FUJI EDS3-116E

CO/O2 GAS ANALYZER FOR STACK GAS
DATA SHEET
ZSQ
This instrument is equipped with a non-dispersive infrared
CO sensor and a zirconia O2 sensor. It is used for measuring CO and O2 in the gas emitted from a garbage incinerator to meet Japanese regulation on emission control of
dioxins (tetrachlorodibenzoparadioxins).
CO instantaneous value, O2 instantaneous value, CO instantaneous value after O2 correction and CO average values
after O2 correction (1-hour and time-variable averagings) can
be output, and a peak count/alarm function is incorporated.
FEATURES
1. Suitable functions for emission control of dioxins.
• Output includes CO instantaneous value, O2 instantaneous
value, CO instantaneous value after O2 correction and
CO average values after O2 correction (1-hour and timevariable averagings).
• Count/alarm function against excess over specified value
2. Space-saving configuration by unitizing the analyzer and
sampling device into 19-inch types.
3. Easy to maintain with auto calibration and with “From
the front maintenance construction”.
4. A microjet recorder (Fuji Electric’s type PHC or PHE) can
be built in for recording measured gas concentration and
temperature values.
SPECIFICATIONS
• Measuring system:
Non-dispersive infrared absorption (NDIR)
method for CO
Zirconia method for O2
• Measurable component and range:
CO ; 0~200/500ppm
0~200/1000ppm
0~500/1000ppm
0~500/2000ppm
O2 ; 0~25vol%
• Repeatability: ± 0.5% of full scale
• Zero drift:
Max. ± 2.0% of full scale/week for CO
Max. ± 2.0% of full scale/month for O2
• Span drift:
Max. ± 2.0% of full scale/week
• Linearity:
Max. ± 1.0% of full scale
• Sample gas extracting rate:
Approx. 2R/min
• Response time: Within 90 seconds for 90% indication (after extracting sample gas through the
inlet)
• Warm-up time: Within 4 hours after power-on
• Output signals: Each signal within range from 4 to 20 mA
DC
Non-isolated (isolated output available at
option)
CO instantaneous value
O2 instantaneous value
CO instantaneous value after O2 correction
1-hour moving average CO value after O2
correction
Moving average CO value after O2 correction (averaging time variable)
Time setting in 1 to 59 minutes (1-minute
increment) or 1 to 4 hours (1-hour increment)
Allowable load resistance 550 Ω or less
• Contact output: 1a no-voltage contact (contact capacity
250 V AC, 2 A)
Peak count alarm, CO instantaneous
value upper limit alarm, auto calibration
and maintenance statuses, calibration error, analyzing block error and range discrimination
• Contact input: No-voltage contact (ON/0V, OFF/5 V DC,
DC 5mA at ON)
Auto calibration remote start (for 500
msec or longer), remote range
changeover
Remote average value resetting (for 500
msec or longer)
• Indication:
LCD with back light for indicating CO instantaneous value, O 2 instantaneous
value, CO instantaneous value after O2
correction and CO average values after
O2 correction (1-hour and time-variable
averagings), peak count value and parameter assignment
EDS3-116e
Date
Jun. 30, 2011
ZSQ
• Recorder (option):
100 mm-width recorder (Fuji Electric’s
type PHC or PHE) built in, max. 6 points
recordable
• Gas extractor: Electrical heating type (filter built in)
• Filter mesh; 40µm mesh of SUS 316
stainless steel
• Probe; SUS 316 stainless steel of 300,
400, 600 or 800 mm length
• Flange; JIS 5K65AFF
• Mass; Approx. 9 kg
• Sample inlet tube:
φ10/φ8 Teflon tube or heating tube (max.
30 m)
• Functions:
1) O2 correction
Conversion of measured CO gas concentration into a value
at standard O2 concentration
Calculating equation: C =
C;
CS;
OS;
ON;
CS(21−ON)
21−OS
CO concentration after O2 correction
Measured CO concentration
Measured O2 concentration (%)
Standard O2 concentration (12% for
garbage incinerator)
2) Auto calibration
• Auto calibration cycle settable range: 1 to 99 hours (1hour step) or 1 to 40 days (1-day step)
• Auto calibration gas flow time settable range: 60 to 599
seconds (1-sec step)
• Auto/manual calibration error contact output: Provided
when calibration quantity exceeds 50% of full scale.
• Contact output during auto calibration and maintenance:
Provided during calibration gas flow and replacement.
Also provided during maintenance.
