Solitel Instruction Manual BE51-640

SOLITEL
Vibrating rod
level switches
for solids
®
INSTALLATION / INSTRUCTION MANUAL
DESCRIPTION
The Solitel vibrating rods are used for high or low level
detection on powders and granulars with a minimum
density of 0.05 kg/dm3 and max. diameter of 10-12 mm.
The electronics and the sensor are an integral unit for easy
installation.
OPERATING PRINCIPLE
A resonant frequency of 350 Hz keeps the vibration on the
resonant part of the sensor ongoing. The medium will
dampen the vibration when reaching the sensor. For a
successful switching operation, the vibrating part of the
sensor should be covered for 20 to 60 mm (depending
upon density). The sensed dampening will cause the
electronics to actuate the relay.
UNPACKING
Unpack the instrument carefully. Inspect all units for
damage. Report any concealed damage to carrier within 24
hours. Check the contents of the packing slip and purchase
order. Check and record the serial and reference numbers
for future reference when ordering parts.
MODEL IDENTIFICATION
A complete measuring system consists of
Order code for SOLITEL sensor and electronics
S
H
Standard SOLITEL - max 110°C (230°F) (Flexible sensor - max 90°C (200°F))
High temperature SOLITEL - max 160°C (320°F) (only for rigid sensors - VHH/VHR)
TYPE OF SENSOR
H
Standard sensor with 235 mm (9.2") insertion length
R
Extended rigid sensor with insertion length from 50 cm up to 3 m (19.60" up to 9.8')
K
Extended flexible sensor with insertion length from 1 m up to 20 m (3.26' up to 65.6')
S
A
Non Ex
Dust Ex zone 10
POWER SUPPLY
1
2
6
0
240
024
024
120
V
V
V
V
AC
DC
AC
AC
INSERTION LENGTH
0 0 0
0 5 0-3 0 0
0 0 1-0 2 0
V
1 B
Standard length = 235 mm (9.2") (VSH/VHH units)
Extended rigid sensors (VSR/VHR units), specify length
in 10 cm (3.9") increments, min 50 cm (19.7") and max 300 cm (9.8').
Extended flexible sensors (VSK), specify length in 1 m (3.28') increments,
min 1 m (3.28') and max 20 m (65.6')
complete order code for SOLITEL – rigid sensor and electronics
GENERAL INSTRUCTIONS
MOUNTING
WIRING
Solitel switches have to be installed at the desired
switching point. The VSH and VSR units may be installed
horizontally (at 45° inclination) as well as vertically. The
VSK units with wire rope extension should only be
mounted vertically. However the mechanical torque on the
sensor and the adhesion properties of the medium also
have to be taken into consideration when installing the unit
in the tank (see Electrical Specifications (page 3) for the
maximum mechanical load on sensor).
1. Screw sensor part into the mounting bushing (G 1 1/2").
Only apply a wrench (width flats = 55 mm) on the
mounting nut.
2. Position the housing manually in the correct direction to
allow easy wiring. The housing can rotate one full turn.
Use cable with an outside diameter of 8 to 15 mm (0.3" to
0.6") _ inner core 0.75 to 2.5 mm2 _ in order to block the
cable sufficiently into the provided cable glands.
1. Make sure that the power supply is turned off.
2. Use a size 5 Allan type wrench to remove the housing
cover _ 4 screws at top of housing.
3. Pass the power cable through the cable glands (conduit
connection) into the housing (power is left cable gland,
relays is right cable gland).
4. Connect power leads to the proper terminals. To allow
access to the terminals, the unit utilises easy to remove
connector plugs (power terminals are 1 and 2, see
Figure 4).
When using 3-wire cable, use the grounding screw at
the inside of the housing base.
When using 2-wire cable, provide a grounding wire to be
connected to the grounding screw at the outside of the
housing base.
Important
1. Install the unit at a location where the level variation is
detectable, taking the arching or caving of the material
into account. See Figure 1.
2. Install the rigid sensors at a tilt angle larger than the
slope angle of the material being filled (or for high level
applications-vertically mounted), build up on the sensor
would prevent the unit from starting the vibration and
thus eliminates the 'self cleaning' effect which can lead
to false output when level decreases. See Figure 1.
