MPSIND MPU60A-108

MPU60A SERIES
63W Desk Top Switching Power Supplies For Medical Equipment
Description:
The MPU60A series of AC/DC switching mode power supplies
provide 63 Watts of continuous output power . All supplies are
UL94V-1 min compliant, include IEC-320-C14 input for worldwide
applications. They are suited for use in hospital instrument and
many other applications. All models meet FCC Part-18 class B and
CISPR-11 EN55011 class B emission Limits and are designed to
comply with UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1), and
new CE requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single Output
Output Voltage Available From 12VDC Thru 36VDC
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Output Voltage Protection(Crowbar Desing)
CEC level V and Engergy Star 2.0 Compliance
Class I
3 year warranty
Safety Approvals :
EN 60601-1
IEC 60601-1
Electrical Characteristics:
Sym.
*
Parameter
Test Conditions
Vin
fin
Po
Vo
Io
Input Voltage
Input Frequency
Output Power Range
Output Voltage Range
Output Current Range
Vin=90 to 264VAC
Iil
Iih
Input Current (Low Line)
Input Current ( High Line)
Io=Full load, Vin=115 VAC
Io=Full load, Vin=230 VAC
I rl
Low Line Inrush Current
I rh
High Line Inrush Current
Io=Full load, 25 C , Cool start,
Vin= 115VAC
Io=Full load, 25 C , Cool start,
Vin= 230VAC
Io=Full Load, Vin=230VAC
Operating Voltage
Min. Typ. Max. Unit
264
63
90
47
63
0
See rating chart
See rating chart
1.62
0.72
A
A
26
30
A
43
47
A
92
%
%
%
Eff
REG-i
REG -o
Efficiency
OVP
112
Over Current Protection
110
150
Time of Transient Response Io=Full Load to Half Load, Vin=100VAC
4
Hold-Up Time
Io=Full Load, Vin=110VAC
16
Io=Full Load, Vi n= 10 0V AC
Start Up Time
0.3
1
2
Ripple & Noise (Peak to Peak) Full Load, Vin=100VAC
1
Safety Ground Leakage C urrent Io = Fu ll Load, Vin=240VAC
0.1
No-Load Power Consumption
No load,Vin=240VAC
0.3 0.4 0.5
Temperature Coefficient
-0.04
All output
0.04
OCP
Ttr
Thold
Ts
Vrn
Ilk
Pno
TC
Line Regulation
Load Regulation
Over Voltage Protection
VAC
Hz
W
V
A
87
Io=Full Load
Vin=230VAC
1
5
132
%
%
mS
mS
S
%
mA
W
%/ C
* Note: The specifics for testing the energy efficiency of MPU60 Series are outlined in a separate document titled "Test Method for Calculating the
Energy Efficiency of Single-Voltage External Ac-Dc and Ac-Ac PowerSupplies (August 11, 2004)," which is available on the ENERGY STAR Website.
Environmental :
Sym.
Toper
Tstg
Hr
MTBF
Pd
Parameter
Test Conditions
Operating Temperature
Storage Temperature
Relative Humidity
Operating Temperature at 25 C, Calculated per MIL-HDBK-217F
Derate linearly from 100% load at 50 C to 50% load at 70 C
- 17 -
Min. Typ. Max.
Unit
50
C
C
%
Hrs
0
-40
5
0.1M
70
85
95
MPU60A SERIES
63W Desk Top Switching Power Supplies For Medical Equipment
Safety Specifications:
Sym.
Parameter
Test Conditions
Vps Dielectric Withstanding Voltage
for Primary to secondary
Dielectric Withstanding Voltage
Vpg for
Primary to Ground
Ris Isolation Resistance
CISPR EMI requirements for CISPR-11
FCC EMI requirements for FCC PART-18
Min. Typ. Max.
Unit
Primary to secondary
5656
VDC
Primary to ground
2800
VDC
50
B
B
M
CLASS
CLASS
Test Voltage = 2100VDC
Vin=220VAC
Vin=110VAC
Output Voltage And Current Rating Chart (Single Output) :
Model Number
*
*
Output Voltage
MPU60A-105
MPU60A-106
MPU60A-107
MPU60A-108
MPU60A-109
MPU60A-110
12
15
18
24
30
36
VDC
VDC
VDC
VDC
VDC
VDC
Output Current
5.25
4.20
3.50
2.62
2.10
1.75
Total Regulation
A
A
A
A
A
A
5%
5%
5%
3%
3%
3%
M aximum
Output Power
63W
63W
63W
63W
63W
63W
Mark " * " means " PSE approval "
The total regulation on model 105~107 is required to use AWG#16/2C/4FT output cable.
The total regulation on model 108~110 is required to use AWG#18/2C/6FT output cable.
The regulation and efficiency will be changed by modified output cable.
Note:
1. Dimensions are shown in inches or mm.
2. Weight: 330-380gs approx.
3. Optional output connector:
See page Appendix.
Mechanical Specifications:
- 18 -