MA-COM PSL0XSS60

CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications
P Series
UL Recognized
CSA Certified
VDE Approved
PS Unfiltered
PS000SS30
PS Filtered
The P series power entry module offers fuller flexibility of
design in a more compact package.
H models provide susceptibility protection without leakage current associated with line-to-ground capacitors and
are designed to allow equipment to meet UL 544 for patient
care and non-patient care equipment.
As the first 10 amp module to provide all five power entry
functions in one compact design, the CHAMELEON
module readily adapts to its environment and the needs of
international markets.
Also available is an interconnection block, B models.
The block connects the voltage selection terminals of an
unfiltered CHAMELEON module with a switch and an IEC
connector to reduce external wiring.
• Snap-in or flange mounting
L Models – High performance medical filter versions that
are designed to help bring most digital equipment (including switching power supplies) into compliance with
EN55022, Level B (as well as FCC part 15J, Class B)
conducted emissions limits while additionally allowing
equipment to meet UL544, UL2601and EN60601 for patient
care equipment and non-patient care equipment. They are
available with current ratings of 6 and 10 amps. High
performance versions are available with horizontal mounting ears, single voltage, complete can shield with options
for switch, fuses and current ratings only. Extenders are
not available.
P Series
• IEC power line connector
• Both North American and European fusing capabilities
• Two voltage selections
• Optional DPST on/off switch
• Filtering options for general purpose, medical and
high-performance applications.
The CHAMELEON modules compact design and modular
construction will allow you to select the power entry
features you require — without altering the panel cutout.
And the Chameleon module, with adapters, will fit any
standard panel cutout.
The CHAMELEON module has four filter options.
S models offer protection for general purpose applications where line-to-ground and line-to-line noise must be
controlled. The filters are designed to meet low leakage
requirements of VDE portable equipment. They are
available in current ratings of 3, 6 and 10 amps.
Z Models – High performance filter versions are designed
to help bring most digital equipment (including switching
power supplies) into compliance with EN55022, Level B
(as well as FCC Part 15J, Class B) conducted emissions
limits. They are available with current ratings of 6 and 10
amps. High performance versions are available with
horizontal mounting ears, no extenders, single voltage,
complete can shield with options for switch, fuses and
current ratings only.
PE Filter with "A" Shield
132
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
P Series Part Number Schematic
Part numbers are constructed by selecting the alphanumeric character which represents the desired feature.
Mounting Extender
Style
Options*
Switch
Voltage
Fuse
Options ‡ Select ** Options
Filter
Type
Filter
Current
Shield
Options*
P S J 0 X S S 6 0
Switch Option
0 = Blank
S = Double pole
single throw
(DPST)
Extender
Option
0 = None
C = FN260
J = J Series
L = L Series
Filter Current
0 = No filter
3 = 3 amps
6 = 6 amps
X = 10 amps
Fuse Options
D = Dual fuse
S = Single fuse
Input Voltage Select
0 = Single voltage
S = Dual voltage **
select (SMPS)
X = Dual voltage **
select (series/parallel)
Notes:
An additional extender which allows the P series to adapt to Corcom's L
series product cutout as well as the panel cutout of Schaffner and Delta
parts is also available.
• Extenders cannot be added to units with B shields
† When using the interconnection block, the last 3 digits of the part number
are BX (0, A, or B)
‡ For alternative switch orientation options, please contact technical sup
port or your Corcom sales representative
** Dual voltage options only available with filter types O, H or S
••• High performance versions are available with horizontal mounting ears
(PE style), no extenders, single voltage, complete can shield with options
for switch, fuses, filter type (L or Z) and 6 or 10 amp current ratings only.
Power Entry Modules
Mounting Style
E = Mounting ears
(Horizontal)
S = Snap-in
M = Mounting ears
(Vertical)
Shield Options
Filter Type
0 = No filter
0 = No shield
B = Interconnection
A = Filter shield
B = Complete
block ✝
Shield
H = Medical Filter
L = High Performance
C = Complete can
Medical
(for L & Z models)
S = General purpose
Z = High Performance ***
Extender for
J-Series or others
Fuse holder
Optional switch
The part number PSJ0XSS60 would represent the equivalent of the Corcom
J Series 6J4:.
