EMERSON-NETWORKPOWER NFS110

Embedded Power for
Business-Critical Continuity
Rev. 06.10.08
NFS110 Series
1 of 5
NFS110 Medical
Series
Single and
quad output
Total Power:
80 - 110 W
Input Voltage: 90 - 253 Vac
127 - 357 Vdc
# of Outputs: Single, quad
Electrical Specifications
Output
Special Features
• 7.0 x 4.25 x 1.8 inch package
• Medical, dental and
laboratory applications
• Overvoltage and short circuit
protection
• 110 W with 20 CFM
• UL, VDE and CSA safety
approvals
• EN60601-1 and UL2601
medical approvals
• Available RoHS compliant
• 2 year warranty
Voltage adjustability
+5.1 V o/p on multi’s
5.1 V single output
12 V single output
15 V single output
24 V single output
±3.0%
±3.0%
12-14 V
15-18 V
24-30 V
Line regulation
LL to HL, FL
All outputs on all units
±0.1% max.
Overshoot/undershoot
Temperature coefficient
Overvoltage protection
At turn-on no lead
All outputs
Multi o/p 5.1 V only
5.1 V single
12 V single
15 V single
24 V single
0%
±0.02%/°C
6.25 V ±0.75 V
6.25 V ±0.75 V
15.75 V ±1.0 V
22 V ±1.5 V
33 V ±2.5 V
Output power limit
Primary power
limited
Pin max. 160 W
Pout min. 110 W
Burst mode operation
Short circuit protection
Safety
VDE0805/EN60601-1/
IEC601/IEC1010
File No. 10401-3336-1049
Licence No. 2874
UL2601 File No. E147937
CSA C22.2 No. 125
File No. LR41062C
Input
Input voltage range
90-253 Vac
127-357 Vdc
Input frequency range
47-440 Hz
Input surge current
110 Vac. 50 Hz
230 Vac. 50 Hz
Safety ground leakage current 132 Vac
264 Vac
All specifications are typical at nominal input, full load at
25°C unless otherwise stated
17 A
35 A
50 μA
100 μA
Embedded Power for
Business-Critical Continuity
Rev. 06.10.08
NFS110 Series
2 of 5
EMC Characteristics
Conducted emissions
EN55022, FCC part 15
Level A
Radiated emissions
EN55022, FCC part 15
Level A
ESD air
EN61000-4-2, level 3
Perf. criteria 1
ESD contact
EN61000-4-2, level 4
Perf. criteria 1
Surge
EN61000-4-3, level 3
Perf. criteria 1
Fast transients
EN61000-4-4, level 3
Perf. criteria 1
Radiated immunity
EN61000-4-5, level 3
Perf. criteria 2
Conducted immunity
EN61000-4-6, level 3
Perf. criteria 2
Hold-up time
110 Vac @ 80 W
110 Vac @ 110 W
230 Vac @ 80 W
230 Vac @ 110 W
35 ms
17 ms
140 ms
100 ms
Efficiency
Multiple outputs
+5.1 V single
12 V and 15 V singles
24 V single
70% typical
70% typical
72% typical
75% typical
Isolation voltage
Input/output
Input/chassis
4000 Vac
1500 Vac
General Specifications
Approvals and standards
(see note 12)
VDE0750, IEC60601,
IEC1010, UL2601
CSA C22.2 No. 125
Weight
Singles
Multiple outputs
550 g (19.4 oz)
600 g (21.2 oz)
MTBF (@25º C)
MIL-HDBK-217E
125,000 hours min.
Environmental Specifications
Thermal performance
Operating, see curve
0° C to +70 °C
(See notes 9, 10)
Non-operating
-40 °C to +85 °C
0 °C to 50 °C amb. convection cooled 80 W
+50 °C to +70 °C,
amb. convection cooled
Derate 2 W/°C
0 °C to +50 °C, 20 CFM forced air
110 W
+50 °C to +70 °C, 20CFM forced air
Derate 2.75 W/°C
Peak, 0 °C to +50 °C, max. 60 seconds 110W
Relative humidity
Non-condensing
5% to 95% RH
Altitude
Operating
10,000 feet max.
Non-operating
40,000 feet max.
