tid report - M.S. Kennedy Corp.

Total Dose Radiation Test Report
MSK5805RH
RAD Hard Ultra Low Dropout
Adjustable Positive Linear Regulator
July 9, 2014 (1st Test, IC Wafer Lot: W10809524.1 #8
Transistor Wafer Lot: CJ302831RC#20)
B. Horton
M. Fleury
M.S. Kennedy Corporation
Page 1 of 9
I.
Introduction:
The total dose radiation test plan for the MSK5805RH series was developed to qualify the device as
RAD Hard to 300 Krad(Si). The testing was performed beyond 300 Krad(Si) to show trends in device
performance as a function of total dose. The test does not classify maximum radiation tolerance of
the device, but simply offers designers insight to the critical parameter-shifts up to the specified total
dose level.
MIL-STD-883 Method 1019 and ASTM F1892-06 were used as guidelines in the development and
implementation of the total dose test plan for the MSK5805RH.
II.
Radiation Source:
This total dose test was performed at the University of Massachusetts, Lowell, using a cobalt 60
radiation source. The dose rate was determined to be 119 rads(Si)/sec. The total dose schedule can
be found in Table I.
III. Test Setup:
All test samples were subjected to Group A Electrical Test at 25°C IAW the device data sheet. In
addition, all devices received 320 hours of burn-in per MIL-STD-883 Method 1015. For test platform
verification, two control devices were tested at 25°C. Ten devices were then tested at 25°C, prior to
irradiation, and were found to be within acceptable test limits.
The devices were vertically aligned with the radiation source and enclosed in a lead/aluminum
container during irradiation.
Five devices were biased during irradiation. The maximum
recommended operating voltage of +7.5V was used for the bias condition. Five devices had all leads
grounded during irradiation for the unbiased condition.
After each irradiation the device leads were shorted together and transported to the MSK electrical
test platform and tested IAW MSK device data sheet. Testing was performed on irradiated devices,
as well as two control devices, at each total dose level. Electrical tests were completed within one
hour of irradiation. Devices were subjected to subsequent radiation doses within two hours of
removal from the radiation field.
IV. Data:
All performance curves are averaged from the test results of the biased and unbiased devices,
respectively. If required, full test data can be obtained by contacting M.S. Kennedy Corporation.
V. Summary:
Based on the test data recorded during radiation testing, the MSK5805RH qualified as a 300Krad(Si)
radiation hardened device. Feedback Voltage, Shutdown Threshold, and Output Current Limit
exhibited the most significant shift due to irradiation, however all parameters remained within
specification up to 450 KRAD (Si) TID.
Page 2 of 9
MSK 5805RH Biased/Unbiased Dose Rate
Schedule
Dosimetry Equipment
Bruker Biospin # 0371
Irradiation Date
7/9/14
Exposure
Incremental Dose
Length
rads(Si)
(min:sec)
14:26
7:13
21:39
21:39
103,054
51,527
154,581
154,581
Cumulative
Dose
rads(Si)
103,054
154,581
309,162
463,743
Biased S/N – 0027, 0028, 0029, 0030, 0031
Unbiased S/N – 0032, 0033, 0034, 0035, 0036
Table 1
Dose Time, Incremental Dose and Total Cumulative Dose
Page 3 of 9
Page 4 of 9
Page 5 of 9
Page 6 of 9
Page 7 of 9
Page 8 of 9
Page 9 of 9