• Auto calibration remote start contact input: Calibration
starts by opening and closing the contact input of 500
msec or longer in duration.
3) Output holding function
Output signal is held during auto calibration and maintenance.
4) Alarm
1 CO instantaneous value upper limit alarm
2 CO instantaneous value peak count
Alarm is issued and indicated when CO instantaneous
value has exceeded the set upper limit by the set number
of times. (Settable number of times: 1 to 99, settable
value range: 0 to 1000 ppm)
5) Other functions
• Remote range changeover input: Low range is selected
when closing the contact input, and high range is selected when opening the contact input
• Average value resetting input: The output and indication
of CO average value after O2 correction is reset by opening the contact after the contact is closed for 1.5 sec or
longer.
• Range discrimination contact output: Contact closed for
low range
• Analyzing block error contact output:
• Temperature input signal: K thermocouple input x 2 (input for recorder)
2
• Standard requirements for sample gas:
• Temperature; 60 to 800˚C
• Dust;
100 mg/Nm3 or less
• Pressure;
–2.94 to +2.94kPa
{–300 to +300mmH2O}
• Components; SO2 500 ppm or less
NOx 1000 ppm or less
CO2 0 to 15%
CO
0 to 2000 ppm
0 to 21%
O2
HCR 1000 ppm or less
N2, H2O Remaining percent
• Rated operating conditions:
• Ambient temperature; –5 to +40˚C
• Ambient humidity; 90% RH or less
• Power supply voltage; 100 V AC ± 10 V
• Power supply frequency;
50 or 60 Hz ± 0.5 Hz
• Power consumption; Max. 400 VA
(without gas extractor)
• Installation requirements:
1) Selection of a place which does not
receive direct sunlight or radiation
from hot substances
2) Avoidance of a place under heavy vibration
3) Clean atmospheric air
• External dimensions (H x W x D):
Indoor type
1550 x 735 x 650 mm
Outdoor type
1640 x 745 x 765 mm
• Mass:
Approx. 140 kg (excluding standard gas)
• Cubicle finish color:
Munsell 5Y7/1 semi-gloss
• Other:
Three standard gas (3.4R) cylinders
accommodable
SCOPE OF DELIVERY
•
•
•
•
•
Gas analyzer system
Specified external drain separator
Specified gas extractor/probe set
Specified gas inlet tube set
Specified standard gas set (with pressure reducing
valve)
• Specified recorder to be accommodated in cubicle
• Standard accessory set
ORDERING INFORMATION
1.
2.
3.
4.
Code symbols
Necessity of spares for 1-year measurement
Type of recorder (option)
Anchor bolt
CODE SYMBOLS
1 2 3 4 5 6 7 8
9 10 11 12 13
4 -
Z SQ1
14 15 16 17
Description
Measurement range (5th code)
0~200 / 500 ppmCO 0~25% O2
0~200 / 1000 ppmCO 0~25% O2
0~500 / 1000 ppmCO 0~25% O2
0~500 / 2000 ppmCO 0~25% O2
A
B
C
D
Power supply (6th code)
3
4
6
7
AC 100V 50Hz
AC 100V 60Hz
AC 115V 60Hz
AC 230V 50Hz
Cubicle structure (7th code)
Indoor type
Outdoor type
Indoor type
Specify when installation place is at –5˚C
Outdoor type
or lower (–10˚C min.)
A
B
C
D
Gas pressure at extraction point (9th code)
– 0.98~+4.90kPa {– 100~+500mmH2O}
– 2.94~+2.94kPa {– 300~+300mmH2O}
– 4.90~+0.98kPa {– 500~+100mmH2O}
1
2
3
External drain separator (10th code)
Not provided
Provided (Note 1)
Y
A
Recorder (11th code)
Not provided
With 6-point recorder PHC
With 2-point recorder PHE
Y
C
E
(Note 2) (Recorder hardware needs to be
prepared separately.)
Isolated output (12th code)
Non-isolated output
Isolated output (5 outputs)
0
5
Jpapn quality assurance organization test (13th code)
No
Yes (Test of individual products needs to be ordered separately.)
