FUJI EDS8-42H

STOP VALVE & CONDENSATE CHAMBER
DATA SHEET
FXJ, FXK
In measuring pressures, flow rates or levels, a process piping
and a pressure piping to feed pressures from a process tank
to a transmitter are required.
Fuji's stop valve and condensate chamber are designed to
be used with a pressure piping system to provide proper
maintenance of transmitters and better measurement
accuracy.
Both the stop valve and condensate chamber are available
in different operating pressure ratings to permit selection
according to applications.
SPECIFICATIONS
Stop Valve (Type:FXJ)
Application:
The stop valve is used to remove a transmitter from a pressure piping when maintenance of the transmitter is required.
It can be readily fitted to the pressure
piping.
Material of main unit:
Materials of other parts:
Operating pressure and temperature:
Symbol Material of
main unit
Temperature
[°C]max.
Materials of Operating
other parts pressure
2
FCD400
1MPa{10.20kgf/cm }
A
SUS 403
FCD450
2MPa{20.39kgf/cm2}
C
SUS 304
2MPa{20.39kgf/cm2}
SUS304
D
SUS 304
15MPa{153.0kgf/cm2}
ASTM A182F11 SUS 403
F
24MPa{244.7kgf/cm2}
ASTM A182F22 SUS 403
G
11MPa{112.2kgf/cm2}
SF440A
H
SUS
403
*
28MPa{285.5kgf/cm2}
SF440A
SUS 403
*J
7MPa{71.38kgf/cm2}
SUS304
K
SUS 304
Note: Symbols A to K denote the 4th digit of the code.
200
250
250
425
525
360
425
300
Type of connection:
Inlet
Rc1/2
Rc1/2
φ 22.2 socket
Outlet
Application
Rc1/2
φ 9.6 union
φ 22.2 socket
FXJA, C, D, K
FXJA, C, D
FXJF, G, J, H
Operating pressure and temperature, and material:
2MPa·350°C max., JIS STPG370
{20.39kgf/cm2}
6MPa·400°C max., JIS STPT370
{61.18kgf/cm2}
15MPa·500°C max., ASTM A182 F11
{153.0kgf/cm2}
24MPa·525°C max., ASTM A182 F22
{244.7kgf/cm2}
Type of connection:
R1/2
Rc1/2
1/2B pipe
1/2B socket
1/2, 1B flange
(JIS 10, 16, 20, 30K)
Note: For the arrangement of inlet and outlet,
refer to Table given in page 2
Water inlet:
Pressure: 6MPa{61.18kgf/cm2} max. for
standard provision
Pressure: 15MPa{153.0kgf/cm2} min. for
non-standard provision
Ranges of Operating Temperature and
Pressure
Mass {weight}:
FXJA 0.5kg FXJC, D 2kg
FXJK
3kg FXJF,G,H,J
3.5kg
Finish color:
Silver (excluding SUS parts)
Range of delivery:
Stop valve
Condensate Chamber (Type:FXK)
Application:
The condensate chamber is used with a
differential pressure transmitter when
measuring boiler drum level or stream
flow rate. It keeps condensation at a
fixed level for better measurement
accuracy.
EDS8-42h
Date
May. 30, 2000
FXJ, FXK
CODE SYMBOLS
Stop Valve
1 2 3 4 5 6 7 8
F XJ
9
Description
3
Material/pressure
Material of
Materials of
main unit
other parts
FCD400
SUS403
FCD450
SUS304
SUS304
SUS304
ASTMA182F11 SUS403
ASTMA182F22 SUS403
SF440A
SUS403
SF440A
SUS403
SUS304
SUS304
A
C
D
F
G
*H
*J
K
Operating pressure
1MPa {10.20kgf/cm2}
2MPa {20.39kgf/cm2}
2MPa {20.39kgf/cm2}
15MPa {153.0kgf/cm2}
24MPa {244.7kgf/cm2}
11MPa {112.2kgf/cm2}
28MPa {285.5kgf/cm2}
7MPa {71.38kgf/cm2}
Temperature
Connection
[°C]max
Rc1/2
200
Rc1/2
250
Rc1/2
250
1/2 socket (φ22.2)
425
1/2 socket (φ22.2)
525
1/2 socket (φ22.2)
360
1/2 socket (φ22.2)
425
Rc1/2
300
Connection
Screw, Rc1/2
Union
Socket
10
20
30
Treatment
None
Oil repulsion (SUS304 only)
Y
A
Y
A
Stainless steel tag plate
None
Yes
Condensate Chamber
1 2 3 4 5 6 7 8
F XK
3
Description
Material, pressure And temperature
Operating
Temperature
End
Drum
pressure
[°C]max
plate
1
2
3
4
STPG370 SPHC-P
STPT370 SF440A
ASTM A182 F11
ASTM A182 F22
350
400
500
525
Type of connection
(1) Type of inlet is entered in the 5th digit of
the code, and the type of outlet in the 6th
digit.
(2) Symbols entered in the 5th and 6th digits
are referred to Table given below.
R1/2
Rc1/2, with union joint (screw type)
Rc1/2, with socket (screw type)
1/2B pipe (welded to socket)
1/2B, with sockert (welded to socket)
JIS 10K 15A, with flange
JIS 16K 15A, with flange
JIS 20K 15A, with flange
JIS 30K 15A, with flange
JIS 10K 25A, with flange
JIS 16K 25A, with flange
JIS 20K 25A, with flange
JIS 30K 25A, with flange
Other
A
B
C
D
E
F
G
H
J
K
L
M
N
Z
Y
A
2
2MPa {20.39kgf/cm2}
6MPa {61.18kgf/cm2}
15MPa {153.0kgf/cm2}
24MPa {244.7kgf/cm2}
Stainless steel tag plate
None
Yes
Arrangement of the 5th and 6th digits of the
code following the symbol in the 4th digit
4th 5th
digit digit
A
1
B
C
F
G
H
K
L
M
A
2
C
D
E
H
J
M
N
E
3
E
4
6th digit
Fig.No.
A B C D E F G H J
Fig. 6
Fig. 7, 8
Fig. 9, 10
Fig. 11
Fig. 12
Fig. 13, 14
Fig. 15, 16
Fig. 17
Fig. 18
OUTLINE DIAGRAM (Unit:mm)
Stop Valve
3
FXJ, FXK
Condensate Chamber
4
5
FXJ, FXK
Caution on Safety
Asterisked (*) items: Non-standard.
*Before using this product, be sure to read its instruction manual in advance.
Head Office
Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome,
Shinagawa-ku, Tokyo 141-0032, Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan
Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187
http://www.fic-net.jp/eng
Information in this catalog is subject to change without notice.