CombiTemp

CombiTemp Temperature Measuring System
Flexible building block concept
All wetted parts in acid-proof, stainless steel
Some hygienic connections conform to 3-A standards,
FDA demands and EHEDG guidelines
Flush mounted surface sensor
ø80 mm stainless steel housing or
DIN head, form B
Pt100 or Pt1000 sensors, 2- or 4-wire
DIN A or B (1/1, 1/3, 1/6) elements
Standard or fast response time
Display and 4...20 mA transmitter
Cable sensors
Standard and Ex versions
Description
CombiTemp comprises a series of basic elements which can be combined to various temperature sensors and transmitters. The system
includes 4-wire Pt100 cable sensors.
Being a building block system CombiTemp offers a great flexibility in
respect to modification, service and maintenance.
A wide range of process connections according to national and international standards, sensor elements and transmitters, can be selected
to meet the actual requirements.
CombiTemp is particularly suitable for use in food, beverage, pharmaceutical and chemical industries.
A complete temperature sensor can be ordered on one type number.
Please refer to CombiTemp Accessories data sheet for a comprehensive range of accessories.
Please refer to the separate data sheets for information and ordering
details for transmitters and displays.
The parts can be assembled by the user or delivered assembled and
calibrated, if relevant.
Provided an 80 mm dia. housing is used, options are also a 4...20 mA
configurable display or, for stand alone indicating purposes, a battery
powered display.
www.baumer.com
Design and specifications subject to change without notice
Data Sheet 2000-1
Page Technical Data
Environmental conditions
Sensor tube and connection
Media temperature, std.
-50...400°C
Material
Acid-proof, stainless steel
Surface sensor
-40...150°C
(AISI 316L/W.1.4404)
Ambient temperature
-40...160°C
Media pressure
Max. 16 bar
(or max. temperature range for display/transmitter)
Time constant t0.5
See table below
Humidity
Mechanical tolerances
ISO 2768-m
< 100% RH, condensing
(or max. humidity for display/transmitter)
Sensor element
Protection class
DIN housing IP 65
ø80 mm housing IP 65 + IP 66
Vibrations
Sensor type
Pt100, Class A or B
Pt1000, Class B GL, test 2
Accuracy
DIN/EN/IEC 60751
(sensor tubes < 200 mm only)
1/1 DIN B 1/3 DIN B
1/6 DIN B
1/1 DIN A
±(0.3 + 0.005 x t) °C
+1/3 x (0.3 + 0.005 x t) °C
+1/6 x (0.3 + 0.005 x t) °C
+(0.15 + 0.002 x t) °C
Disposal of product and packing
According to national laws or by returning to Baumer
Time Constant τ0.5
Sensor type
Liquids
Dimension
Response
Insert
ø6 mm tube
fast ø8 mm tube
fast ø10 mm tube
fast ø12 mm tube
fast ø6 mm tube
normal ø8 mm tube
normal ø10 mm tube
normal
ø12 mm tube
normal ø8 mm tube
normal 5.6 mm
ø10 mm tube
normal 5.6 mm
ø12 mm tube
normal 5.6 mm
Surface sensor, flush mounted
0.4 m/sec.
< 1.5 sec.
< 1.5 sec.
< 1.5 sec.
< 1.5 sec.
< 6.1 sec.
< 7.6 sec.
< 11.1 sec.
< 16.2 sec.
< 13.6 sec.
< 28.1 sec.
< 31.3 sec.
< 1.0 sec.
Air
3 m/sec.
< 21.4 sec.
< 33.6 sec.
< 46.8 sec.
< 59.9 sec.
< 27.2 sec.
< 47.7 sec.
< 57.8 sec.
< 70.8 sec.
< 51.1 sec.
< 67.0 sec.
< 82.3 sec.
0 m/sec.
< 135.6 sec.
< 181.0 sec.
< 238.9 sec.
< 311.4 sec.
< 137.8 sec.
< 200.9 sec.
< 270.6 sec.
< 319.8 sec.
< 253.1 sec.
< 271.1 sec.
< 289.3 sec.
Dimensional Drawings - Mounting Details
[mm]
76
55
85
58
71
G1/4A
7.5
12.5
Clamp ISO 2852, DN38 mounted on a
Stainless steel housing, ø55 mm
www.baumer.com
Design and specifications subject to change without notice
Male Nipple G1/2A
Male Nipple G1/2A
mounted on DIN B housing
11
Male Nipple G1/2A
mounted on ø80 mm housing
Data Sheet 2000-1
Page Examples of Application
[mm]
R
74 - 03
8141 4211 0500 110 xxxx
8141 H241 07F0 110 xxxx
8142 D232 0500 110 xxxx
8142 C252 0000 000 xxxx
+CombiConnect connection
R
74 - 03
R
74 - 03
8140 0007 03A2 010 0000
Cable sensor 8155-3xx-xxxx
BattTemp 86 30-511
8142 D231 02A2 213 xxxx
Display 81 46-525
Transmitter 22xx-000x
8142 3212 02A1 110 xxxx
8141 3211 0272 010 xxxx
BattTemp 86 30-511
R
PTFE
Thermal
isolation
74 - 03
8140 E010 02A2 213 0000
Display 81 46-525
Transmitter 22xx-000x
8142 D231 07F0 210 xxxx
Transmitter 22xx-000x
8141 3212 12A2 121 xxxx
Sensor Tube and -Element Selection Table
Sensor tube
[mm]
Embedded sensor,
max. [mm]
6
6000
8
6000
10
12
Cable sensor,
max. [mm]
Single,
2-wire
Duplex,
2-wire
Single,
4-wire
x
x
x
1000
x
x
x
6000
6000
x
x
x
6000
6000
x
x
x
www.baumer.com
Design and specifications subject to change without notice
Duplex,
4 wire
x
Data Sheet 2000-1
Page Ordering Details - CombiTemp Series
Sensor tip
Not specified
Normal response sensor tip
Fast response sensor tip, Note {1}, {6}
Connection
Not specified
Sensor tube without thread or connection
R1/2 Male nipple, ISO 7/1
G1/2A Male nipple, ISO 228/1
G1/2 Rotating female nut, ISO 228/1
G1/2A Rotating male nipple, ISO 228/1
G3/4 Rotating female nut, ISO 228/1
G3/4A Rotating male nipple, ISO 228/1
G1 Rotating female nut, ISO 228/1
G1A Rotating male nipple, ISO 228/1
1/2“ - 14 NPT Male nipple, ANSI/ASME B1.20.1
G1/2A Male nipple for CombiConnect, ISO 228/1
3A/DN38 Hygienic, sensor tube, 3-A conform, Note {8}
3A/DN38 Hygienic, flush mounted surface sensor, 3-A conform, Note {8}
GEA Tuchenhagen Varivent flange, DN40/DN50
Clamp DN25/38, ISO 2852, 3-A conform, Note {8}
Clamp DN51, ISO 2852, 3-A conform, Note {8}
1/2" - 3/4" Tri-clamp®
M20 x 1.5 Male nipple, DS/ISO 261
SMS 1145 Rotating female nut, DN38
SMS 1145 Rotating female nut, DN51
As customer specification
4´ digit
5´ digit
0
1
2
3
4
5
6
7
8
9
A
C
D
E
G
H
I
K
M
P
Q
S
Sensor tube dimension
6´ digit
Not specified
ø6 mm x 1 mm, AISI 316L, see table page 3
ø8 mm x 1 mm, AISI 316L, see table page 3
ø10 mm x 1 mm, AISI 316L, see table page 3
ø12 mm x 1 mm, AISI 316L, see table page 3
As customer specification
Sensor element
7´ digit
Not specified or empty sensor tube
Pt100, 1/1 DIN B, single, specified accuracy -50...400°C
Pt100, 1/1 DIN B, duplex, specified accuracy -50...400°C
Pt100, 1/3 DIN B, single, specified accuracy 0...150°C
Pt100, 1/3 DIN B, duplex, specified accuracy 0...150°C
Pt100, 1/6 DIN B, single, specified accuracy 0...100°C
Pt100, 1/6 DIN B, duplex, specified accuracy 0...100°C
Pt100, 1/1 DIN A, single, specified accuracy -50...400°C
Pt100, 1/1 DIN A, duplex, specified accuracy -50...400°C
Pt100, 1/1 DIN B, single, specified accuracy -50...600°C, Note {7}
Pt1000, 1/3 DIN B, single, specified accuracy -50...400°C
Pt1000, 1/1 DIN B, single, specified accuracy -50...400°C
As customer specification
Sensor insert type
Not specified
Sensor tube with embedded sensor element, 2-wire
Sensor tube with embedded sensor element, 4-wire
ø5.6 x 0.5 mm insert AISI 316, 2-wire, max. length 600 mm, Note {3}, {7}
ø5.6 x 0.5 mm insert AISI 316, 4-wire, max. length 600 mm, Note {3}, {7}
ø5.6 x 0.5 mm insert spring-loaded AISI 316, 2-wire, max. length 600 mm, Note {3}, {7}
ø5.6 x 0.5 mm insert spring-loaded AISI 316, 4-wire, max. length 600 mm, Note {3}, {7}
Cable sensor, 4-wire, Note {4}
www.baumer.com
Design and specifications subject to change without notice
814x xxxx - - - - - - - - - - - - - 0
1
2
0
1
2
3
4
S
0
1
2
3
4
5
6
7
8
9
A
B
S
8´ digit
0
1
2
3
4
5
6
7
Data Sheet 2000-1
Page Ordering Details - CombiTemp Series (continued)
- - - - - - xxxx xxx
Cooling neck
Not specified
1 off cooling neck (71 mm)
2 off cooling neck (142 mm)
3 off cooling neck (213 mm)
As customer specification
0
1
2
3
S
Housing
Not specified
ø80 mm stainless steel housing, Note {5}
ø80 mm stainless steel housing with ø110 mm wall bracket, Note {5}
DIN B head with cover and M20 gland
DIN B head with cover and M20 gland, shield connection terminal
TE1 Field housing (no display)
As customer specification
10´ digit
Electrical connection
Not specified or std. DIN B head selected in the previous choice
ø80 mm housing: 1 off M20 x 1.5 adaptor and 1 off blind plug (no gland incl.)
ø80 mm housing: 2 off M20 x 1.5 adaptor (no gland incl.)
ø80 mm housing: 2 off blind plugs
ø80 mm housing: 1 off M16 gland and 1 off blind plug
ø80 mm housing: 2 off M16 gland
ø80 mm housing: 1 off M16 with shield connection terminal and 1 off blind plug
DIN B head: 1 off M20 gland for marine applications
TE1 housing: M12 plug, 4 pole
TE1 housing: M16, plastic
TE1 housing: M16 gland, nickel plated brass
TE1 housing: M16 gland, nickel plated brass, with earth connection terminal
ø80 mm housing: M12 plug, 4 pole
As customer specification
11´ digit
Cover or display
Not specified
Cover for ø80 mm housing
Display, Note {4}
Terminal block or transmitter
Not specified
Ceramic terminal block
Transmitter, Note {4}
12´ digit
Assembling
Not specified
Assembled, process connection at the base
Assembled, process connection at the rear (ø80 mm housing only)
Welded, process connection at the base (ø80 mm housing only)
Welded, process connection at the rear (ø80 mm housing only)
14´ digit
Approvals
Standard version
Ex ia (zone 0) II 1 G EEx ia IIC
Ex n (zone 2) II 3 G EEx nA II
15´ digit
Sensor tube length (L)
Length in mm. Please observe max. lengths
xxxx
9´ digit
0
2
3
5
6
7
S
0
5
6
7
A
B
C
D
E
F
G
H
I
S
0
1
2
13´ digit
0
1
2
0
1
2
3
4
0
4
5
16...19´ digit
xxxx
Calibration certificate, -30...150°C, 4 points, order type number 5509-142
Material certificate, 3.1 (EN10 204), order type number 5509-227
Cable type no.:
132548
4 pol M12 plug with 5m cable
127804
4 pol M12 angle plug with 5m cable
www.baumer.com
Design and specifications subject to change without notice
Data Sheet 2000-1
Page Notes:
{1} Single sensor element only. Max. sensor length 300 mm.
{3} ø80 mm housing: Insert only if sensor tube is mounted at the rear.
Observe length, refer to „Sensor Insert Type“.
{4} Refer to data sheet. Specify separate type number and configuration.
{5} Assembling is mandatory. Valid for 3A applications.
{6} Embedded sensors only.
{7} Not valid for a sensor with fast response tip.
