SLPOWER XPFM200B

XPFM200A+ Medical Power Supply
FEATURES:
•
•
•
•
•
Medical version of ATX200 power supply
Approved to IEC601-1 and UL2601
Includes power factor correction to IEC1000-3-2
Conducted emissions to CISPR11 class B/IEC601-1-2
marked to LVD
Remote Sense
Provided as a standard feature on V2 (+) lead of all
models. Capable of compensating for 0.25 V total of
cabling losses in output voltages.
SPECIFICATIONS:
Ac Input
90-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at minimum input voltage and full
load is 3.5 A. Active Power Factor Correction circuitry
assures compliance with IEC1000-3-2, Class A.
Inrush Current
Inrush is limited by internal thermistors. The inrush at 240
Vac, averaged over the first ac half-cycle under cold start
conditions will not exceed 37 A. Worst case current under
hot start conditions will be 60 Amps.
Input Protection
Internal ac fuse provided on all units. The fuse is designed to blow only if a catastrophic failure occurs in the
unit. The fuse does not blow on unsustained overload or
short circuit.
Efficiency
75% typical at full rated load, nominal input voltage,
depending on load distribution.
Output Power
Continuous output power 200 W.
Overload Protection
Fully protected against short circuit and output overload.
Short circuit protection is cycling type power limit on V1, 2
& 3; thermal foldback type on V4, 5, 6. Power limit factory
set to cycle outputs off at 240 W typical. Recovery after
removal of fault is automatic.
Over Voltage Protection
Built in on V1, 2 and 6 with firing point set per Table 1.
OVP firing shuts down power supply.
Voltage Adjustment
Factory set to specified voltage. Potentiometer on V1 not
considered user accessible.
No Load Turn-On/Standby
A minimum load of 1 A on V2 is required for proper
regulation. If not met, no degradation of reliability will
occur.
Output Regulation
Regulation for all outputs is the maximum deviation from
initial set point under all line and load conditions. Initial
set tolerance is measured with all outputs at 50 % of full
rated load.
110
Output Noise
0.5% RMS, 1% Pk-Pk, 20 MHz Bandwidth, differential
mode. Measured using a differential noise probe. Probe
should be placed directly across the power supply output
terminals terminated with a 1uF low Z capacitor.
Transient Response
All outputs stay within their specified regulation limits with
a 20% load change. ∆i / ∆t < 0.2 A / µs.
Hold-Up Time
25 ms total from loss of ac input at full load until loss of
regulation.
Inhibit
Inhibit signal is pulled to the dc output common to inhibit
V1 - 5 voltages. (V = < 0.4V, sink > 1.5mA)
Power Good
TTL / CMOS compatible output rises high 100 to 400 ms
after V2 reaches regulation and should assure that
sufficient energy is stored in the input section to provide
normal power fail / shutdown.
Temperature Coefficient
0.03% / °C typical on all outputs.
Turn-On Time
Less than 2 s at 120 Vac 25°C
Storage
-40 to +85 °C.
Leakage Current
90µA under normal conditions (120 Vac @ 60 Hz).
Maximum under single-fault conditions (264 Vac @ 50
Hz) is 300 µA.
EMI/EMC Compliance
All models include built-in EMI filtering to meet the
following emissions requirements:
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients / Bursts
Surge Susceptibility
Conducted RF Susceptibility
Voltage Sags & Surges
EN55011, Class B; FCC Class B
EN61000-4-2, 6 kV contact 8 kV air
EN61000-4-3, 3V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-4-6, 3V
EN61000-4-11
Safety Agency
All models are certified to be in compliance with the
applicable requirements of UL2601-1, CSA-C22.2 No.
601.1, IEC601-1/60601-1.
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
Model
Output No.
Output
Minimum
Current
XPFM200A+
XPFM200B+
Maximum
Current
Peak (E)
Current
Total Regulation
OVP
Trip
Ripple and
Noise
Notes
D
1(C)
+3.3 V
0.1 A
14 A
16 A
± 4%
4 V± 0.3 V
2%
2 (C)
+5.1 V
1A
22 A
28 A
± 4%
6.2 V 0.5 V
1%
3 (A)
+12 V
0.1 A
7A
9A
± 5%
1%
4 (B)
-12 V
0A
0.8 A
0.8 A
± 8%
1%
5
-5 V
0A
0.5 A
0.5 A
± 6%
1%
6 (E)
+5 VSB
0A
1.5 A
--
± 4%
6.2 V ±0.8 V
2%
1 (C)
+2.7 V
0.1 A
14 A
16 A
± 4%
4 V± 0.3 V
2%
2 (C)
+5 V
1A
22 A
28 A
± 4%
6.2 V 0.5 V
1%
3 (A)
+12 V
0.1 A
7A
9A
± 5%
1%
4 (B)
-12 V
0A
0.8 A
0.8 A
± 8%
1%
5
-5 V
0A
0.5 A
0.5 A
± 6%
1%
6 (E)
+5 VSB
0A
1.5 A
--
± 4%
6.2 V ±0.8 V
D
2%
A. To maintain regulation on V3, the +5.1 V current must be at least 1/5 of V3 and not greater than 5 times the V3 current. Required +5 V to be adjusted
to within ±1% with at least a 1 A load to maintain regulation on this output.
B. Combined current of V4, 5 must not exceed 0.8 A.
C. Combined load for V1, 2 must not exceed 125 W.
D. 200 Series has external fan, for internal fan replace 200 in model number with 201.
E. V6 maximum current is 1.25 A at initial start up.
XPFM200+ MECHANICAL SPECIFICATIONS:
XPFM200
External Fan
Environmental
Specification
Temperature (A)
Humidity (A)
Shock (B)
Altitude
Vibration (C)
Operating
XPFM201
Internal Fan
Non-operating
0 to 40ºC
-40 to +85ºC
0 to 95% RH
0 to 95% RH
20 gpk
40 gpk
-500 to 10,000 ft
-500 to 40,000 ft
1.5 grms, 0.003 g2/Hz
5 grms, 0.026
g2/Hz
A. Units should be allowed to warm up/operate under non-condensing conditions
before application of power. Cooling provided by internal fan— heatsink
temperatures should not exceed 90ºC for extended periods in the installation.
B. Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
C. Shock testing—half-sinusoidal, 10 ± 3 ms duration, ± direction, 3 orthogonal axes,
total 6 shocks.
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com
111