0
1
Standard gas (14th code)
Not provided (space for accommodating three 3.4Rcylinders of air + O2/N2 + CO/N2)
Not provided (space for accommodating two 3.4Rcylinders of O2/N2 + CO/N2)
Air + O2/N2 + CO/N2 (3.4R, 3 cylinders)(incorporated) with official organization's test
O2/N2 + CO/N2 (3.4R, 2 cylinders)(incorporated)
Air + O2/N2 + CO/N2 (3.4R, 3 cylinders)(incorporated)
1
2
3
4
5
Gas extractor flange size (15th code)
Without gas extractor
JIS 5K 65A
Y
A
Gas extractor insertion length (16th code)
Without gas extractor
300mm
400mm
600mm
800mm
Y
A
B
C
D
Y
A
B
C
D
E
F
G
H
J
K
L
Sample inlet tube type
Not provided
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
φ10/ φ8 Teflon tube
Heating tube
Heating tube
Heating tube
Heating tube
Heating tube
Length (17th code)
Not provided
5m
10m
15m
20m
25m
30m
10m
15m Specify when the ambient temperature
20m may fall to –5˚C or lower (–10˚C at the
25m lowest level).
30m
(Note 1) Specify this code when the downward inclination of the sample inlet
tube from the gas extraction point to the analyzer gas inlet is less than 15˚,
or when the water contents at the gas extraction point exceed 30 vol%.
(Note 2) Recorder assignment is as follows.
In 6-point recording:
In 2-point recording:
CH1; CO instantaneous value
CH1; CO instantaneous value
CH2; O2 instantaneous value
CH2; O2 instantaneous value
CH3; CO instantaneous value after O2 correction
CH4; 1-hour average CO value after O2 correction
CH5; K thermocouple 1
CH6; K thermocouple 2
3
ZSQ
Gas extractor
Sampling module
Gas inlet tube
Membrane filter
15˚ min.
Solenoid
valve
Gas aspirator
Sample gas inlet,
φ10/φ8 joint
(Rc 3/8)
CO
Flowmeter
Infrared gas
sensor
(Note 1)
Electronic
gas cooler
O2
Solenoid
valve
Gas conditioner
ZERO
Zirconia
O2 sensor
SPAN 1 SPAN 2
Pressure
reducing valve
Gas outlet
Air
Drain
(Note 2)
CO
/N2
O2
/N2
Standard gas (3.4r)
Gas Sampling System Diagram
Note 1: Solenoid valve for using atmospheric air as zero gas
Note 2: Unnecessary when atmospheric air is used as zero gas
Functions of Individual Components
• Gas extractor: Gas extraction, with heating type stainless steel filter having a standard diameter of 40µm
• Gas conditioner:
For separation of drain, prevention of
drain from being sucked through secondary filter and composite operation of constant-pressure bubbler.
• Gas aspirator: For aspiration of sample gas (sample gas
flow rate approx. 2R/min).
• Sampling module:
Incorporates electronic gas cooler, solenoid valve, membrane filter and flowmeter.
• Electronic gas cooler;
Dries the moisture in sample gas to a
dew point of approx. 2˚C.
• Solenoid valve;
Used for introducing calibration gas.
• Membrane filter;
Glass fiber and PTFE filter used to eliminate fine dust particles and permit
monitoring of dust adhering condition
on the front panel of the gas analyzer.
• Flowmeter;
Adjusts and monitors the flow rate of
sample gas.
4
• Standard gas: Reference gas used for calibrating zero
and span of the analyzer; air and CO for
span, and O2 for zero.
• Zirconia O2 sensor:
Zirconia oxygen sensor used for measuring the oxygen concentration (0 to 25%)
in sample gas.
• Solenoid valve for atmospheric air:
Can be built in for using the atmospheric
air instead of standard air.
Standard Accessories
No.
Name
Type/part No.
