ETC 2108I

2108i
MODEL
EUROTHERM
ON TROL S
Model 2108i
1/8 DIN (96 x 48mm)
Indicator and Alarm Unit
Ideal for:
• Temperature indication
• Process indication
• Over-temperature
protection
• Process value trip unit
The 2108i provides accurate measurement and display of
temperature and other process variables in a standard 1/8 DIN panel
format. Two alarm relay outputs are provided for operator alert and
process protection.
A universal input allows selection of nine internally stored
thermocouple types. Other input linearisations may be factory
downloaded. Linear, process inputs are scaleable to the desired
display range. Keylock and passwords prevent unauthorised access to
configuration and alarm settings.
A large, red or green LED display gives good visibility in low and
high ambient lighting.
New alarms can be acknowledged and old alarms reset by a front
panel button or an external contact closure input.
Three internal alarm setpoints are provided. They are configurable
as high, or low, or rate of change alarms. Any combination of the
three internal alarms can activate the two relay outputs. Latching or
non-latching operation can be selected and alarm delays can be
applied. A special mode known as ‘alarm blocking’ is available. In
this mode, after power on, the process value must first enter a good
state before the alarm becomes active. This is particularly useful for
low alarms which can be ‘blocked’ while the process is warming-up.
Calibration accuracy is preserved with a self-correcting input circuit,
while a User Calibration facility allows offsets to be applied to
compensate for sensor and other system errors.
Features:
• Universal input
Covers a wide range of temperature and process inputs
• Two alarm relay outputs
For operator alert and plant protection
• Wide-range 85-264V ac,
supply
Can be installed world-wide
• Plug-in from front
Rapid replacement - reducing downtime
• Front panel configuration Easy, on-site set-up
• Tactile buttons
Friendly operation
• Three year warranty
Low ownership cost
Indicator
and Alarm
Unit
Dimensions
Specifications
ε
48 mm
(1.89in)
E
U
R
O
T
H
E
R
M
Display:
4 digit, red or green. 15.9mm high characters
103 mm (4.01in)
96 mm (3.78in)
Supply voltage:
100 to 240Vac -15%, +10%
2108i
OP1
Operating ambients:
0-55°C, 5-95%RH non-condensing
OP2
ACK/
RESET
Panel sealing:
IP54, plug-in from front panel
Relay ratings:
2A, 240Vac resistive
Panel cut-out
- 0.0
+ 0.6
45
- 0.0
+ 0.8
x 92
1.77
Calibration accuracy:
±0.25% of reading, or ±1°C, or ±1LSD
whichever is the greater
mm
-0
+ 0.03
-0
+ 0.02
x 3.62
in
Cold junction compensation:
>15 to 1 rejection of ambient temperature
change
Electrical connections
Input filtering:
Off-999.9 seconds
Relay Output 2
3C
EMC standards:
EN50081-2 and EN50082-2 generic standards
for industrial environments
Model 2108i
3A
3B
Safety standard:
Meets EN61010, installation category II.
Pollution degree 2
Relay Output 1
Sensor inputs
V-
V+
VI
AC
AB
AA
LC
LB
LA
Atmospheres:
Not suitable for use above 2000m or in
explosive or corrosive atmospheres
L
N
Neutral
Ground
Line
85-264Vac
T/C
PT100
Link for
keylock
Alarm
acknowledge/reset
mA Input
2.49Ω
Ordering Code
Model
Number
Function
2108i
AL
Function
AL Alarm unit
Display Colour
GN
RD
Green
Red
Supply Voltage
VH 85-264Vac
Display
Colour
Relay Output 1
RF Unconfigured
Non-latched alarm
FH High alarm 1
FL Low alarm 1
AL High alarm 1 & low
alarm 3
RA Rate of change
alarm 1
Latched alarm
HA High alarm 1
LA Low alarm 1
AA High alarm 1 & low
alarm 3
RT Rate of change
alarm 1
NW New alarm status
Supply
Voltage
Relay
Output 1
Relay
Output 2
Manual
VH
Relay Output 2
RF Unconfigured
Non-latched alarm
FH High alarm 2
FL Low alarm 2
RA Rate of change
alarm 2
Latched alarm
HA High alarm 2
LA Low alarm 2
RT Rate of change
alarm 2
NW New alarm status
Manual
XX
ENG
FRA
GER
NED
SPA
SWE
ITA
No manual
English
French
German
Dutch
Spanish
Swedish
Italian
Sensor Input
Sensor
Input
Setpoint
Min.