P Series (P)
with a snap-in mount (S)
with the J extender to fit the 6J4 cutout(J), and includes no switch (0)
dual voltage select (X)
single fusing (S),
general purpose filter (S)
for 6 amps (6)
with no shield (0).
Optional filter
Additional extender which allows the P series to adapt to CORCOM's L
series product cutout, as well as the panel cutout of Schaffner and Delta
parts is available.
* Extenders cannot be added to units with B shields.
133
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
Voltage Selection
Input Voltage Selection Schemes
The P series of power entry modules provides a dual voltage selector which is integrated with the fuseholder. To
simplify the wiring in each individual application, Tyco
Electronics provides dedicated voltage selectors for different voltage selection schemes. The two most popular
schemes are the parallel and serial for dual primary transformers and the “DPST switch” for switching mode power
supplies (jumper types.) P series products with an “S” as
the fifth digit are specifically designed for “jumper” type
applications associated with switching mode power supplies. P series products with an “X” as the fifth digit are
specifically designed for dual primary transformer applications. The voltage selector installed will allow proper wiring
from these applications. No matter what voltage selection
scheme is used, wiring is always made to the same four
terminals.
X - Dual Primary Transformer Type
115V
230V
Power on/off
Hot
G
Note 1
A
Note 2
B
Ground
Note 1
C
Note 2
Neutral
115V
115V
D
H
Transformer
primary
Provision for
single fusing
S - Jumper Type
115V
230V
Power on/off
Hot
G
Note 1
Note 1
Note 1
Note 1
B
Ground
Neutral
A
C
H
Power
input
Sel.
Jumpers
D
SMPS
Provision for
single fusing
Note 1: Jumpers are required if a filter or interconnection module is not
used.
Note 2: Location of optional filter. Jumpers are required if a filter or
interconnection block is not used.
Shield
A new feature available on the P series is an RF shield. The
metal shield, available on filtered models, provides shielding from radiated emissions and provides an additional RF
ground for the filter to the case. The shield is available in
two versions, a shield of the filter components (designated
by A as the ninth digit) and a complete shield (designated
by B as the ninth digit). The A shield covers the filter portion of the module and increases performance of the filter
by protecting the components from magnetic coupling.
This shield improves RF ground connection to the case
while still allowing the use of the P Series extender. The B
shield covers the entire power entry module with metal,
protecting the equipment from all radiated noise. Used
with the filter, the shield provides the most secure protection from RFI noise problems. The B shield cannot be used
with any extender.
A Shield
B Shield
134
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
Fuseholder
Power Entry Modules
Another feature of the P series power entry module is a versatile fusing arrangement. Its fuseholder can hold two 1/4”
x 1-1/4” or 5 x 20mm fuses. It can also be converted to
accept one fuse with a conversion clip that bridges one of
the two fuse chambers. The conversion clip is installed on
the P Series power entry modules set for single fusing
(those with part numbers with an “S” as the sixth digit).
Units with a “D” as the sixth digit do not include a conversion clip.
Fuseholder shown with 5x20mm fuse
Conversion clip included for use in single
fuse applications.
Interconnection Block
Also available is an interconnection block (B models). The
block connects the voltage selection terminals of an unfiltered CHAMELEON module with a switch and an IEC connector to reduce external wiring. Installation of the unfiltered versions of the P series involves wiring of the IEC
socket to the switch and the switch to the fuseholder. Labor
can be eliminated by ordering the product with an interconnection block. This feature, designated by a BX in the seventh and eighth digits, pre wires the module for easier
installation. The wiring is protected with the plastic case to
better prevent access to the terminals and connections.
The dimensions of this alternative are the same as the
filtered versions.
Interconnection Block installed in place of filer.
