FUJI EDS3-124C

IN-SITU
ZIRCONIA OXYGEN ANALYZER
ZSB
DATA SHEET
This oxygen analyzer is used to continuously measure
oxygen concentration in combustion exhaust gas of
industrial boilers or furnaces, and is ideally suited for
combustion moni-toring and control.
The analyzer consists of a flow guide tube with a
detector, which is directly inserted into a flue, etc. to
introduce the gas to be analyzed, a transducer, which
performs control, signal processing, display, input to
and output from outside, and communication, and an
auto calibration/blow unit. The detector of a unit sensor
structure facilitates on-site maintenance. The transducer
provided with an unconventional sensor diagnostic
function and sensor renewal function ensures long-term
stable detecting operation.
FEATURES
1.No need for gas sampling devices
Since the sensor unit is directly inserted into a flue,
etc., gas-sampling devices such as gas aspirator and
dehumidifier are not required, which ensures high
response speed.
2.Easy maintenance
The sensor in a unit structure mounted to the detector can be replaced easily. Since the detector and
the flow guide tube are installed separately, the filter
at the tip of the detector can be replaced easily, and
detector and the flow guide tube can be maintained
separately depending on the degree of formation of
rust, for example.
3.High reliability ensured by the sensor diagnostic and
sensor renewal functions
Depending on the components in the gas to be
measured, the sensor electrode may degrade. The
transducer diagnoses the degree of degradation of
the sensor and renews it electrically, thus ensuring
high reliability and long-term stable measurement.
4.Improved safety
Detecting a break of the thermocouple for heater control installed at the sensor, power supply to the detector is stopped. Furthermore, external contact input
stops power supply to the detector in an emergency.
Those functions along with the key lock function are
equipped as standard to ensure improved safety.
5.Simple operation
The transducer can be operated and various settings
can be made using an interactive system, and display
language is available in English, Japanese, and Chinese.
SPECIFICATIONS
General
Measuring object: Oxygen in noncombustible gas
Measuring method:
Directly insert type zirconia system
Measuring range:
Repeatability:
0 to 2 ···Setting range at option 2 in
50 vol% O2 (in 1 vol% O2
steps)
Within ±0.5%FS
Linearity:
Within ±2%FS
Response time:
Within 4 to 7 sec, for 90% (from calibration gas inlet)
More than 10 min.
Warmup time:
Analog output:
Power supply:
4 to 20mA DC (allowable load resistance 500Ω or less) or 0 to 1V DC (output resistance more
100Ω or more)
Rated voltage;
100 to 120V AC (operating voltage
90 to 132V AC)
Rated frequency; 50/60Hz
EDS3-124c
Date
Apr 1, 2011
ZSB
Power consumption:
Material:
Maximum 240VA (Detector: approx.
200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx.
50VA, Converter: approx. 20VA)
Steel plates (main frame)
Mass:
Approx. 1.6kg
Finish color:
Silver and SUS metallic color
Calibration gas flow:
1.5 to 2 L/min
Blowdown air inlet pressure:
200 to 300kPa
Outer dimensions (H x W x D):
Mass:
Finish color:
Wall-mounting type: 700 X 400 X 180mm
Self-standing type: 1500 X 530 X 550mm
Wall-mounting type:
Approx. 27 kg (excluding cable,
detector and standard gas)
Self-standing type:
Approx. 48 kg (excluding cable,
detector and standard gas)
Munsell 5Y7/1, semi-gloss
Mounting method: Wall-mounting or self-standing
Detector (ZFK8)
Measured gas temperature:
Flow guide tube system; −20 to +600ºC (for general-use,
corrosive gas)
Measured gas pressure:
−3 to +3kPa
Flow guide tube:
• With or without blowdown nozzle
Flange; JIS5K 65A FF
Insertion length; 0.3, 0.5, 0.75, 1m
• For high particulate gas (with blowdown nozzle)
With or without cover
Flange; JIS5K 80A FF
Insertion length; 0.3, 0.5, 0.75, 1m
Operating temperature:
−10 to +60ºC for Primary detecting
element −5 to +100ºC for detector
125ºC or less at detector flange surface with power applied
Storage temperature:
−20 to +70ºC
Structure:
Dust/rain-proof structure (IEC IP66
equivalent, except the filter part at
the tip)
Use a heat insulation cover in cold
climates (as specified)
Filter:
Alumina (filtering accuracy 50µm) and
quartz paper
Main materials of gas-contacting parts:
Zirconia, SCS14 (SUS316 equivalent),
platinum, SUS304
Calibration gas inlet:
φ6mm tube join
Reference air inlet: φ6mm tube join
Detector mounting:
Horizontal plane ±45º, ambient air
should be clean.
