Accessories for infrared detector
Temperature
controllers
Heatsinks for
TE-cooled detector
Chopper, etc.
Wide lineups of accessories for infrared detector
HAMAMATSU provides temperature controllers, heatsinks for TE-cooled detector, chopper and cables, etc as accessories for infrared detectors.
Connection example of accessories for infrared detectors
*1
Power supply (±15 V)
Chopper
C4696
TE-cooled
detector*3
*2
POWER
OUT
Measuring instrument
Amplifier for
infrared detector*4
C4159 series, C5185-02
Heatsink for
TE-cooled
detector
A3179 series
Power supply
(100 V, 115 V, 230 V)
ûC
GND
+12 V
POWER
CHOPPER
TRIG
*3
Chopper driver circuit
(accessory of the C4696)
Temperature
controller
C1103 series
Power supply (+12 V)
KACCC0321ED
*1: Attach the bare wire ends to a 3-pin or 4-pin connector or to a banana jack, and then connect them to the power supply.
*2: Soldering is needed. When using the C5185-02 amplifier, a BNC connector (prepared by the user, example: one end of the E2573) is required.
*3: No socket is available. Soldering is needed.
*4: Refer to the datasheet “Amplifiers for infrared detectors” for detailed information.
Note: Refer to page 6 for details on cables.
www.hamamatsu.com
1
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Accessories for infrared detector
Temperature controllers C1103 series
The C1103 series is a temperature controller designed for thermoelectrically cooled infrared detectors. The C1103 series allows easy
but accurate temperature setting for the thermoelectric cooler mounted in an infrared detector.
Specifications
Parameter
C1103-04
C1103-05
C1103-07
One-stage/two-stage TE-cooled type
Two-stage/three-stage TE-cooled type
One-stage TE-cooled type
Applicable detector*5
InAsSb, InAs photovoltaic detector,
InSb photoconductive detector
InSb photoconductive detector
InGaAs, Si photodiode
Setting element temperature
-30 to +20 °C
-75 to -25 °C
-30 to +20 °C
Temperature stability
Within ±0.1 °C
Output current for temperature control
1.1 A min., 1.2 A typ., 1.3 A max.
Power supply
100 V ± 10 % · 50/60 Hz*6
Power consumption
30 W
Dimensions
107 (W) × 84 (H) × 190 (D) mm
Weight
1.9 kg approx.
Instruction manual
Accessories
4-conductor cable (with a connector, 3 m) A4372-05*7, power supply cable
*5: It doesn't correspond to TE-cooled type infrared detector module with preamp.
*6: Please specify power supply requirement (AC line voltage) from among 100 V, 115 V and 230 V when ordering.
*7: When used in combination with an A3179 series heatsink, do not use the 4-conductor cable supplied with the A3179 series, but use
the A4372-05 instead.
Absolute maximum ratings
Parameter
Operating temperature
Operating humidity
Storage temperature
Value
+10 to +40 °C
90% Max.*8
-20 to +40 °C
*8: No dew condensation
When there is a temperature difference between a product and the surrounding area in high humidity environment,
dew condensation may occur on the product surface. Dew condensation on the product may cause deterioration in characteristics
and reliability.
Note: Exceeding the absolute maximum ratings even momentarily may cause a drop in product quality. Always be sure to use the
product within the absolute maximum ratings.
Block diagram
C1103 series
TE-cooled
detector
Thermistor
Comparator
Amp circuit
Current
circuit
TE-cooled element
Power
supply
AC input
KIRDC0008EB
2
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Accessories for infrared detector
Heatsinks for TE-cooled detector (TO-8, TO-3 package) A3179 series
These heatsinks are designed for use with thermoelectrically cooled detector sealed in a 6-pin TO-8, TO-3 package. The cooling (heat
dissipation) capacity of the A3179 and A3179-03 is about 35 °C relative to the ambient temperature 25 °C, the A3179-01 is about 40
°C, and that of the A3179-04 is about 85 °C. The A3179-03 is designed only for two-color detector K3413 series, the A3179, A3179-01
for TO-8, the A3179-04 for TO-3 (heatsink for TO-66 is available as a custom product.) .
