FUJI EDS3-139B

IN-SITU
DATA SHEET
ZIRCONIA OXYGEN ANALYZER
FOR HAZARDOUS LOCATION
ZFKE, ZKME
This oxygen analyzer is used to continuously measure
oxygen concentration in noncombustible exhaust gas of
industrial boilers or furnaces, and is ideally suited for combustion management and control.
The analyzer system is comprised of the detector and
converter coupled together as a complete system. Detector setting configuration includes the flow guide tube and
detector sensor. The flow guide tube is inserted directly
into the gas and directs gas to the sensor for measurement.
The converter (ZKME) is comprised of the signal processor, input/output and communications, display and system
controls.
The converter is equipped with advanced functionality such as performing the sensor diagnostics and sensor
recovery function, so the detector can be used within long
term stability.
Detector (ZFKE)
FEATURES
1. Gas sampling device is unnecessary
For quick response, insert the detector directly into the
staek. Gas sampling functions such as a gas aspirator
and a dehumidifier are not required.
2. Easy maintenance
The sensor equipped with the detector, has unit construction, it is easy to replace.
By separating the detector and the flow guide tube, filter
replacement is easy.
3. More reliable than sensor diagnosis, sensor recoverable
function
Depending on the components in the measurement
gas, the characteristics of the sensor might deteriorate.
The equipment includes sensor recovery function electronically, checking the deterioration status of the sensor
depletion.
Therefore, it has high reliability and long-lasting stability.
4. Safe and secure
System detects thermocouple break for heater control on
the sensor side. Safety functions of isolating power supply to the detector or isoralting power via external contact
input are also.
5. Easy operation
The operation and setting for the converter can be performed interactively, and available as English, Japanese
or Chinese for language display.
Converter (ZKME)
SPECIFICATIONS
General Specifications
Measuring object: Oxygen in noncombustible gas
Measuring method: Directly insert type zirconia system
Measuring range: 0 to 2 ··· 0 to 50 vol% O2
2 ranges available in 1 vol% O2 steps
Repeatability: Within ±0.5%FS
Linearity: Within ±2%FS
Response time: Within 4 to 7 sec, for 90% (from calibration gas inlet)
approx. 10 min
Warmup time: Analog output: 4 to 20mA DC (allowable load resistance less than 500Ω) or 0 to 1V DC
(output resistance more than 100Ω)
Power supply: Rated voltage;
100 to 120V AC (operating voltage 90
to 132V AC)
200 to 240V AC (operating voltage
190 to 264V AC)
Rated frequency; 50/60Hz
Power consumption:
Maximum 240VA (Detector: approx.
200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx. 50VA,
Converter: approx. 20VA
EDS3-139b
Date
Apr. 1, 2011
ZFKE, ZKME
Detector Specifications (ZFKE)
Measured gas temperature:
Flow guide tube system; −10 to +600ºC
(for general-use, corrosive gas)
Measured gas pressure:
−3 to +3kPa (−306 to +306mmH2O)
Flow guide tube: With or without blow-down nozzle
Flange; JIS5K 65A FF
(JIS5K-80AFF for high particulate gas)
Insertion length; 0.3, 0.5, 0.75, 1m
Other: See. Code Symbols
Ejector (general-use):
Probe for vacumming up measured gas
to detector (option)
Operating temperature:
−10 to +60ºC for Primary detecting element
125ºC or less at detector flange surface
with power applied
Storage temperature:
Sensing element: −20 to +70ºC
Structure: Dust/rain-proof structure(IEC IP66
equivalent)
Flame proof: See Table 1.
SUS316 (filtering accuracy 60µm)
Filter: Main materials of gas-contacting parts:
Detector; Zirconia, SUS316, platinum
Flow guide tube; SUS316
Calibration gas inlet:
ø6mm tube join or ø1/4-inch tube join
(as specified)
Reference gas inlet (option):
ø6mm tube join or ø1/4-inch tube join
(as specified)
Detector mounting:
Horizontal plane ±45º, ambient surrounding air should be clean.