Caution:
24 V DC units: the negative terminal of the power
supply is connected internally to ground.
5. Connect control circuit leads to relay terminals.
(7, 8 and 9, see Figure 4)
Caution:
For alarm applications, it is recommended to use deenergised status of the relay as alarm status to allow
detection of power failure as an alarm status, the
relay contacts will show:
POWER
MATERIAL
LEVEL
A position
B position
A position
low
B position
high/low A/B
high
on
fails
High level indication
Low level indication
Figure 1
3. Install the unit in this way that the vibrating part is entirely
free to vibrate. When the vibrating rod is fully blocked by
medium, the unit will not change status of the relay. See
Figure 2.
4. When the falling material can damage the rod by its
nature (rough, heavy or with sharp angles) then protect
the sensor. See Figure 3.
FAILSAFE
JUMPER
RELAY COIL
Figure 3
5. Provide extra damping materials such as rubber joints,
when the tank is exposed to intense vibrations.
2
TERMINALS
7 and 8 8 and 9
de-energised
de-energised
de-energised
de-energised
de-energised
open
closed
closed
open
open
closed
open
open
closed
closed
6. Reposition the connector plugs, pull excess cable back
and tighten the cable gland. Finally screw cover on
housing base.
7. Wiring is completed. Turn power on.
High/low level
selector
LED
Adjustment screw
Relay
terminal
Power supply
terminal
Figure 2
RELAY
Grounding screw
Figure 4
CALIBRATION
After being properly installed the unit is fully operational.
The LED will show the relay status. By touching the rod by
hand and thus performing a damping of the vibration the
LED will reverse colour. See Figure 4.
The LED shows the status of the relay:
RED = relay energised
GREEN = relay de-energised.
screw (see Figure 4) counterclockwise until the unit
switches in that specific material.
When adjusted too sensitive, the unit may detect even
slight residues of the material instead of level.
Caution:
It is recommended to perform a test set up of the unit
in a sample of the material to be detected before
installation.
DENSITY ADJUSTMENT
The unit is factory preset for media with densities higher
than 0.3 g/cm3. For lower densities turn the adjustment
MAINTENANCE
The Solitel vibrating rods do not require maintenance on a regular basis. In some instances however it may be advised to
clean the sensor of deposited material.
STORAGE CONDITIONS
Ambient temperature: -20°C up to +60°C (-4°F up to 140°F)
Relative humidity:
up to 98%
F
M
M
F
VSH
VSR
VSK
REPLACEMENTS PARTS
Connector plugs for power supply: 2 pins
Connector plugs for control circuits: 3 pins
Cable gland PG 16
Sensor: VSH = standard rigid sensor
VSR = extended rigid sensor, specify length from bottom mounting nut to end of sensor _ min. 0.5 m, max. 2 m
(min. 1.6 ft, max. 6.5 ft)
VSK = extended wire rope sensor, specify length from bottom mounting nut to end of sensor _ min. 1 m, max. 20 m
(min. 3.2 ft, max. 65 ft)
Electronics module
3
ELECTRICAL SPECIFICATIONS
Description
Specifications
Input voltage
240 V AC
(-15 to +10% tolerance on nominal voltage)
110 V AC
024 V DC / V AC
less or equal than 3 V A
± 350 Hz
8 A, SPDT (max. load 250 V AC)
1 A, SPDT (max. load 024 V DC)
6 seconds
1 1/2" BSP thread
IP65, cast aluminium
Non Ex
Dust Ex Zone 10
-40°C to +110°C (-40°F to +230°F) -40°C to +110°C (-40°F to +230°F)
-40°C to +160°C (-40°F to +320°F) -40°C to +160°C (-40°F to +320°F)
-25°C to + 90°C (-13°F to +200°F) -20°C to + 90°C ( -4°F to +200°F)
-40°C to +60°C (-40°F to +140°F)
Non Ex
Dust Ex Zone 10
25 bar abs (360 PSIG)
6 bar abs (87 PSIG)
6 bar abs ( 87 PSIG)
6 bar abs (87 PSIG)
all wetted parts of 316 Ti SST