Outer dimensions (L × max. dia.):
210mm × 100mm
2
Converter (ZKM3)
Concentration value indication:
Digital indication in 4 digits
Contact output signal:
(1)Contact specification; 6 points, 1a 250V AC/3A or 30V DC/3A
(2)Contact function;
•Under maintenance
•Instrument anomalies Note1)
•Alarm Note2)
•Zero calibration gas valve
•Span calibration gas valve
•Under blowdown Note3)
Note1) The following Instrument errors (1) Thermocouples break (2) Sensor break (3) Temperature fault (4) Calibration fault (5) Zero/span fault (6) Output error the contact ON
Note2) Alarm selects just one as mentioned
below (1) High (2) Low (3) H/L (4) HH (5)
LL, it turns ON while operating.
Note3) Under blow down, it is available in case of
option, and it turns ON while operating.
Contact input signal:
(1)Contact specification; 3 points (the following option)
ON; 0V (10mA or less), OFF; 5V
(2)Contact function;
•External hold
•Calculation reset
•Heater OFF
•Blow down (option)
•Inhibition of calibration
•Calibration start
•Range change
Calibration method:
Calibration gas:
(a) Manual calibration with key operation
(b) Auto. calibration (option)
Calibration cycle; 00 day 00 hour
to 99 days 23 hours
(c) All calibration
• Settable range
Zero gas; 0.010 to 25.00% O2
Span gas: 0.010 to 50.00% O2
• Recommended calibration gas concentration
Zero gas; 0.25 to 2.0% O2
Span gas; 20.6 to 21.0% O2 (oxygen
concentration in the air)
Blowdown:
A function for blowing out dust with
compressed air that has deposited
in the flow guide tube. Blowdown
can be performed only for a predetermined time and at prede-termined
intervals.
Blowdown cycle; 00 hour 00 minute to 99 hours 59 minutes
Blowdown time; 0 minute 00 second to 0 minutes 999 seconds
Output signal hold: Output signal is held during manual/
auto calibration, blowdown, sensor
recovery proc-essing, warm-up, PID
auto tuning, and while maintenance
mode setting is "available". The hold
function can also be released.
Communication function:
RS232C (MODBUS) standard specification
RS485 (MODBUS) (option)
Combustion efficiency display:
When you select this display, "rich
mode display" will be simultaneously
displayed.
This function calculates and displays
combustion efficiency from oxygen
concentration and measured gas
temperature.
Thermocouple (R) is required for temperature measurement.
Operating temperature:
−20 to +55ºC
Operating humidity:
95% RH or less, non condensing
Storage temperature:
−30 to +70ºC
Storage humidity: 95% RH or less, non condensing
Outer dimensions (H × W × D):
182 × 163.5 × 70.6mm
Mass:
Approx. 2kg (excluding cable and
detector) (IP66)
Mounting method: Mounted flush on panel
3
ZSB
CODE SYMBOLS
1 2 3 4 5 6 7 8
2 -
Z S B
9 10 11 12 13
14 15
Description
Converter mounting (4th digit)
Wall-mounting type (with indication window)
Self-stand type (with indication window)
4th digit 1
2
Sensor, Power supply, Cal. Gas inlet (5th digit)
For general use without protection cover, 100/115 V AC, 50/60 Hz and inlet for φ6 mm tube
For general use with protection cover, 100/115 V AC, 50/60 Hz and inlet for φ6 mm tube
* Protection cover should be selected when ambient temperature is -10°C or lower.
5th digit A
B
Sensor cable + calibration gas tube (6th digit)
None
6 m cable (without conduit) + 6 m Teflon tube (φ6/φ4)
10 m cable (without conduit) + 10 m Teflon tube (φ6/φ4)
15 m cable (without conduit) + 15 m Teflon tube (φ6/φ4)
20 m cable (without conduit) + 20 m Teflon tube (φ6/φ4)
6 m cable (with conduit) + 6 m Teflon tube (φ6/φ4)
10 m cable (with conduit) + 10 m Teflon tube (φ6/φ4)
15 m cable (with conduit) + 15 m Teflon tube (φ6/φ4)
20 m cable (with conduit) + 20 m Teflon tube (φ6/φ4)
6 m cable (4-core cable only. Heater cable should be prepared by customer.) + 6 m Teflon tube (φ6/φ4)
10 m cable (4-core cable only. Heater cable should be prepared by customer.) + 10 m Teflon tube (φ6/φ4)
15 m cable (4-core cable only. Heater cable should be prepared by customer.) + 15 m Teflon tube (φ6/φ4)
20 m cable (4-core cable only. Heater cable should be prepared by customer.) + 20 m Teflon tube (φ6/φ4)
* When exclusive-special conduit and heater cable are not used, the following conduit should be used.
Conduit with outer diameter φ23 and inner diameter φ16 as specified in JIS C8411
For preparing two heater cables, their rating should be 3 A or more.
6th digit Y
A
B
C
D
E
F
G
H
J
K
L
M
Calibration gas unit (7th digit)
Automatic calibration unit + zero gas + filter regulator
Automatic calibration unit + zero gas + air standard gas
Manual calibration unit + zero gas + filter regulator
Manual calibration unit + zero gas + air standard gas
7th digit 1
2
3
4
9th, 10th,
11th digits
0
5
5
5
5
5
5
5
5
6
6
6
6
Z
Y
A
A
A
A
B
B
B
B
D
D
D
D
Z
0
3
5
7
1
3
5
7
1
3
5
7
1
Z
12th digit B
E
13th digit J
E
C
14th digit 1
2
15th digit Y
1
4
Flow guide tube (9th, 10th and 11th digit)
(Flange)
(Application)
None
None
SUS304
For general use
SUS304
For general use
SUS304
For general use
SUS304
For general use
SUS316
For corrosive
SUS316
For corrosive
SUS316
For corrosive
SUS316
For corrosive
SUS316
Without high particulate cover
SUS316
Without high particulate cover
SUS316
Without high particulate cover
SUS316
Without high particulate cover
Others
Others
Output signal (12th digit)
4 to 20mA DC
0 to 1V DC
Language (13th digit)
Japanese
English
Chinese
Communication functions (14th digit)
RS232
RS485
Reference gas (15th digit)
None
φ6mm tube joint
(Length)
None
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
Others
(Note)
With blow-back unit
With blow-back unit
With blow-back unit
With blow-back unit
DEVICE CONFIGURATION
The device to be combined differs according to the conditions of the gas to be measured. Please select the flow guide
tube, referring to the table shown below.
Measured gas
Application
Temperature
Boiler,
refuse incinerator,
sludge incinerator, 600ºC or less
etc.
Gas Flow
5 to 20m/s
Device configuration
Dust
Type at
4th to 7th digits
Note
Flow guide tube
at 9th to 11th digits
Less than
0.2g/Nm3
For boiler. Fuel; gas, oil
ZSB
2–
5A
Less than
1g/Nm3
When moisture content in measured
gas is low.
