REMOTE SEAL TYPE PRESSURE TRANSMITTER FKB...5 DATA SHEET The FCX -AIII pressure transmitter accurately measures gauge pressure and transmits a proportional 4 to 20mA signal. The transmitter utilizes a unique micromachined capacitance silicon sensor with state-of-the-art microprocessor technology to provide exceptional performance and functionality. Totally welded construction of the seals assures excellent reliability in high temperature and highly corrosive process conditions. FEATURES 1. High accuracy 0.2% accuracy for all calibrated spans is a standard feature for all GP models covering 1.3kPa {0.013bar} range to 50000kPa {500bar} high pressure range. 0.1% accuracy is available as option. Fuji’s micro-capacitance silicon sensor assures this accuracy for all elevated or suppressed calibration ranges without additional adjustment. 2. Minimum environmental influence The “Advanced Floating Cell” design which protects the pressure sensor against changes in tempera ture, and overpressure substantially reduces total measurement error in actual field applications. 3. Fuji/HART® bilingual communications protocol FCX -AIII series transmitter offers bilingual communications to speak both Fuji proprietary protocol and HART®. Any HART® compatible devices can communicate with FCX -AIII. 4. Application flexibility Various options that render the FCX -AIII suitable for almost any process applications include: − Full range of hazardous area approvals − Built-in RFI filter and lightning arrester − 5-digit LCD meter with engineering unit − Stainless steel electronics housing − Wide selection of materials − High temperature, high vacuum seals 5. Burnout current flexibility (Under Scale: 3.2 to 4.0mA, Over Scale: 20.0 to 22.5mA) Burnout signal level is adjustable using Model FXW Hand Held Communicator (HHC) to comply with NAMUR NE43. 6. Dry calibration without reference pressure Thanks to the best combination of unique construction of mechanical parts (Sensor unit) and high performance electronics circuit (Electronics unit), reliability of dry calibration without reference pressure is at equal level as wet calibration. SPECIFICATIONS Functional specifications Service: Liquid, gas, or vapour Span, range, and overrange limit: Span limit [kPa]{bar} Type F KB 1 FKB 2 FKB 3 FKB 4 FKB 5 Min. Max. 1.3 {0.013} 5 {0.05} 30 {0.3} 100 {1} 500 {5} 130 {1.3} 500 {5} 3000 {30} 10000 {100} 50000 {500} Range limit [kPa]{bar} −100 to +130 {−1 to +1.3} −100 to +500 {−1 to +5} −100 to +3000 {−1 to +30} −100 to +10000 {−1 to +100} −100 to +50000 {−1 to +500} Overrange limit [MPa] {bar} 1 {10} 1.5 {15} 9 {90} 15 {150} 75 {750} Remark: To minimize environmental influence, span should be greater than 1/40 of the max. span in most applications. − Lower range limit (vacuum limit) ; Silicone fill sensor: See Fig. 1, Fig. 2 Fluorinated fill sensor: Atmospheric pressure − Conversion factors to different units; 1MPa=103kPa=10bar=10.19716kgf/cm2=145.0377psi 1kPa=10mbar=101.9716mmH2O=4.01463inH2O Output signal: 4 to 20mA DC with digital signal superimposed on the 4 to 20mA signal. Power supply: Transmitter operates on 10.5V to 45V DC at transmitter terminals. 