
None
Cooling neck
Cooling neck + 1 spacer
Cooling neck + 2 spacers
Process connection
Tube without connection
G½
Male DIN 3852 form E
G½
Male DIN 3852 form A
R½
Male ISO 7/1
½"-14 NPT Male ANSI/ASME B1.20.1
Seal
No seal
Seal NBR
For G½ DIN 3852 - E
Seal EPDM
For G½ DIN 3852 - E
Seal FKM (Viton®) For G½ DIN 3852 - E
Sensor diameter
No sensor
ø 6 mm
AISI 316
ø 8 mm
AISI 316
Sensor tip
No sensor
Standard Normal response
Fast
Fast response
ø 4 mm tip
Max sensor length : 300 mm
Approvals
None
Sensor tube length
Length in mm (min. 20 mm)