MPSIND MPU15B-102

MPU15B SERIES
15W Desk Top Switching Power Supplies For Medical Equipment.
Description:
The MPU15B series of AC/DC switching mode power supplies
provide 15 Watts of continuous output power and is well-suited for
a variety of applications. All supplies are UL 94V-1 min compliant.
All models comply with UL/c-UL(UL 60601-1:2ndEdition), TUV/Tmark(EN 60601-1:2ndEdition) and new CE requirements. All units
are 100% burned-in and tested.
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C8 Input Inlet
Single Output
Energy Star 2.0, Efficiency level V
Over Voltage and Over Load protection.
Over temperature Detection
Class II
3 year warranty
Safety Approvals:
EN 60601-1
IEC 60601-1
Electrical Characteristics:
Sym.
Parameter
Vin
Safety Approvals Input
Voltage Range
Operate Voltage Range
fin
Input Frequency
Po
Output Power Range
Vo
Output Voltage Range
CB
Agree to apply for the PSE
if order on hand
Test Conditions
Vin=90 to 264VAC
Min. Typ. Max.
Unit
100
240
VAC
90
264
VAC
47
63
Hz
0
15
W
See rating chart
V
See rating chart
A
Io
Output Current Range
Iil
Input Current (Low Line)
Io=Full load, Vin=115VAC
0.25 0.33
A
Iih
Input Current (High Line)
Io=Full load, Vin=230VAC
0.17 0.18
A
Irl
Low Line Inrush Current
Io=Full load, 25c, Cool start,
Vin=115VAC
18
23
A
Irh
High Line Inrush Current
Io=Full load, 25c, Cool start,
Vin=230VAC
38
45
A
Eff
Efficiency
Io=Full Load, Vin=230VAC
85
%
1
%
REG-i Line Regulation
Io=Full Load
REG-o Load Regulation
Vin=230VAC
73
0.5
5
%
OVP
1
Over Voltage○
Protection
112
132
%
OCP
1
Over Current○
Protection
110
150
%
Ttr
Time of Transient Response
4
mS
Thold Hold-Up Time
Ts
Vp-p
Start Up Time
Io=Full Load, Vin=110VAC
10
Io=Full Load, Vin=100VAC
Ripple & Noise (Peak to Peak) Full Load, Vin=90VAC
TC
Temperature Coefficient
All output
No-Load Power Consumption
No load, Vin=230VAC
The parameter is not subject
to production test-verified
by design/characterization
of integrated controller.
1
Thermal Shutdown○
by
Junction Temperature
Controller
3
Io=Full Load to Half Load, Vin=100VAC
Pno
Tjsd
1
16
1
S
0.5
1
%
0.04
%/c
0.3
W
125
c
-0.04
0
mS
0.5
0.25
-20
1 When the power system interruption is isolated, the product would re-start after recovering by hand.
○
Environmental :
Sym.
Parameter
T e s t Conditions
Min. Typ. Max.
Unit
50
70
C
Toper
Operating Temperature
Tstg
Storage Temperature
-40
85
C
Ho
Operating Humidity
0
95
%
Hr
Storage Humidity
0
75
%
MTBF
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Pd
Derate linearly from 100% load at 50 C to 50% load at 70 C
0
0.1M
Hrs
MPU15B SERIES
15W Desk Top Switching Power Supplies For Medical Equipment.
Safety Specifications:
Sym.
Vps
Parameter
Min. Typ. Max.
Unit
5656
VDC
Vin=230VAC,50HZ
B
CLASS
Vin=120VAC,60HZ
B
CLASS
T e s t Conditions
Dielectric Withstanding Voltage
for Primary to secondary
E N 5 5 0 2 2 EMI requirements for EN55022
EMI requirements for FCC PART-18
FCC
Primary to secondary
Output Voltage And Current Rating Chart ( Single Output ) :
①
*
*
*
*
Model Number
Output Voltage
Output Current
Total
Regulation
Maximum
Output Power
MPU15B-102
5 ~ 6 VDC
2.60~2.16 A
5%
13W
6 ~ 8 VDC
2.16~1.62 A
5%
13W
MPU15B-104
8 ~ 11 VDC
1.87~1.36 A
5%
15W
MPU15B-105
11 ~ 13 VDC
1.36~1.15 A
5%
15W
MPU15B-106
13 ~ 16 VDC
1.15~0.93 A
5%
15W
MPU15B-107
MPU15B-103
16 ~ 21 VDC
0.93~0.71 A
5%
15W
MPU15B-108
21 ~ 27 VDC
0.71~0.55 A
3%
15W
MPU15B-109
27 ~ 33 VDC
0.55~0.45 A
3%
15W
MPU15B-110
33 ~ 36 VDC
0.45~0.41 A
3%
15W
Mark " * " means " PSE approval "
①
MPU15B-102.103 are required to use AWG#16 / 4FT output cable.
MPU15B-105~110 are required to use AWG#18 / 4FT output cable.
The regulation and efficiency will be changed by modified output cable.
Mechanical Specifications :
Note:
1. Dimensions are shown in mm.
2. Weight: 170gs approx.
3. Optional output connector:
See page Appendix.