REM10-S_D(W)

Features
• Reinforced Insulation for 250VAC Working Voltage
• Clearance and Creepage Distance: 8mm
• 5kVAC I/P to O/P 2MOPP Isolation
• 2µA Patient Leakage Current
• Industry Standard Pinout
• 2:1 and 4:1 Wide Input Range
Regulated
Converters
Description
Selection Guide
Part
Number
Input
Voltage Range
(VDC)
Output
Output
Voltage Current
(VDC)(mA)
10 Watt
5
2:1 & 4:1
DIP24
Single and Dual
Output
3.3
REM10-xx05S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75
5
2000 84 / 85.5 / 86.5 / 86.5
2500
REM10-xx12S/
4.5-9 / 9-18 / 18-36 / 36-75
12
830
86.5 / 88 / 89 / 89
430
REM10-xx15S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75
15
670
87 / 89 / 89 / 89
350
(3,4)
2500
80 / 83 / 83 / 82.5
4.5-9 / 9-18 / 18-36 / 36-75
24
416
85.5 / 89 / 89 / 88.
125
±5
±1000
83 / 84 / 85 / 85
±1440
REM10-xx12D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75
±12
±416
85.5 / 89 / 89 / 88 ±250
REM10-xx15D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75
±15
±333
86.5 / 88 / 89 / 88
±180
REM10-xx3.3SW/
(3,4)
9-36 / 18-75
3.3
2500
83 / 82.5
3000
REM10-xx05SW/ (3,4)
9-36 / 18-75
5
2000
86.5 / 86.5
2500
REM10-xx12SW/
9-36 / 18-75
12
830
89 / 89
430
REM10-xx15SW/ (3,4)
9-36 / 18-75
15
670
89 / 89
350
(3,4)
9-36 / 18-75
24
416
89 / 88.5
125
REM10-xx05DW/ (3,4)
9-36 / 18-75
±5
±1000
85 / 85 ±1440
REM10-xx12DW/ (3,4)
9-36 / 18-75
±12
±416
89 / 88
±250
REM10-xx15DW/ (3,4)
9-36 / 18-75
±15
±333
88 / 88
±180
(3,4)
rranty
3000
REM10-xxx05D/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75
(3,4)
wa
EAR
Efficiency Max. Capacitive
typ.
Load
(%)
(µF)
REM10-xx3.3S/ (3,4) 4.5-9 / 9-18 / 18-36 / 36-75
REM10-xx24SW/
REM10
5Y
The REM10 series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute
isolation with low 2µA leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The
compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions.
The outputs are short circuit and overload protected. The converters are available in two different pinning options and optionally with an external control pin for standby consumption as low as 12.5mW. The
converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry the CE and UL marks.
REM10-xx24S/
DC/DC Converter
E314885
2MOPP
250VAC
Model Numbering
REM10- __ __ __ __ / __ __
Output Power
CTRL Option (4)
Input Voltage (1)
Pinning (3)
Output Voltage
Input Type (2)
UL-60950-1 Certified
EN-55011 Certified
EN-55022 Certified
IEC-60601-1 Certified
ES-60601-1 Certifed
Single or Dual
Notes:
Note1: for 4:1 Input Voltage Type add “W”, see Note 2.