5-500 Hz
2.4 G rms peak
Vibration (See Note 11)
Embedded Power for
Business-Critical Continuity
Rev. 06.10.08
NFS110 Series
3 of 5
Ordering Information
Output
Voltage
Output Currents
Max
(1)
Peak (2)
Fan (3)
3
4
5
6
7
8
9
10
11
12
13
14
Total
Regulation (5)
Model Numbers (13, 14, F)
+5.1 V
8A
20 A
10 A
50 mV
±2.0%
+12 V
4.5 A
9A
5A
120 mV
±3.0%
–12 V
0.5 A
1.5 A
1A
120 mV
±3.0%
–5 V
0.5 A
1.5 A
1A
50 mV
±3.0%
+5.1 V (IA)
8A
20 A
10 A
50 mV
±2.0%
+24 V (IB) (6)
3.5 A
4.5 A
4.5 A
240 mV
+10/–5.0%
+12 V
4.5 A
9A
5A
120 mV
±3.0%
–12 V
0.5 A
1.5 A
1A
120 mV
±3.0%
+5.1 V
8A
20 A
10 A
50 mV
±2.0%
+15 V
4A
7.5 A
5A
150 mV
±4.0%
–15 V
0.5 A
1.5 A
1A
150 mV
±3.0%
–5 V
0.5 A
1.5 A
1A
50 mV
±3.0%
12 V
7A
9A
9A
120 mV
±2.0%
NFS110-7912J (7,8)
15 V
5A
7.3 A
7.3 A
150 mV
±2.0%
NFS110-7915J (7,8)
24 V
3.5 A
4.5 A
4.5 A
240 mV
±2.0%
NFS110-7924J (7,8)
Notes
1
2
Ripple (4)
Convection cooled, 80 W maximum.
Peak outputs lasting less than 60 seconds with duty cycle less than 10%. Total
peak power must not exceed 110 W.
Forced air, 20 CFM at 1 atmosphere, 110 W maximum.
Figure is peak-to-peak. Output ripple is measured across a 50 MHz bandwidth
using a 12 inch twisted pair terminated with a 47 μF capacitor.
Total regulation is defined at the static output regulation at 25 °C, including
initial tolerance, line voltage within stated limits and output voltages adjusted
to their factory settings. Also for NFS110-7902PJ, for 24 V output stated
regulation IA / IB ² 5. This output will maintain ±5.0% regulation if
IA ² 5 A, where IA = +5.1 V output current and IB = +24 V output current.
Single output models have floating outputs which may be referenced as either
positive or negative. Higher voltage supplies, may be adjusted over a wide
output voltage range, as long as the total output power does not exceed 80
Watts (natural convection) or 110 Watts (forced air).
Power fail detect not available on single output models.
Derating curve is application specific for ambient temperatures > 50 °C, for
optimum reliability no part of the heatsink should exceed 90 °C and no
semiconductor case temperature should exceed 100 °C.
Caution: Allow a minimum of 1 second after disconnecting the power when
making thermal measurements.
The user should read the PSU installation instructions in conjunction with the
relevant national safety regulations in order to ensure compliance.
Three orthogonal axes, random vibration, 10 minute test for each axis.
This product is only for inclusion by professional installers within other
equipment and must not be operated as a stand alone product.
The ‘J’ suffix indicates that these parts are Pb-free (RoHS 6/6) compliant. TSE
RoHS 5/6 (non Pb-free) compliant versions may be available on special request,
please contact your local sales representative for details.
NOTICE: Some models do not support all options. Please contact your local
Emerson Network Power representative or use the on-line model number
search tool at http://www.powerconversion.com to find a suitable alternative.
NFS110-7901PJ
NFS110-7902PJ
NFS110-7904PJ
TRANSIENT RESPONSE
NFS110-7901PJ
+5.1 V (7.5-10 A)
+12 V (2.5-5 A)
-12 V (0.5-1 A)
-5 V (0.5-1 A)
NFS110-7902PJ
+5.1 V (7.5-10 A)
+12 V (2.5-5 A)
-12 V (0.5-1 A)
24 V (1.5-3 A)
NFS110-7904PJ
+5.1 V (7.5-10 A)
+15 V (2.5-5 A)
-15 V (0.5-1 A)
-5 V (0.5-1 A)
150 mV peak,
1 ms recovery
100 mV peak,
0.5 ms recovery
100 mV peak,
0.5 ms recovery
100 mV peak,
0.5 ms recovery
150 mV peak,
1 ms recovery
100 mV peak,
0.5 ms recovery
100 mV peak,
0.5 ms recovery
300 mV peak,
1 ms recovery
150 mV peak,
1 ms recovery
100 mV peak,
0.5 ms recovery
100 mV peak,
0.5 ms recovery
100 mV peak,
0.5 ms recovery
NFS110-7905J
+5.1 V (10-20 A)
250 mV peak,
1 ms recovery
NFS110-7912J
+12 V (4.5-9 A)
360 mV peak,
1 ms recovery
NFS110-7915J
+15 V (3.65-7.3 A)
450 mV peak,
1 ms recovery
NFS110-7924J
+24 V (2.25-4.5 A)
720 mV peak,
Embedded Power for
Business-Critical Continuity
Rev. 06.10.08
NFS110 Series
4 of 5
OPTIONAL POWER FAIL DETECT TIMING DIAGRAM
AC (J1) mating connector
Molex 09-50-3051 or Molex 09-91-0500 mating connector with 2478
or equivalent crimp terminals.
Power fail detect signal (Note 8)
50ms ≤ T1 ≤ 200ms
T2 will vary with line and load
T3 ≥ 3ms
Pout: 110W
PFD output is an open collector which
will sink ≤ 40mA in the low state.
AC INPUT
4.75V
4.75V
5V OUTPUT
HIGH
DC (J2) mating connector
PFD SIGNAL
Molex 09-50-3131 or Molex 09-91-1300 mating connector with 2478
or equivalent crimp terminals.
3.5V MIN
LOW
0-0.4 MAX
T3
T2
T1
0.475
0.010
(12.07)
1.80 (45.72) MAX
1.00 0.005
(25.4)
DERATING CURVE (See Notes 9, 10)
0.050 0.020
(1.27)
6-32 UNC
(4PL)
1.55 0.005
(39.37)
3.35 0.015 (85.09)
Output Power (Watts)
110W
PIN 1
Voltage Adjust
20 CFM FORCED AIR COOLING
80W
Optional
Power Fail
Detect Circuit
(Key = Pin 2)
4.250
0.020
(107.95)
J2
NATURAL
CONVECTION
COOLING
T1
T1
3.750
0.005
(95.25)
55W
40W
Pin 1
PIN 1
L1
11W
J1
C6
F1
0.156
DIA. HOLE
(4PL)
11W MINIMUM LOAD REQUIRED AT 230VAC
0W
0C
10C
20C
30C
40C
50C
60C
70C
5A, 250 VAC
0.250
0.010
(6.35)
6.50 0.005 (165.1)
Pin Connections
7.00 0.020 (177.8)
0.250
0.010
(6.35)
ALL DIMENSIONS IN INCHES (mm)
Mechanical Notes
A Metallic or non-metallic stand-offs (maximum diameter 5.4mm) can be used in
all four mounting holes without effecting safety approval.
B The ground pad of the mounting hole near J1, allows system grounding
through a metal stand-off to the system chassis.
C The heat sink is grounded, and allows system grounding by mechanical
connection to the system chassis.
D The supply must be mechanically supported using the PCB mounting holes and
may be additionally supported by the heatsink mounting holes.
E It is always advisable to attach the power supply heat sink to another thermal
dissipator (such as a chassis or finned heatsink etc). The resulting decrease in
heat sink mounted component temperatures will improve power supply
lifetime.
F A standard L-bracket and cover is available for mounting which contains all
screws, connectors and necessary mounting hardware. The kit is available,
order part number “NFS110CJ”.
J1
-7901PJ
-7902PJ
-7904PJ
SINGLES
Pin 1
AC Ground
AC Ground
AC Ground
AC Ground
Pin 2
AC Neutral
AC Neutral
AC Neutral
AC Neutral
Pin 3
AC Line
AC Line
AC Line
AC Line
Pin 1
+5.1 V
+5.1 V
+5.1 V
Vout
Pin 2
+5.1 V
+5.1 V
+5.1 V
Vout
Pin 3
+5.1 V
+5.1 V
+5.1 V
Vout
Pin 4
Return
Return
Return
Return
Pin 5
Return
Return
Return
Return
Pin 6
Return
Return
Return
Return
Pin 7
Return
Return
Return
Return
Pin 8
+12 V
+12 V
+15 V
Vout
Pin 9
+12 V
+12 V
+15 V
Vout
Pin 10
PFD
PFD
PFD
N/C
Pin 11
-12 V
-12 V
-15 V
N/C
J2
Pin 12
Pin 13
Removed for Key
-5 V
+24 V
-5 V
N/C
N/C = no connection.
Embedded Power for
Business-Critical Continuity
Americas
5810 Van Allen Way
Carlsbad, CA 92008
USA
Telephone: +1 (760) 930 4600
Facsimile: +1 (760) 930 0698
Rev. 06.10.08
NFS110 Series
5 of 5
Europe (UK)
Waterfront Business Park
Merry Hill, Dudley
West Midlands, DY5 1LX
United Kingdom
Telephone: +44 (0) 1384 842 211
Facsimile: +44 (0) 1384 843 355
Asia (HK)
14/F, Lu Plaza
2 Wing Yip Street
Kwun Tong, Kowloon
Hong Kong
Telephone: +852 2176 3333
Facsimile: +852 2176 3888
For global contact, visit:
www.powerconversion.com
technicalsupport@
powerconversion.com
While every precaution has been taken to ensure
accuracy and completeness in this literature, Emerson
Network Power assumes no responsibility, and disclaims
all liability for damages resulting from use of this
information or for any errors or omissions.
Emerson Network Power.
The global leader in enabling
business-critical continuity.
AC Power
Connectivity
DC Power
Embedded Computing
Embedded Power
Monitoring
Outside Plant
Power Switching & Controls
Precision Cooling
Racks & Integrated Cabinets
Services
Surge Protection
EmersonNetworkPower. com
Emerson Network Power and the Emerson
Network Power logo are trademarks and
service marks of Emerson Electric Co.
©2008 Emerson Electric Co.