650-700 WATT MEDICAL & ITE POWER SUPPLIES DESCRIPTION PM650 SERIES (LVD) The PM650 series comprising single and multiple output models for 650-700 watts of continuous output power is RoHS specially designed for medical and ITE applications, not for life-support. They operate at 90-264 VAC input voltage without the need of a selector strap. The units are constructed on a printed circuit board with a U-bracket for mechanical support and heat sinking. A cover-and-fan assembly can be added during manufacturing. FEATURES z z z z z z z z z z EN61000-3-2 class A and D compliant Power Factor 0.98 typical Overvoltage protection Short-circuit protection Thermal protection Power Fail Detect (PFD) signal 100% burn-in at full rated load Remote sense on output #1 and output #2 Remote inhibit – TTL high to disable output Compliant with RoHS requirements SAFETY STANDARD APPROVALS UL 60601-1, CSA C22.2 No. 601.1 File No. E178020 TÜV EN 60601-1 UL 60950-1, CSA C22.2 No. 60950-1 TÜV EN 60950- 1 INPUT SPECIFICATIONS Input voltage: Input frequency: Input current: Earth leakage current: 90-264 VAC 47-63 Hz 10 A (rms) for 115 VAC 5 A (rms) for 230 VAC 240 μA max. @ 264 VAC, 63 Hz ENVIRONMENTAL SPECIFICATIONS Operating temperature: Storage temperature: Relative humidity: Derating: 0℃ to +70℃ -40℃ to +85℃ 5% to 95% non-condensing Derate from 100% at +50℃ , linearly to 50% at +70℃ OUTPUT SPECIFICATIONS GENERAL SPECIFICATIONS Output voltage/current: Maximum output power: Ripple and noise: Overvoltage protection: Switching frequency: 70 KHz ±10 KHz Power factor: 0.98 typical Efficiency: 80% minimum on all models Hold-up time: 12 ms minimum at 110 VAC Overcurrent protection: Temperature coefficient: Transient response: Fan power: See rating chart. See rating chart. 1% peak to peak maximum Provided on output #1 only; set at 115-140% of its nominal output voltage All outputs protected to short circuit conditions All outputs ±0.04% / ℃ maximum Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change 12 V at 400 mA maximum for B version, 12 V at 100 mA maximum for C version ±0.2% maximum at full load 50 A @115 VAC or 100 A @ 230 VAC at 25 ℃ cold start 4000 VAC from input to output, Withstand voltage: 1500 VAC from input to ground, 500 VAC from output to ground 300,000 hours minimum at full load at MTBF: 25 ℃ ambient, calculated per MIL-HDBK- 217F EMC Performance (IEC60601-1-2) Class B conducted, Class A radiated EN55011: Harmonic distortion, Class A and D EN61000-3-2: Line regulation: Inrush current: INTERFACE SIGNALS EN61000-3-3: Line flicker PFD: EN61000-4-2: ESD, ±8 KV air and ±6 KV contact EN61000-4-3: EN61000-4-4: EN61000-4-5: EN61000-4-6: EN61000-4-8: EN61000-4-11: Radiated immunity, 3 V/m Fast transient/burst, ±2 KV Surge, ±1 KV diff., ±2 KV com. Conducted immunity, 3 Vrms Magnetic field immunity, 3 A/m Voltage dip immunity, 30% reduction for 500 ms, 60% reduction for 100 ms and >95% reduction for 10 ms Inhibit: TTL logic high for normal operation and TTL logic low upon loss of input power. This signal appears at least 1 ms prior to master output dropping 5% below its nominal value. This signal also provides a minimum delay of 100 ms after master output is within regulation. Requires an external TTL high level signal to inhibit outputs for standard models 37 UNIVERSAL INPUT PM650 MEDICAL & ITE SERIES OUTPUT VOLTAGE/CURRENT RATING CHART (1) Model PM650-12B PM650-13B PM650-14B PM650-15B PM650-16B PM650-17B PM650-18B PM650-20B PM650-21B PM650-22B PM650-23B PM650-24B PM650-30B PM650-31B PM650-32B PM650-33B PM650-34B PM650-35B PM650-36B PM650-37B V1 12 V 15 V 24 V 27 V 30 V 36 V 48 V 24 V 24 V 48 V 48 V 48 V 24 V 24 V 24 V 24 V 48 V 48 V 48 V 48 V Output #1 (3)(5) Imin. 0A 0A 0A 0A 0A 0A 0A 1.50 A 1.50 A 0.75 A 0.75 A 0.75 A 1.50 A 1.50 A 1.50 A 1.50 A 0.75 A 0.75 A 0.75 A 0.75 A Imax. 54.2 A 43.4 A 27.1 A 24.1 A 21.7 A 18.1 A 14.6 A 18.0 A 18.0 A 9.0 A 9.0 A 9.0 A 18.0 A 18.0 A 18.0 A 18.0 A 9.0 A 9.0 A 9.0 A 9.0 A (5) Output #2 Tol. ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2% V2 12 V 15 V 24 V 12 V 15 V 12 V 15 V 12 V 15 V 12 V 15 V 12 V 15 V Imin. (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) N/A 1.0 A 0.6 A 1.2 A 1.0 A 1.2 A 1.0 A 1.2 A 1.0 A 1.2 A 1.0 A 1.2 A 1.0 A Output #3 Imax. Tol. V3 22 A 18 A 12 A 22 A 18 A 22 A 18 A 22 A 18 A 22 A 18 A 22 A 18 A ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% ±5% 3.3 V 3.3 V 5.1 V 5.1 V 3.3 V 3.3 V 5.1 V 5.1 V NOTES: 1. Suffix “B” in model numbers denotes U-bracket form. Change “B” to “C” for enclosed form with cover-and-fan assembly, e.g. PM650-14C. 2. All outputs are floating. They can be connected externally for positive or negative output. 3. Output #1 can be adjusted within +/- 5% of their nominal voltage. 4. Output #3 can be adjusted within +/- 15% of their nominal voltage. 5. 650-700 watts for "C" version with cover-and-fan assembly. 325-350 watts for "B" version without moving air (maximum current of output #1 and #2 derated to 70%), or 650-700 watts with 50 CFM forced air provided by user. 6. All models may be operated at no-load. At no-load, output voltage tolerance increases to +/- 10%. 7. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output. MECHANICAL SPECIFICATIONS Single Output Models U-bracket Form Enclosed Form 38 Imin. (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) (N/A) 0A 0A 0A 0A 0A 0A 0A 0A (4) Imax. 10 A 10 A 10 A 10 A 10 A 10 A 10 A 10 A Tol. ±3% ±3% ±3% ±3% ±3% ±3% ±3% ±3% Max. Output (5) Power 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 350 W / 700 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W 325 W / 650 W OUTPUT POWER DERATING CURVE UNIVERSAL INPUT PM650 MEDICAL & ITE SERIES MECHANICAL SPECIFICATIONS Multiple Output Models U-bracket Form Enclosed Form NOTES: 1. Dimensions shown in inches [mm] 2. Tolerance 0.02 [0.5] maximum 3. Input connector P1 is Dinkle DT-4C-B01W-03 with M3, nickel-plated screws. 4. Connector P4 mates with Molex housing 50-37-5103 and pins 5263. 5. Connector P2-1, P2-2, P3-1 & P3-2: M3*0.5 screw connections. 6. Connectors P2, P3: M3*0.5 screw connections 7. Output connector P5 is Dinkle DT-35-B01W-06. Screws are M3, nickel plated. 8. Weight: 2.0 Kgs. (4.4 lbs.) approx. for U-bracket form, 2.2 Kgs. (4.84 lbs.) approx. for enclosed form. 9. Maximum penetration depth of fixing screws is 4 mm from the outer surface of chassis. PIN CHART CONN MODEL PM650-12B PM650-13B PM650-14B PM650-15B PM650-20B PM650-21B PM650-22B PM650-30B PM650-31B PM650-32B PM650-33B MODEL PM650-12B PM650-13B PM650-14B PM650-15B PM650-20B PM650-21B PM650-22B PM650-30B PM650-31B PM650-32B PM650-33B PIN PM650-16B PM650-17B PM650-18B PM650-23B PM650-24B PM650-34B PM650-35B PM650-36B PM650-37B CONN PIN PM650-16B PM650-17B PM650-18B PM650-23B PM650-24B PM650-34B PM650-35B PM650-36B PM650-37B P1 (AC) P2 P5 P3 1 2 3 1 2 Live Neutral Ground +V1 V1 Return Live Neutral Ground +V1 V1 Return +V2 Live Neutral Ground +V1 V1 Return 3 4 5 6 V2 Return N.A. N.A. +V2 V2 Return +V3 V3 Return N.A. P4 1 2 3 4 5 6 7 8 9 10 PFD Return +V1 Sense -V1 Sense PFD Inhibit +V Inhibit -V N.C. N.C. Fan Return +12V Fan PFD Return +V1 Sense -V1 Sense PFD Inhibit +V Inhibit -V +V2 Sense -V2 Sense Fan Return +12V Fan PFD Return +V1 Sense -V1 Sense PFD Inhibit +V Inhibit -V +V2 Sense -V2 Sense Fan Return +12V Fan 39