2:1
4:1 “W”
xx= 4.5-9 Vin = “05”
xx= 9-36Vin =
xx= 9-18Vin = “12”
xx= 18-75Vin = xx= 18-36Vin = “24”
xx= 36-75Vin = “48”
“24”
“48”
Note2: Blank for Standard 2:1 Input Voltage Range; „W“ suffix for 4:1 Input Voltage Range
Note3: „A“ suffix for A pinning; „C“ suffix for C pinning, for more details refer to Package Style and Pinning
Note4: „CTRL“ suffix for control pin option, for A pinning only, for C pinning not aviable Examples:
REM10-0512D/A
REM10-1215S/C
REM10-4815SW/A/CTRL
REM10-243.3SW/C
www.recom-power.com
=
=
=
=
2:1 Input,
2:1 Input,
4:1 Input,
4:1 Input,
4.5-9Vin,
9-18Vin,
36-75Vin,
9-36Vin,
±12Vout,
15Vout,
15Vout,
3.3Vout,
pinout „A“,
pinout „C“,
pinout „A“
pinout „C“,
without control pin
without control pin
with control pin
without control pin
REV.: 03/2015
Refer to Applications Notes
MED-1
REM10
DC/DC Converter
Series
Specifications (measured at TA= 25°C, nominal input voltage, full load and after warm-up)
BASIC CHARACTERISTICS
Parameter
Condition
2:1
Absolute Maximum Input Voltage
( 3sec max.)
4:1
2:1
Typ.
Max.
16VDC
25VDC
50VDC
100VDC
50VDC
100VDC
5Vin nom.
12Vin nom.
24Vin nom.
48Vin nom.
24Vin nom.
48Vin nom
Under Voltage Lockout
4:1
Start-up Time
Min.
5Vin nom.
12Vin nom.
24Vin nom.
48Vin nom.
24Vin nom.
48Vin nom
4VDC
8VDC
16VDC
33VDC
8VDC
16VDC
4.5VDC
9VDC
18VDC
36VDC
9VDC
18VDC
constant resistive load, Power up or Remote ON/OFF
30ms
DC-DC ON
DC-DC OFF
Open or 0-1.2VDC
2.2-12VDC
Remote ON/OFF
(referenced to -Vin Pin)
Current of CTRL Pin
-0.5mA
1mA
Remote OFF Input Current
2.5mA
Operating Frequency Range
270kHz
Output Ripple and Noise
(20MHz BW limited)
300kHz
10µF/25V 7XR MLCC for 3.3, 5Vout
10µF/25V 7XR MLCC for 12, 15Vout
4.7µF/50V X7R MLCC for 24Vout
330kHz
30mVp-p
40mVp-p
50mVp-p
Efficiency
Table1: Efficiency Crosstable
Graph1: Efficiency Factor vs. Load
1.2
Efficiency Crosstable (%) @ full load
Input Voltage
24
48
24W
48W
80
83
83
82.5
83
82.5
05S
84
85.5
86.5
86.5
86.5
86.5
12S
86.5
88
89
89
89
89
15S
87
89
89
89
89
89
24S
85.5
89
89
88.5
89
88.5
05D
83
84
85
85
85
85
12D
855
89
89
88
89
88
15D
86.5
88
88
88
88
88
Efficiency Factor
12
3.3S
Output Voltage
1.0
5
0.8
0.6
0.4
0.2
0.0
0
10
20
30
40
50
60
70
80
90 100
Load (%)
continued on next page
MED-2
REV.: 03/2015
www.recom-power.com
REM10
DC/DC Converter
Series
Specifications (measured at TA= 25°C, nominal input voltage, full load and after warm-up)
Calculation Example:
Efficiency (%)
100
90
choose your model:
80
REM10-1212D
70
• Efficiency from Table1 (= 89% @ max Load / nom Vin)
60
50
• Loading conditions in application (= 50%)
10 20 30 40 50 60 70 80
• use Eff factor from Graph1 (= 0.99)
Load (%)
Calculation:
Eff50%=Eff100% * Efffactor50% = 89 * 0.991 = 88.19%
Vin = 12V
Pout100% * 0.5
Iout = 50%
PDIS50%= Pin50% - Pout50% =
- (Pout100% * 0.5) = 5.67 - 5 = 0.67W
Eff50%
Eff100% = 89%
Efffactor50% = 0.99
TOVER =Rth * PDIS50% = 18 * 0.67 = 12.1°C
Rth = 18°C/W
TCASEmax = 105°C
T
= T
- T = 105 - 12.1 = 92.9°C
AMBmax
CASEmax
90 100
OVER
REGULATIONS
Parameter
Condition
Type
Value
Output Voltage Accuracy
±1%
Line Voltage Regulation
Load Voltage Regulation
Cross Regulation
Transient Response
LL to HL
LL to HL
Single
Dual
±0.2%
no load to full load
no load to full load
Single
Dual
±0.2%
±0.5%
±1%
asymmetrical load 25% / Full Load only Dual Output
±5%
25% load step change
250µs
PROTECTIONS
Parameter
Condition
Type
Short Circuit Protection (SCP)
Over Load Protection (OLP)
Value
continuous, auto-recovery
% of Iout rated
Hiccup mode, 150% typ.