Outer dimensions: (L × max. dia.) 215mm × 164mm (detector)
Mass (approx.) {weight}: Detector; 3.0kg
Flow guide tube (for corrosive gas, 1m);
6kg
Finish color: Case: Silver and SUS metallic color
Cover: Blue
Ejector air inlet flow rate:
5 to 10 L/min
Calibration gas flow:
1.5 to 2 L/min
Blowdown air inlet pressure:
200 to 300kPa {2 to 3 kgf/cm2}
Table 1
Detector
TIIS
(pending)
NEPSI
2
Exd IIB T4
EExd IIC T5 Ex II2G
Converter specification (ZKME)
Concentration value indication:
Digital indication in 4 digits
Contact output signal:
(1) Contact specification; 6 points, 1a 250V AC/3A or 30V
DC/3A
(2) Contact function;
• Under maintenance
• Under blowdown Note3)
• Span calibrating gas
• Zero calibration gas
• Instrument anomalies Note1)
• Alarm Note2)
Note1) The following Instrument errors (1) Thermocouples break (2) Sensor break (3) Temperature fault
(4) Calibration fault (5) Zero/span adjustment fault
(6) Output error turn the contact-ON
Note2) Alarm selects just one as mentioned below (1)
High (2) Low (3) Upper and Lower (4) High-high
(5) Low-low, it turns ON while operating.
Note3) Under blow down is available in case of option,
and it turns ON while operating.
Contact input signal:
(1) Contact specification; 3points (the following option)
ON; 0V (10mA or less), OFF; 5V
(2) Contact function;
• External hold
• Calculation reset
• Heater OFF
• Blow down (option)
• Inhibition of calibration
• Calibration start
• Range change
Calibration method:
(a) Manual calibration with key operation
(b)Auto. calibration (option)
Calibration cycle; 00 day 00 hour to 99 days 23 hours
(c) All calibration
Calibration gas: • Range settings
Zero gas; 0.010 to 25.00% O2
Span gas: 0.010 to 50.00% O2
• Recommended calibration gas concentration
Zero gas; 0.25 to 2.0% O2
Span gas; 20.6 to 21.0% O2
(oxygen concentration in the
air)
Blowdown:
A function for blowing out with compressed air dust that has deposited in
the flow guide tube. Blowdown can
(option)
be performed for a predetermined time
and at predetermined intervals.
Blowdown cycle;00 hour 00 minute to
99 hours 59 minutes
Blowdown time; 0 minute 00 second
to 0 minutes 999
seconds
Output signal hold:
Output signal is held during calibration, processing recoverable
sensor,processing diagnosis of sensor,
warm-up, PID auto tuning, under set
up maintenance mode " available" and
blowdown. The hold function can also
be released.
Valve and flow Selects zero or span gas during manual
meter (option):
zero or span calibration.
Communication function:
RS232C (MODBUS) standard specification
RS485 (MODBUS) (option)
Combustion efficiency display (option): When you select this display, "rich mode
display" will be an simultaneous display.
This function calculates and displays
combustion efficiency from oxygen
concentration and measured gas temperature.
Thermocouple (R) is required for temperature measurement.
Operating temperature:
−20 to +55ºC
Operating humidity:
95% RH or less, non condensing
Storage temperature:
−30 to +70ºC
Storage humidity: 95% RH or less, non condensing
Construction: Dust-proof, rainproof construction
(corresponding to IP65)
Explosion proof: See Table 2
Aluminum case
Material: Outer dimensions (H x W x D):
470 X 326 X 211mm (IP65)
Mass {weight}: IP65:Approx.22kg (excluding cable and
detector)
Finish color: Case: Silver
Cover: blue
Mounting method:Mounted flush on panel
Electrical Safety:
Overvoltage category
; II power supply input
; I relay interfaces
(IEC1010-1)
External overcurrent protective device
; 10A
Equipment interfaces are safety
separated (SELV)
ZFKE, ZKME
Table 2
Converter
TIIS
(pending)
NEPSI
Exd IIB T6
EExd IIC T6 Ex II2G
3
ZFKE, ZKME
CODE SYMBOLS
(Detector)
1 2 3 4 5 6 7 8
Z F K ER
(Converter)
9 10 11 12 13
5 -
Y
14 15 16 17
-
1 2 3 4 5 6 7 8
Description
1
Cal. gas inlet
Non (G3/8 female screw)
For φ6mm tube
For φ1/4 inch tube
Ejector for φ6mm tube
Ejector for φ1/4 inch tube
Y
1
2
A
B
Power supply
100 to 120VAC 50/60Hz
200 to 240VAC 50/60Hz
1
3
