FUJI EDS3-135E

cross stacK laser Gas analYZer
DATA SHEET
ZSS
Cross stack laser gas analyzer (ZSS) provides continuous
measurement of density of HCl contained in flue gas in an
incineration system, NH 3 in denitration equipment, combustion gas, CO, CO 2 and O 2 with quick response speed.
Due to in-situ method, Measured gas does not require the
preparation by sampling system. This analyzer can be
used under high dust conditions, so it is possible to install
upstream of the bagfilter where gas sampling is normally
difficult.
Control unit
Features
1. Maintenance is minimized thanks to no need of gas sampling devices.
2. Running cost is negligible due to having few expendable
parts.
3. Zero drift is within ±2.0%FS, and an excellent stability is
ensured for a long period of time.
4. Almost no interference by other gas component because an infrared semiconductor laser, which has exact
absorption wavelength as the measured component, is
used.
5. High speed response is available since the gas sampling
device is unnecessary.
6. High vibration resistance because the optical adjustment mechanism is separated from the airtight compartment, and it is easy to make a fine adjustment and large
adjustment angle.
specIFIcatIons
Measurement principle:
Non-dispersive infrared absorbance system (NDIR)
Measuring method:
Cross-stack system (path system)
Use application: Incineration facilities, denitration
equipment, etc.
Light source:
Near-infrared laser
Laser class:
Class 1 (excluding O2 meter for use in
high dust)
Outline (D × W × H) mm:
Receiver unit (180 × 400 × 200 mm)
Transmitter unit (240 × 400 (432) ×
200 mm)
Control unit (135 × 240 × 320 mm)
Mass:
Receiver/Transmitter unit: Approx.
10kg
Control unit: Approx. 8kg
Structure:
Outdoor use type, dustproofing/rainproofing structure (IP65)
Box finish color: Receiver/Transmitter box: gray
Control unit cover: blue
Control unit case: silver
Receiver unit
Transmitter unit
Measurable component and range:
HCl
Min. measuring
range
10 ppm
Max. measuring
range
5000 ppm
HCl+H2O (*1)
50 ppm (HCl)
1000 ppm (HCl)
NH3
15 ppm
5000 ppm
NH3+H2O (*1)
50 ppm (NH3)
1000 ppm (NH3)
O2 (Class 1 Laser)
4 vol%
100 vol%
O2 (For use in high
dust)
4 vol%
50 vol%
CO
2.0 vol%
50 vol%
CO (For use in high
temperature)
10 vol%
50 vol%
Component
CO2
2.0 vol%
50 vol%
CO2 (For use in high
temperature)
10 vol%
50 vol%
CO+CO2
2.5 vol%
50 vol%
CO+CO2 (For use in
high temperature)
10 vol%
50 vol%
CH4
100 ppm
50 vol%
(*1) Range for H2O is fixed at 50vol%·m.
(Range in the case stack (optical path length) diameter is 1m)
Max. and min. ranges for each stack diameter
= (Max. and min. ranges when stack diameter is 1m)
÷ Stack diameter
* Stack diameter = Optical path length
(Example: In case stack diameter is 2m for HCl measurement,
Max.range: 2500ppm and min. range: 5ppm).
Material:
Receiver unit / Transmitter unit:
Aluminum, SUS316
Control unit: Aluminum
Materials of gas-contacting parts:
SUS316, BK7, FKM, PTFE, glass-cloth,
silicone
EDS3-135e
Date
Apr. 1, 2011
ZSS
Air purge connection diameter:
RC1/4 (tube φ8 × 6)
Power supply: Rated voltage 100 to 240V AC ±10%
Rated frequency 50/60 Hz
Power consumption:
Max. rated power: Approx. 75VA or
less
Calibration interval:
Once every six months (Maintenance
cycle may vary depending on the operating environment.)
Indicator (control unit): LCD with back light
Communication functions:
USB (for loader), RS-485 (Modbus® protocol)
Cable length:
Receiver unit to Transmitter unit:
Standard 2m (Maximum 25m)
Receiver unit to control unit:
Standard 5m (Maximum 100m)
Analog output: 4 to 20mADC or 0 to 1VDC x 2 (4)
Isolated output indicated in the parenthesis is optional.
(1 to 5V DC, 0 to 5V or 0 to 10V is available.)
Allowable load: 4 to 20mA DC 550Ω or
less, 0 to 1V DC 100kW or more
(Output measurement value and O2 corresponding value. Average value and
instantaneous value are switchable by
the settings.)
