FUJI EDS10-11C

FC SERIES
MOVING COIL TYPE INDICATOR
DATA SHEET
PBA
This is an indicator compatible with various types of inputs
such as FC series signal, DC voltage and current signals as
well as thermocouple and resistance bulb signals.
FEATURES
1. Various types of signals can be input directly to the indicator.
2. High input impedance
The instrument has an input impedance of 1 MΩfor input
of 1 to 5V DC. No circuit adjusting resistor is required for
thermocouple input.
3. The indicator is compact and light weight.
SPECIFICATIONS
Measuring input: FC series signal
Input
resistance
Allowable input
signal source
resistance
Voltage
Slide resistance (resistance span 100Ω)
J: Three wire type
K: Two-wire type
(with circuit adjusting resistor)
Opening angle transmitter L:
Three-wire type, resistance
span 100Ω (Zero point and span
adjustable)
Input resistance and allowable input signal source resistance;
1 to 5V
1MΩ
2kΩ or less
Others
40kΩ to 1MΩ
100Ω or less
Current
A; 1 to 5V DC
B; 4 to 20mA DC
DC current
D; 0 to 100 µA
...200mA DC
DC voltage
E; 0 to 10mV ... 5V DC
Thermocouple F; 0 to 10mV DC or more
(With reference junction compensator)
(Upper limit overshoot
burnout circuit is available on request)
Thermocouple G;10mV DC span or
more
(With reference junction compensator)
(Upper limit overshoot
burnout circuit is available on request)
Resistance bulb H;
JPt100Ω
50°C span or more,
three-wire type
Resistance bulb W;
Pt100Ω
50°C span or more,
three-wire type
4 to 20mA
5Ω
—
Others
1 to 100Ω
—
Input signal
Thermocouple
Resistance
40 to 600kΩ
—
100Ω or less
(50Ω or less
when burnout circuit is provided)
6Ω or less per wire
EDS10-11c
Date
Fab. 28, 1991
PBA
Allowance:
± 1.0%
Scale length: 100mm
Response time:Within 3 sec (time required for stabilizing the
indicator pointer within ± 1.5% after applying
input corresponding to 2/3 (approx.) of full
scale)
Power supply: 24V (20 to 30V) DC or
100V± 10% AC, 50/60Hz
Power indicator lamp:
Rectangular green LED
Power consumption:
Approx. 4W (24V DC)
Approx. 4.5VA (100V AC)
Ambient temperature:
0 to 45°C
Ambient humidity:
90% RH max.
Enclosure:
Steel case
External dimensions (H x W x D):
144 x 48 x 437mm
Weight:
Approx. 3kg
Finish color:
Munsell 7Y 7.3/1.4
Optional devices (optional items):
• Alarm device
Upper limit (H or A), lower limit (L or B),
upper and lower limits (K or C, with one
common contact)
Setting accuracy: ± 1.0%
H, L, K:
Excitation alarm
A, B, C:
Non-excitation alarm
Contact:
On alarm type, off alarm is
available on request
Contact capacity:
100V, 0.3A AC
(resistive load)
24V, 0.2A DC
• Alarm indicator lamp
Alarm signal displayed with red LED
Scope of delivery:
Indicator and mounting bracket
Mounting method:
Panel flush mounting
Standard: Vertical mounting on vertical
panel
Non-standard: Titled mounting
(angle α to be specified)
}
α
α = 90 to 0°
CODE SYMBOLS
1 2 3 4 5 6 7 8
9 10
P BA
2
1
Description
Input signal
1 to 5V DC
4 to 20mA DC
0 to 100µA ...200mA DC
0 to 10mV ....5V DC
0 to 10mV DC or more (with reference
junction compensator)
10mV DC span or more (with reference
junction compensator)
Resistance bulb JPt 100Ω, 50°C span or
more,three-wire type
Resistance bulb Pt 100Ω, 50°C span or
more,three-wire type
Slide resistant, three-wire type
Slide resistant, two-wire type
(with circuit adjusting resistor)
For opening angle indicator (resistance
100Ω, zero point and span adjustable)
Other inputs
A
B
D
E
F
G
H
W
J
K
L
*Z
Y
Alarm device
Upper limit
Lower limit
Upper and
lower limits
Upper limit
Lower limit
Upper and
lower limits
Without alarm
1
* 3
Power supply
24V DC
100V AC 50/60Hz
H
L
K
*A
*B
*C
Excitation alarm
(ON alarm)
Non-excitation alarm
(ON alarm)
Application
For general use
For connecting Zener barrier:
0
3
"3" to be specified for connecting the
indicator with thermocouple, resistance bulb
input to Zener barrier (PWZB3 or PMZB4).
2
Mounting-mehtod
Vertical mounting
0
* 1
Alarm indicator lamp
None
provided
Notes: Symbols of resistance bulbs are as follows.
Jpt 100 ... Previous JIS standard
pt 100 ... New JIS standard
OUTLINE DIAGRAM (Unit : mm)
Resistance
bulb
Resistor
Power supply
24V DC (or
100V AC)
Input
Voltage
Current
L C H I+ I- PC VP G IO I+ I11 10 9 8 7 6 5 4 3 2 1
CONNECTION DIAGRAM
Alarm
PBA
ORDERING INFORMATION
1. Product name
2. Code symbol
3. Input
4. Scale
5. Whether or not alarm device and
alarm indicator lamp are required
6. Power supply
7. Application
8. Other necessary information
Caution on Safety
Asterisked (*) items; Non-standard.
*Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng
Information in this catalog is subject to change without notice.