Model AK - Honeywell Test and Measurement Sensors

Model AK
Digital Pressure Test Gauge
DESCRIPTION
The hand-held Model AK is an in-progress gauge or por-
ond. These units are manufactured from stainless steel and use
table calibrator and delivers long-lasting, enhanced accuracy
sensor technology which has no moving parts, requires little
performance. Accuracy of 0.05 % coupled with an enhanced
recalibration and has excellent overpressure tolerance. Optional
resolution, 4½-digit display makes this unit an excellent transfer
NEMA 4 rating extends the operating temperature to -18 °C to
standard for calibrating pressure measuring equipment.
82 °C [0 °F to 180 °F]. Units of measure are field-selectable and
The AK is rugged and features high/low pressure capture with
a field-adjustable update rate from three to 10 times per sec-
the unit is offered with gage, absolute, vacuum, or compound
reference.
* with optional NEMA 4 rating
FEATURES
•
High accuracy: 0.05 % full scale
•
0 °F to 180 °F operating temperature*
•
Enhanced resolution
•
Stainless steel construction
•
Field-selectable units of measure
•
4½ digit display with 0.5 in height
•
Customer recalibration
•
On/Off switch
•
CE compliant (certain models, consult factory)
•
NEMA 4 rating (optional)
•
Optional carrying case and panel mounting ring
•
Optional pressure ports
•
NIST traceable (optional)
•
9 V battery power
•
High/low capture and store
Model AK
PERFORMANCE SPECIFICATIONS
Characteristic
Measure
Accuracy
0.05 % of full scale BFSL
Pressure ranges 0 psi to
15, 50, 100, 200, 500, 750, 1000, 1500,
2000, 5000, 7500, 10000 psi
ADDITIONAL SPECIFICATIONS
ENVIRONMENTAL SPECIFICATIONS
Pressure range (psi)
Maximum safe over
pressure (psi)1
Incremental display
steps
0 psi to 15 psi = 15
75
0.005
0 psi to 50 psi = 50
250
0.01
0 psi to 100 psi = 100
500
0.02
0 psi to 200 psi = 200
1000
0.05
0 psi to 500 psi = 500
1500
0.1
Characteristic
Measure
0 psi to 750 psi = 750
1500
0.2
Temperature, operating
(standard)
-1 °C to 71 °C [30 °F to 160 °F]
0 psi to 1000 psi = 1K
2000
0.2
Temperature, operating
(optional)
-18 °C to 82 °C [0 °F to 180 °F] (NEMA 4
option required)
0 psi to 1500 psi = 1.5K
3000
0.5
0 psi to 2000 psi = 2K
4000
0.5
0 psi to 5000 psi = 5K
7500
1
0 psi to 7500 psi = 7.5K
12000
2
0 psi to 10000 psi = 10K
15000
2
ELECTRICAL SPECIFICATIONS
Characteristic
Measure
Rating
NEMA 2 (NEMA 4 optional, except AKT)
Display
4½ LCD digits - 12,7 mm [0.5 in] high
Power, Model AKW
One or two 9 V alkaline batteries (included)
Power, Model AKT
110 Vac adapter @ 60 Hz (included)
Power, Model AKV
11 Vdc to 32 Vdc @ 100 mA (3 ft cable
included)
1
Maximum safe overpressure is the pressure which the unit can experience occasionally without the loss of accuracy or permanent damage
ORDER CODES
AK
ORDER CODE
W 200 G
Z
* = Example
SERIES
AK*
POWER
MECHANICAL SPECIFICATIONS
Characteristic
Measure
W
T
V
Physical diameter
93,98 mm [3.7 in]
PRESSURE RANGE (PSI)
Pressure port
9/16-18 UNF-2B, female straight thread
with SAE spec J514 0-ring boss
Wetted parts
Stainless steel
15
50
100
200
Case material
Stainless steel
REFERENCE
Low battery indication
Standard
High and low capture
Standard
Update speed
Adjustable from 3 to 10 per second
Face membrane
Tactile feedback raised memory chips
ELECTRICAL CONNECTION
Calibration data
Stored on memory chip
Z = No electrical connection (Model AKW - 9 V battery)*
C = 3 foot cable (Model AKV)
T = ac adapter (Model AKT)
G
A
V
C
= Battery Powered (9 V alkaline)*
= 110 Vac adapter (included)
= 11 Vdc to 32 Vdc (3 ft. cable)
=
=
=
=
=
=
=
=
0-15
0-50
0-100
0-200*
500 =
750 =
1K =
1.5K =
0-500
0-750
0-1000
0-1500
2K =
5K =
7.5K =
10K =
0-2000
0-5000
0-7500
0-10000
Gage (std) reads zero at atmosphere *
Absolute (opt) reads zero at vacuum (15 psi or higher)
Vacuum (opt) scaled in In Hg (15 psi or lower)
Compound (opt) reads both positive & negative
(vacuum) pressure
Call factory for availability of other electrical connections & cable lengths.
OPTIONAL
4 = NEMA 4 (Not available on AKT)
NEMA 4 option also provides temperature range of 0-180°F
FIELD SELECTABLE STANDARD UNITS OF MEASURE
Each gauge has the following scaling built in for field selection:
Bar
mBar
kPa
MPa
= Bar
= Millibar
= Kilopascals
= Megapascals
Ft H20
In H20
In Hg
mmHg
=
=
=
=
Feet of water
Inches of water
Inches of mercury
mm of mercury
Consult factory for other units of measure not listed
2 Honeywell • Sensing and Control
4
Model AK
Digital Pressure Test Gauge
Mounting dimensions and characteristics
Warranty. Honeywell warrants goods of its manufacture as
being free of defective materials and faulty workmanship.
Honeywell’s standard product warranty applies unless agreed
to otherwise by Honeywell in writing; please refer to your
order acknowledgement or consult your local sales office for
specific warranty details. If warranted goods are returned to
Honeywell during the period of coverage, Honeywell will repair
or replace, at its option, without charge those items it finds
defective. The foregoing is buyer’s sole remedy and is in lieu
of all warranties, expressed or implied, including those of
merchantability and fitness for a particular purpose. In no
event shall Honeywell be liable for consequential, special, or
indirect damages.
While we provide application assistance personally, through our
literature and the Honeywell web site, it is up to the customer to
determine the suitability of the product in the application.
For reference only
Specifications may change without notice. The information we
supply is believed to be accurate and reliable as of this printing.
However, we assume no responsibility for its use.
Notes
1. Ure faci bla feuipit nostin vullam zzril
2. Ure faci bla feuipit nostin vullam zzril dolesto consent nis del ipsuscilis nismodiamcor sustinit am eugait velestinit ut nibh elisim eugait nim
iril ut del duisi.
3. Ure faci bla feuipit nostin vullam zzril dolesto consent nis del4. Ure faci bla feuipit nostin vullam zzril dolesto consent nis del5. Ure faci bla feuipit nostin vullam zzril dolesto consent nis del
6. Ure faci bla feuipit nostin vullam zzril dolesto consent nis del ipsuscilis nismodiamcor sustinit am eugait velestinit ut nibh elisim eugait nim
iril ut del duisi.
For more information about Sensing and Control products, visit
www.honeywell.com/sensing or call +1-815-235-6847
Email inquiries to [email protected]
WARNING
PERSONAL INJURY
•DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
Failure to comply with these instructions could result in
death or serious injury.
WARNING
MISUSE OF DOCUMENTATION
•The information presented in this catalogue is for
reference only. DO NOT USE this document as product
installation information.
•Complete installation, operation and maintenance
information is provided in the instructions supplied with
each product.
Failure to comply with these instructions could result in
death or serious injury.
Sensing and Control
Automation and Control Solutions
Honeywell
1985 Douglas Drive North
Golden Valley, MN 55422 USA
+1-815-235-6847
www.honeywell.com/sensing
008689-1-EN IL50 GLO May 2008
Copyright © 2008 Honeywell International Inc. All rights reserved.