Setpoint
Max.
note 1
note 1
Input
Adaptor
Units
Setpoint Min/Max
Standard Sensor Inputs
Min °C Max
J
J Thermocouple
-210 to 1200
K
K Thermocouple
-200 to 1372
T
T Thermocouple
-200 to 400
L
L Thermocouple
-200 to 900
N
N Thermocouple-Nicrosil/Nisil
-200 to 1300
R
R Thermocouple-Pt/Pt13%Rh
-50 to 1768
S
S Thermocouple-Pt/Pt10%Rh
-50 to 1768
B
B Thermocouple-Pt/Pt30%Rh-6%Rh
0 to 1820
P
Platinell II Thermocouple
0 to 1369
Z
RTD/PT100 DIN 43760
-200 to 850
Factory Downloaded Input
Min °C Max
C
C Thermocouple - W5%Re/W26%Re(Hoskins)
0 to 2319
D
D Thermocouple - W3%Re/W25%Re
0 to 2399
E
E Thermocouple
-200 to 1000
1
Ni/Ni18%Mo Thermocouple
0 to 1399
2
Pt20%Rh/Pt40%Rh Thermocouple
0 to 1870
3
W/W26%Re (Engelhard) Thermocouple
0 to 2000
4
W/W26%Re (Hoskins) Thermocouple
0 to 2010
5
W5%Re/W26%Re (Engelhard) Thermocouple
10 to 2300
6
W5%Re/W26%Re (Bucose) Thermocouple
0 to 2000
7
Pt10%Rh/Pt40%Rh Thermocouple
200 to 1800
8
Exergen K80 I.R. pyrometer
-45 to 650
Process Inputs (Scaled to setpoint min and max)
Min
Max
M
-9.99 to +80mV linear
-1999 to 9999
Y
0 to 20mA linear (note 2)
-1999 to 9999
A
4 to 20mA linear (note 2)
-1999 to 9999
V
0 to 10Vdc linear (Input Adapter required)
-1999 to 9999
Units
C
F
K
X
°C
°F
Kelvin
Linear input
Input Adaptor
XX Not fitted
V1 0-10Vdc
A1 0-20mA current sense
resistor (2.49Ω 0.1%)
Note 1. Setpoint min and
max: include the decimal
points required in the display
value; up to one for
temperature inputs, up to
two for process inputs.
Note 2. A 1% 2.49Ω current
sense resistor is supplied as
standard. If greater accuracy
is required a 0.1% resistor
can be specified in the input
adaptor field.
EUROTHERM CONTROLS LIMITED
http://www.eurotherm.co.uk
UK SALES OFFICE
Eurotherm Controls Limited. Faraday Close, Durrington, Worthing, West Sussex BN13 3PL
Telephone Sales: (01903) 695888; Technical: (01903) 695777; Service: (01903) 695444
Fax (01903) 695666
Sales and support in over 30 countries worldwide
Enquiries/orders to: Eurotherm Controls Limited Export Dept., Faraday Close, Durrington, Worthing. West Sussex BN13 3PL.
Telephone (01903) 268500
Fax (01903) 265982
© Copyright Eurotherm Controls Limited 1998
All rights strictly reserved. No part of this document may be stored in a retrieval system, or any form or by any means without prior written permission from Eurotherm Controls Limited. Every effort has been
taken to ensure the accuracy of this specification. However in order to maintain our technological lead we are continuously improving our products which could, without notice, result in amendments or
omissions to this specification. We cannot accept responsibility for damage, injury, loss or expenses resulting therefrom.
Part No. HA026159 Issue 3