Z K ME
9 10 11 12 13
1 1 -
1
B
E
1
2
Y
1
2
3
4
14
Description
-
Output signal
4 to 20mA DC
0 to 1V DC
Communication function
RS-232C
RS-485
Optional Functions
None
Combustion efficiency display function Note1)
Blowdown
Auto calibration
Combustion efficiency indication
+ Blowdown Note1)
Combustion efficiency indication
+ Auto calibration Note1)
Blowdown + Auto calibration
Combustion efficiency indication + Blowdown
+ Auto calibration Note1)
Instruction manual language
Japanese
English
Chinese
Mounting Option
None (Mounting on panel surface)
With valve
With valve + flowmeter
Specification name plate
Standard
Number of Cable Gland
3
4
5
6
7
Ex Standard
NEPSI
TIIS (pending)
Flow guide tube
None
<Flange size>
JIS 5K 65A
7
5
JIS 5K 80A
8
6
JIS 5K100A
9
7
JIS 10K 65A
A
JIS 10K 80A
B
JIS 10K 100A
C
J
ANSI 150LB 2B
E
D
C
ANSI 150LB 3B
E
ANSI 150LB 4B
F
Y
DIN DN50 PN10
G
1
DIN DN80 PN10
H
2
<Application / meterial>
For corrosive gas / SUS316
F
1
With blow-down nozzle / SUS316
G
For
high
particular
/
SUS316
H
3
For high particular with cover / SUS316
J
4
For high particular / SUS310S
K
5
For high particular with cover / SUS310S
L
6
For high particular / titanium
M
7
For high particular with cover / titanium
N
<Length>
N
3
300mm
T
5
500mm
7
Note1) When you select this display, rich mode
750mm
will be a simultaneous display.
1
1000mm
Reference gas inlet
Y
Non (G1/8 female screw)
A
for φ6mm tube
(Exclusive-special cable)
B
For φ1/4 inch tube
1 2 3 4 5 6 7 8
9
Filter
Description
Z
R
Z
E
R
1
Standard
5
For high particular
7
Connectable devices
Instruction manual language
E
For ZKME
J
Japanese
Types
E
English
R
For R thermocouple
C
Chinese
Cable length
Specification name plate
YA
6m
1
Standard
YB
10m
Ex. Standard
YC
15m
N
NEPSI
YD
20m
T
TIIS (pending)
YE
30m
YF
40m
YG
50m
YH
60m
YJ
70m
(Replacement Detector element)
YK
80m
YL
90m
Power supply
Code symbols
YM
100m
0Y 0
4
100 to 120V AC
ZFK8YY15-0Y0YY-0YY
200 to 240V AC
ZFK8YY35-0Y0YY-0YY
0
1
2
Cable end treatment
None
One side (detector side)
Both sides
SCOPE OF DELIVERY
CAUTIONS
Detector:
Detector main unit × 1, Viton Packing
× 1, thermo seal × 1, mounting screw
(M5mm × 25) × 6, flow guide tube (as
specified) × 1, Wrench × 1, Instruction
manual × 1
Converter:
Converter main unit × 1, mounting
screw (M12 × 50) × 4, Cock (option) ×
1, flowmeter (option) × 1,
Accessories (AC250V 500mA T fuse ×
2, AC250V 2.5A T fuse × 2), Wrench ×
1, Instruction manual × 1
Ejector: With detector main unit (option)
Items to be prepared separately:
(1)Standard gas for calibration
Type ZBM NSH4-01 (up to 5% O2 range)
Type ZBM NSJ4-01 (over 5% O2 range)
(2)Reduction valve for standard gas (type ZBD61003)
(3)Flowmeter
Type; ZBD42203, 0.2 to 2L/min (for calibrating gas)
(unnecessary when the code 11th of ZKME is 2)
Type; ZBD42403, 1 to 10L/min (for ejector)
(4)Opner
Type; ZZP*TK7N9329P2 (for detector; ZFKE)
Type; ZZP*TK7N9329P1 (for converter; ZKME)
• If combustible gas (CO, H2 etc.) exists in the measured
gas, error will occur due to burning at the sensor section.
The inclusion of corrosive gas (Si vapor, alkaline metal, P,
Pb etc.) will shorten the life of the sensor.
• When the measured gas temperature is high (+300°C
or higher), the flange should be separated from the
furnace wall in order to bring the detector flange surface
temperature below the specified value +125°C). The
flow guide should be attached in the direction in which
the gas flow to the detector decreases.
• When much dust is included in the gas, the flow guide
tube should be attached at an inclination so that the flow
goes from below to above. And the flow guide should
be attached in the direction in which the gas flow to the
detector decreases.
• In the case of a refuse incinerator, automatic blow down
of the flow guide should not be performed (to prevent
corrosion of the flow guide tube due to drainage). Blowdown should be performed manually when change in
the indication has become very little with the furnace
stopped.