ZSB
2–
5B
Less than
25g/Nm3
When moisture content in measured
gas is low. With blow-down
ZSB
2–
6D
Note (1) Dust volume is approximate value.
(2) Reference air port specification should be selected when standard gas concentration changes.
(3) For optional specifications, contact the manufacturer separately.
SCOPE OF DELIVERY
1)Standard delivery
1 set x zirconia oxygen analyzer 1 set x instruction manual 1 set x standard accessories
·1 x O-ring (for detector)
·1 x ceramic filter (for detector)
·2 each x fuse (3.15 A, 0. 5 A)
·1 set x inlet (for standard gas connection)
·1 x polyethylene tube (for standard gas connection)
·1 x Toaron tube (for standard gas connection)
·1 set x flow guide tube accessories (bolt, nut, spring washer x 4 pieces each)
2)Items to be prepared separately
·Replacement detector
·Spare (detector)
Spares for 1-year operation 1 x O-ring
2 x ceramic filter
1 x Filter frame Spares for 2-year operation 2 x O-ring
4 x ceramic filter
2 x Filter frame
CAUTIONS
·If combustible gas (CO, H2 etc.) exists in the
measured gas, error will occur due to burning at
the sensor section. The inclusion of corrosive gas
(Si vapor, alkaline metal, P, Pb etc.) will shorten the
life of the sensor.
·When the measured gas temperature is high
(+300ºC or higher), the flange should be separated
from the furnace wall in order to bring the detector
flange surface temperature below the specified
value (+125ºC). The flow guide tube should be
attached in the direction in which the gas flow to
the detector decreases.
·If much dust is contained in the gas, the flow
guide tube should be connected obliquely from
the top. This tube should also be connected in the
direction where gas flow into the detector section
is reduced.
·In application to refuse incineration, automatic
blow-down of the flow guide tube should be avoided (in order to protect the tube from corrosion due
to drain formation). Blow-down should be carried
out manually when indication change becomes
small during stop of the incinerator.
5
ZSB
CONFIGURATION
Rc1/4
(upper side blow port should be selected)
Blow-down, 200 to 300kPa
Reduction valve
(unnecessary when supply air pressure is 200 to 300kPa)
Solenoid valve
(not supplied)
Supply air
15ASGP
tube equivalent
or φ10/φ8mm Teflon
pipe or steel pipe
(not supplied)
Heat insulation cover
Flow guide tube
15ASGP
tube equivalent
or φ10/φ8mm Teflon
pipe or steel pipe
(not supplied)
Detector
(ZFK8)
Gas temperature.
600ºC max.
6
1
3
5
Ref. air
2
4
White
Black
White
Yellow
Blue
Rainproof flexible
conduit.
(Max. 20m)
Red
Gas outlet
Gas temperature.
600ºC max.
Flow guide tube
for high particulate
Cable gland
Gas outlet
T1
Converter (ZKM)
R1
1
2
+ O2 –
Gas temperature.
600ºC max.
3
4
+ TC1 –
O2 input
Flow guide tube for high
particulate with cover
Thermocouple
input
6
AO
7
Thermocouple
for combustion
control
10
11
12
13
TM-1
14
ZERO
GAS
Fault
contact
output
Blow
HEATER
contact Calibrating
output gas contact
output
Alarm
Maintenance
contact
Calibrating gas
contact
output
contact output
output
Contact input
ALARM
+ TC2 –
DI1
DI2
DI3 DICOM
15
17
18
19
16
9
ZV
Analog output
4 to 20mA DC
or 0 to 1V DC
*1
8
–
or
20
SPAN
GAS
MAINTE
COM
21
22
23
24
25
SV
26
AC power
supply
N AC L
27
28
1
φ6mm Teflon tube
for calibration gas
(not supplied)
5
+
2
TM-2
3
Converter
Auto
breaker
R
T
E
1
2
3
4
Maintenance signal during
calibration/maintenance
6
5
7
8
9
10
11
T1
12
R1
Earth
3
AC power
supply
*1 Use a screw for mounting of
converter main unit
TGS
2
When auto calibration
is selected
(1.5L/min)
Zero
Solenoid valve
Joint
Pressure regulator
Zero
O2/N2
Flowmeter
Solenoid valve
(0.2L/min)
Standard gas for
zero point
Span
Reference air
Flowmeter
Container box
Span
Supply air
Reduction valve
(mounted on the
main unit)
6
Pressure regulator
Air
or
O2/N2
Standard gas for
span point
When air standard gas is selected
OUTLINE DIAGRAM
(Unit : mm)
Self-standing type: ZSB2
400
180
100
External
connection
terminal
(converter)
{490}
{350}
Instrumentation air inlet
(for blow-back)
Rc1/2
G1/2
700
Rc1/2
Instrumentation air outlet
(for blow-back)
Regulator + solenoid valve
(selectable by
code symbol)
Anchor plan,
anchor bolts
(option)
M12×160×50
1500
External
connection
terminal
(power supply)
Rc1/4
Instrumentation air outlet
(for span calibration gas)
Filter regulator
(selectable by code symbol)
760
Rc1/4
Instrumentation air inlet
(for span calibration gas)
4-φ15
30
490
530
350
550
Front view
(Front) Cable gland
exclusive for detector
Right side view
EXTERNAL CONNECTION DIAGRAM
(POWER SUPPLY)
1
2
3
R
T
E
4
Power Ground
supply
5
6
7
8
9
10 11 12
Maintenance signal
Calibration status/
maintenance status
Inlet (for φ6 mm tube)
(Rear)
Cable gland A15A
(Box bottom view)
7
ZSB
Wall-mounting type: ZSB2
Instrumentation air inlet
(for blow-back)
Rc1/2
400
255
180
300
External
connection
terminal
(converter)
G1/2
700
Rc1/2
Rc1/4
Instrumentation air outlet
(for span calibration gas)
Filter regulator
(selectable by code symbol)
Rc1/4
Instrumentation air inlet
(for span calibration gas)
Right side view
Front view
(Front)
500
Instrumentation air outlet
(for blow-back)
Regulator + solenoid valve
(selectable by
code symbol)
External
connection
terminal
(power supply)
Rear view
Cable gland
exclusive for detector
EXTERNAL CONNECTION DIAGRAM
(POWER SUPPLY)
1
2
3
R
T
E
4
Inlet (for φ6 mm tube)
(Rear)
Cable gland A15A
(Box bottom view)
8
4-φ15
Power Ground
supply
5
6
7
8
9
Maintenance signal
Calibration status/
maintenance status
10 11 12
Converter: ZKM3
163.5
10
70.6
143.5
Oxygen Analyzer
OXYGEN TRANSM ITTER
Made in Japan
50/60Hz 60VA
Fuji Electric Co.,Ltd.
TYPE
OUTPUT DC
POWER AC
SER. NO.
MFD
172
ESC
ENTER
POWER
注 意
182
MODE
POWER
変換器へ供給する電源電圧は、接続される
検出器の電源電圧に合わせてください。
FUSE T 0.5A
HEATER
CAUTION
FUSE T 2.5A
Power supply must be selected
the same voltage of detector.
POWER
ON
OFF
+ O2 − + TC1 − + AO − FAULT
BLOW ZV HEATER
1
2
3
4
5
6
7
8
9
10 11 12 13 14
15 16 17 18 19 20 21 22 23 24 25 26 27 28
ALARM MAINAC
COM SV
+ TC2 − DI1 DI2 DI3 DIN
L
TENANCE
COM
4-φ4.5
MTG. holes
External terminal (TM1)
O2 sensor
input
O2 sensor
thermocouple
input
Exclusive
cable
Analog
output
4-20mA DC
or 0-1V DC
1
+
O2
2
3
−
+
4
TC1
−
5
+
AO
6
Fault
contact output
contact capacity
250V AC/3A
30V DC/3A
7
8
9
+
15
TC2
11
12
13
14
ZV
FAULT
BLOW
ALARM
Ground
terminal
M4
10
−
or
O2 sensor
heater power
output
Blow contact
output
250V AC/3A
30V DC/3A
−
DI1
DI2
DI3 DICOM
16
17
18
19
22
Alarm
contact output
250V AC/3A
30V DC/3A
Thermocouple
input type R
Contact input
(combustion control) Short circuit : on (Current 5mA)
Open : off (Voltage 5V)
HEATER
SPAN
GAS
MAINTE
21
20
ZERO
GAS
23
COM
24
25
Maintenance
contact output
250V AC/3A
30V DC/3A
SV
26
Calibrating gas
contact output
contact capacity
250V AC/3A
30V DC/3A
N
27
AC
L
28
AC
Power supply
input
(100 to 240V AC)
COMMUNICATION TERMINAL (TM2) /INSERTION TERMINAL
Terminal number
1
2
3
Remarks
RS232C
TXD
RXD
GND
Standard
RS485
TRX+
TRX–
GND
Option
9
ZSB
Detector: ZFK8
Approx.132
φ 80
Approx.62
Approx.130
φ 67
5
6-φ 6
Temp.Proof cover
(To order)
Terminal box
φ 85
Heater
125.5
φ 26
φ 79
EXTERNAL CONNECTION DIAGRAM
Approx.30
°
60
Filter
Thermocouple
1
3
2
4
Black / White/ red / White/
2-core wire
Ground-wire
Screw: M4
Calibration gas inlet
Ref. Air inlet (to order) (To order)
SUS316,for φ 6/φ 4 tube SUS316,for φ 6/φ 4 tube
Exclusive
Cable gland
Sensor unit (ZFK8YY)
(132.7)
20
(?17.5)
3.5
4-φ 3.5(Mount for detector)
5
S-
H
2
4
T-
45°
TK4J4664R0
3
made in japan
40
T+
S+ H
6
1
45°
φ 35
10
5
6
yellow / blue
4-core wire
Exclusive-special cable
Replacement Detector element
(88.4)
Element output
Flow guide tube (for general-use)
φ155
Approx. L
12
Approx. 20
φ130
φ67
φ34
Gas inlet
20
ZSB
6-M5
detector side
Gas outlet
Code 11th
3
L (m)
0.3
Oxygen detector
MASS
2.7
Approx (kg)
5
7
1
0.5 0.75 1.0
3.3
2-5A 3
5
7
1
4.1
4.8
Z
8-φ15
MTG. holes
L=
(to order)
Flow guide tube (for corrosive gas)
φ155
12
Approx. 40
Approx. L
φ130
3
φ67
φ60.5
φ67
Gas inlet
ZSB
6-M5
detctor side
Gas outlet
Oxygen detector
2-5B 3
5
7
1
Code 11th
3
L (m)
0.3
MASS
3.3
Approx (kg)
5
7
1
0.5 0.75 1.0
4.5
6.1
7.6
Z
L=
(to order)
8-φ15
MTG. holes
11
ZSB
45
φ1
4-Rc 1/4
with plug
φ180
Flow guide tube (for high particulate)
Blow down air inlet
At installation, select upper blow port
at one position (as a measure for drain).
8-φ19
MTG. hole
φ67
φ29
.5
Code 11th
3
L (m)
0.3
MASS
4.5
Approx (kg)
Flange: JIS 5K80A FF
5
7
1
0.5 0.75 1.0
5.6
7.0
8.3
Z
L=
(to order)
ZSB MTG. position
Gas outlet
Approx. L
185
Packing
ZSB
50
Oblong
hole
2-6D 3
5
7
1
30
Gas flow
Applicable companion flange height about 160 mm
(Closing gas outlet should be avoided.)
Tube (50A SCH40)
Caution on Safety
*Before using this product, be sure to read its instruction manual in advance.
International Sales Div