Power consumption
Operation frequency
Output relay
Time delay
Process connection
Housing
Max. process
temperature
VSH/VSR
VHH/VHR
VSK
Ambient temperature electronics
Max. process pressure
VSH/VSR - VHH/VHR
VSK
Rod material
Materials:
Vibrating rod
Process connection
Cable for extended flexible Solitel:
Max. mechanical
load on rod
Weight of unit
APPROVALS
4
Agency
Approval
DMT
Dust Ex Zone 10
316 SS Ti (Standard electropolished)
14408 x 6 Cr Ni Mo 1810
PE coated steel
standard
extended
version
rigid version
flexible
version
85 Nm
86 Nm
45 kN
2.5 kg
2.5 kg +
1.4 kg/m
2.5 kg +
0.6 kg/m
DIMENSIONS IN mm (inches)
Standard VSH sensor
Flexible VSK sensor
120 (4.72)
120 (4.72)
105
(4.13)
105
(4.13)
PG 16
2 entries
PG 16
2 entries
1 1/2" BSP
process
connection
1 1/2" BSP
process connection
Ø 44 (1.73)
Insertion length
1000 - 20000
(39.37 - 787.40)
Insertion
length
235 (9.25)
Ø 44 (1.73)
161
(6.34)
161
(6.34)
Extended VSR rigid sensor
120 (4.72)
105
(4.13)
PG 16
2 entries
1 1/2" BSP
process connection
Ø 44 (1.73)
Insertion length
500 - 3000
(19.68 - 118,11)
161
(6.34)
5
IMPORTANT
SERVICE POLICY
Owners of Magnetrol products may request the return of a control; or, any part of a control for complete rebuilding or
replacement. They will be rebuilt or replaced promptly. Magnetrol International will repair or replace the control, at no cost to
the purchaser, (or owner) other than transportation cost if:
a. Returned within the warranty period; and,
b. The factory inspection finds the cause of the malfunction to be defective material or workmanship.
If the trouble is the result of conditions beyond our control; or, is NOT covered by the warranty, there will be charges for labour
and the parts required to rebuild or replace the equipment.
In some cases, it may be expedient to ship replacement parts; or, in extreme cases a complete new control, to replace the
original equipment before it is returned. If this is desired, notify the factory of both the model and serial numbers of the
control to be replaced. In such cases, credit for the materials returned, will be determined on the basis of the applicability of
our warranty.
No claims for misapplication, labour, direct or consequential damage will be allowed.
RETURNED MATERIAL PROCEDURE
So that we may efficiently process any materials that are returned, it is essential that a “Return Material Authorisation” (RMA)
form will be obtained from the factory. It is mandatory that this form will be attached to each material returned. This form is
available through Magnetrol’s local representative or by contacting the factory. Please supply the following information:
1.
2.
3.
4.
5.
6.
Purchaser Name
Description of Material
Serial Number and Ref Number
Desired Action
Reason for Return
Process details
All shipments returned to the factory must be by prepaid transportation. Magnetrol will not accept collect shipments.
All replacements will be shipped FOB factory.
BULLETIN N°:
EFFECTIVE:
SUPERSEDES:
UNDER RESERVE OF MODIFICATIONS
BELGIUM
®
DEUTSCHLAND
FRANCE
ITALIA
UNITED
KINGDOM
INDIA
BE 51-640.3
FEBRUARY 1998
July 1996
Heikensstraat 6, 9240 Zele
Tel. (052) 45.11.11
Fax. (052) 45.09.93
Schloßstraße 76, D-51429 Bergisch Gladbach-Bensberg
Tel. (02204) 9536-0
Fax. (02204) 9536-53
Le Vinci 6 - Parc d’activités de Mitry Compans,
1, rue Becquerel, 77290 Mitry Mory
Tél. 01.60.93.99.50
Fax. 01.60.93.99.51
Via Arese 12, I-20159 Milano
Tel. (02) 607.22.98 (R.A.)
Fax. (02) 668.66.52
Unit 1 Regent Business Centre
Jubilee Road Burgess Hill West Sussex RH 15 9TL
Tel. (01444) 871313
Fax (01444) 871317
B4/115 Safdurjung Enclave, New Delhi 110 029
Tel. 91 (11) 6186211
Fax 91 (11) 6186418