10.5V to 32V DC for the units with optional arrester. EDSX5-94h Date Aug. 10, 2011 FKB...5 Load limitations: see figure below [Ω] Load resistance 2000 1500 1474Ω R= E-10.5 (Imax+0.9)x10-3 1000 577Ω Operating area 500 24V 250 0 10 16.35 10.5V Power voltage (DC) 45V 50 E[ V ] Temperature limit: Ambient: −40 to +85°C (−20 to +80°C for LCD indicator) (−40 to +60°C for arrester option) (−10 to +60°C for fluorinated oil fill transmitter) (−10 to +85°C for silicone oil “H”, “S”, “K”) *) (+20 to +85°C for silicone oil “J”, “T”) *) *) In case of capillary length is more than 7m, max. temperature is +55°C. For explosionproof units (flameproof or intrinsic safety), ambient temperature must be within the limits specified by each standard. Process: required. Hazardous locations: SEE TABLE2 Zero/span adjustment: Zero and span are adjustable from the HHC(1). Zero and span are also adjustable externally from the adjustment screw. Damping: Adjustable from HHC or local adjustment unit with LCD display. The time const ant is configurator between 0.06 to 32 seconds. Zero elevation/suppression: Zero can be elevated or suppressed within the specified range limit of each sensor model. Normal/reverse action: Selectable from HHC(1). Indication: Analog indicator or 5-digit LCD meter, as specified. Burnout direction:Selectable from HHC(1) If self-diagnostic detect transmitter failure, the analog signal will be driven to either “Output Hold”, “Output Overscale” or “Output Underscale” modes. “Output Hold”: Output signal is hold as the value just before failure happens. “Output Overscale”: Adjustable within the range 20.0mA to 22.5mA from HHC(1) “Output Underscale”: Adjustable within the range 3.2mA to 4.0mA from HHC(1) 3.2 4 20 22.5 [mA] Over scale Burnout Normal operating range Under scale Probable under range Burnout Probable over range Output Limits comforming the NAMUR NE43 by order. Loop-check output: Transmitter can be configured to provide constant signal 3.2mA through 22.5mA by HHC(1). EMC Conformity: EN61326-1: 2006 (Note) 2 (1) HHC: Hand Held Communicator Fluorinated oil Code in the 13th digit of “Code symbols” W, A and D −20 to 120°C Silicone oil H −15 to 250°C Fill fluid Note: For communication with HHC(1) (Model: FXW), min. of 250 is Lower limit of static press. Process temperature Atmospheric pressure 20 to 300°C J Y and G −40 to 180°C S −15 to 250°C T 20 to 300°C K −15 to 200°C 2.7kPa abs {20mmHg abs} 0.13kPa abs {1mmHg abs} or more Storage: −40 to +90°C Humidity limit: 0 to 100% RH Communication: With HHC(1) (Model FXW, consult Data Sheet No. EDS8-47), following items can be remotely displayed or configured. Note: HHC’s version must be higher than 7.0 (or FXW 1– 4), for FCX AIII. Local configurator with LCD display (option): Local confi gurator with 3 push button and LCD display can support following items. Items By communication By local configurator with FXW (with 3 push button) Display Set Display Set Tag No. v v v v Model No. v v v v — Serial No. & Software Version v — v Engineering unit v v v v Range limit v — v — Measuring range v v v v Damping v v v v Output mode v — v — Burnout direction v v v v Calibration v v v v Output adjust — v — v Data v — v — Self diagnoses v — v — Printer (In case of FXW with printer option) v — — — External switch lock v v v v Transmitter display v v v v Linearize v v — — Rerange v v v v Saturate current v v v v Write protect v v v v History – Calibration history – Ambient temperature history v v v — v v v — Performance specifications Physical specifications Reference conditions, silicone oil fill, 316SS isolating diaphragms, 4 to 20mA analog output in linear mode. Accuracy