Analog input:
4 to 20mA DC × 6
Measured gas pressure, measured gas
temperature, measured gas velocity, O2
gas concentration, water concentration,
air purge pressure
(Measurement concentration correction,
O2 correction or alarm output is performed
according to the input signal.)
Contact output: Relay contact output (contact capacity
24V DC 1A 1a or 1b × 5 (Standard: 1a,
1b: for power supply off only)
Low light transmission, outside the range
of upper/lower limits, device failure, during hold/during calibration, power supply
off.
Contact input (option):
Photo coupler receiver contact input
(operating voltage 12 to 24V DC / 5 to
20mA) × 3
Average value reset signal, switching instantaneous value/moving average value
and remote hold
Alarm output (screen-displayed):
LD failure, LD temperature error, high
gas temperature, air purge (low pressure), box temperature warning, low light
transmission, PD over range, connection
error, AI under, Range (H-Limit) or Range
(L-Limit)
Display contents:
Component, concentration (instantaneous value, average value O2 correction
instantaneous value and O2 correction
average value), alarm (fault status)
2
contact output contents (1a contact)
Low light transmission:
Contact output is performed (close) when
the amount of light transmission is insufficient.
Outside the range of upper/lower limits:
According to the preset upper or lower
limit alarm value, contact output is performed (close) when it becomes lower
than alarm upper/lower limit.
Device failure: Contact output is performed (close) when
laser failure, laser temperature control
failure, exceeding the amount of light
transmission or communication error are
occurred.
During hold/during calibration:
While AO output is held by the hold setting, the values, output (close) of which is
held, is the ones just before the holding
or arbitrary set values.
Output (close) during calibration
Power supply off:
Output (close) while power supply is
off
Contact Input Contents (option)
Average value reset signal:
Output of converted average value is
started from the initial state by applying
rectangular-wave voltage (pulse width
2 sec or more) to the input terminal of
average value resetting. Output is reset
by inputting and restarted by opening.
Switching instantaneous value/moving average value:
Switching to and from the instantaneous
value and the average value of the analog
output is performed by applying rectangular-wave voltage (pulse width 2 sec or
more) to the input terminal for switching
between the instantaneous value and the
moving average values.
The analog output is held by applying
Remote hold:
rectangular-wave voltage (pulse width
2 sec or more) to the remote hold input
terminal, and restarted by applying it
again.
Functions
O2 correction:
Conversion of measured HCl gas concentrations into values at standard O 2
concentration
Correction formula:
21 – On
C=
× Cs
21 – Os
C: Converted concentration
Cs: Measured concentration of sample
gas
Os:Measured O2 concentration (Upper
limit settable 1 to 20% O2)
On:Standard O 2 concentration (value
changeable by setting; 0 to 19%
O 2)
The result of calculation is indicated and
output in an analog output signal.
Ambient temperature:
–20 to 55°C (Receiver unit/Transmitter
unit), –5 to 45°C (control unit)
Ambient humidity: 90% R.H. or less
Measurable optical path length (stack diameter):
0.5 to 10m
Standard flange:JIS10K 50A flange (JIS B 2212)
Air purge:
Instrument air (compressor must be
installed when power supply cannot
be provided.) N 2 for an O 2 meter less
than 20 vol% Pressure 0.5 to 0.7MPa
or more
Air purge flow rate:
20L/min or more (depending on measured
gas temperature, velocity, pressure, moisture and dust)
One-side air purge flow rate: (L/min) = Gas flow rate (m/s)×10
Measured gas condition:
Temperature:1200°C or less (O2, CO, CO2,
CO+CO2)(Note)
130°C or more, and 450°C
or less (NH3+H2O)
130°C or more, and 450°C
or less (HCl+H2O)
300°C or more (CH4)
Pressure:±10kPa
Moisture:50vol% or less (Should not be
saturated water vapor.)
Velocity: 25m/s or less (Consult us for
use in high dust or moisture
environment.)
Note) When measuring high-temperature gas at 500°C or
higher, install a pressure sensor on the air-purge unit to
monitor air purge. If measurement is taken while purge
is not performed, the device may be damaged.
Dust: 5 to 30g/Nm3 (depending on environmental
conditions such as measuring component,
equipment specification, light path length
and particle diameter).
If dust exceeds 30g/Nm3-m, consult with
Fuji.
Vibration: 0.5G or less (0.2G or less when the frequency range is 20 to 40Hz) (when optical
path length is 1m)
Mounting angle: Horizontally ±5 degrees or less (No dew
condensation should accumulate on the
window.)
Performance (Except for H2O)
Repeatability:
±1.0%FS (depending on measuring
component and measuring range)
Linearity: ±1.0%FS (depending on measuring
component and measuring range)
Zero drift: ±2.0% FS
(NH3 ±3.0%, FS when range is 20ppm
or less)
Interference from other gas components:
±2.0%FS (The gasses which have no
absorption within the wavelength range
of measured gas (several tens pm) do
not interfere in principle).
Minimum detectable limit:
1% of minimum detectable limit
Response time (90% FS response): 1 to 5 seconds
Warm up time: 90 minutes or less
(* Excluding influenced range with absorptive gases such as HCl
and NH3)
EC directive compliance
Low voltage safety:EN61010-1(2001)
Electromagnetic: "Installation Category II"
EN61326-1(2006)
compatibility
EN61326-2-3(2006)
EN61000-3-2(2006)
EN6100-3-3(1995), A1(2001), A2(2005)
Basic principle
An infrared semiconductor laser is adopted for the light
source and photo-diode for photo-detection. Each measured gas component has its own wavelength range where
the gas absorbs the light (see the figure below).
The wavelength range consists of a set of many absorption
lines, and one of these lines is used for measurement.
As measurement is performed in this very narrow wavelength range, no other gas does not have an interference
in principle.
To detect the gas concentration, used is not a variation
of light intensity, but intensity of modulated signal amplitude.
N 2O
N 2O
NO
H 2S
O2
H 2S
CH4
HF
HF
HCl
NH3
750nm
Absorption intensity
Installation Environment
1000nm 1250nm 1500nm
Absorption line used for
measurement
1720nm 1740nm
1760nm
1780nm
1750nm
NH3
NH3
CO2
CO
2000nm 2250nm
Wavelength
Absorbing spectrum map of
measuring gases
1800nm 1820nm 1840nm
Wavelength
Absorption lines of HCl
3
ZSS
CODE SYMBOLS
When ordering, be sure to submit the order sheet on the last page of this Data Sheet.
Note 1
Digit
4 Measurable components
5
Unit
6
Measuring range
7
8
9
Modification No.
Flange rating
Specification
CO
CO (For use in high temperature)
HCl
HCl+H2O
CO2
CO2 (For use in high temperature)
CO+CO2
CO+CO2 (For use in high temperature)
O2 (Class 1 Laser)
O2 (For use in high dust)
CH4
NH3
NH3+H2O
ppm
mg/m3
vol%
0 to 2
0 to 2.5
0 to 4
0 to 5
0 to 10
0 to 15
0 to 20
0 to 25
0 to 50
0 to 100
0 to 200
0 to 250
0 to 400
0 to 500
0 to 1000
0 to 2000
0 to 5000
0 to 6000
Others
10 Number of analog output points
11 Number of analog input points
12 Analog output
13 Contact output/input
14 Cable length between receiver
and control unit
15 Cable length between receiver
and transmitter
16 Display and operation manual
18 Measuring optical path length
(unit: 1m)
19 Measuring optical path length
(unit: 0.1m)
20 Measuring optical path length
(unit: 0.01m)
21 Dust proof specification
4
ZSS
Note
Note 1
Note 1, 5
Note 1
Note 1, 9
Note 5
Note 5
Note 1
Note 1, 9
Note 2
10K 50A (JIS B 2212)
10K 100A
DN50/PN10
ANSI #150 2B
2 points
4 points
2 points
6 points
4 to 20mA DC
0 to 20mA DC
0 to 1V DC
0 to 5V DC
1 to 5V DC
5 output points, No input
5output points, 3 input points
Note 3
5m
Note 6
10m
20m
30m
40m
50m
80m
100m
Others
Note 4
2m
Note 7
5m
10m
15m
20m
25m
Others
Japanese
English
Chinese
0m
1m
2m
3m
4m
5m
6m
7m
8m
9m
0.0m
0.1m
0.2m
0.3m
0.4m
0.5m
0.6m
0.7m
0.8m
0.9m
0.00m
0.05m
(Used only when 10m is specified)
Standard
Note 8
Dust proof
4 5 6 7 8
9 10 11 12 13
14 15 16 17 18 19 20
Y4
A
B
C
F
G
H
K
L
P
Q
R
W
X
21
0
1
3
5
K
Q
S
L
V
0
1
T
A
B
C
D
J
E
F
G
H
M
X
Y
4
A
B
C
D
0
1
A
B
1
2
3
4
5
0
1
A
B
C
D
E
F
G
H
X
Note 1) When O2 conversion is required,
specify the reference O2
concentration (settable within
0 to 19vol%, O2: Integer).
(HCl meter, NH3 meter, CO meter)
Note 2) Specify the range within the
max/min range calculated from
path length.
If the measuring range x stack
length (optical path length)
exceeds 1000ppm/m,
consult Fuji.
Note 3) Cable length between receiver
and control unit: Max. 100m
Note 4) Cable length between receiver
and transmitter: Max. 25m
Note 5) Specified to use in high
temperature gas: 500°C or more,
and 1200°C or less
Note 6) Others (specified with X) are
possible only for 10 m or more.
Note 7) Others (specified with X) are
possible only for 5 m or more.
Note 8) Specify ‘D’ when dust exceeds
5g/Nm3-m.
Note 9) If H2O is contained in measured
component, be sure to write the
measured gas temperature into
“Specification sheet” at the last
page of this catalog, or contact
directly to Fuji’s service
department.
A
B
C
D
E
F
X
J
E
C
0
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
0
5
9
N
D
How to specify measuring optical path
length
Example: 2.25m 2 for 18th digit, 2 for
19th digit and 5 for 20th digit
10m 9 for 18th digit, 9 for
19th digit and 9 for 20th digit
LIST FOR COMBINATIONS OF MEASURABLE COMPONENTS, UNITS AND MEASUREMENT RANGES
Measurable component
4th digit
CO
CO (For use in high temperature)
HC
HC+H2O
CO2
CO2 (For use in high temperature)
CO+CO2
A
B
C
F
G
H
K
CO+CO2
(For use in high temperature)
O2(Class 1 Laser)
O2(For use in high dust)
CH4
NH3
NH3+H2O
L
P
Q
R
W
X
Measurable range 6th digit
(Unit: When 5th digit is 1 or 3)
Not feasible
Not feasible
V, 0, 1, T, A, B, C, D, J, E, F, G, H, Z
A, B, C, D, J, E, F
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
Not feasible
B, C, D, J, E, F, G, H, M, Z
0, 1, T, A, B, C, D, J, E, F, G, H, Z
A, B, C, D, J, E, F
Measurable range 6th digit
(Unit: When 5th digit is 5)
K, Q, S, L, V, 0, 1, T, A, X
V, 0, 1, T, A, X
Not feasible
Not feasible
K, Q, S, L, V, 0, 1, T, A, X
V, 0, 1, T, A, X
Q, S, L, V, 0, 1, T, A, X (CO)
Q, S, L, V, 0, 1, T, A, X (CO2)
V, 0, 1, T, A, X (CO)
V, 0, 1, T, A, X (CO2)
S, L, V, 0, 1, T, A, B, X
S, L, V, 0, 1, T, A, X
K, Q, S, L, V, 0, 1, T, A, X
Not feasible
Not feasible
Scope of Delivery
• Receiver unit
• Transmitter unit
• Control unit
• Cable between receiver unit and transmitter unit (specified
length)
• Cable between receiver unit and control unit (specified length)
• Standard accessory set, instruction manual
Optional Items
• Spare parts for one year (ZBN1SS12)
• Calibration gas cell (*1) (*2)
• Cable between receiver unit and transmitter unit (for
calibration) (*1)
• Cable between receiver unit and control unit (for calibration) (*1)
• Standard gas (ZBM), pressure regulator (ZBD)
• Recorder (when necessary, Fuji's product type PHL/PHR, etc.)
• Others
*1: One set of the cables and calibration gas cell are necessary for installation and annual maintenance.
*2: Standard length 1m (200mm or 500mm when the measuring range is low concentration)
Standard accessories
Name
Quantity
Specification
Bolt
8 (16)
M16×55 (70) SUS (*)
Nut
8 (16)
M16
SUS (*)
Spring washer
8 (16)
M16
SUS (*)
Flat washer
8 (16)
M16
SUS (*)
Companion flange packing or
flange packing specified for
use in high temperature
2
According to flange specifications
Bolt for angle adjustment
6
Hex socket bolt M8 × 70
Power supply fuse
2
Instruction manual
1
(*When "B" or "C" is specified in the 9th digit in a code symbol, quantity is 16 pieces. 8 pieces are attached in other cases.)
(*When "B", "C" or "D" is specified in the 9th digit, Bolt length is 70mm. It is 55mm in other cases. Inch-sized bolts are not
applicable.)
Spare Parts for One Year (ZBN1SS12)
Name
Quantity
Remarks
Silicon packing A
2 pieces
For bellows (ZZP*ZSSTK7N3508P1)
O-ring
2 pieces
(ZZP*ZSSTK7P2530P3)
5
ZSS
Calibration/installation fixtures list (Option)
Parts name
(1)
(2)
Quantity
Cable between receiver unit and
control unit (for calibration)
Cable between receiver unit and
transmitter unit (for calibration)
Calibration gas cell (*3)
(3)
(HCl, NH 3 , CO,CO 2, CO+CO 2,CH 4)
1
ZZP*ZSSTK4J1271C2
1
ZZP*ZSSTK4J0641C3
1
ZZP*ZSSTK4J3676C1
Calibration gas cell (*3)
(3) Calibration gas cell / about 1000mm
ZZP*ZSSTK4J5026C1
(HCl+H 2 O, NH 3+H 2 O, O 2)
(4)
Type
Optical axis adjusting tool (laser
pointer, target)
1
ZZP*ZSSTK4J1274C1
(5)
IR card (for NH 3)
1
ZZP*ZSSTK7N4505P1
(6)
Check cell
1
ZZP*ZSSTK4J2605C1
(7)
Filter regulator
1
ZZP*ZSSTK7F9554P2
(8)
Mist separator
1
ZZP*ZSSTK7H8049P1
(9)
R1/4 stopper (plug) for mist separator
1
ZZP*ZSSTK738114P7
Flow meter (20 to 100L/min) (*4)
1
ZZP*ZSSTK7N4624P1
1
ZZP*ZSSTK7N4624P2
Flow meter (30 to 300L/min) (*4)
1
ZZP*ZSSTK7N8849P1
Air-set box (20 to 100L/min)
1
ZZP*ZSSTK7N6685C1
(10) Flow meter (4 to 50L/min) (*4)
(11) Air-set box (4 to 50L/min)
1
ZZP*ZSSTK7N6685C2
Air-set box (30 to 300L/min)
1
ZZP*ZSSTK7N6685C3
Air purge mechanism (20 to 100L/min)
1
ZZP*ZSSTK7P1433C1
(12) Air purge mechanism (4 to 50L/min)
1
ZZP*ZSSTK7P1433C2
Air purge mechanism (30 to 300L/min)
1
ZZP*ZSSTK7P1433C3
(13) BNC cable for optical axis adjustment
1
ZZP*ZSSTK7P2524C1
(2) Cable between receiver and
transmitter (approx. 1600mm)
Parts used for calibration
(1) Cable between receiver unit and control unit
Stack
(4) Optical axis adjusting tool (target)
* 3: Standard length 1m (200mm or 500mm when the
measuring range is low concentration)
* 4: Usually, 2 units are required to adjust each air purge
for receiver unit and transmitter unit.
(4) Optical axis adjusting tool
(laser pointer)
Parts for adjusting the optical axis
Transmitter unit
(5) IR card
IR CARD
Laser beam (visible)
Invisible light
Invisible light / visible fixtures(for NH3)
Receiver unit
(6) Check cell
Transmitter unit
Stack
Control unit
Laser gas analyzer with check cell (ZSS)
SV3
Atmospheric
release
SV2
SV1
(AO)
Sequencer
Remote check input (AI)
N2 cylinder Standard gas
cylinder
6
Supplies for checking span value on site
(11) Air set box
4-φ11
(Mounting hole)
Parts for air purge
(7) Filter regulator
Box
(8) Mist separator
Mist separator
(OUT)
(IN)
(OUT)
(IN)
350
(OUT)
380
(OUT)
(IN)
measuring
instrument
air inlet
Receiver
purge outlet
Transmitter
purge outlet
(10) Flow meter
420
500
110
65
Measuring instrument
air inlet
Rc1/4
Transmitter
purge outlet
Rc1/4
120
200
43
Receiver
purge outlet
Drain outlet
(φ10 piping)
(not supplied by us)
Piping system diagram
150
223
Filter regulator
Pipe with
external
diameter φ10
(out of supply
scope)
300
(IN)
Flow
meter
58
Rc1/4
Drain outlet
Piping retainer for φ10
(12) Air purge mechanical part
total 20m
1m
Piping material: φ8/6 Teflon pipe
(OUT)
(OUT)
(OUT)
(IN)
(IN)
0.5m
0.5m
(IN)
for φ8 tube
(metallic tube
recommendable)
to Transmitter Mist separator
purge inlet
Filter regulator
to Receiver
purge inlet
to Receiver
purge inlet
Measuring
instrument air inlet
(Coupling joint)
Piping system diagram
(13) BNC cable for optical axis adjustment
7
ZSS
OUTLINE DIAGRAM (Unit : mm)
Or air-set box
2-Flowmeter
(option)
Copper or SUS pipe (not supplied)
Instrumentation airline joint (not supplied)
Filter regulator (option)
20 to 100L/min
φ8/6 Teflon pipe
(not supplied)
Instrumentation
air inlet
Mist separator
(option)
150 to 200
Approx. 400
Flange of angle
adjustment
Purging gas inlet
Receiver box
(with bell-and-spigot
(120×180×100)
joint for φ8/6)
16
100
150 to 200
With bolt and nut
(M16×55)
Flange pipe to
mount stack
(not suppried)
Note 1
Approx. 400 (432)
Flange of angle Transmitter box
adjustment
(160×240×100)
Purging gas inlet
100
Stack
420
130
74
160
JIS10K50A FF
(with packing)
Receiver unit
Transmitter unit
φ34
(Connector)
2-φ12
Cable between receiver unit
and transmitter unit (standard : 2m)
240
2-φ12
74
Note 1: Be sure to set the R1/4 sealing
plug (ZZP*ZSSTK738114P7)
on the drain outlet of the mist
separator.
φ10.5
420
(Connector)
φ14
φ34
320
120
φ155
0.5 to 10m
Control unit
LASER GAS ANALYZER
ZSS
φ30
134
AI
AO
Power
supply
Cable length
(200mm)
Connector
Cable between receiver unit and control unit
(standard : 5m)
8
3
DI/O
7-Waterproof grand
(Internal diameter φ12)
74
12
Mounting diagram
Be sure to prepare instrument air (N2 for an O2 meter).
Note) Air purge should be performed if the equipment is
installed in a flue whether or not it is operated.
EXTERNAL CONNECTION DIAGRAM (When number of analog output is 2 points)
There are two types of output/input terminal.
Power supply terminal
1
2
3
100 to 240 V AC
(50/60Hz)
FG
Screw diameter : M4
AO terminal
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20
Screw diameter : M2 or the equivalent
1 AO1+
2 AO1-
21 AO2+
22 AO2-
Analog output 1
Analog output 2
Al terminal
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20
Screw diameter : M2 or the equivalent
3
4
5
6
7
8
AI1+
AI1AI2+
AI2AI3+
AI3-
Analog input 1
Analog input 2
Analog input 3 (AI extension board is required)
23
24
25
26
27
28
AI4+
AI4AI5+
AI5AI6+
AI6-
Analog input 4 (AI extension board is required)
Analog input 5 (AI extension board is required)
Analog input 6 (AI extension board is required)
DI/DO terminal
21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40
1
2
9
10
11
12
13
14
15
16
17
18
19
20
DO1
DO1
DO2
DO2
DO3
DO3
DO4
DO4
DO5
DO5
DO6
DO6
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20
Screw diameter : M2 or the equivalent
Low light transmission
Device failure
During hold / during calibration
Outside the range of upper /
lower limits (component 1)
Range identification (component 1)
29
30
31
32
33
34
35
36
37
38
39
40
DO7
DO7
DO8
DO8
DI1
DI1
DI2
DI2
DI3
DI3
DI4
DI4
Power supply off
Average value reset signal (option)
Switching instantaneous value/moving average value (option)
Remote hold (option)
Note) Unassigned terminal may be connected to internal circuit. So they should not be used as repeating terminals.
9
ZSS
External connection diagram (when number of analog output is 4 points)
Power supply terminals
1
2
3
AC100 to 240V
50/60Hz FG
Screws diameter: M4
AO terminal
AO4 Unassigned Unassigned
+
-
7
8
9
10 11 12
1
2
3
4
+
AO1
+
AO2
5
6
1
2
3
4
AO1+
AO1AO2+
AO2-
1
2
3
4
5
6
AI1+
AI1AI2+
AI2AI3+
AI3-
Analog output 1
Analog output 2
5
6
7
8
AO3+
AO3AO4+
AO4-
7
8
9
10
11
12
AI4+
AI4AI5+
AI5AI6+
AI6-
Analog output 3
Analog output 4
+
AO3
AI terminal
AI4
+
AI5
-
+
AI6
-
+
-
7
8
9
10 11 12
1
2
3
4
+
AI1
5
+
AI2
6
+
AI3
Analog input 1
Analog input 2
Analog input 3
(AI extension board
is required)
Analog input 4
(AI extension board is required)
Analog input 5
(AI extension board is required)
Analog input 6
(AI extension board is required)
DI/DO terminal
25 26 27 28 29 30 31 32 33 34 35 36
13 14 15 16 17 18 19 20 21 22 23 24
1
1
2
3
4
5
6
7
8
9
10
13
14
15
16
17
18
2
DI1
DI1
DI2
DI2
DI3
DI3
DI4
DI4
DI5
DI5
DO1
DO1
DO2
DO2
DO3
DO3
3
4
5
6
7
8
9 10 11 12
Average value reset signal (option)
Switching instantaneous value/moving
average value (option)
Remote hold (option)
Low light transmission
Device failure
During hold/during callibration
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
DO4
DO4
DO5
DO5
DO6
DO6
DO7
DO7
DO8
DO8
DO9
DO9
DO10
DO10
DO11
DO11
Outside the range of upper/lowerlimits (component 1)
Range identification (component 1)
Power supply off
Note) Unassigned terminals may be connected to internal circuit. So they should not be used as repeating terminals.
10
MOUNTING DIMENSIONS (Unit: mm)
Do not install the device in a stack if air purge is not performed.
It may become unusable because the window will get dirty.
Stack
c
d
a
a: Inner diameter of φ50mm or more
b: 20 to 70mm (*5)
c: Approximately 200 to 250mm (*6)
d: 400mm or more (*7)
e: 500 to 1500mm (*7)
b
Flange pipe to mount stack (*9)
e
Scaffold (*8)
*5: When used in an enviroment with high dust, reserve at least 50mm.
*6: When internal gas temperature is high, reserve the distance of 250mm or more. Note that, adjustable angle
range narrows down, so mounting angle of flange pipe to mount stack should be adjusted accurately.
*7: Sufficiently secure a space for installation which permits easy operation.
*8: Scaffold is required for the both side of receiver unit and transmitter unit.
*9: Use a flange provided with a valve where there is a risk of gas injection or the operator may be exposed to
dangerous situation.
*10: Tighten the flange with a torque of 118±114N-m. When tightening the flange, apply grease to the bolts
(recommendable to use grease containing molybdenum).
Rough guide to mount the companion flange
When θ is larger than 5°, mount it within the γ angle (≈5°) inside the circle with the A radius.
γ
Y
X
A
θ
Z
X: Inner diameter of flange of receiver unit and transmitter unit
Y: Length of flange of receiver uit and transmitter unit
Z: Length of stack
γ, θ = ATAN
X
2 (Y + 125)
A ≈ 0.087 × (Z + Y + 125)
11
ZSS
Please specify the following items when you contact us for inquiry.
1.Code symbol ZSS
2. Parameter of measured gas
Item
Min. value
Typical value
Max. value
Measured gas concentration
Temperature (°C)
Pressure (kPa)
Flow (m/s)
Moisture (vol%)
Dust (mg/Nm3)
Other gas component (vol%/ppm)
Other gas component (vol%/ppm)
Other gas component (vol%/ppm)
3.Stack diameter (Inner diameter) (measured optical path length)
m
A
4.C able between receiver unit and control unit. B
m
5.C able between receiver unit and transmitter unit. C
m
A
Instrumentation air
Instrumentation air
Stack
Receiver unit
Transmitter unit
C
Control unit
B
6.Spares for 1-year measurement
no need necessary
7.Optional items
no need necessary
8.Output of O2 correction value (for only HCl and NH3 meters) no need necessary
9.Oxygen correction reference value (%) (When "necessary" is selected in 8)
10.Vibration
no need necessary (
%
G)
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div
Sales Group
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fujielectric.com
Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425
http://www.fjielectric.com/products/instruments/
Information in this catalog is subject to change without notice.
Printed in Japan