Accessories
*9: When used in combination with a C1103 series temperature
controller, do not use the 4-conductor cable supplied with the
A3179 series, but use the 4-conductor cable A4372-05 (sold
separately, with a connector).
*10: No socket is supplied for connection to infrared detectors.
Connect infrared detectors by soldering.
Cover the soldered joints and detector pins with vinyl insulating tubes.
Instruction manual
4-conductor cable (2 m): for TE-cooler and thermistor*9, *10
Coaxial cable (2 m): for signal*10
Dimensional outlines (unit: mm, tolerance unless otherwise noted: ±0.3)
A3179
ϕ32
3
3
ϕ46
0.4 ± 0.3*2
ϕ26 ± 0.2
*1
Photosensitive
surface*3
Detector
metal package
32
(4 ×) ϕ3.5
ϕ40
60°
32.6
Weight: 50 g approx.
*1: Bottom surface (reference surface) of detector metal package
*2: When the detector is installed
*3: The position of the photosensitive surface differs according
to the detector used.
Refer to the dimensional outline for the detector.
KIRDA0018ED
3
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Accessories for infrared detector
A3179-01, A3179-03
(4 ×) ϕ3.5
32
ϕ40
ϕ46
ϕ32
3
Photosensitive
surface*3
Detector
metal package
0.4 ± 0.3*2
ϕ26 ± 0.2
*1
B
60°
36
A3179-01: B=6
A3179-03: B=6.4
Weight: 53 g approx.
*1: Bottom surface (reference surface) of detector metal package
*2: When detector is installed
*3: The position of the photosensitive surface differs according
to the detector used.
Refer to the dimensional outline for the detector.
KIRDA0019ED
3 ± 0.5
A3179-04
Fixation board
40
56
1
+ 0
Photosensitive
surface
79.
ϕ1
ϕ38 ± 0.5
(4 ×) ϕ3.5 ± 0.5
56
6.7 ± 0.5
68
45
80
Weight: 320 g approx.
KIRDA0149EB
4
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Accessories for infrared detector
Chopper C4696
Specifications
Parameter
Chopping frequency*11
Power supply (VD)
Duty ratio
Rotational stability
Min.
Synchronous signal
(high level)
Max.
Operating temperature
Maximum current consumption*12
Accessory
Specification
115 to 380 Hz, 345 Hz typ.*12
DC 5 to 13 V, 12 V typ.
1:1
0.06 %/°C
VD - 0.5 V
VD - 0.2 V
0 to 50 °C
90 mA
Magnet stand A1447
*11: Chopping frequency will be 230 to 760 Hz when an optional disc is used.
*12: VD=12 V
Dimensional outline (unit: mm, tolerance unless otherwise noted: ±1)
<Chopper>
Output
window
ϕ8.0
19
Input window
ϕ4.0
23.0 45.0
202.3 to 272.3 (adjustable)
8
ϕ8
A1447
(Magnet stand)
6-pin receptacle
cord length: 2 m
(For connection
to driver circuit)
<Driver circuit>
38.5
COUNTER
13.5
12.5
25
37
GND
+12 V
ON-OFF
TRIG
BNC
18
12.5
17.5
85
30
60
6-pin connector
KIRDA0022EA
5
Accessories for infrared detector
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Chopping frequency vs. operating voltage
(Typ.)
800
Chopping frequency (Hz)
700
600
500
When used with
optional disk
400
300
200
C4696
100
0
2
4
6
8
10
12
14
Operating voltage (V)
KIRDB0376EA
Cables A4372 series
Cable No.
Cable
Length approx.

Coaxial cable (for signal)
2m

4-conductor cable (with a connector)
A4372-05
3m
‘
4-conductor cable (with a connector)
A4372-02
2m
’
“
”
•
BNC connector cable E2573
Power supply cable
(for temperature controller)
Chopper driver cable
(connected to chopper)
2-conductor cable or coaxial cable
(for chopper power supply)
1m
1.9 m
2m
2 m or less
Note
Supplied with heatsink A3179 series.
When using this cable, make it as short as possible
(preferably about 10 cm).
Supplied with temperature controller C1103 series.