rating: (including linearity, hysteresis, and repeatability) (Standard) For spans greater than 1/10 of URL: ±0.2% of span For spans below 1/10 of URL: 0.1 URL ± 0.1+0.1 % of span Span ( ) (Option) (Code; 21th digit H,K) Not available for Max span 50000kPa model. For spans greater than 1/10 of URL: ±0.1% of span For spans below 1/10 of URL: 0.1 URL ± 0.05+0.05 % of span Span ( ) Stability: ±0.2% of upper range limit (URL) for 10 years. Temperature effect: Effect per 28°C change between the limits of - 40°C and +85°C URL (Standard) Zero shift: ± 0.35 % x ( ) URL Total effect: ± (0.5 % x ) (Option) (Code; 21th digit J,K) Zero shift: ±0.3% (x 1/4 URL) 0.25 URL ± 0.1 + 0.2 % (x < 1/4 URL) x Total effect: ±0.4% (x 1/4 URL) 0.25 URL ± 0.2 + 0.2 % (x < 1/4 URL) x ( ) ( ) Overrange effect: Zero shift; 0.2% of URL for any overrange to maximum limit Supply voltage effect: Less than 0.005% of calibrated span per 1V Update rate: 60 msec Step response: Time constant: 0.3s (at 23°C) Dead time: 0.12s (without electrical damping) Dielectric strength: 500V AC, 50/60Hz 1 min., between circuit and earth. Insulation resistance: More than 100M/500V DC. Internal resistance for external field indicator: 12 or less Electrical connections: G1/2, 1/2-14 NPT, Pg13.5, or M20 1.5 conduit, as specified. Process connections: JIS, ANSI, or DIN raised face flanges or screw connection JIS/ISO G1 external thread. Refer to “Code symbols.” Process-wetted parts material: Diaphragm: 316L stainless steel, Hastelloy-C Monel, Tantalum, Titanium or Zirconium Flange face: 316 stainless steel, Hastelloy-C Monel, Tantalum, Titanium or Zirconium Extension: 316 stainless steel, Hastelloy-C (Refer to “Code symbols”) Non-wetted parts material: Electronics housing: Low copper die-cast aluminum alloy finished with polyester coating (standard), or 316 stainless steel (ASTM CF8M), as specified. Capillary: In case of 11th code “D, E, L, F, M, N, P”, PE armored stainless steel. In case of 11th code “Q, R, S, T, V, W, X”, stainless steel armored stainless steel. Mounting flange: 304 stainless steel or carbon steel, as specified Fill fluid: Silicone oil (standard) or fluorinated oil Mounting bracket: 304 stainless steel Environmental protection: IEC IP67 and NEMA 6/6P Mounting: On 60.5mm (JIS 50A) pipe using mounting bracket, direct wall mounting Mass {weight}: Transmitter approximately 8.2 to 11.2kg without options. Add; 0.5kg for mounting bracket 4.5kg for stainless steel housing option 1.5kg per 50mm extension of diaphragm 3 FKB...5 Optional features ACCESSORIES Indicator: A plug-in analog indicator (2.5% accuracy). An optional 5-digit LCD meter with engineering unit is also available. Local configurator with LCD display: An optional 5 digits LCD meter with 3 push buttons can support items as using communication with FXW. Arrester: A built-in arrester protects the electronics from lightning surges. Lightning surge immunity: 4kV (1.2 50µs) Oxygen service: Special cleaning procedures are followed throughout the process to maintain all process wetted parts oil-free. The fill fluid is fluorinated oil. Chlorine service: Oil-free procedures as above. Includes fluorinated oil for fill. Degreasing: Process-wetted parts are cleaned, but the fill fluid is standard silicone oil. Not for use on oxygen or chlorine measurement. Vacuum and high temperature service: Special silicone oil and filling procedure are applied. See Fig.1 and Fig.2. Optional tag plate: An extra stainless steel tag for customer tag data is wired to the transmitter. Coating of cell: Cell’s surface is finished with epoxy/polyurethane double coating. Specify if environment is extremely corrosive. Hand-held communicator: (Model FXW, refer to Data Sheet No. EDS8-47) ORDERING INFOMATION When ordering this instrument, specify. 1. CODE SYMBOLS 2. Measuring range. 3. Output orientation (burnout direction) when abnormality is occurred in the transmitter. Hold / Overscale / Underscale Unless otherwise specified, output hold function is supplied. 4. Indication method (indicated value and unit) in case of the actual scale (code D, H, P, S 2,5 on 9th digit). 5. Tag No. (up to 14 alphanumerical characters), if required. Silicone (Code Y,G) [kPa abs] {mbar abs} [kPa abs] {mbar abs} Silicone (Code S) Silicone (Code T) Sealing liquid:Silicone (Code K) 101 {1010} 53.3 {533} Operating pressure Operating pressure 101 {1010} Operating area 13.3 {133} Non-operating area 0.27 {2.7} 0.13 {1.3} 2.7 {27} –40 –15 20 60 85 120 180 Process temperature[°C ] 230 250 Fig. 1 Relation between process temperature and operating pressure 4 Operating area 300 –15 100 Process temperature[°C ] 200 Fig. 2 Relation between process temperature and operating pressure CODE SYMBOLS 1 2 3 4 5 6 7 8 Digit 4 5 <Conduit connection> Case type Conduit connection T type G1/2 1/2-14NPT T type T type Pg13.5 T type M20×1.5 L type G1/2 1/2-14NPT L type L type Pg13.5 L type M20×1.5 <Flanges> Mounting flange Flange size and rating 7 Ranges 1 2 * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * Note1: (*1) Note2: (*2) Note3: (*3) Note4: (*4) Diaph. extension [mm] 0 50 100 (*3) 150 200 0 Hastelloy-C 316 stainless steel 0 316 stainless steel Monel Tantalum Titanium Zirconium 0 0 0 0 0 9 10 11 12 13 14 15 21 5 Digit No. of code 0 1 3 4 6 7 9 M S T A B D E G H K V W X Y U P Q Note 2 Note 1 Note 1 <Material/diaphragm extension> Diaphragm Flange face 316L stainless 316 stainless steel steel Hastelloy-C 316L stainless +Au coating Hastelloy C Monel Tantalum Titanium Zirconium F KB 5 6 7 8 S T V W 3 4 5 JIS 10K 80A JIS 10K 100A ANSI/JPI 150LB "3" ANSI/JPI 150LB "4" DIN PN16/40 DN80 DIN PN16 DN100 JIS 20K 80A * JIS 30K 80A * ANSI/JPI 300LB 3B * ANSI/JPI 600LB 3B * Carbon steel JIS 10K 80A JIS 10K 100A ANSI/JPI 150LB "3" ANSI/JPI 150LB "4" DIN PN16/40 DN80 DIN PN16 DN100 Screw type, JIS/ISO G1 * * 316 stainless ANSI/JPI 150LB 3B * * * steel ANSI/JPI 150LB 4B * * * ANSI/JPI 300LB 3B * * * ANSI/JPI 300LB 4B * * * ANSI/JPI 600LB 3B * * * None 3 inch wafer * * (wafer type) 4 inch wafer * * <Span limit (*2) [kPa]{bar}> 1.3…130 {0.013…1.3} 5…500 {0.05…5} 30…3000 {0.3…30} 100…10000 {1…100} 500…50000 {5…500} Available only with material code “V” (*1) 304 stainless steel 6 Note Description (*4) Note 3 Note 4 1 2 3 4 5 V (*1) A B C D H J W M T P R If range 4 or 5 is selected, specify material “V” in any cases. 100: 1 turn down is possible, but should be used at a span greater than 1/40 of the maximum span for better performance. Available for 5th digit code “1”,“4”,“7”,“B”,“E”,“H”,“Q”, “W”, “Y”, when 13th digit code “S”, “T”, “K”. Available for 6th digit code “2”, “3” and 5th digit “0”, “3”, “6”, “9”, “A”, “D”, “G”, “P”, “M”, “S”, “T”, “U”, “V”, “X”. 5 FKB...5 1 2 3 4 5 6 7 8 Digit 9 10 11 12 Description <Indicator and arrester> Indicator Arrester None None Analog, 0 to 100% linear scale None Analog, custom scale None None Yes Analog, 0 to 100% linear scale Yes Analog, custom scale Yes Digital, 0 to 100% linear scale None Digital, custom scale None Digital, 0 to 100% linear scale Yes Digital, custom scale Yes Digital, 0 to 100% linear scale (Local configurator unit with LCD display) None Digital, custom scale (Local configurator unit with LCD display) None Digital, 0 to 100% linear scale (Local configurator unit with LCD display) Yes Digital, custom scale (Local configurator unit with LCD display) Yes <Approvals for hazardous locations> None (for ordinary locations) TIIS, Flameproof (Cable gland seal) (*11) TIIS, Intrinsic safety FM, Flameproof (or explosionproof) (*14) FM, Intrinsic safety and nonincentive FM Combined of flameproof and intrinsic safety (*14) ATEX Flameproof (*13) ATEX Intrinsic safety ATEX Type n ATEX Combined of flameproof and intrinsic safety (*13) IECEx Scheme, Flameproof (*13) IECEx Scheme, Intrinsic safety CSA, Flameproof (or explosionproof) (*14) CSA, Intrinsic safety and nonincentive NEPSI, Flameproof (or exprosionproof) NEPSI, Intrinsic safety NEPSI, Combined of flameproof and intrinsic safety <Capillary and mounting bracket> Capillary Mounting bracket Armor of capillary 1.5 m 304 Stainless steel PE (*5) 3 304 Stainless steel PE (*5) 5 304 Stainless steel PE (*5) 6 304 Stainless steel PE (*5) 7 304 Stainless steel PE (*5) 8 304 Stainless steel PE (*5) 10 304 Stainless steel PE (*5) 1.5 304 Stainless steel Stainless steel (*6) 3 304 Stainless steel Stainless steel (*6) 5 304 Stainless steel Stainless steel (*6) 6 304 Stainless steel Stainless steel (*6) 7 304 Stainless steel Stainless steel (*6) 8 304 Stainless steel Stainless steel (*6) 10 304 Stainless steel Stainless steel (*6) <Options> Extra SS tag plate Stainless steel elec. housing Coating of cell None None None Yes None None None (*7) None Yes Yes None Yes None Yes Yes (*8) Yes Yes Yes Note5: (*5) Note6: (*6) Note7: (*7) Note F KB A B D E F H L P Q S 1 2 4 5 Note 11 Note 14 Note 14 Note 13 Note 13 Note 13 Note 14 Note 5 Note 5 Note 5 Note 5 Note 5 Note 5 Note 5 Note 6 Note 6 Note 6 Note 6 Note 6 Note 6 Note 6 Note 7 Note 8 Note 8 Available for 13th digit code “Y, W, G, A, D”. Inquire about in case of 13th other code. Available for all 13th digit code. Customer tag number can be engraved on standard stainless steel name plate. If extra tag plate is required, select “Yes”. Note8: (*8) Not available for 10th digit code "C". Note11: (*11) Available for 4th digit code "S". Note13: (*13) Available for 4th digit code "6", "8", "T", "W". Note14: (*14) Available for 4th digit code "6", "T". 6 9 10 11 12 13 5 A C G D H V X K P M R T E J F S U D E L F M N P Q R S T V W X Y B M N P Q 14 15 21 Digit No. of code 1 2 3 4 5 6 7 8 Digit 13 14 15 21 Description Note <Special applications and fill fluid> Treatment Fill fluid Standard Silicone oil Standard Fluorinated oil Degreasing Silicone oil Oxygen service Fluorinated oil (7th digit code "V", "W", "A", "B", "C" and "D") Chlorine service Fluorinated oil (7th digit code "H", "T") High temp. 250°C Silicone oil Available for 6th digit code "1", "2" or "3", High temp. 300°C Silicone oil and available for 7th digit code "V", "W", "A", "B", "C", "D", "H" (*15) Note 15 High temp. and vacuum (250°C) Silicone oil Available for 6th digit code "2", "3" only, and High temp. and vacuum (300°C) Silicone oil available for 7th digit code "V", "W", "A", "B", High temp. and high vacuum Silicone oil "C" and "D" <Teflon membrane> None Yes (Available for the 5th digit code "0", "3", "6", "9", "M", "S", "T", "A", "D", "G", "P", "V", "X", "U", and 7th digit code "V", "W", "J". Not available for the 13th digit code "H", "J", "S", "T", "K".) Note 9 <Bolt/nut> (*9) None (6th digit code "1", "2", "3") Cr-Mo alloy hexagon socket head cap screw/carbon steel nut (6th digit code Cr-Mo alloy hexagon bolt/carbon steel unit "4", "5") 304 stainless steel/304 stainless steel (6th digit code "4") 630 stainless steel/304 stainless steel (6th digit code "5") 316 stainless steel/316 stainless steel (6th digit code "4") Note 10 <Other options> (*10) High accuracy type Instruction manual attached Low temperature effect type Instruction manual attached H+J Instruction manual attached Instruction manual unattached High accuracy type Instruction manual unattached Low temperature effect type Instruction manual unattached T+U Instruction manual unattached F KB 9 10 11 12 13 14 15 21 5 Digit No. of code Y W G A D H J S T K Y C Y A B E F U H J K L T U V Note9: (*9) In case of tropical use, select stainless bolts and nuts. Note10: (*10) If other option is not necessary, 21st digit code is blank. In case of 21st digit code is blank, instruction manual attached. Note15: (*15) Treatment; Standard 7 FKB...5 OUTLINE DIAGRAM (Unit:mm) <AMP. case: L type> < Flange type > IN CASE OF 4TH DIGIT CODE "S, V, W" IN CASE OF 4TH DIGIT CODE "T" φ81 103 φ81 103 8 10 ZERO ADJ. SCREW MTG. PIPE JIS 50A (2B) 38 U-BOLT M8 57 37 "B" CONDUIT CONN. 172 (21) (11) 4 HOLES 9 DIA. N HOLES h DIA WITH ANALOG IND. WITH DIGITAL IND. φ E P φH±1 φF φG φE φD FLANGE 20 L 162 t INNER OF GASKET 126 43 41 DETAIL "B" (CONDUIT CONN.) 198 φ15 EARTH TERMINAL J SEE TABLE1 4th digit of the code symbols S T V W 7th digit of the code symbols V, W J, H, M, T, P, R A B C D Conduit conn. J K M 18 2 G 1/2 4 1/2-14NPT 16 Pg13.5 10.5 4.5 M20×1.5 16 4 S(m) 1.5 3 5 6 7 8 10 2-M8X12 S K M 11th digit of the code symbols D, Q E, R L, S F, T M, V N, W P, X MASS. APPROX(kg) 8.2 9.2 10.2 10.7 11.2 L 0 50 100 150 200 H.P. 62 70 120 INNER OF GASKET TABLE 1 2-M6 SCREW DEPTH 9 0, A 9 M 1, B 3, D, V S, X T, U 4, E, W Y 6, G 7, H 185 200 210 210 191 210 210 229 254 200 220 CorrosionStandard resistant materials materials specification specification Note 1 Note 2 134 150 126 160 126 134 170 126 134 175 151 158 152.5 126 134 168 126 134 168 126 134 190.5 151 158 200 151 158 160 126 134 180 151 158 φE φG φH±1 t P N-φh FLANGE 100 100 100 103 100 100 100 103 103 100 103 73 73 73 96 73 73 73 96 96 73 96 38 42 48 38 44 49 52 44 52 44 40 116 116 116 141 116 116 116 141 141 116 141 8-19 8-23 8-23 8-19 4-20 8-23 8-23 8-20 8-23 8-18 8-18 JIS-10K-80A JIS-20K-80A JIS-30K-80A JIS-10K-100A ANSI 150LB 3B ANSI 300LB 3B ANSI 600LB 3B ANSI 150LB 4B ANSI 300LB 4B DIN PN40 DN80 DIN PN16 DN100 Q φD φG 5th digit of the code symbols φF φF 20 <WAFER TYPE> Q 80A, 3B 100A, 4B Note1) In case of the 7th digit of type code is V, A, B, C, D, J or W. Note2) In case of the 7th digit of type code is H, M, T, P or R. 116 141 OPTION PARTS FOR FLAMEPROOF OF TIIS (JAPAN) CONNECTION DIAGRAM <SS TAG PLATE> 10th digit code C (Cable gland type) NOTE1) G1/2 DETAIL C φ13 φ33.5 −/CK− CK+ 75 + 106 17 19 20 "C" CONDUIT CONN. G1/2 CABLE GLAND NOTE1) IN CASE OF 10TH CODE“C” , φ11 CABLE IS SUITBLE. 8 <AMP. case: T type> < Flange type > IN CASE OF 4TH DIGIT CODE "5, 7, 8" EARTH TERMINAL 8 10 IN CASE OF 4TH DIGIT CODE "6" 100 (21) (11) 38 U-BOLT M8 57 φ81 φ81 ZERO ADJ. SCREW 100 MTG. PIPE JIS 50A (2B) WITH ANALOG IND. 37 "B" CONDUIT CONN. 185 4 HOLES 9 DIA. WITH DIGITAL IND. N HOLES h DIA φ E P φH±1 φF φG φE φD FLANGE 20 L t 185 INNER OF GASKET 127 20 226 DETAIL "B" (CONDUIT CONN.) K 11th digit of the code symbols D, Q E, R L, S F, T M, V N, W P, X J SEE TABLE1 5 6 7 8 Conduit conn. J K M 18 2 G 1/2 4 1/2-14NPT 16 Pg13.5 10.5 4.5 4 M20×1.5 16 1.5 3 5 6 7 8 10 2-M8X12 MASS. APPROX(kg) 8.2 9.2 10.2 10.7 11.2 L 0 50 100 150 200 H.P. 62 70 120 TABLE 1 INNER OF GASKET φF 0, A 9 M 1, B 3, D, V S, X T, U 4, E, W Y 6, G 7, H 185 200 210 210 191 210 210 229 254 200 220 CorrosionStandard resistant materials materials specification specification Note 1 Note 2 150 134 126 134 160 126 170 126 134 175 151 158 134 152.5 126 134 168 126 134 168 126 158 190.5 151 158 200 151 134 160 126 158 180 151 φE φG φH±1 t P N-φh FLANGE 100 100 100 103 100 100 100 103 103 100 103 73 73 73 96 73 73 73 96 96 73 96 38 42 48 38 44 49 52 44 52 44 40 116 116 116 141 116 116 116 141 141 116 141 8-19 8-23 8-23 8-19 4-20 8-23 8-23 8-20 8-23 8-18 8-18 JIS-10K-80A JIS-20K-80A JIS-30K-80A JIS-10K-100A ANSI 150LB 3B ANSI 300LB 3B ANSI 600LB 3B ANSI 150LB 4B ANSI 300LB 4B DIN PN40 DN80 DIN PN16 DN100 Note1) In case of the 7th digit of type code is V, A, B, C, D, J or W. Note2) In case of the 7th digit of type code is H, M, T, P or R. CONNECTION DIAGRAM 20 <WAFER TYPE> Q 80A, 3B 100A, 4B 116 141 <SS TAG PLATE> −/CK− CK+ 75 + 2-M6 SCREW DEPTH 9 φF φD φG 5th digit of the code symbols Q 4th digit of the code symbols 7th digit of the code symbols V, W J, H, M, T, P, R A B C D S(m) S M φ15 20 9 FKB...5 <AMP. case: L type> < Screw type > IN CASE OF 4TH DIGIT CODE "S, V, W" ZERO ADJ. SCREW 103 φ81 103 φ81 8 IN CASE OF 4TH DIGIT CODE "T" 10 MTG. PIPE JIS 50A (2B) 38 37 "B" CONDUIT CONN. 172 4 HOLES 9 DIA. (21) (11) 20 U-BOLT M8 S 57 WITH ANALOG IND. 100 WITH DIGITAL IND. 14 45 G1 53 2 PACKING 162 126 41 φ23.6 43 φ30 EARTH TERMINAL K M 198 DETAIL "B" (CONDUIT CONN.) φ15 2-M8X12 J SEE TABLE1 11th digit of the code symbols D, Q E, R L, S F, T M, V N, W P, X S(m) 62 1.5 3 5 6 7 8 10 70 120 4th digit of the Conduit conn. code symbols J K M S G 1/2 18 2 T 1/2-14NPT 16 4 V Pg13.5 10.5 4.5 W M20×1.5 16 4 TABLE 1 OPTION PARTS FOR FLAMEPROOF OF TIIS (JAPAN) CONNECTION DIAGRAM <SS TAG PLATE> 10th digit code C (Cable gland type) NOTE1) G1/2 φ13 φ33.5 −/CK− CK+ 75 + DETAIL C 106 17 19 20 "C" CONDUIT CONN. G1/2 CABLE GLAND NOTE1) IN CASE OF 10TH CODE“C” , φ11 CABLE IS SUITBLE. 10 <AMP. case: T type> < Screw type > IN CASE OF 4TH DIGIT CODE "6" 8 EARTH TERMINAL MTG. PIPE JIS 50A (2B) 10 IN CASE OF 4TH DIGIT CODE "5, 7, 8" (21) (11) 38 WITH ANALOG IND. 37 S 57 φ81 φ81 100 100 ZERO ADJ. SCREW 4 HOLES 9 DIA. 185 WITH DIGITAL IND. U-BOLT M8 100 20 "B" CONDUIT CONN. PACKING 2 53 14 45 G1 185 127 φ23.6 20 φ30 226 DETAIL "B" (CONDUIT CONN.) K M φ15 2-M8X12 J SEE TABLE1 11th digit of the code symbols D, Q E, R L, S F, T M, V N, W P, X 4th digit of the code symbols 5 6 7 8 S(m) 62 70 1.5 3 5 6 7 8 10 120 Conduit conn. J K G 1/2 18 1/2-14NPT 16 Pg13.5 10.5 M20×1.5 16 M 2 4 4.5 4 TABLE 1 CONNECTION DIAGRAM −/CK− CK+ 75 + <SS TAG PLATE> 20 11 FKB...5 TABLE 2 Authorities ATEX Intrinsic safety Authorities Ex II 1 G Ex ia IIC T5 Tamb = –40°C to +50°C Ex ia IIC T4 Tamb = –40°C to +70°C Entity Parameters: Ui=28V, Ii=94.3mA, Pi=0.66W, Ci=26nF (Without Arrester), Li=0.6mH (Without analog indicator), Ci=36nF (With Arrester), Li=0.7mH (With analog indicator) Factory Mutual 13th digit Y,G,H,J,S,T,K Y,G,H,J,S,T,K Y,G,H,J,S,T,K Y,G,H,J,S,T,K W,A,D Factory Mutual Class I Div.1 Groups B, C, D T6 Type 4X Class II III Div.1 Groups E, F, G T6 Type 4X Tamb max = +60°C CSA Class I Div.1 Groups C, D Class II Div.1 Groups E, F, G Class III Div.1 Note) “Seal Not Required” enclosure is allowed. Class I Div.1 Groups A, B, C, D Class II Div.1 Groups E, F, G Class III Div.1 Temp Code T5 Tamb max = +50°C Temp Code T4 Tamb max = +70°C Entity Parameters: Vmax=28V, Imax=94.3mA, Ci=25nF (Without Arrester), Ci=36nF (With Arrester),Li=0.6mH (Without analog meter), Li=0.7mH (With analog meter) TIIS Ex do IIB+H2 T4 Tamb max = +60°C Maximum process temp. = +120°C IECEx Scheme Ex d IIC T5 IP66/67 Tamb = −40°C to +85°C Ex d IIC T6 IP66/67 Tamb = −40°C to +65°C NEPSI Ex d IIB+H2 T6 Tamb = −40°C to +60°C Ex ia IIC T4 Tamb max = +60°C Entity Parameters: Ui=28V, Ii=94.3mA, Pi=0.66W, Ci=38.4nF, Li=0.694mH IECEx Scheme Ex ia IIC T4 Tamb = –40°C to +70°C Ex ia IIC T5 Tamb = –40°C to +50°C Entity Parameters: Ui=28V, Ii=94.3mA, Pi=0.66W, Ci=26nF (Without Arrester), Li=0.6mH (Without analog indicator), Ci=36nF (With Arrester), Li=0.7mH (With analog indicator) 13th digit Y,G,H,J,S,T,K Y,G,H,J,S,T,K Y,G,H,J,S,T,K Y,G,H,J,S,T,K A,W,D Tamb –40°C to +85°C –20°C to +80°C –20°C to +60°C –40°C to +60°C –10°C to +60°C Factory Mutual (pending) Entity Parameters: Ui=42.4V, Ii=113mA, Pi=1W, Ci=35.98nF, Li=0.694mH CSA Caution on Safety Ex II 3 GD EEx nL IIC T5 Tamb = –40°C to +50°C EEx nL IIC T4 Tamb = –40°C to +70°C Specific Parameters: Model without arrester: Ui=42.4V, Ii=113mA, Pi=1W, Ci=25.18nF, Li=0.694mH Model with arrester: Ui=32V, Ii=113mA, Pi=1W, Ci=35.98nF, Li=0.694mH EEx nAL IIC T5 Tamb = –40°C to +50°C EEx nAL IIC T4 Tamb = –40°C to +70°C Specific Parameters: Model without arrester: Umax=42.4V, Imax=113mA, Pmax=1W Model with arrester: Umax=32V, Imax=113mA, Pmax=1W Ex ia IIC T4 Ex d IIB+H2 T6 / Ex ia IIC T4 Model code 9th digit A,B,D L,P,1,2 Q,S,4,5 E,F,H – Type n Nonincendive Authorities ATEX TIIS NEPSI Ex II 2 GD Ex d IIC T6 IP66/67 T85°C Tamb = −40°C to +65°C Ex d IIC T5 IP66/67 T100°C Tamb = −40°C to +85°C Tamb –40°C to +85°C –20°C to +80°C –20°C to +60°C –40°C to +60°C –10°C to +60°C Entity Parameters: Vmax=28V, Imax=94.3mA, Pi=0.66W, Ci=35.98nF, Li=0.694mH CSA ATEX Class I II III Div.1 Groups A, B, C, D, E, F, G T4 Entity Type 4X Model code 9th digit A,B,D L,P,1,2 Q,S,4,5 E,F,H – Flameproof Class I II III Div.2 Groups A, B, C, D, F, G T4 Entity Type 4X Model code 9th digit A,B,D L,P,1,2 Q,S,4,5 E,F,H – 13th digit Y,G,H,J,S,T,K Y,G,H,J,S,T,K Y,G,H,J,S,T,K Y,G,H,J,S,T,K W,A,D Tamb –40°C to +85°C –20°C to +80°C –20°C to +60°C –40°C to +60°C –10°C to +60°C Class I Div.2 Groups A, B, C, D Class II Div.2 Groups E, F, G Class III Div.2 Temp Code T5 Tamb max = +50°C Temp Code T4 Tamb max = +70°C Entity Parameters: Vmax=28V, Ci=25.18nF (Without Arrester), Ci=35.98nF (With Arrester), Li=0.694mH *Before using this product, be sure to read its instruction manual in advance. International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fjielectric.com/products/instruments/ Information in this catalog is subject to change without notice. Printed in Japan