Quantity
1
Filter paper for membrane filter/as spare
TK7H6750P1
2 sheets
2
Filter paper for membrane filter/as spare
TK700735P3
2 sheets
3 Fuse (2A)/as spare
75718N9
2 fuses
4 Fuse (3.2A)/as spare
75718N10
2 fuses
5 Standard gas joint R1/4–φ6
TK7M6047P1
3 joints
(2 pieces for using only 2 standard gases)
6 Hose band for fixing standard gas
cylinder
TK7M2005P2
12bands
(8 pieces for using only 2 standard gases)
7 Toalon tube for standard gas connection,
0.45 m and φ9/φ5
TK727528P1
1 tube
8 Polyethylene tube for standard gas
connection, 5 m and φ6/φ4
415966P4
1 tube
9 Water bottle for injection into gas
conditioner
TK7H3493P1
1
10 Gas extractor flange packing
TK7H4240P1
1 sheet
1 set
11 Gas extractor fastening bolt and nut
1
12 O-ring (G50) for gas extractor/as spare
13 Heating tube support
TK7N0820C1
1 set
(only when the heating tube has been specified)
14 Instruction manual
INZ-TN4ZSQ
1 copy
Spare Parts for 1-Year Measurement
Type: ZBNISQ12 (without gas extractor)
Type: ZBNISQ42 (with gas extractor)
Contents
• 25-sheet filter paper for membrane filter × 1
• 2-sheet filter paper for membrane filter × 1
• Membrane filter O-ring (P49) × 2
• Membrane filter O-ring (P3) × 2
• Fuse (2 A) × 3
• Fuse (3.2 A) × 4
• Filter for gas conditioner × 2
• O-ring (G65) for gas conditioner × 2
• Diaphragm for gas aspirator × 1
• Valve for gas aspirator × 1
• Fixed restrictor (φ1) × 1
• Fixed restrictor (φ1.4) × 1
• O-ring (G50) for gas extractor × 1
• O-ring (G45) for gas extractor × 1
• Packing for gas extractor mesh filter × 1
when equipped
with gas extractor
• Mesh filter for gas extractor × 1
5
ZSQ
Terminal block 3
<TN3>
Contact output,
for atmospheric a
ir solenoid valve
Contact output,
range discrimination
signal (CO)
1
11
2
12
3
13
4
14
5
15
6
16
Contact output,
upper limit alarm (CO)
7
17
8
18
Contact output,
peak count alarm
9
19
10
20
Terminal block 2
<TN2>
For internal Analog input,
connection O2 sensor
Contact output,
analyzing block error
Contact output,
calibration error
For internal connection
Contact output,
pump ON/OFF
Contact output,
auto calibration status
and maintenance status
(For internal connection)
CONNECTION DIAGRAM
(M3.5 screw)
Analog output,
CO instantaneous value
1
11
2
12
–
3
13
+
4
14
–
5
15
+
6
16
–
7
17
+
8
18
–
9
19
+
10
20
(M3.5 screw)
11
–
+
2
12
+
3
13
–
4
14
+
5
15
–
6
16
+
7
17
8
18
9
19
–
10
20
+
Terminal block
TB1
1
100 V AC
main power supply
(For internal connection)
Analog output,
CO instantaneous value
after O2 correction
Analog output,
O2 instantaneous value
–
+
1
Contact input,
average value resetting
Contact input,
auto calibration remote start
(Output holding during
maintenance)
For internal connection
Contact input,
remote range (CO)
(M3.5 screw)
Terminal block 1
<TN1>
Analog output,
1-hour average CO value
after O2 correction
Analog output,
CO average value after
O2 correction
(averaging time variable)
–
2
3
Ground
4
Terminal block
TB2 (option for isolated output)
1
– Analog output (1-hour
average CO value after
2
+ O2 correction)
3
– Analog output, average CO
value after O2 correction
4
+ (averaging time variable)
Gas extractor
power supply
5
5
6
6
Heating tube
power supply
7
7
8
8
Temperature input 1 +
(option) –
Temperature input 2 +
(option) –
– Analog output, CO
instantaneous value after
+ O2 correction
– Analog output, O2
+ instantaneous value
9
9
10
10
11
(M4 screw)
– Analog output, CO
+ instantaneous value
12
(M4 screw)
OUTLINE DIAGRAM (Unit: mm)
Sample gas inlet
(with joint for
φ10/φ8 tube)
(Rc 3/8)
Notes: 1. Connection to the parenthesized terminals has already been made.
2. Avoid connection to the terminals for internal connection.
765 (for outdoor installation)
615
Anchor plan and door
opening diagram
420
For outdoor
installation
Gas conditioner
1550
1450
610
3 wiring ports
with φ34
grommet
1640 (for outdoor installation)
Rear side
Front side
Door
Anchor bolt
Left side view
6
(135)
420
600
Front view
4-φ15
35
100
610
650
32
(4-M12 × 160 × 50) (option)
Drain, gas outlet
(Rc 1/4)
OUTLINE DIAGRAM (Unit: mm)
<Gas extractor>
Rc1/2 Gas outlet with joint for φ10/φ8
UP
Handle exchange filter
Heating tube support
(added for heating
tube specifications)
PL name plate
357
Cable hole
G1/2
212
293
Cover
(76)
Terminal box
Rc3/4
Name plate
Flange
JIS 5K 65A FF
4-φ15
By order
(750)
f155
Sampling pipe 20A
φ130
φ155
φ27
Gas inlet
7
ZSQ
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan