Technical Data

Cable Sensor, Pt100/Pt1000
Sensors with 4-wire silicone cable
for pockets or tubes
4-wire air sensors
Elements in
1/1 DIN B,
1/3 DIN B
and 1/6 DIN B
The sensors can be used with the CombiTemp,
Building Block Temperature Measuring System
Description
A platinum resistor is built into the temperature sensor. Changes in
temperature are reflected by changes in the electrical resistance, so
that measuring the value of the resistance gives an analogue expression for the actual temperature.
The Pt100 element has a resistance of 100 Ohms at 0°C and a welldocumented working curve (DIN/EN/IEC 60751) within the nominal
working range.
The Pt100 sensor fits as an insert in sensor tubes with an internal
diameter of 6 mm such as the CombiTemp temperature measuring
system.
Air sensors are used without sensor pockets to ensure fast response
time.
The sensors are used in a wide range of applications within the
marine, industrial, energy and food sectors.
www.baumerprocess.com
Design and specifications subject to change without notice
Data Sheet 2902-1
Page 1
Technical Data
Standard sensor
Common data for both types
Measuring range
-50...205°C
Ambient temperature
-50...205°C
Case material
Acid-proof, stainless steel
AISI 316 Ti (W 1.4571)
Marking label
-30...105°C
Case dimensions
ø5.8 mm x 60 mm
Pressure range
< 25 bar (water flow 3m/sec.)
Time constant τ0,5
See table
Humidity
< 98% RH, condensing
Accuracy
DIN/EN/IEC 60751
Protection class
IP 65
1/1 DIN B
+(0.3 + 0.005 x t) °C
Cable type
High-flexible silicone, grey
1/3 DIN B
+1/3 x (0.3 + 0.005 x t) °C
Wires
4 (2 x Red, 2 x white)
1/6 DIN B
+1/6 x (0.3 + 0.005 x t) °C
Length
up to 99.99 metres
Vibrations
Lloyds Register, test 2
Mechanical tolerances
ISO 2768-m
Air sensor
Measuring range
-50...205°C
Disposal of product and packing
Ambient temperature
-50...205°C
According to national laws or by returning to Baumer
Marking label
-30...105°C
Environment
Non-aggressive air
Air gap
8 holes, ø3 mm
Humidity
< 98% RH, condensing
Protection class
IP 65
Cable type
High-flexible silicone, grey
Wires
4 (2 x Red, 2 x white)
Length
up to 99.99 metres
Time Constant τ0.5
If a pocket or sensor tube is used, the response time is extended, i.e. the time duration for the sensor to reach the correct temperature when
the temperature of the medium changes suddenly.
Environment
Sensor type
Standard Sensor
Air Sensor
Fluid, 0.4 m/sec.
Fluid, 0.4 m/sec. (Stainless steel pocket/tube with silicone paste)
Air, 3 m/sec.
Air, still
8 sec.
17 sec.
35 sec.
135 sec.
25 sec.
105 sec.
Ordering Details - Cable sensors
8141 3xx
Sensor type
Standard sensor
Air sensor
6´ digit
Sensor element (DIN/EN/IEC 60751)
Pt100, 1/1 DIN B, single, specified accuracy -50...400°C
Pt100, 1/3 DIN B, single, specified accuracy 0...150°C
Pt100, 1/6 DIN B, single, specified accuracy 0...100°C
Pt100, 1/1 DIN A, single, specified accuracy -50...400°C
Pt1000, 1/3 DIN B, single, specified accuracy -50...400°C
Pt1000, 1/1 DIN B, single, specified accuracy -50...400°C
7´ digit
Cable length (L)
Length in cm
www.baumerprocess.com
Design and specifications subject to change without notice
xxxx
3
5
1
3
5
7
A
B
8...11´ digit
xxxx
Data Sheet 2902-1
Page 2
Example of Application
230 Vac
Pt100/Pt1000 sensor
5
3
7
3 4
5
1 2
1 2
4
3
6
2
-
+
1
4...20 mA
FlexTop 2211
+
+
-
Trigger module
24 Vdc
-
Power supply
Dimensional Drawing
{mm]
Red
Red
White
White
60
Cable length (L) + 20 mm
4-wire Sensors
White
White
Red
Red
Warning:
One of the wires may not be
connected in case of a 3-wire
connection to the temperature
transmitter
Ordering Details - Pocket for Cable Sensors
2909 0001 xxx
Pocket type
Stainless steel, W 1.4404 (AISI 316L), R1/2 with M12 gland
Pocket length (L)
Length in mm
[mm]
M12
20
9...11´ digit
xxx
AF22
25
R1/2
L
ø8
www.baumerprocess.com
Design and specifications subject to change without notice
Data Sheet 2902-1
Page 3
UK/2008-06-02 This data sheet may only be reproduced in full.
www.baumerprocess.com
Design and specifications subject to change without notice
Data Sheet 2902-1
Page 4