Accessories
A full range of accessories are available for the P Series.
To simplify customer installation wiring can be simplified by
ordering a PA100 connector assembly which is pre-wired
with 6" of wire lead.
See page 143 for a complete list of all accessories
available for this series.
PA100 Assembly
135
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
Electrical Schematics
Minimum insertion loss in dB:
H Model
Line-to-ground in 50 ohm circuit
S Model
Current
.03
Rating
H Models
03A
7
06A
10A
3
S Models
03A
7
06A
10A
-
.1
.15
Frequency-MHz
.5 1
3
5
10
30
17
8
5
21
11
8
27 30
15 17
10 12
29
19
11
26
18
11
23
16
10
15
13
10
17
8
3
21
12
5
27 33
17 23
10 13
40
32
23
44
36
27
50
44
35
32
30
27
.1
Frequency-MHz
.15 .5
1
5
10
30
27
22
30 37
24 27
43
32
49
52
52
47
42
40
27
22
29 34
23 24
35
25
25
21
21
18
16
14
10
30
48
40
40
41
35
32
50
55
55
45
45
45
L Model
LINE
LOAD
Z Model
LINE
LOAD
Specifications
Maximum leakage current, each line-to-ground: †
H & L Models
S & Z Models
@ 120 VAC 60 Hz:
2 µA
.25 mA
.50 mA
@ 250 VAC 50 Hz:
5 µA
†Applies to Filtered models only
Hipot rating (one minute):
line-to-ground
1500 VAC
line-to-line
1450 VAC
Operating frequency:
50/60 Hz
Rated voltage:
250 VAC
Operating voltages:
115/230 VAC
Selectable or Fixed
Rated current:
Unfiltered – 10A
Filtered – 3, 6 or 10A
Switch:
Double-insulated, rated
for 10,000 operations at full load at
51 Amp inrush current.
Fuseholder:
Accepts one or two fuses
1/4 x 1-1/4 or 5 x 20 mm
Conversion clip provided on fuseholder
for single fuse models.
Current
Rating
.01 .05
Z Models
06A
8
21
10A
5 17
L Models
06A
8 21
10A
5
17
Line-to-line in 50 ohm circuit
Current
Frequency-MHz
Rating
.10 .15 .5 1
3
5
H Models
03A
2
4 12 18 31 40
06A
2
4 12 16 26 35
10A
2
4 12 16 26 33
S Models
03A
2
4 12 15 30 48
06A
2
4 12 15 22 42
10A
2
4 12 15 22 42
Current
Rating
.01 .05
Z Models
06A
10 15
10 20
10A
L Models
10 15
06A
10A
10 20
.1
Frequency-MHz
1
5
.15 .5
10
30
34
20
44 75
35 67
75
75
75
75
70
70
60
60
34
20
44 75
35 67
75
75
75
75
70
70
60
60
Terminals:
Standard versions:
.187 x 0.032
[4.8 x .81] terminal tabs
High performance versions:
.250 x 0.032
[6.4 x .81] terminal tabs
Recommended for use with mating connector - no solder
136
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
Case Styles
PS/PSL
.06 R Max
1.52
C
E
C
(Note 1)
.225
5.72
A
.225
5.72
B1
D
B2
Unfiltered
Power Entry Modules
.020 R Min
.508
Panel Cutout
Filtered
Note1: For snap-in applications, the D sides of the cutout must have a .02 [.508] radius on the installation side.
Snap-in models allow front mounting only. Not recommended for plastic panels.
.06 R Max
1.52
PE
E
C
C
.020 R Min
.508
.225 Max.
5.72
D
B1
.225 Max.
5.72
D
B2
F
F
Unfiltered
A
PM
Panel Cutout
Filtered
A
D
D
F
E
C
.225
5.72
Max.
B1
Unfiltered
Typical dimensions:
Terminals: .187 x 0.032 [4.8 x .81] terminal tabs
Slot (Ground Terminal): .06 x .115 [1.52 x 2.92] ± .004 [.10]
.225
5.72
Max.