DEVICE CONFIGURATION
The device to be combined differ according to the conditions of the gas to be measured. Select the devices to be combined with reference to the following table.
Measured gas
Application
Temperature Gas Flow
General-use
600ºC or
(boiler)
less
For corrosive
600ºC or
gas (refuse
less
Device configuration
Note
Detector type
Fuel; gas, oil
ZFKER
5- F Y -
Fuel: coal
ZFKER
5- G Y -
with blow down
Included low moisture
ZFKER
5- F Y -
Less than 10g/Nm3
Included low moisture
ZFKER
5- G Y -
Less than 25g/Nm3
with blow down
Included low moisture
DUST
5 to 20m/s Less than 0.2g/Nm3
Less than 10g/Nm3
5 to 20m/s Less than 1g/Nm3
incinerator)
with blow down
Less than 25g/Nm3
Included high moisture
with blow down
ZFKER
ZFKER
H
5- K Y M
J
5- L Y N
Converter
type
ZKME
ZKME
ZKME
ZKME
ZKME
ZKME
Note (1) Dust volume is approximate value.
(2) Instrument quality air or bottled air is available as reference air by selecting detector with reference air inlet.
5
ZFKE, ZKME
CONFIGURATION
Flow guide tube system
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Reduction valve
(unnecessary when supply
air pressure is 200 to 300kPa
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
(not supplied)
Flow guide tube
15ASGP
tube equivalent
(not supplied)
Detector
(ZFKE)
from (1)
Gas temperature.
600ºC max.
exhaust
6
Sample gas outlet
1
5
2
3
AC power
supply
4
Black
Black
Red
Blue
White
Yellow
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Converter (ZKME)
Sample gas outlet
1
2
+ O2 −
O2 sensor
input
*4
Gas temperature.
600ºC max.
Flow guide for high
particulate with cover
3
4
5
+ TC1 −
+
AO
6
7
8
9
4 to 20mA DC
or
0 to 1V DC
12
13
DI1
DI2
DI3 DICOM
15
17
18
19
20
Calibrating gas
contact output
Maintenance
contact
output
ALARM
Contact input
Calibrating gas
contact output
SPAN
GAS
MAINTE
COM
21
22
23
TM-1
14
O2 sensor
heater power input
ZERO
GAS
Blow
contact
output
Alarm
contact
output
+ TC2 −
16
11
ZV
O2 sensor or
Fault contact
thermocouple
output
input
Analog output
Thermocouple
for combustion
control
10
−
24
25
SV
26
AC power
supply
PE
*3
N AC L
27
28
1
2
TM-2
3
RS232C
or
RS485
(option)
AC power
supply
φ 6mm or φ 1/4 inch tube
for calibration gas
(not supplied)
Joint
*2 Ref. gas
Solenoid valve
(not supplied)
AC power
supply
Reduction
valve (ZBD6)
Standrad gas
for zero point
(ZBM)
Flowmeter
(ZBD4)
(1.5 to 2.0L/min)
Reduction valve
(unnecessary when supply air pressure is 50 to 100kPa)
(1) to Detector
*1 Air supply
φ 6mm or φ 1/4 inch tube
for calibration gas
(1.5 to 2.0L/min)
(not supplied)
Solenoid valve
(not supplied)
Reduction vale
(ZBD6)
Span point
*1
Standard gas for
span point (ZBM)
Note: *1
*2
*3
*4
6
Standard gas or instrumentation air can be used in place of span gas.
Instrument quality air or bottled air is available as reference gas instead of ambient air.
Protective earth.
Connect the shield of a exclusive cable with the ground terminal in the converter.
Flow guide tube system (with valve)
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Reduction valve
(unnecessary when supply
air pressure is 200 to 300kPa
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
(not supplied)
Flow guide tube
15ASGP
tube equivalent
(not supplied)
Detector
(ZFKE)
from (1)
Gas temperature.
600ºC max.
exhaust
6
Sample gas outlet
1
5
2
3
AC power
supply
4
Black
Black
Red
Blue
White
Yellow
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Converter (ZKME)
Sample gas outlet
1
2
+ O2 −
O2 sensor
input
*4
Gas temperature.
600ºC max.