Single
3.3Vout
5Vout
12Vout
15Vout
24Vout
3.7VDC min. / 5VDC max.
5.6VDC min. / 7VDC max.
13.5VDC min. / 16VDC max.
18.3VDC min. / 22VDC max.
29.1VDC min. / 34.5VDC max.
Dual
5Vout
12Vout
15Vout
5.6VDC min. / 7VDC max.
13.5VDC min. / 18.2VDC max.
17VDC min. / 22VDC max.
Output Over Voltage Protection (OVP)
continued on next page
www.recom-power.com
REV.: 03/2015
MED-3
REM10
DC/DC Converter
Series
Specifications (measured at TA= 25°C, nominal input voltage, full load and after warm-up)
Isolation Voltage
I/P to O/P
working voltage
5kVAC / 1 minute
250VAC / continuous
Means of Protection
2MOPP
Leakage Current
240VAC, 60Hz
2µA
Medical Device Classification
Type CF applied device (design to meet)
Internal Clearance
Creepage
I/P to O/P
I/P to O/P
External Clearance
I/P to O/P, Single Output
I/P to O/P, Dual Output
I/P to O/P, Single Output
I/P to O/P,
Dual Output
Creepage
>8mm
>8mm
>19.72mm
>14.64mm
>19.72mm
>14.64mm
Isolation Capacitance
12pF typ. / 17pF max.
Insulation Grade
Notes:
Note5:
Reinforced Insulation
This Power module is not internally fused. A input line fuse must be always used.
Recomended Fuse:
2:1 Input Voltage
Fuse (slow blow)
4:1 Input Voltage
Fuse (slow blow)
5V
T5A
24V
T2A
12V
T2A
48V
T1A
24V
T1A
48V
T0.5A
ENVIRONMENTAL
Parameter
Condition
Relative Humidity
Temperature Coefficient
Thermal Impedance
MTBF (+25°C)
Value
5% to 95% RH
±0.02% / °C
natural convection (20LFM)
according to MIL-HDBK-217F, full load
max. Case Temperature Range
max. Ambient Temperature Range
18°C / W
3849 x 103 hours
-40°C to +105°C
see calculation example
SAFETY AND CERTIFICATIONS
Certificate Type
CB Medical Safety
Report / File Number
E314885-A6
1409015
Standard
IEC-60601-1
Medical Report + ISO14971 Risk Assessment
ANSI/AAMI
E314885-A6
ES60601-1
CAN/CSA Medical
E314885-A6
C22.2 No. 60601-1:08
Certificate Type (Environmental)
Conditions
Standard / Criterion
EMI Standard
(7)
Conducted
EN-55011 (EN-55022), Class A, B
Radiated
EN-55011 (EN-55022), Class A, B
Conducted and Radiated
FCC18
Air ±8kV; Contact ±6kV
EN61000-4-2, Criteria A
Radiated Immunity
10V/m
EN61000-4-3, Criteria A
Fast Transient (6)
±2kV
EN61000-4-4, Criteria A
±2kV
EN61000-4-5, Criteria A
20Vr.m.s
EN61000-4-6, Criteria A
ESD
Surge
(6)
Conducted Immunity
continued on next page
MED-4
REV.: 03/2015
www.recom-power.com
REM10
DC/DC Converter
Series
Specifications (measured at TA= 25°C, nominal input voltage, full load and after warm-up)
Power Frequency Magnetic Field
10A/m
EN61000-4-8, Criteria A
Thermal Shock
MIL-STD-810F
Vibration
MIL-STD-810F
Notes:
Note6: An external input filter capacitor is required if the model has to meet EN61000-4-4 or/and EN61000-4-5.