B2
F
.06
1.52
R Max
Filtered
Panel Cutout
Holes: .055 [1.5] Dia. ± .004 [.10]
Mounting Holes (PE, PM only) : .135 [3.43] Dia. (2) countersunk
137
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
Case Styles (continued)
PSC/PSJ
.06 R Max
1.52
C
E
(Note 1)
.020
R Min
.508
.225
5.72
Max.
A
B1
Unfiltered
D
.225
5.72
B2
Filtered
Panel Cutout
Note 1: For snap-in applications, the D sides of the cutout must have a .02 [.508] radius on the installation side. Snap-in models, allow front mounting only.
Typical dimensions:
Terminals: .187 x 0.032 [4.8 x .81] terminal tabs
Slot (Ground Terminal): .06 x .115 [1.52 x 2.92] ± .004 [.10]
Holes: .055 [1.5] Dia. ± .004 [.10]
Mounting Holes (PE, PM only) : .135 [3.43] Dia. (2) countersunk
PE High Performance
3.93 Max.
C
F
.225
5.72
A
2.39 Max.
2.17 Max.
B2
D
Rear View
Note 1: Use cutout for PE Style.
Typical dimensions:
Terminals: .250 x 0.032 [6.4 x .81] terminal tabs
Slot (Ground Terminal): .07 x .16 [1.8 x 4.1]
Holes: .07 [1.8] Dia. (2)
Mounting Holes: .135 [3.43] Dia. (2) countersunk
138
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com
CORCOM Product Guide
The CHAMELEON Adaptable Module for General, High Performance or
Medical Applications (Continued)
P Series
Shielding
Case Dimensions
Part No.
A
B1
B2
C
(max) (max) (max) (max)
PS
PSL
2.19 Max
PE
0.225
.508
Max.
2.19 Max
A Shield
B Shield
Shields can be used only with filtered models.
Shields add approximately 0.06" to depth.
B shield may not be used with J or C extender.
PM
1.93
1.24
2.13
1.93
PE
High
A B C D
A B C D
N
L
GND
2.31
1.12 2.201*
54.1 58.67 28.45 55.91
1.98
2.13
1.93
2.31
F
—
55.91
31.49 49.02
—
1.122 2.201* 1.575
50.29 49.02
54.1 58.67
28.5
55.91 40.0
1.24
2.13
1.06
2.13 2.496
1.93
1.24
1.93
1.24
1.93
31.49 49.02
Rear View
∆ 2.201*
1.06∆
54.1 58.67 26.93
31.49 49.02
PSJ
2.31
31.49 49.02
31.49 49.02
PSC
2.13
E
1.98
50.29
Performance
—
2.9
54.1 73.66 26.93
54.1
63.4
2.13
2.52
—
2.51
1.06
54.1 63.75 26.93 64.01
2.13 2.72 1.06
2.6+
54.1 69.09 26.93
4.42
Power Entry Modules
0.225
.508
Max.
1.24
D
+.008/-.000 +.008/-.000
+.20/-.00
+.20/-.00
—
66.0
2.31
1.12
2.201 1.575
112.27 58.67
28.5
55.91 40.0
* Panel cutout for thickness of .031 - .079 [0.8 - 2.0]. For panel thickness
at .083 - .114 [2.1 - 2.9] use 2.213 [56.21].
+ Panel cutout for thickness of 0.06 - 0.09 [0.76 - 2.29].
∆ For shielded models use D=1.12 +.008/-.000.
GND
Filtered
Unfiltered
139
Catalog 1654001
Revised 05-07
Dimensions are in inches and
millimeters unless otherwise
specified. Values in italics
are metric equivalents.
Dimensions are shown for
reference purposes only.
Specifications subject
to change.
USA Cust. Svc.: 1-800-468-2023
CORCOM Prods: 1-847-680-7400
Germany:
49-89-6089-0
www.corcom.com
www.tycoelectronics.com