Flow guide for high
particulate with cover
3
4
+ TC1 −
5
+
AO
6
7
8
9
4 to 20mA DC
or
0 to 1V DC
12
13
DI1
DI2
DI3 DICOM
15
17
18
19
20
Calibrating gas
contact output
Maintenance
contact
output
ALARM
Contact input
Calibrating gas
contact output
SPAN
GAS
MAINTE
COM
21
22
23
TM-1
14
O2 sensor
heater power input
ZERO
GAS
Blow
contact
output
Alarm
contact
output
+ TC2 −
16
11
ZV
O2 sensor or
Fault contact
thermocouple
output
input
Analog output
Thermocouple
for combustion
control
10
−
24
25
SV
26
AC power
supply
PE
*3
N AC L
27
28
1
2
TM-2
3
RS232C
or
RS485
(option)
AC power
supply
φ 6mm or φ 1/4 inch tube
for calibration gas
(not supplied)
Flowmeter
(option)
Valve
(option)
MEASURE
Reduction
valve (ZBD6)
ZERO
SPAN
*2 Ref. gas
Flowmeter
(ZBD4)
(1.5 to 2.0L/min)
Standrad gas
for zero point
(ZBM)
(1) to Detector
*1 Air supply
φ 6mm or φ 1/4 inch tube
for calibration gas
(1.5 to 2.0L/min)
(not supplied)
Note: *1
*2
*3
*4
Standard gas or instrumentation air can be used in place of span gas.
Instrument quality air or bottled air is available as reference gas instead of ambient air.
Protective earth.
Connect the shield of a exclusive cable with the ground terminal in the converter.
Reduction vale
(ZBD6)
Span point
*1
Standard gas for
span point (ZBM)
7
ZFKE, ZKME
Flow guide tube system (with ejector)
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Reduction valve
(unnecessary when supply
air pressure is 200 to 300kPa
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
(not supplied)
Flow guide tube
15ASGP
tube equivalent
(not supplied)
Detector
(ZFKE)
from (1)
Gas temperature.
600ºC max.
exhaust
Calibrating gas
Flowmeter
(ZBD4)
(5 to 10L/min)
exhaust
6
Sample gas outlet
1
5
2
3
from (2)
AC power
supply
4
Black
Black
Red
Blue
White
Yellow
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Converter (ZKME)
Sample gas outlet
1
2
+ O2 −
O2 sensor
input
*4
Gas temperature.
600ºC max.
Flow guide for high
particulate with cover
3
4
5
+
+ TC1 −
AO
6
7
8
9
4 to 20mA DC
or
0 to 1V DC
Blow
contact
output
Alarm
contact
output
ALARM
Contact input
+ TC2 −
DI1
DI2
DI3 DICOM
15
17
18
19
16
11
12
13
ZV
O2 sensor or
Fault contact
thermocouple
output
input
Analog output
Thermocouple
for combustion
control
10
−
20
Maintenance
contact
output
O2 sensor
heater power input
ZERO
GAS
Calibrating gas
contact output
Calibrating gas
contact output
SPAN
GAS
MAINTE
COM
21
22
23
TM-1
14
24
25
SV
26
AC power
supply
PE
N AC L
27
28
1
2
TM-2
3
RS232C
or
RS485
(option)
Joint
φ 6mm or φ 1/4 inch tube
for calibration gas
(not supplied)
*2 Ref. gas
Flowmeter
(ZBD4)
(1.5 to 2.0L/min)
Solenoid valve
(not supplied)
AC power
supply
AC power
supply
Standrad gas
for zero point
(ZBM)
Reduction valve
(unnecessary when supply air pressure is 50 to 100kPa)
(2) to ejector
(1) to Detector
φ 6mm or φ 1/4 inch tube
for calibration gas
(0.2 to 0.5L/min)
(not supplied)
Reduction
valve (ZBD6)
*1 Air supply
Joint
Solenoid valve
(not supplied)
Reduction vale
(ZBD6)
Span point
*1
Standard gas for
span point (ZBM)
Note: *1
*2
*3
*4
8
Standard gas or instrumentation air can be used in place of span gas.
Instrument quality air or bottled air is available as reference gas instead of ambient air.
Protective earth.
Connect the shield of a exclusive cable with the ground terminal in the converter.
*3
Flow guide tube system (with ejector+valve)
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Reduction valve
(unnecessary when supply
air pressure is 200 to 300kPa
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
(not supplied)
Flow guide tube
15ASGP
tube equivalent
(not supplied)
Detector
(ZFKE)
from (1)
Gas temperature.
600ºC max.
exhaust
Calibrating gas
Flowmeter
(ZBD4)
(5 to 10L/min)
exhaust
6
Sample gas outlet
1
5
2
3
from (2)
AC power
supply
4
Black
Black
Red
Blue
White
Yellow
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Converter (ZKME)
Sample gas outlet
1
2
+ O2 −
O2 sensor
input
*4
Gas temperature.