This cable is also sold separately.
This cable is supplied with the C4159 series, C5185-02 amplifiers
for infrared detector, and infrared detector modules with preamp
(room temperature operation type).
This cable is also sold separately. Besides this cable, the A437203, which is a 6-conductor cable (with connector) supplied with
“infrared detector module with preamp”, is also sold separately.
Option
Supplied with temperature controller C1103 series
Connected to chopper
Prepared by user
6
Accessories for infrared detector
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Magnet stand A1447
This is a magnet stand (sold separately) designed for the B749 photon drag detector.
Dimensional outline (unit: mm, tolerance unless otherwise noted: ±1)
1/4-20UNC
100
49.5
160 min. 230 max.
5.5
ϕ20
KIRDA0017EA
7
Temperature controllers,
heatsinks for TE-cooled detector, chopper, etc.
Accessories for infrared detector
Valve operator for metal dewar A3515
With this valve operator, metal dewars can be re-evacuated to maintain the desired vacuum level. Refer to the instruction manual for details.
Please be aware that the detector performance is not guaranteed after re-evacuation is performed with the valve operator.
Vacuum pump
Metal dewar type infrared detector
Valve operator
Dimensional outline (unit: mm)
Pump tube
ϕ9.5 ± 0.5
Leak mount
Gland nut
O-ring
Knob
80.0 ± 1 (close)
115.0 ± 1 (open)
KIRDA0021EC
Related information
www.hamamatsu.com/sp/ssd/doc_en.html
Precautions
∙ Disclaimer
Technical information
∙ Infrared detector
Information described in this material is current as of November, 2015.
Product specifications are subject to change without prior notice due to improvements or other reasons. This document has been carefully prepared and the
information contained is believed to be accurate. In rare cases, however, there may be inaccuracies such as text errors. Before using these products, always
contact us for the delivery specification sheet to check the latest specifications.
The product warranty is valid for one year after delivery and is limited to product repair or replacement for defects discovered and reported to us within that
one year period. However, even if within the warranty period we accept absolutely no liability for any loss caused by natural disasters or improper product use.
Copying or reprinting the contents described in this material in whole or in part is prohibited without our prior permission.
www.hamamatsu.com
HAMAMATSU PHOTONICS K.K., Solid State Division
1126-1 Ichino-cho, Higashi-ku, Hamamatsu City, 435-8558 Japan, Telephone: (81) 53-434-3311, Fax: (81) 53-434-5184
U.S.A.: Hamamatsu Corporation: 360 Foothill Road, Bridgewater, N.J. 08807, U.S.A., Telephone: (1) 908-231-0960, Fax: (1) 908-231-1218
Germany: Hamamatsu Photonics Deutschland GmbH: Arzbergerstr. 10, D-82211 Herrsching am Ammersee, Germany, Telephone: (49) 8152-375-0, Fax: (49) 8152-265-8
France: Hamamatsu Photonics France S.A.R.L.: 19, Rue du Saule Trapu, Parc du Moulin de Massy, 91882 Massy Cedex, France, Telephone: 33-(1) 69 53 71 00, Fax: 33-(1) 69 53 71 10
United Kingdom: Hamamatsu Photonics UK Limited: 2 Howard Court, 10 Tewin Road, Welwyn Garden City, Hertfordshire AL7 1BW, United Kingdom, Telephone: (44) 1707-294888, Fax: (44) 1707-325777
North Europe: Hamamatsu Photonics Norden AB: Torshamnsgatan 35 16440 Kista, Sweden, Telephone: (46) 8-509-031-00, Fax: (46) 8-509-031-01
Italy: Hamamatsu Photonics Italia S.r.l.: Strada della Moia, 1 int. 6, 20020 Arese (Milano), Italy, Telephone: (39) 02-93581733, Fax: (39) 02-93581741
China: Hamamatsu Photonics (China) Co., Ltd.: B1201, Jiaming Center, No.27 Dongsanhuan Beilu, Chaoyang District, Beijing 100020, China, Telephone: (86) 10-6586-6006, Fax: (86) 10-6586-2866
Cat. No. KACC1062E15 Nov. 2015 DN
8