Model TJE - Honeywell Test and Measurement Sensors

Model TJE
Precision Gage/Absolute
Pressure Transducer
DESCRIPTION
Model TJE pressure transducers are all-welded stainless steel
The gage Model TJE is a strain gage based transducer and
sensors built for rugged industrial applications that require high
features a unique “true gage” design which utilizes a second
accuracy and measurement stability. The Model TJE is available
welded stainless steel diaphragm that hermetically seals the
with a variety of options for extended temperature operation,
strain gage circuitry from atmospheric contamination. This
electrical terminations and high-level outputs including 5 Vdc or
design references the primary pressure sensing diaphragm to
10 Vdc and 4 mA to 20 mA. Most high-level output models have
the atmosphere, and provides a stable zero regardless of the
internal shunt calibration circuits as a standard feature to allow
transducer environment.
easy set-up of the sensor to the data system. An optional internal
signature calibration chip provides calibration information for
automatic set up with the Model SC four-or-twelve channel digital
indicator.
FEATURES
•
0.1 % accuracy
•
0.0025 % F.S./°F temperature effect
•
1 psig/a to 60000 psig/a range
•
mV/V, 4 mA to 20 mA, 0 Vdc to 5 Vdc, or 0 Vdc to 10 Vdc
output
•
All-welded, stainless steel construction
•
Intrinsically safe available (2N option only)18
•
CE19
The absolute Model TJE has an all-welded vacuum reference
chamber assuring long-term stability.
Model TJE
PERFORMANCE SPECIFICATIONS
Characteristic
Measure
Accuracy
±0.10 % full scale
Linearity
±0.10 % full scale
Hysteresis
±0.05 % full scale
Media
All gases/liquids compatible with wetted
parts
Resolution
Infinite
Calibration
5-point calibration: 0 %, 50 %, and 100
% of full scale
1
MOUNTING DIAGRAM AND CHARACTERISTICS
ENVIRONMENTAL SPECIFICATIONS
Characteristic
Measure
Temperature compensated
15 °C to 71 °C [60 °F to 160 °F]
Temperature effect, zero
0.0025 % full scale/°F
Temperature effect, span
0.0025 % reading/°F
Sealing
Hermetically sealed IP68/NEMA 6P
For reference only
TYPICAL SYSTEM DIAGRAM
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Strain gage type
Bonded foil
Insulation resistance
5000 mOhm @ 50 Vdc
Bridge resistance
350 ohm
Shunt calibration data
Included
Electrical termination (std)
PTIH-10-6P or equiv.
(hermetic stainless)
Mating connector (not incl)
PT06A-10-6S or equiv. (AA111)
MECHANICAL SPECIFICATIONS
Characteristic
Measure
Wetted parts material
number here
< 2000 psig/a
17-4 PH stainless steel
> 2000 psig/a
15-5 PH stainless steel
Weight
10 oz
Case material
304 stainless steel
Marking
Permanent metal name plate MILSTD130F 4.3; Individual sequential
serial number per sensor; Country of
origin and date of manufacture
2 Honeywell • Sensing and Control
Special requirements (consult factory)
Have a special requirement? New case pressure, different cable lengths,
electrical connectors, or materials? Consult our factory by calling +1
614-850-5000 (800-848-6564). Customization is key to our test and
measurement business. Special outputs, wiring codes, and calibrations
are all standard to us.
Precision Gage/Absolute Pressure Transducer
RANGE CODES
Pressure
range (psi)
1
RANGE CODE AP
2
5
10
15
25
50
75
100
150
200
300
500
750
1000
1500
AR
AT
AV
BJ
BL
BN
BP
BR
CJ
CL
CP
CR
CT
CV
DJ
3k
3.5 k
4k
D mm [in]
psia
57 [2.25]
38 [1.50]
D mm [in]
psig
57 [2.25]
45 [1.75]
L mm [in]
psia
65 [2.54]
60 [2.35]
L* mm [in]
psia
96 [3.79]
91 [3.60]
L mm [in]
psig
46 [1.81]
51 [2.00]
51 [2.02]
L* mm [in]
psig
78 [3.06]
83 [3.25]
83 [3.27]
38 [1.50]
Over pressure
150 % full scale
(test) (psi)
150 % full scale
Over pressure
50
(burst) (psi)
100
Port volume
cm3 [in3]
4,1 [0.25]
Natural
frequency
(Hz)
Pressure
range
(psi)
5,2 [0.32]
500
550
1000
200
1.3 k
2.1 k
400
2k
2,8 [0.17]
2.5 k
2.9 k
3.5 k
2000
3000
5000
7500
10000
DL
DN
DR
DT
DV
RANGE CODE
800
4.6 k
6k
7k
60000
EJ
EL
EN
EP
ES
48 [1.90]
56 [2.21]
L* mm [in]
psia
80 [3.15]
89 [3.46]
Over pressure
(test) (psi)
150 % full scale
Consult factory
Over pressure
(burst) (psi)
8k
Port volume
cm3 [in3]
3,1 [0.12]
Natural
35 k
frequency (Hz)
20 k
50000
L mm [in] psia
25 k
17 k
30000
D mm [in] psia
25 k
12 k
20000
38 [1.50]
20 k
9.5 k
15000
38 [1.50]
12 k
9k
40 k
45 k
60 k
80 k
80 k
>100 k
>100 k
>100 k
>100 k
1,5 [0.06]
40 k
54 k
60 k
80 k
100 k
* Length of pressure transducer with amplified option (see option codes)
** 0.5 psi is available for gage only
Honeywell • Sensing and Control 3
Model TJE
OPTION CODES
Range Code
Many range/option combinations are available in our quick-ship and fast-track manufacture programs. Please see http://sensing.honeywell.
com/TMsensor-ship for updated listings.
Pressure
ranges
1, 2, 5, 10, 15, 25, 50, 75, 100, 150, 200 300, 750, 1500, 2000, 15000, 20000, 30000, 50000, 60000 psig/a
500, 1000, 3000, 5000, 7500, 10000 psia
500, 1000, 3000, 5000, 7500, 10000 psig
1a. 60 °F to 160 °F
Temperature
compensation 1b. 30 °F to 130 °F
1f. 70 °F to 250 °F 11
1g. 70 °F to 325 °F 11
1h. 70 °F to 400 °F 11
1i. -65 °F to 250 °F 11
Internal
amplifiers 10
2n. 4 mA to 20 mA (two wire) intrinsically safe8, 17
2t. 0 Vdc to 10 Vdc output
2u. Unamp., mV/V output
1c. 0 °F to 185 °F
1d. -20 °F to 130 °F
1e. -20 °F to 200 °F
2a. 0 Vdc to 5 Vdc (four wire) output15
2c. 0-5 Vdc output15
2j. 4 mA to 20 mA (three wire) output
2k. 4 mA to 20 mA (two wire) output8, 17
Internal ampli- 3a. Input/output isolation15
3d. Remote buffered shunt calibration
fier enhancements
Pressure
ports9
5a. 1/4-18 NPT female (3000 psig to
10000 psig)
5b. 1/4-18 NPT male (15 psig/a to 10000
psig/a)
5c. 7/16-20 UNF female (per
MS33649-4)
5d. 7/16-20 UNF male
5g. G 1/4 male
Electrical
termination
6a. Bendix PTIH-10-6P (or equiv), 6 pin
(max 250 °F)
6b. MS type connector to mate with
MS3106-14S 6S (max 160 °F)13
6e. Integral cable: Teflon (0 °F to 180
°F)
6f. Integral cable: PVC (-20 °F to 160 °F)
6g. Integral cable: Neoprene (0 °F to
180 °F)
6h. Integral cable: Silicone (-65 °F to
300 °F)
6i. Integral underwater cable (max
180 °F)
6j. 1/2-14 conduit fitting with 5 ft of 4
conductor PVC cable
Shunt calibration
8a. Precision internal resistor11
Special
calibration9
9a. 10 point (5 up/5 down) 20 % increments @ 70 °F (gage)
9b. 20 point (10 up/10 down) 10 % increments @ 70 °F (gage)
Wetted
diaphragm9
10a. 316 stainless steel9
10b. Crucible A-286
Bridge
resistance9
12a. 1000 ohm (foil) (max 400 °F)
12b. 5000 ohm (foil) (max 400 °F)
Zero and span 14a. No access to pots
14b. Top access to pots
adjustment
Shock and
vibration
44a. Shock and vibration resistance
Interfaces
53e. Signature calibration11
53t. TEDS IEEE1451.4 module16
4 Honeywell • Sensing and Control
10c. Hastelloy-C
10d. Monel K-500
6m. DIN 43650
6q. Molded integral cable: Polyurethane (max
180 °F)
6t. Integral cable with Heyco spring strain
relief (5 ft)
Precision Gage/Absolute Pressure Transducer
iNTERNAL AMPLIFIERS
Amplifier
specifications
mV/V output
standard
Voltage output: Option
2a4
Vehicle voltage output:
Option 2c4
Vehicle voltage Current
output: Option three-wire:
2t4
Option 2j4
Current twowire:
Option 2k4
Intrinsically
safe amp: Option 2N (2n)
Output signal
3 mV/V 2
0 Vdc to 5 Vdc
0-5 Vdc or ±5
Vdc @ 5 mA
0-10 Vdc or ±10
Vdc @ 5 mA
4 mA to 20
mA
4 mA to 20 mA
4 mA to 20 mA
Input power
(voltage)
10 Vdc regulated
±15 Vdc or
26-32 Vdc
11 Vdc to 28
Vdc
15 Vdc to 28 Vdc
22 Vdc to 32
Vdc3
9 Vdc to 32
Vdc3
9 Vdc to 28
Vdc3
Input power
(current)
28.5 mA @ 10
Vdc
45 mA
40 mA
40 mA
65 mA
4 mA to 28 mA
4 mA to 24 mA
Freq. resp
(amp)
Natural frequency
2000 Hz
3000 Hz
3000 Hz
2500 Hz
2500 Hz
2000 Hz
Power supply
rej.
NA
60 db
60 db
60 db
60 db
60 db
60 db
Operating
temp.
-73 °C to 121
°C [-100 °F to
250 °F]
-28 °C to 85 °C
[-20 °F to 185
°F]
-40 °C to 93 °C
[-40 °F to 200
°F]
-40 °C to 85 °C
[-40 °F to 185 °F]
-40 °C to 85
°C [-40 °F to
185 °F]
-40 °C to 85 °C
[-40 °F to 185
°F]
-28 °C to 85 °C
[-20 °F to 185
°F]
Reverse volt.
prot.
NA
Yes
Yes
Yes
Yes
Yes
Yes
Short cir.
protection
NA
Momentary
Momentary
Momentary
Yes
Yes
Yes
Wiring code:
connector
(std)5
A (+) Excitation
B (+) Excitation
C (-) Excitation
D (-) Excitation
E (-) Output
F (+) Output
A (+) Supply
B Output com.
C Supply ret.
D (+) Output
E Shunt Cal 1
F Shunt Cal 2
A (+) Supply
B Output com **
C Supply ret **
D (+) Output
E Shunt Cal 1
F Shunt Cal 2
A (+) Supply
B Output com **
C Supply ret **
D (+) Output
E Shunt Cal 1
F Shunt Cal 2
A (+) Supply
B Output com
**
C Supply
ret **
D (+) Output
E Shunt Cal 1
F Shunt Cal 2
A (+) Supply
B No conn.
C No conn.
D (+) Output
E Case ground
F No conn.
A (+) Supply
B No conn.
C No conn.
D (+) Output
E Case ground
F No conn.
Wiring code:
cable5,6,7
R (+) Excitation
Bl (-) Excitation
G (-) Output
W (+) Output
R (+) Supply
Bl Output com.
G Supply ret.
W (+) Output
B Shunt Cal 1
Br Shunt Cal 2
R (+) Supply
Bl Output com*
G Supply ret.*
W (+) Output
B Shunt Cal 1
Br Shunt Cal 2
R (+) Supply
Bl Output com*
G Supply ret.*
W (+) Output
B Shunt Cal 1
Br Shunt Cal 2
R (+) Supply
Bl Output
com*
G Supply ret.*
W (+) Output
B Shunt Cal 1
Br Shunt Cal 2
R (+) Supply
Bl (+) Output
W Case ground
R (+) Supply
Bl (+) Output
W Case ground
Reference application sheet
#008-0356-00
Reference application sheet
#008-0357-00
Reference application sheet
#008-0360-00
Reference application sheet
#008-0361-00
Reference application sheet
#008-0361-60
See Honeywell
Web site for info
on intrinsically
safe approvals.
#008-0547-00
For current
information
* Black and green wires are internally connected. • ** Pins B and C are internally connected.
Honeywell • Sensing and Control 5
Model TJE
NOTES
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
Precision Gage/Absolute Pressure Transducer
Accuracies stated are expected for best fit straight line for all errors
including linearity, hysteresis & non-repeatability thru zero.
Output for 0.5 psig/a, 1 psig/a, 2 psig/a units is 1 mV/V to 2 mV/V.
Input power (voltage) for internal amplifier options 2j, 2k, 2n(2N)
depends on load resistance.
CE mark requires options 6a & 3d.
Interconnecting shunt cal. 1 with shunt cal. 2 terminal provides 50
% (unamplified units), 75% (4 mA to 20 mA three-wire units), or
80 % (voltage amp. units) of full scale output for quick calibration. Shunt calibration comes standard with internal amplifier options 2a,
2b, 2c, 2t and 2j.
G=Green; B=Blue; W=White; Bl=Black; Br=Brown; Y=Yellow;
R=Red; O=Orange. Color specifying cable and number or letter
specifying connector.
No mating connector necessary with cable option.
Options 2k, 2n(2N) only available with option 12b.
Availability varies according to range.
Not available with temperatures below -29 °C [-20 °F] or above 85
°C [185 °F].
Cannot be used with amplified option.
Gage pressure units greater than 500 psi are sealed at atmospheric
pressure.
No pot access available with MS type connector.
Temperature 82 °C [180 °F] max., non-shielded standard, shielded
available.
Input/output isolation only available with voltage output (options 2a,
2b, 2c).
Consult factory for TEDS availability with amplified models.
5000 ohm bridge required.
Range dependent; consult factory. Termination dependent; consult
factory.
Internal amp and termination dependent; consult factory.
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to [email protected]
WARNING
PERSONAL INJURY
•DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
•The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
•Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008606-1-EN IL50 GLO May 2008
Copyright © 2008 Honeywell International Inc. All rights reserved.