600ºC max.
Flow guide for high
particulate with cover
3
4
+ TC1 −
5
+
6
AO
7
8
4 to 20mA DC
or
0 to 1V DC
11
12
13
DI1
DI2
DI3 DICOM
15
17
18
19
20
Calibrating gas
contact output
Maintenance
contact
output
ALARM
Contact input
Calibrating gas
contact output
SPAN
GAS
MAINTE
COM
21
22
23
TM-1
14
O2 sensor
heater power input
ZERO
GAS
Blow
contact
output
Alarm
contact
output
+ TC2 −
16
10
ZV
O2 sensor or
Fault contact
thermocouple
output
input
Analog output
Thermocouple
for combustion
control
9
−
24
25
SV
26
AC power
supply
PE
*3
N AC L
27
28
1
2
TM-2
3
RS232C
or
RS485
(option)
Flowmeter
(option)
φ 6mm or φ 1/4 inch tube
for calibration gas
(not supplied)
*2 Ref. gas
Valve
(option)
MEASURE
AC power
supply
Reduction
valve (ZBD6)
ZERO
Flowmeter
(ZBD4)
(1.5 to 2.0L/min)
SPAN
Standrad gas
for zero point
(ZBM)
(1) to Detector
*1 Air supply
φ 6mm or φ 1/4 inch tube
for calibration gas
(0.2 to 0.5L/min)
(not supplied)
Reduction vale
(ZBD6)
Span point
*1
Standard gas for
span point (ZBM)
Note: *1
*2
*3
*4
Standard gas or instrumentation air can be used in place of span gas.
Instrument quality air or bottled air is available as reference gas instead of ambient air.
Protective earth.
Connect the shield of a exclusive cable with the ground terminal in the converter.
9
ZFKE, ZKME
OUTLINE DIAGRAM
(Unit:mm)
Y
Detector (ZFKER 1 )
2
Ref. gas inlet (to order)
SUS316, for φ 6mm tube
(the 13th digit is “A”)
or 1/4 inch tube
Approx. 152
(the 13th digit is “B”)
Approx.
140
Approx. 215
77
17.5
Approx. φ 100
Lock screw M6
View A
External ground
terminal M4
Explotion-proof
name plate
Exclusive Cable gland
(For power supply)
Approx.
32
Filter
Approx. 162
View B
Approx. 126
P
Calibration gas inlet (To order)
SUS316, for φ 6mm tube
(the 6th digit is “1”)
Or 1/4 inch tube
(the 6th digit is “2”)
(Not provided if the 6th digit is “Y”)
Exclusive Cable gland
(For signal)
Approx. 242
Approx. 180
7
φ6
6
φ
6-
(13th digt code A, B)
(14th digt code 7)
VIEW A
VIEW B
EXTERNAL
CONNECTION DIAGRAM
Heater
Thermocouple
Element output
1
3
2
4
5
Black Black Red White Yellow
2-core wire
4-core wire
Exclusive-special cable
VIEW FROM P
Sensor unit (ZFK8YY)
(132.7)
3.5
20
(φ 17.5)
(88.4)
4-φ 3.5(Mount for detector)
S-
H
2
T-
4
45°
TK4J4664R0
3
made in japan
40
5
10
T+
S+ H
6
1
45°
φ 35
6
Blue
Detector (ZFKER
A
)
B
Ref. gas inlet (to order)
SUS316, for φ 6mm tube (the 13th digit is “A”)
Or 1/4inch tube (the 13th digit is “B”)
Approx. 215
Approx.
155
17.5
77
Approx. 140
Approx. 100
Lock screw M6
P
View B
Approx.
126
External ground
terminal M4
Approx. 47
Approx. 57
Approx. 185
Filter
View A
Exclusive Cable gland
(For power supply)
Calibration
Air out
Calibration gas inlet
SUS316, for φ 6mm tube
(the 6th digit is “A.”)
1/4 inch tube (the 6th digit is “B.”)
Air in
Approx. 37
Approx. 21
Approx. 78
Exclusive Cable gland
(For signal)
Ejector gas inlet
SUS316, for φ 6mm tube
(the 6th digit is “A.”)
1/4 inch tube (the 6th digit is “B.”)
Ejector gas outlet
SUS316, for φ 6mm tube
(the 6th digit is “A.”)