Dimensional Drawings
[mm]
12.5
12.5
AF22
27.5
G1/2A
G3/4A
G1A
L
G1/2A
G3/4A
G1A
27.5
32
35
L
ø6...12
ø6...12
ø4...12
ø4...12
AF27
AF32
AF36
Male nipple G1/2A, R1/2 and M20 x 1.5
Rotating male nipple G1/2A, G3/4A and G1A
R
57
74 - 03
DN25/38
DN51
L
ø6...12
ø4...12
Clamp DN25/38/51
EPDM gasket included, Note {8}
www.baumer.com
Design and specifications subject to change without notice
G1/2
G3/4
G1
51
ø50.5
ø64
L
AF27
AF32
AF41
ø6...12
ø4...12
Rotating female nut G1/2, G3/4 and G1
Data Sheet 2000-1
Page Dimensional Drawings
R
R
36
74 - 03
36
AF22
74 - 03
52
31.5
31.5
ø6...12
L
ø6...12
L
ø4...12
ø4...12
Male nipple G1/2A for CombiConnect
3A/DN38 hygienic sensor tube
EPDM O-ring included, Note {8}
3A/DN38 hygienic flush mounted
surface sensor
EPDM O-ring included, Note {8}
[mm]
12.5
12.5
AF22
6
51.5
34.4
ø6...12
L
ø6...12
L
ø4...12
ø6...12
L
ø4...12
ø4...12
DN38
DN51
Male nipple 1/2“ - 14 NPT
Sensor tube without connection
ø25
30
ø55
ø65
SMS 1145 rotating female nut
25
19
L
ø6...12
ø4...12
L
ø6...12
ø4...12
Tri-Clamp® 1/2 - 3/4 www.baumer.com
Design and specifications subject to change without notice
GEA Tuchenhagen Varivent DN40/DN50
Data Sheet 2000-1
Page Sensor inserts
Calculation of insert length:
When ordering an insert the length must be calculated from the
formula:
Housing shaft {8}
Total body length
Insert length = Sensor length + total body length + housing shaft
(Total body length can be calculated from the dimensional drawings on
pages 6 and 7)
Sensor length (L)
Example.
Insert for a G1/2A male nipple sensor in a DIN housing, 100 mm
sensor tube, normal response sensor tip:
Housing shaft = 11.5 mm
Total body length = 27.5 + 12.5 = 40 mm (see page 6)
Sensor length = 100 mm
DIN housing with G1/2A male nipple connection
and sensor insert
Note {8}
Housing shaft
DIN housing: 11.5 mm
ø80 housing: 6.5 mm
This insert must be ordered with a 151.5 mm sensor tube.
Notes
Single sensor element only. Max. sensor length 300 mm.
ø80 mm housing: Insert only if sensor tube is mounted at the
rear. Observe length, refer to „Sensor Insert Type“.
Refer to data sheet.
Specify separate type number and configuration.
Assembling is mandatory. Valid for 3-A applications.
Embedded sensors only.
Not valid for a sensor with fast response tip.
{8} The 3-A mark is valid only when the product is mounted in a 3-A
marked counter part and installed according to the installation manual.
Use also a 3-A marked O-ring or gasket if relevant. The 3-A marked
products conforms to the 3-A Sanitary Standard criteria. Materials and
surfaces fulfill the FDA demands and follow the EHEDG guidelines
regarding design, materials and finishing.
EPDM O-rings supplied with 3-A marked products are conform to
Sanitary Standard Class II (8% milk fat max.)
EPDM gaskets supplied with 3-A marked products are conform to
Sanitary Standard Class I (8% milk fat max.)
4-wire Sensors
White
White
Duplex Elements
Red
Red
White
White
Red
Red
Warning:
One of the wires may not be
connected in case of a 3-wire
connection to the temperature
transmitter
Application Photos
ø80 mm housing with FlexView
Gland (blind plug at the rear)
Clamp connection
Fast response sensor
Cut through DIN - B housing
FlexTop 2201 temperature transmitter
G1/2A male nipple connection
Normal response sensor
www.baumer.com
Design and specifications subject to change without notice
UK/2010-02-02 This data sheet may only be reproduced in full.
{1}
{3}
{4}
{5}
{6}
{7}
ø80 mm housing with 2 glands
Cover not mounted
3A/DN38 connection at the rear
Fast response sensor
Data Sheet 2000-1
Page