Recommended components:
5Vin
aluminium capacitor (Nippon Chemi-con KY series, 1000µF/25V) and a
reverse diode (Vishay V10P45) to connect in parallel
12Vin, 24Vin
aluminium capacitor (Nippon Chemi-con KY series, 470µF/50V)
48Vin
aluminium capacitor (Nippon Chemi-con KY series, 330µF/100V)
Note7:
The whole REM10 series can meet EMI Class A with no external filter. And Class B only with external components.
EMC Filter Suggestion for Class B
L1
L2
C1
C2
DC/DC
Converter
+VIN
C3
-VIN
MODEL
C1
REM10-05xxS_D
N/A
REM10-12xxS_D
REM10-24xxS_D
C3
22µF/16V
22µF/16V
MLCC
MLCC
4.7µF/50V
4.7µF/50V
MLCC
MLCC
1µF/100V
4.7µF/100V
MLCC
MLCC
REM10-24xxS_D/W
REM10-48xxS_D
C2
REM10-48xxS_D/W
Notes:
Note8:
L1
L2
3.3µH; 3.3A
52µH (8)
N/A
10µH; 2.3A
175µH (8)
N/A
10µH; 2.3A
419µH (8)
For more details, please contact our technical support service at [email protected]
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Type
Case Material
Potting Material
Value
non-conductive black plastic
Silicone (UL94-V0)
Package Dimension (LxWxH)
31.80 x 20.30 x 10.40mm
Package Weight
14g
continued on next page
www.recom-power.com
REV.: 03/2015
MED-5
REM10
DC/DC Converter
Series
Specifications (measured at TA= 25°C, nominal input voltage, full load and after warm-up)
Dimension Drawing (mm)
“C” Pinning
“A” Pinning (Standard)
20.3
REM10-Series
MEDICAL GRADE DC/DC CONVERTER
20.3
MEDICAL GRADE DC/DC CONVERTER
REM10-Series
31.8
15.24
2.54
3.8
3.8
10.4
10.4
31.8
2
15.24
20.32
2.54
Pin Connections
Pin # Single
Dual
1 +Vin+Vin
11
No Pin
Com
12
-Vout No Pin
13 +Vout-Vout
15
No Pin +Vout
23 -Vin-Vin
24 -Vin-Vin
Bottom View
24 23
15
20.32
Pin Connections
Pin # Single
Dual
1 CTRL*CTRL*
2 -Vin-Vin
11
No Pin -Vout
14 +Vout+Vout
16 -VoutCom
22 +Vin+Vin
23 +Vin+Vin
11 12
1
2 5.08
2.54
0.6 dia.
13
1 2
0.6 dia.
11
Bottom View
23 22
16 14
* If don´t choose CTRL option, there is
no pin on the corresponding pin number
Tolerance: xx.x= ±0.5mm
xx.xx= ±0.25mm
Tolerance: xx.x= ±0.5mm
xx.xx= ±0.25mm
PACKAGING INFORMATION
Parameter
Type
Value
Packaging Dimension (LxWxH)
Tube
255 x 21.8 x 16.5mm
Packaging Quantity
7pcs
Storage Temperature Range
-55°C to +125°C
16.5±0.75
11.5±0.75
0.5±0.2
Tube Dimension Drawing (mm)
11±0.75
21.8±0.75
The product information and specifications are subject to change without prior notice. RECOM products are not authorized for use in safety-critical applications (such as life support) without RECOM’s explicit written
consent. A safety-critical application is defined as an application where a failure of a RECOM product may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The buyer
shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications.
MED-6
REV.: 03/2015
www.recom-power.com