Catalog: 1654001 Issue Date: 06.2011 Corcom Product Guide High Performance Power Line Filter for Medical Applications HZ Series UL Recognized CSA Certified VDE Approved 1 RFI Power Line Filters HZ Series Specifications • Designed to provide good attenuation to RFI noise in the frequency range from 10kHz to 30MHz • Size and cost-effective Maximum leakage current each Line to Ground: @ 120 VAC 60 Hz: 2 A @250 VAC 50 Hz: 5 A Hipot rating (one minute): Line to Ground: Line to Line: • Low leakage current • New versions up to 30A Ordering Information 3 E HZ 1 2250 VDC 1450 VDC Rated Voltage (max): 250 VAC Operating Frequency: 50/60 Hz Rated Current: Input / Output Style 1 – .250 [6.3] spade terminals 6 – Threaded bolt HZ Series Leakage current designation E – Low leakage (<0.5 mA) Current Rating 3, 4, 6, 10, 15, 20 or 30A 3 to 30A Operating Ambient Temperature Range (at rated current Ir): -10°C to +40°C In an ambient temperature (Ta) higher than +40°C the maximum operating current (Io) is calculated as follows: Io = Ir (85-Ta)/45 Case Styles 3EHZ1 Available Part Numbers B 3EHZ1 6EHZ1 15EHZ1 30EHZ6 4EHZ1 10EHZ1 20EHZ1 D E A C Electrical Schematic LINE LOAD 43 Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change. For email, phone or live chat, please go to te.com/help corcom.com Catalog: 1654001 Issue Date: 06.2011 Corcom Product Guide High Performance Power Line Filter for Medical Applications (continued) HZ Series Case Styles (continued) 30EHZ6 4EHZ1 B B D E A C D E A C 6EHZ1 B Case Dimensions C D E A Part No. 3EHZ1 10, 15 & 20EHZ1 4EHZ1 6EHZ1 10EHZ1 15EHZ1 20EHZ1 B 30EHZ6 C A B C D E D E A 44 Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change. For email, phone or live chat, please go to te.com/help corcom.com Catalog: 1654001 Issue Date: 06.2011 Corcom Product Guide High Performance Power Line Filter for Medical Applications (continued) Performance Data Typical Insertion Loss Measured in closed 50 Ohm system 3EHZ1 4EHZ1 6EHZ1 db 100 db 100 db 100 90 90 90 80 80 80 70 70 70 60 60 60 50 50 50 40 40 40 30 30 30 20 20 20 10 10 0 0 0.1 1 10 30 Frequency in MHz 10 0.01 0.1 1 0 10 30 Frequency in MHz 0.01 0.1 10EHZ1 15EHZ1 20EHZ1 db 100 db 100 db 100 90 90 90 80 80 80 70 70 70 60 60 60 50 50 50 40 40 40 30 30 30 20 20 20 10 10 0 0 0.01 0.1 1 10 30 Frequency in MHz RFI Power Line Filters 0.01 1 1 10 30 Frequency in MHz 1 10 30 Frequency in MHz 10 0.01 0.1 1 0 10 30 Frequency in MHz 0.01 0.1 30EHZ6 db 100 90 80 70 Common Mode / Asymmetrical (L-G) Differential Mode / Symmetrical (L-L) 60 50 40 30 20 10 0 0.01 0.1 1 10 30 Frequency in MHz Minimum Insertion Loss Common Mode / Asymmetrical (Line to Ground) Differential Mode / Symmetrical (Line to Line) 45 Dimensions are in inches and millimeters unless otherwise specified. Values in italics are metric equivalents. Dimensions are shown for reference purposes only. Specifications subject to change. For email, phone or live chat, please go to te.com/help corcom.com