PM450 - PROTEK POWER

450-480 WATT MEDICAL & ITE POWER SUPPLIES
DESCRIPTION
PM450 SERIES
(LVD)
The PM450 series comprising single and multiple output
models for 450-480 watts of continuous output power is
RoHS
specially designed for medical and ITE applications, not for
life-supporting equipment. They operate at 90-264 VAC input
voltage without the need of a selector strap. The units are
constructed on a printed circuit board with a U-bracket for
mechanical
support
and
heat
sinking.
A cover-and-fan
assembly can be added during manufacturing.
FEATURES
z
z
z
z
z
z
z
z
z
z
SAFETY STANDARD APPROVALS
EN61000-3-2 class A and D compliant
Power Factor 0.98 typical
Overvoltage protection
Short-circuit protection
Thermal protection
Power Fail Detect (PFD) Signal
100% burn-in at full rated load
Remote sense on output #1 and output #2
Remote inhibit – TTL high to disable output
Compliant with RoHS requirements
UL 60601-1, CSA C22.2 No. 601.1
File No. E178020
TÜV EN 60601-1
UL 60950-1, CSA C22.2 No. 60950-1
TÜV EN 60950-1
ENVIRONMENTAL SPECIFICATIONS
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
90-264 VAC
47-63 Hz
7.1 A( rms) for 115 VAC
3.5 A( rms) for 230 VAC
240 μA max.@ 264 VAC, 63 Hz
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
GENERAL SPECIFICATIONS
OUTPUT SPECIFICATIONS
Output voltage/current:
See rating chart.
Maximum output power:
Ripple and noise:
Overvoltage protection:
See rating chart.
1% peak to peak maximum
Provided on output #1 only; set
at 115-140% of its nominal
output voltage
All outputs protected to short
circuit conditions
All outputs ±0.04% /℃ maximum
Maximum excursion of 4% or
better on all models, recovering
to 1% of final value within 500us
after a 25% step load change
Overcurrent protection:
Temperature coefficient:
Transient response:
Fan power:
0 ℃ to +70 ℃
-40 ℃ to +85 ℃
5% to 95% non-condensing
Derate from 100% at +50℃
linearly to 50% at +70℃
Switching frequency:
60 KHz ±10 KHz
Power factor:
0.98 typical
Efficiency:
80% minimum on all models
Hold-up time:
12 ms minimum at 110 VAC
Line regulation:
±0.2% maximum at full load
40 A @ 115 VAC or 80 A @ 230 VAC, at
25 ℃ cold start
4000 VAC from input to output,
Withstand voltage:
1500 VAC from input to ground,
500 VAC from output to ground
300,000 hours at full load at 25 ℃
MTBF:
ambient, calculated per MIL-HDBK-217F
EMC Performance (IEC60601-1-2)
Inrush current:
12 V at 400 mA for B version,
12 V at 100 mA for C version
EN55011 / EN55022
Class B conducted, Class A radiated
FCC:
Class B conducted, Class A radiated
VCCI:
Class B conducted, Class A radiated
INTERFACE SIGNALS
EN61000-3-2:
Harmonic distortion, Class A and D
PFD:
EN61000-3-3:
Line flicker
EN61000-4-2:
ESD, ±8 KV air and ±6 KV contact
EN61000-4-3:
Radiated immunity, 3V/m
Inhibit:
TTL logic high for normal operation and
TTL logic low upon loss of input power.
This signal appears at least 1ms prior to
master output dropping 5% below its
nominal value. This signal also provides a
minimum delay of 100 ms after master
output is within regulation.
Requires an external TTL high level signal
to inhibit outputs for standard models.
EN61000-4-4:
Fast transient/burst, ±2 KV
EN61000-4-5:
Surge, ±1 KV diff., ±2 KV com
EN61000-4-6:
Conducted immunity, 3 V/ms
EN61000-4-8:
Magnetic field immunity, 3 A/m
Voltage dip immunity, 30% reduction
for 500 ms, 60% reduction for 100 ms,
and >95% reduction for 10 ms
EN61000-4-11:
33
UNIVERSAL INPUT
PM450 MEDICAL & ITE SERIES
OUTPUT VOLTAGE/CURRENT RATING CHART
(3)(5)
(5)
Output #1
(1)
Model
PM450-12B
PM450-13B
PM450-14B
PM450-15B
PM450-16B
PM450-17-1B
PM450-18B
PM450-19B
PM450-20B
PM450-21B
PM450-22B
PM450-23B
PM450-24B
PM450-30B
PM450-31B
PM450-32B
PM450-33B
PM450-34B
PM450-35B
PM450-36B
PM450-37B
V1
12 V
15 V
24 V
27 V
30 V
40 V
48 V
55 V
24 V
24 V
48 V
48 V
48 V
24 V
24 V
24 V
24 V
48 V
48 V
48 V
48 V
Imin.
0A
0A
0A
0A
0A
0A
0A
0A
1.0 A
1.0 A
0.5 A
0.5 A
0.5 A
1.0 A
1.0 A
1.0 A
1.0 A
0.5 A
0.5 A
0.5 A
0.5 A
Imax.
37.5 A
30.0 A
18.75 A
16.7 A
15.0 A
12.0 A
10.0 A
8.73 A
12.0 A
12.0 A
6.0 A
6.0 A
6.0 A
12.0 A
12.0 A
12.0 A
12.0 A
6.0 A
6.0 A
6.0 A
6.0 A
Tol.
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
±2%
V2
12 V
15 V
24 V
12 V
15 V
12 V
15 V
12 V
15 V
12 V
15 V
12 V
15 V
Output #3
Imin.
Imax.
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
1.00 A
17 A
0.75 A
15 A
0.50 A
10 A
1.00 A
17 A
0.75 A
15 A
1.00 A
17 A
0.75 A
15 A
1.00 A
17 A
0.75 A
15 A
1.00 A
17 A
0.75 A
15 A
1.00 A
17 A
0.75 A
15 A
NOTES:
1.
Suffix “B” in model numbers denotes U-bracket form. Change “B”
to “C” for enclosed form with cover-and-fan assembly, e.g.
PM450-14C.
2.
All outputs are floating. They can be connected externally for positive or
negative output.
3.
Output #1 can be adjusted within ±5% of its nominal voltage.
4.
Output #3 can be adjusted within ±15% of its nominal voltage.
5.
450-480 watts for “C” version with cover-and-fan assembly. 225-240
watts for “B” version without moving air (maximum current of output #1 &
2 derated to 50%), or 450 watts with 40 CFM forced air provided by user.
6.
All models may be operated at no-load. At no-load, output voltage
tolerance increases to ±10%.
MECHANICAL SPECIFICATIONS
Single Output Models
U-bracket Form
(4)
Output #2
Tol.
V3
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
3.3 V
3.3 V
5.1 V
5.1 V
3.3 V
3.3 V
5.1 V
5.1 V
Tol.
±3%
±3%
±3%
±3%
±3%
±3%
±3%
±3%
Max. Output
(5)
Power
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
240 / 480 W
240 / 480 W
240 / 480 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
225 / 450 W
OUTPUT POWER DERATING CURVE
Enclosed Form
34
Imin. Imax.
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
(N/A)
0A
8A
0A
8A
0A
8A
0A
8A
0A
8A
0A
8A
0A
8A
0A
8A
UNIVERSAL INPUT
PM450 MEDICAL & ITE SERIES
Multiple Output Models
U-bracket Form
NOTES:
1.
2.
3.
4.
5.
6.
7.
8.
Enclosed Form
Dimensions shown in inches [mm]
Tolerance 0.02 [0.5] maximum
Input connector P1 is Dinkle DT-35-B01W-03 with M3, nickel plated screws.
Connector P4 mates with Molex housing 50-37-5103 and pins 5263.
P2, P2-1, P2-2, P3, P3-1 & P3-2: M3*0.5 screw connections
Output connector P5 is Dinkle DT-35-B01W-04 with M3, nickel plated screws.
Weight: 1.8 Kgs (3.96 lbs.) approx. for U-bracket form, 2.0 Kg (4.4 lbs.) approx. for enclosed form
Maximum penetration depth of fixing screws is 4 mm from the outer surface of chassis.
PIN CHART
CONN
MODEL
PM450-12B
PM450-13B
PM450-14B
PM450-15B
PM450-20B
PM450-21B
PM450-22B
PM450-30B
PM450-31B
PM450-32B
PM450-33B
PIN
PM450-16B
PM450-17-1B
PM450-18B
PM450-19B
PM450-23B
PM450-24B
PM450-34B
PM450-35B
PM450-36B
PM450-37B
1
P1 (AC)
2
3
Live
Neutral
Live
P3
Ground
+V1
V1 Return
Neutral
Ground
+V1
V1 Return
+V2
Live
Neutral
Ground
+V1
V1 Return
+V2
1
2
3
4
5
PIN
PM450-16B
PM450-17-1B
PM450-18B
PM450-19B
PM450-23B
PM450-24B
6
7
8
9
10
N.C.
N.C.
Fan
Return
+12V
Fan
PM450-34B
PM450-35B
PM450-36B
PM450-37B
1
2
3
4
V2
Return
N.C.
N.C.
V2
Return
+V3
V3
Return
N.A.
P4
CONN
MODEL
PM450-12B
PM450-13B
PM450-14B
PM450-15B
PM450-20B
PM450-21B
PM450-22B
PM450-30B
PM450-31B
PM450-32B
PM450-33B
P5
P2
PFD
Return
+V1 Sense -V1 Sense
PFD
Inhibit
+V
Inhibit
-V
PFD
Return
+V1 Sense
-V1 Sense
PFD
Inhibit
+V
Inhibit
-V
+V2 Sense -V2 Sense
Fan
Return
+12V
Fan
PFD
Return
+V1 Sense
-V1 Sense
PFD
Inhibit
+V
Inhibit
-V
+V2 Sense -V2 Sense
Fan
Return
+12V
Fan
35