1/4 inch tube (the 6th digit is “B.”)
Approx. 242
Approx. 180
(13th digt code A, B)
7
φ6
(14th digt code 7)
VIEW A
6
VIEW B
φ
6-
Explotion-proof
name plate
EXTERNAL
CONNECTION DIAGRAM
Heater
Thermocouple
Element output
1
3
2
4
5
Black Black Red White Yellow
2-core wire
6
Blue
4-core wire
Exclusive-special cable
VIEW FROM P
Flow guide tube (with blow-down nozzle) (ZFKE: 10th digit code. G)
Approx. L
D
C
67
Approx. 20
f
60.5
t
Gas inlet
6-M5
detector side
Gas outlet
ZFKE MTG. position
Code 11th
3
L (m)
0.3
Mass
3.0
Approx.(kg)
5
7
1
0.5 0.75 1.0
3.8
4.8
5.7
Z
L=
(to order)
Blow down
N– h
air inlet
MTG. holes
4-Rc1/4
with plug int.thd.
Flange size
JIS 5K 65A
JIS 5K 80A
JIS 5K 100A
JIS 10K 65A
JIS 10K 80A
JIS 10K 100A
ANSI 150LB 2B
ANSI 150LB 3B
ANSI 150LB 4B
DIN DN50 PN10
DIN DN80 PN10
Code
9th
7
8
9
A
B
C
D
E
F
G
H
D
C
t
f
155
180
200
175
185
210
150
190
230
165
200
130
145
165
140
150
175
120.7
152.4
190.5
125
160
14
14
16
18
18
18
17.5
22.3
22.3
18
20
2
2
2
2
2
2
2
2
2
0
0
110
121
141
116
126
151
92.1
127
157.2
0
0
N
h
4
4
8
4
8
8
4
4
8
4
4
15
19
19
19
19
19
19.1
19.1
19.1
18
18
11
ZFKE, ZKME
Blow down air inlet
Select an upper port to
avoid the condensation
in the piping
C
4 - Rc 1/4
with plug int thd.
D
Flow guide tube (for high particulate) (ZFKE: 10th dight code. H, K, M)
N- h
MTG. hole
67
38.
5
ZFKE MTG.
position
t
Packing
188
Code 11th
3
0.3
50
Mass
4.5
Approx.(kg)
5
1
7
0.5 0.75 1.0
5.6
8.3
7.0
Z
L=
(to order)
L
Gas
outlet
L (m)
30
Gas flow
Flange size
JIS 5K 65A
JIS 5K 80A
JIS 5K 100A
JIS 10K 65A
JIS 10K 80A
JIS 10K 100A
ANSI 150LB 2B
ANSI 150LB 3B
ANSI 150LB 4B
DIN DN50 PN10
DIN DN80 PN10
Code
9th
7
8
9
A
B
C
D
E
F
G
H
D
C
t
f
155
180
200
175
185
210
150
190
230
165
200
130
145
165
140
150
175
120.7
152.4
190.5
125
160
14
14
16
18
18
18
17.5
22.3
22.3
18
20
2
2
2
2
2
2
2
2
2
0
0
110
121
141
116
126
151
92.1
127
157.2
0
0
N
h
4
4
8
4
8
8
4
4
8
4
4
15
19
19
19
19
19
19.1
19.1
19.1
18
18
Blow down air inlet
Select an upper port to
avoid the condensation
in the piping
C
4 - Rc 1/4
with plug int thd.
D
Flow guide tube (for high particulate with cover) (ZFKE: 10th dight code. J, L, N)
N- h
MTG. hole
67
38.
5
ZFKE MTG. position
(178)
t
L
(L-10)
Gas
outlet
50
Packing
Gas flow
Production
tube
12
(38)
30
Code 11th
3
L (m)
0.3
Mass
7.1
Approx.(kg)
5
7
1
0.5 0.75 1.0
Z
L=
9.0 11.4 13.6 (to order)
Flow guide tube (ZFKE: 10th dight code. F)
f
D
C
67
Approx.
40
Approx. L
Gas inlet
60.5
t
6-M5
detector side
Gas outlet
N- h
MTG. holes
ZFKE MTG. position
3
Code 11th
5
7
0.3
0.5 0.75
MASS
3.3
Approx.(kg)
4.5 6.1
7.6
L (m)
Z
1
1.0
Flange size
L=
(to order)
JIS 5K 65A
JIS 5K 80A
JIS 5K 100A
JIS 10K 65A
JIS 10K 80A
JIS 10K 100A
ANSI 150LB 2B
ANSI 150LB 3B
ANSI 150LB 4B
DIN DN50 PN10
DIN DN80 PN10
Code
9th
7
8
9
A
B
C
D
E
F
G
H
D
C
t
f
155
180
200
175
185
210
150
190
230
165
200
130
145
165
140
150
175
120.7
152.4
190.5
125
160
14
14
16
18
18
18
17.5
22.3
22.3
18
20
2
2
2
2
2
2
2
2
2
0
0
110
121
141
116
126
151
92.1
127
157.2
0
0
N
h
4
4
8
4
8
8
4
4
8
4
4
15
19
19
19
19
19
19.1
19.1
19.1
18
18
Coverter (ZKME)
Mounting hole 4-ø14
Lock screw M6
326
211
286
17
ZKM
Oxygen Analyzer
Oxygen Ana
MODE
ESC
ENTER
POWER
注 意
POWER
変換器へ供給する電源電圧は、接続される
検出器の電源電圧に合わせてください。
362
HEATER
FUSE T 2.5A
CAUTION
Power supply must be selected
the same voltage of detector.
442
402
FUSE T 0.5A
ON
POWER
OFF
+O 2 - +TC1 - +AO - FAULT
1
2
15
3
16
4
17
5
18
6
19
7
20
+TC2 - DI1 DI2 DI3 DI-
COM
ZV HEATER
BLOW
9
10 11 12 13 14
21 22 23 24 25 26 27 28
MAINALARM
AC
COM SV
N
L
TENANCE
8
CAUTIO
WHEN
POWER OPEN THIS
EXPLOSIVEOFF AND COVER,
KEEP
ATMOSPHERE
OUT
N
Fuji Electr
ic
Fuji Electric Co.,Ltd.
ESC
ENT
68
MODE
Fuji Electric Co.,Ltd.
Co.,Ltd.
Cable gland for flame proof ø9 - ø12 /
closeup plug (To order)
Ground terminal M4
For communication
For contact In/Out
For Other
60
60
60
40
60
52
186
60
For sensor signal
For sensor power
For output
For power
supply
13
ZFKE, ZKME
EXTERNAL TERMINAL (TM1) /M3
Exclusive
cable
O2 sensor
input
O2 sensor
thermocouple
input
Analog
output
4-20mA DC
or 0-1V DC
1
2
3
4
+
O2 −
+
TC1 −
5
+
AO
6
Fault
contact output
250V AC/3A
30V DC/3A
7
8
9
+
15
FAULT
TC2 −
DI1
DI2
DI3 DICOM
16
17
18
19
Thermocouple
input type R
(not supplied)
11
12
BLOW
ZERO
GAS
SPAN
GAS
22
Alarm
contact output
250V AC/3A
30V DC/3A
23
COM
24
25
Maintenance
contact output
250V AC/3A
30V DC/3A
Terminal number
2
3
Remarks
RS232C
TXD
RXD
GND
Standard
RS485
TRX+
TRX–
GND
Option
SV
26
Calibrating gas
contact output
250V AC/3A
30V DC/3A
COMMUNICATION TERMINAL (TM2) /INSERTION TERMINAL
14
14
HEATER
MAINTE
21
20
Contact input
Short circuit : on (Current 5mA)
Open
: off (Voltage 5V)
1
13
ZV
ALARM
Ground
terminal
M4
COM
10
−
or
O2 sensor
heater power
output
Blow contact
output
250V AC/3A
30V DC/3A
N
27
AC
L
28
AC
Power supply
input
100 to120V AC
200 to 240V AC
OUTLINE DIAGRAM
(Unit:mm)
<Option>
SELECTOR VALVES + FLOWMETER (IN CASE OF 11TH DIGIT CODE “2” )
114
102
MEA RE
ZERO CALIB. SPAN
6
2-φ 5.5
Mounting hole
Zero gas inlet
(NPT1/8)
20
ZERO
SPAN
293
126
20 20
15
30
83
Span gas inlet (NPT1/8)
65
Calibration gas outlet (NPT1/8)
SELECTOR VALVES (IN CASE OF 11TH DIGIT CODE “1” )
6
2-φ 5.5
Mounting hole
ZERO CALIB. SPAN
15
30
Zero gas inlet
(NPT1/8)
20 20
65
83
Span gas inlet (NPT1/8)
Calibration gas outlet (NPT1/8)
65
114
102
MEA RE
20
ZERO
SPAN
101
15
ZFKE, ZKME
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan