MPSIND HPU63A-111

HPU63A SERIES
63W Desk Top Switching Power Supplies For Health Care Applications.
Description:
The HPU63A series of AC/DC switching mode power supplies
provide 63 Watts of continuous output power. All supplies are
UL94V-1 min compliant. They are suited for use in Health care
applications. All models meet FCC Part-18 class B and CISPR-11
EN55011 class B emission Limits and are designed to comply
with UL/c-UL(UL 60601-1:2ndEdition) ,TUV/T-mark (EN 606011:2ndEdition), and new CE requirements. All units are 100%
burned in and tested.
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single Output
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Output Voltage Protection(Crowbar Desing)
Energy Star 2.0, Efficiency level V
Class I
2 year warranty
Safety Approvals :
EN 60601-1
IEC 60601-1
Agree to apply for the PSE
if order on hand
Electrical Characteristics:
Sym.
Vin
*
Parameter
Test Conditions
Min. Typ. Max. Unit
Safety Approvals Input
Voltage Range
100
240
VAC
Operate Voltage Range
90
264
VAC
fin
Po
Vo
Io
Input Frequency
Output Power Range
Output Voltage Range
Output Current Range
47
63
Hz
W
V
A
Iil
Iih
Input Current (Low Line)
Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
I rl
Low Line Inrush Current
I rh
High Line Inrush Current
Io=Full load, 25 C , Cool start,
Vin= 115VAC
Io=Full load, 25 C , Cool start,
Vin= 230VAC
Io=Full Load, Vin=230VAC
Vin=90 to 264VAC
63
0
See rating chart
See rating chart
1.62
0.72
A
A
26
30
A
43
47
A
92
%
%
%
Eff
REG-i
REG -o
Efficiency
OVP
112
Over Current Protection
110
150
Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
4
Hold-Up Time
Io=Full Load, Vin=110VAC
12
Io=Full Load, Vi n= 10 0V AC
Start Up Time
0.3
1
2
Ripple & Noise (Peak to Peak) Full Load, Vin=100VAC
1
Safety Ground Leakage C urrent Io = Fu ll Load, Vin=240VAC
0.1
No-Load Power Consumption
No load,Vin=240VAC
0.3 0.4 0.5
Temperature Coefficient
-0.04
All output
0.04
OCP
Ttr
Thold
Ts
Vp-p
Ilk
Pno
TC
Line Regulation
Load Regulation
Over Voltage Protection
87
Io=Full Load
Vin=230VAC
1
5
132
%
%
mS
mS
S
%
mA
W
%/ C
* Note: The specifics for testing the energy efficiency of HPU63 Series are outlined in a separate document titled "Test Method for Calculating the
Energy Efficiency of Single-Voltage External Ac-Dc and Ac-Ac PowerSupplies (August 11, 2004)," which is available on the ENERGY STAR Website.
Environmental :
Sym.
Toper
Parameter
T e s t Conditions
Operating Temperature
Min. Typ. Max.
Unit
40
70
C
Storage Temperature
-40
85
C
Operating Humidity
0
95
%
0
75
Tstg
Ho
Hr
Storage Humidity
MTBF
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd
Derate linearly from 100% load at 40 C to 50% load at 70 C
0
0.1M
%
Hrs
HPU63A SERIES
63W Desk Top Switching Power Supplies For Health Care Applications.
Safety Specifications:
Sym.
Parameter
Test Conditions
Dielectric Withstanding Voltage
for Primary to secondary
Dielectric Withstanding Voltage
Vpg for
Primary to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Vps
Min. Typ. Max.
Unit
Primary to secondary
5656
VDC
Primary to ground
2828
VDC
50
B
B
M
CLASS
CLASS
Test Voltage = 500VDC
Vin=220VAC
Vin=110VAC
Output Voltage And Current Rating Chart (Single Output) :
Model Number
HPU63A-105
HPU63A-106
HPU63A-107
HPU63A-108
HPU63A-109
HPU63A-110
HPU63A-111
Output Voltage
12
15
18
24
30
36
48
VDC
VDC
VDC
VDC
VDC
VDC
VDC
Output Current
5.25
4.20
3.50
2.62
2.10
1.75
1.31
A
A
A
A
A
A
A
Total Regulation
5%
5%
5%
3%
3%
3%
3%
M aximum
Output Power
63W
63W
63W
63W
63W
63W
63W
HPU63A-105~109 had been approved by PSE .
HPU63A-105~107 are required to use AWG#16/2C/4FT output cable.
HPU63A-108~110 are required to use AWG#18/2C/6FT output cable.
The regulation and efficiency will be changed by modified output cable.
Mechanical Specifications:
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 330-380gs approx.
3. Optional output connector:
See page Appendix.