Download Data Sheet

PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
Rev A
Wall Industries, Inc.
FEATURES
• RoHS Compliant
• Remote Setting Multiple PSU via RS232, RS485 & I2C
• 1200 Watts Output Power
• Programmable Output Voltage (0~105%)
• High Efficiency up to 93%
• Programmable Output Current (0~105%)
• 4000VAC I/O Isolation (2 x MOPP)
• Universal Input Voltage Range: 90~264VAC (127~370VDC)
• Constant Current Limiting
• Single Outputs Ranging from 12VDC to 60VDC
• Global Control via RS232
• Selectable +5V/0.5A or +9V/0.3A Auxiliary Output
• Power OK Signal (Power Good, Logic Low)
• Forced Current Sharing at Parallel Operation
• Remote ON/OFF, Remote Sense Functions
• ANSI/AAMI ES60601-1, TUV EN60601-1: 2006, IEC60601-1
• Protection: OLP, OVP, OTP, Fan Failure
Size:
10.256 x 5.000 x 2.500 inches
(260.50 x 127.00 x 63.50 mm)
Weight
5.73 lbs (2.6kg)
Medical Approvals
DESCRIPTION
The PSME1200 series of medical AC/DC switching power supplies provides 1200 Watts of output power in
a 10.256” x 5.000” x 2.500” enclosed case. This series consists of single output models ranging from 12VDC
to 60VDC with a universal input voltage range of 90~264VAC (127~370VDC). Standard features include
high efficiency up to 93%, programmable output voltage and output current, remote on/off, and power OK
signal. This series also has over temperature, over voltage, and over load protection. These supplies also
feature a low leakage current of less than 300μA (earth) and less than 100µA (patient) at 264VAC and are
designed to withstand 4000VAC input to output isolation (2 x MOPP). All models are RoHS compliant and
have ANSI/AAMI ES60601-1, TUV EN60601-1: 2006, and IEC60601-1 medical approvals.
MODEL SELECTION TABLE
Model Number
PSME-1200-12
PSME-1200-15
PSME-1200-24
PSME-1200-30
PSME-1200-36
PSME-1200-48
PSME-1200-60
Input Voltage (2) Output Voltage Output Current
12 VDC
100A
15 VDC
80A
24 VDC
50A
90~264 VAC
30 VDC
40A
(127~370 VDC)
36 VDC
33.4A
48 VDC
25A
60 VDC
20A
Line Regulation
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
Load Regulation Output Power Ripple & Noise (1)
±1.0
1200W
120mVp-p
±1.0
1200W
150mVp-p
±1.0
1200W
150mVp-p
±1.0
1200W
150mVp-p
±1.0
1202.4W
150mVp-p
±1.0
1200W
150mVp-p
±1.0
1200W
150mVp-p
Efficiency
89%
90%
91%
92%
92%
93%
93%
NOTES
1. Ripple & noise is measured at 20MHz limited bandwidth and using a 12" twisted pair-wire terminated with a 0.1µF & 47µF capacitors in parallel.
2. For voltages near the low end of the input voltage range, see the derating curve for the power supply output rating.
3. When in parallel operation only one unit might operate if the total output load is less than 5% of the rated load condition.
4. The power supply is considered a component which will be installed into final equipment. The final equipment must be re-confirmed that it still meets
EMC directives.
DERATING CURVES
Load vs Ambient Temperature
Load vs Input Voltage
120
120
Ta=25°C
100
100
90
Load (%)
Load (%)
80
60
40
80
70
60
50
20
-20
40
0
10
20
30
40
50
Ambient Temperature (°C)
06/05/2014
60
70
80
90
100
110
120
264
Input Voltage (VAC)
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 1 of 7
PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
Rev A
Wall Industries, Inc.
SPECIFICATIONS: PSME1200 SERIES
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
264
370
63
VAC
VDC
Hz
INPUT SPECIFICATIONS
Input Voltage (See Note 2)
AC input voltage range
DC input voltage range
90
127
47
Input Frequency
AC Current
Inrush Current
Power Factor
OUTPUT SPECIFICATIONS
Output Voltage
Voltage Tolerance
Voltage Adjustability
Line Regulation
Load Regulation
Output Power
Output Current
Ripple & Noise (20MHz BW)
Hold-up Time
Setup Time
Rise Time
Temperature Coefficient
PROTECTION
Over Voltage Protection (see page 7)
Over Load Protection
Over Temperature Protection
GENERAL SPECIFICATIONS
Efficiency
Input to Output
Isolation Voltage
Input to FG
Output to FG
Input to Output
Isolation Resistance Input to FG
Output to FG
Earth
Leakage Current
Patient
FUNCTIONS
Auxiliary Power
Remote ON/OFF Control (see page 4)
Power OK Signal
Output Voltage Trim
Output Current Trim
Parallel Operation (Current Sharing)
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Cooling
Vibration
PHYSICAL SPECIFICATIONS
Weight
Dimensions (W x H x D)
SAFETY & EMC (See Note 4)
Safety Approvals
EMI (Conduction & Radiation)
Harmonic Current
EMS Immunity
06/05/2014
At 115VAC and full load
At 230VAC and full load
At 115VAC and cold start
At 230VAC and cold start
At 115VAC and full load
At 230VAC and full load
14.5
6
30
45
0.99
0.95
Includes set-up tolerance, line regulation, and load regulation
Typical adjustment by potentiometer (VR1)
Low Line to High Line
0% to 100% full load
-2.0
-5.0
-1.0
-1.0
A
A
See Table
+2.0
+5.0
+1.0
+1.0
See Table
See Table
See Table
%
%
%
%
Measured with 0.1µF and 47µF capacitors in parallel
At 230VAC and full load
full load
full load
0~50°C
-0.02
Protection type: latch-style. Recovery after reset AC power ON or inhibit
Protection type: constant current limit
Protection type: shut down o/p voltage; auto-recovery after temp. goes down
Variable OVP, 120%±7% Vout
105% rated output power
100°C±5°C detect on heatsink of sec. side
2 x MOPP
4000
1800
500
100
100
100
16
800
100
+0.02
ms
ms
ms
%/°C
See Table
Test is done without enclosure
500VDC
500VDC
500VDC
At 264VAC
At 264VAC
Isolated from output
Sink Current: 20mA max.; Drain Voltage: 40V max.
VAC
MΩ
300
100
µA
Selectable +5V/0.5A or +9V/0.3A aux. output
By external switch
Open drain signal low when PSU turns on
0
105
%Vo
0
105
%Io
See page 5
See derating curve
-20
+60
°C
-40
+85
°C
Non-condensing
20
90
% RH
10
95
% RH
Load and temperature control fan
10~500Hz, 2G 10 min./1 cycle, period for 60 min. each along X, Y, Z axes. Certified IEC60068-2-6-2007; IEC60068-2-64-2008
5.73 lbs (2.6kg)
10.256 x 5.000 x 2.500 inches (260.50 x 127.00 x 63.50 mm)
UL/cUL (ANSI/AAMI ES60601-1; CAN / CSA-C22.2 No.60601-1); NFPA 99; TUV (EN60601-1; IEC60601-1); MOPP approval
EN55011
EN61000-3-2; EN61000-3-3
EN60601-1-2; IEC61000-4-2, 3, 4, 5, 6, 8, 11
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 7
Rev A
Wall Industries, Inc.
PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
MECHANICAL DRAWING
10.236 [260.50]
Unit: inches [mm]
2.500 [63.50]
6.378 [162.00]
POS (+)
NEG (-)
1
V-Adj
CN5
LED
Addr. SW
4-M4 L=4 both sides
5.000 [127.00]
2.638 [67.00]
N
L
1.161 [29.50]
2.264 [57.50]
11.555 [293.50]
23
21
19
17
15
13
11
9
7
5
3
1
CN5
7.028 [178.50]
1.181
[30.00]
0.768
[19.50]
24
22
20
18
16
14
12
10
8
6
4
2
1.417 [36.00]
AC Input Terminal
Pin
Function
1
ACL
2
ACN
3
4-M4 L=4 both sides
Pin
1
2
3
4
5
6
7
8
9
10
11
12
Function
NC
NC
AUX
GND
SCL
SDA
AUX
GND
VCI
GND
ACI
GND
Control Pin Number Assignment (CN5): JST S24B-PHDSS or Equivalent
Description
Pin
Function
Description
For RS232 Receiver Function
13
EN+
Inhibit ON/OFF (+)
For RS232 Transmission Function
14
AUX
+5V/0.5A or +9V/0.3A Auxiliary Power
+5V/0.5A or +9V/0.3A Auxiliary Power
15
ENInhibit ON/OFF (-)
Ground
16
GND
Ground
Serial Clock used in the I2C Interface
17
PAR
Parallel Operation Current Sharing
Serial Data used in the I2C Interface
18
VSET
AUX Output Set
+5V/0.5A or +9V/0.3A Auxiliary Power
19
POK
Power OK
Ground
20
GND
Ground
V Program
21
VSRemote Sense (-)
Ground
22
VONegative Output Voltage
I Program
23
VS+
Remote Sense (+)
Ground
24
VO+
Positive Output Voltage
LED STATUS
LED
Solid (Green)
Solid (Orange)
Slow Blink (Green)
Fast Blink (Red)
Solid (Red)
Slow Blink (Red)
Intermittent Blink (Red)
Interlace Blink (Red)
LED Signal
Status
Power OK (Local Mode)
Power OK (Remote Mode)
Power Standby
Over Voltage Protection (OVP)
Over Load Protection (OLP)
Over Temperature Protection (OTP)
Fan Failure
Power Failure
* Local mode: Use ACI/VCI to control output current and voltage
* Remote Mode: Use RS232 or I2C command to control output current and voltage
06/05/2014
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 7
PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
Rev A
Wall Industries, Inc.
REMOTE ON/OFF
(1) Using Internal 5V Auxiliary Source
(2) ON/OFF Control by NPN Transistor
AUX
AUX
SW
2KΩ
(3) Using External Voltage Source
2KΩ
EN+
AUX
2KΩ
EN+
EN+
SW
AUX
AUX
AUX
EN-
EN-
GND
GND
AUX
EN-
Tr
Ctrl
sw
GND
OFF Power OFF
ON
Ctrl
Power ON
L
Power OFF
H
Power ON
GND
sw
OFF Power OFF
ON
Power ON
Default setting
POWER OK SIGNAL
* Place an additional capacitor to have a better performance of auxiliary power operation.
P.OK
Good
Failure
*
DC / DC
5V/0.5A
or
9V/0.3A
AUX
VSET
GND
AUX
LOAD
* The grounding of “AUX” power should be connected to “GND” port. If “V-“ is connected as
Grounding make sure to short the “GND” and “V-“ ports.
Internal Circuit
VSET
AUX and P.OK Signal
470µF
16V
Open (Default Setting)
Short to GND
5V
9V
REMOTE SENSE
1) Remote Sense
2) Local Sense (Default Setting)
+
+
+
06/05/2014
-
+
-
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 7
PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
Rev A
Wall Industries, Inc.
CURRENT SHARING
3) Current Sharing with Remote Sense
+
+
24
23
VS+
VO+
+
24
23
VO+
VS+
+
24
23
VO+
VS+
24
23
VS+
VO+
VO-
VS-
VO-
VS-
VO-
VS-
VO-
VS-
GND
POK
GND
POK
GND
POK
GND
POK
VSET
PAR
VSET
PAR
VSET
PAR
VSET
PAR
GND
EN-
GND
EN-
GND
EN-
GND
EN-
AUX
EN+
AUX
EN+
AUX
EN+
AUX
EN+
GND
ACI
GND
ACI
GND
ACI
GND
ACI
GND
VCI
GND
VCI
GND
VCI
GND
VCI
GND
AUX
GND
AUX
GND
AUX
GND
AUX
SDA
SCL
SDA
SCL
SDA
SCL
SDA
SCL
GND
AUX
GND
AUX
GND
AUX
GND
AUX
NC
NC
NC
NC
NC
NC
NC
NC
4) Current Sharing with Local Sensing
+
+
24
23
VO+
06/05/2014
VS+
+
24
23
VO+
VS+
+
24
23
VO+
VS+
24
23
VO+
VS+
VO-
VS-
VO-
VS-
VO-
VS-
VO-
VS-
GND
POK
GND
POK
GND
POK
GND
POK
VSET
PAR
VSET
PAR
VSET
PAR
VSET
PAR
GND
EN-
GND
EN-
GND
EN-
GND
EN-
AUX
EN+
AUX
EN+
AUX
EN+
AUX
EN+
GND
ACI
GND
ACI
GND
ACI
GND
ACI
GND
VCI
GND
VCI
GND
VCI
GND
VCI
GND
AUX
GND
AUX
GND
AUX
GND
AUX
SDA
SCL
SDA
SCL
SDA
SCL
SDA
SCL
GND
AUX
GND
AUX
GND
AUX
GND
AUX
NC
NC
NC
NC
NC
NC
NC
NC
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 5 of 7
PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
Rev A
Wall Industries, Inc.
INSTALLATION INSTRUCTIONS
1. Mounting Directions
1-1 Recommended Standard Mounting Methods
(a)
(b)
(c)
+
+
+
4-M4
L=4mm max.
4-M4
L=4mm max.
4-M4
L=4mm max.
2. Mounting Method
2-1 There are ventilating holes on the front and back side panels. Do not obstruct; allow at least 50mm for airflow
2-2 The maximum allowable penetration for the screw is 4mm. Incomplete threading should not be penetrated.
2-3 Recommended torque of mounting screw: M4 screw: 1.27N • m (13.0kgf • cm)
More than 50mm
More than 50mm
FAN
Airflow
FAN
06/05/2014
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 6 of 7
Rev A
Wall Industries, Inc.
PSME1200 SERIES
1200 Watts, Medical AC/DC Switching Power Supplies
CURVES
CMD vs Output Curve
VCI vs OVP Curve
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will provide
a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one example of our
commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its
customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing support for
your project above and beyond the design and production phases. Give us a call today to discuss your future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
06/05/2014
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 7 of 7