FUJI EDS10-45E

MOVING COIL TYPE INDICATOR
DATA SHEET
PAJ
The moving coil type indicator receives various types of
input signals, DC voltage, DC current, and signals from
thermocouple and resistance bulbs.
FEATURES
1. Input signal can be directly connected.
2. High input impedance (1MΩ at 1 to 5V DC input). No
adjusting resistor is required for thermocouple input.
3. Compact and lightweight design.
SPECIFICATIONS
Input resistance and allowable input signal source resistance:
Input
resistance
Allowable input signal
source resistance
1 to 5V
1MΩ
2kΩ or less
Others
40kΩ to 1MΩ
100Ω or less
4 to 20mA
5Ω
–
Others
1 to 100Ω
–
Input signal
Measuring input: Signal
A; 1 to 5V DC
B; 4 to 20mA DC
DC current
D; 0 to 100µA ....... 200mA DC
DC voltage
E; 0 to 10mV ........ 5V DC
Thermocouple
F; 0 to 10mV DC or more
(with reference junction compensator)
(Upper limit overshoot burnout
circuit is available on request)
Thermocouple
G; 10mV DC span or more
(with reference junction compensator)
(Upper limit overshoot burnout
circuit is available on request)
Resistance bulb
H; JPt100Ω
50°C span or more 3-wire type
W; Pt100Ω
50°C span or more 3-wire type
Slide resistant (resistance span; 100Ω)
J; 3-wire type
K; 2-wire type (with adjusting resistor)
Opening angle transmitter (resistance
span: 100Ω)
L; 3-wire type
(zero point and span adjustable)
Voltage
Current
Thermocouple
Resistance
100Ω or less
40 to 600kΩ
–
(50Ω or less when burnout
is provided)
6Ω or less per wire
± 1.0% of full scale
100mm
Within 3 sec (time required for stabilizing
the indicator pointer within ± 1.5%
after applying input corresponding
to 2/3 (approx.) of full scale)
Power supply:
100V ± 10% AC, 50/60Hz
*200V ± 10% AC, 50/60Hz, or
+15%
*24V –10% AC, 50/60Hz
Power consumption:
Approx. 1.8VA max.
Ambient temperature:
–10 to +50°C
Ambient humidity:
30 to 90%RH
Enclosure:
Steel case
External dimensions (H × W × D):
48 × 144 × 210mm
Mass{weight}:
Approx. 1.5kg
Finish color:
Munsell 7.5BG 3.2/0.8 or equivalent
Scope of delivery:Indicator and mounting bracket
Allowance:
Scale length:
Response time:
EDS10-45e
Date
July. 15, 1993
PAJ
Mounting method:
Panel flush mounting
Standard;
Horizontal mounting on vertical panel
Non-standard;
Tilted mounting, vertical mounting
Specify ∠α when mounting on tilted
panel
OUTLINE DIAGRAM (Unit : mm)
CODE SYMBOLS
1 2 3 4 5 6 7 8
P AJ
0
9
Description
2
Input signal
1 to 5V DC
4 to 20mA DC
0 to 100µA.....200mA DC
0 to 10mV.....5V DC
0 to 10mV DC or more
(with reference junction compensator)
10mV DC span or more
(with reference junction compensator)
Resistance bulb JPt100Ω
50°C span or more, 3-wire type
Resistance bulb Pt100Ω
50°C span or more, 3-wire type
Slide resistant, 3-wire type
Slide resistant, 2-wire type
(with circuit adjusting resistor)
For opening angle indicator
(resistance; 100Ω, zero point and span
adjustable)
Other inputs
A
B
D
E
F
G
H
W
J
K
L
*Z
CONNECTION DIAGRAMS
Power supply
24V AC 50/60Hz
100V AC 50/60Hz
200V AC 50/60Hz
*1
3
*5
Application
For general use
For zener barrier connection
Specify "3" only when connecting to
zener barrier (PWZB3 or PWZB4) with
thermocouple or resistance bulb input
used.
0
3
1
2
Mounting method
Horizontal mounting
Vertical mounting
Notes: Symbols of resistance bulbs are as follows.
JPt100 Previous JIS standard
Pt100 New JIS standard
ORDERING INFORMATION
1.
2.
3.
4.
Product name
Code symbols
Input
Scale
5. Power supply
6. Application
7. Other necessary information
Caution on Safety
• Asterisked (*) items; Non-standard
*Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng