Features • • • • • • • • Regulated Converters Long 5 Year Warranty 2MOPP/250VAC Suitable for built in Class II Applications Wide Input Voltage Range (85-264VAC) Low Leakage Current (<100µA) 5000m Operation Active Power Factor Correction Connector Set available AC/DC Converter RACM150 Description 150 Watt 5 Enclosed Case Style Single Output 5Y The RACM150-S(/F) is a compact 4.6” x 2.44” high efficiency AC/DC power supply with 2xMOPP safety approval for medical applications. These space saving enclosed power supplies have a universal input voltage range (85-264VAC), 4kVac isolation, require no minimum load and can be used at ambient temperatures of between -25°C and +80°C. The 12V, 15V, 24V or 48V output voltages are fully protected and have tolerances of less than ±0.2% over the entire input voltage range and less than ±0.5% over the entire load range. The RACM150-S(/F) series is certified to medical safety standard IEC/ES/EN-60601-1 3rd Edition and feature BF rated outputs with less than 100µA leakage current. It has a built-in Class B EMI filter and comes with a five year warranty. wa EAR rranty Selection Guide Part Number Input Output Output Efficiency Peak Output Voltage Range Voltage Current (A) typ. Power (W) (VAC) (VDC)115/230VAC(%)115/230VAC RACM150-12S 85-264 12 10.0 / 10.84 91 120 / 130 RACM150-15S 85-264 15 8.33 / 9.0 92 125 / 135 RACM150-24S 85-264 24 5.2 / 5.63 92 125 / 135 RACM150-48S 85-264 48 2.5 / 2.71 91 120 / 130 RACM150-12S/F 85-264 1212.591 150 (1) RACM150-15S/F (1) 85-264 1510.092 150 RACM150-24S/F (1) 85-264 246.2592 150 RACM150-48S/F (1) 85-264 483.1391 150 Notes: Note1: Please note that removing the fan from the /F version will not give the same performance as the equivalent fanless type. The two versions are not identical. Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up) 2MOPP 250VAC BASIC CHARACTERISTICS Parameter Condition Min. Typ. 85VAC 120VDC Input Voltage Max. 264VAC 370VDC Input Current 115VAC, fulll load 230VAC, full load 1.7A 0.8A Inrush Current cold start, 115VAC cold start, 230VAC 30A 60A 230VAC, with fan 230VAC, without fan No Load Input Power Input Frequency Range AC Input 0.6W 0.25W 47Hz 0.7s Rise Time 20ms Hold up Time 30ms Minimum Load 1s 0% Power Factor 0.95 Operating Frequency (measured @ 20MHz BW) 1W 0.3W EN-55011 Certified EN-55022 Certified ES-60601 Certified IEC/EN-60601 Certified 63Hz Start-up Time Output Ripple and Noise E314885 60kHz 12VDC, with 1µF/25V MLCC 15VDC, with 1µF/25V MLCC 24VDC, with 1µF/50V MLCC 48VDC, with 0.1µF/100V MLCC 120mVp-p 150mVp-p 220mVp-p 250mVp-p continued on next page www.recom-power.com Refer to Applications Notes REV.: 2/2015 MED-1 RACM150(/F) AC/DC Converter Series Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up) Efficiency vs. Load 230VAC 90 80 115VAC 70 60 50 10 20 30 40 50 60 Efficiency vs. Input Voltage 90 70 80 89 Efficiency (%) Efficiency (%) 100 88 87 86 85 84 85 90 110 130 150 170 190 230 240 250 264 90 100 Load (%) Load vs. Input Voltage 125 100 80 60 40 20 85 100 115 130 145 160 Deviation vs. Load 1 Deviation (%) % of Full Load Input Voltage (V) 175 0.5 0 -0.5 264 -1 0 10 20 30 20 50 60 70 80 90 100 Input Voltage (V) Load (%) REGULATIONS Parameter Condition Value Output Voltage Accuracy 230VAC, full load Line Voltage Regulation low line to high line, full load ±0.1% typ. / ±0.2% max. Load Voltage Regulation 0% to 100% load ±0.1% typ. / ±0.5% max. Output Voltage Trim ±0.1% typ. / ±1% max. ±10% Transient Peak Deviation load step from 50% - 75% change at 2.5A/µs 3% Vout max. Transient Recovery Time load step from 50% - 75% change at 2.5A/µs 500µs typ. PROTECTIONS Parameter Input Fuse Condition internal line and neutral Short Circuit Protection (SCP) Over Load Protection (OLP) Over Voltage Protection (OVP) Isolation Voltage Value T3.15A / 250VAC, slow blow type continuous, auto-recovery % of Iout rated Hiccup Mode, 115% min. / 150% max. % of Vout nominal Latch Mode, 115% min. / 135% max. I/P to O/P I/P to Chassis O/P to Chassis working voltage 4kVAC / 1 minute 1.5kVAC / 1 minute 1.5kVAC / 1 minute 250VAC / continuous continued on next page MED-2 REV.: 2/2015 www.recom-power.com RACM150(/F) AC/DC Converter Series Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up) Means of Protection 2MOPP Leakage Current 264VAC 100µA max. Medical Device Classification Type BF applied device Internal Clearance Creepage I/P to O/P I/P to O/P Isolation Resistance 500VDC 8mm min. 8mm min. 100MW min. Insulation Grade Reinforced Insulation ENVIRONMENTAL Parameter Condition Relative Humidity Value non-condensing 5% to 95% RH Temperature Coefficient ±0.02% / °C Operating Temperature Range (refer to derating graph) without fan and with derating with fan and with derating -25°C to +80°C -25°C to +80°C Operating Altitude 5000m max. MTBF (+25°C) according to MIL-HDBK-217F, full load 786.1 x 103 hours Derating Graph Vin= 230VAC 150 150 140 140 Output Current (%) Output Current (%) Vin= 115VAC 120 100 80 60 RACM150-xxS/F RACM150-xxS (xx=15, 24) RACM-xxS (xx=12 48) 40 20 0 -25 0 120 100 80 60 RACM150-xxS/F RACM150-xxS (xx=15, 24) RACM-xxS (xx=12 48) 40 20 0 -25 10 20 30 40 50 60 70 80 90 100 Ambient Temperature (°C) 0 10 20 30 40 50 60 70 80 90 100 Ambient Temperature (°C) SAFETY AND CERTIFICATIONS Certificate Type (Safety) IEC/EN Medical Safety Report / File Number T141119W01-E-1 1408016005 Standard IEC/EN-60601-1 Medical Report + ISO14971 Risk Assessment ANSI/AAMI Medical Safety E314885 ES60601-1 CAN/CSA Medical Safety E314885 C22.2 No. 60601-1 Certificate Type (Others) EMI Standard Conditions Standard / Criterion Conducted CISPR 11, EN-55011, Class B Radiated CISPR 11, EN-55011, Class A Conducted and Radiated FCC18, Class B Air ±8kV; Contact ±6kV EN61000-4-2, Criteria A Radiated Immunity 20V/m EN61000-4-3, Criteria A Fast Transient ±2kV EN61000-4-4, Criteria A ESD continued on next page www.recom-power.com REV.: 2/2015 MED-3 RACM150(/F) AC/DC Converter Series Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up) Surge Conducted Immunity Power Frequency Magnetic Field Voltage Dip and Interruptions 100/230VAC, 50Hz Harmonic Current L-N ±1kV and L-PE/N-PE ±2kV EN61000-4-5, Criteria A 20Vr.m.s EN61000-4-6, Criteria A 10A/m EN61000-4-8, Criteria A 30% 500ms EN61000-4-11, EN60601-1-2, Criteria A >95% 10ms EN61000-4-11, EN60601-1-2, Criteria A >95% 5000ms EN61000-4-11, EN60601-1-2, Criteria B full load EN61000-3-2; Class D Voltage Flicker EN61000-3-3, PASS Thermal Shock MIL-STD-810F Shock IEC68-2-27 Vibration IEC68-2-6 DIMENSION and PHYSICAL CHARACTERISTICS Parameter Type Value Case Material Aluminum Package Dimension (LxWxH) with fan without Fan 116.8 x 62.0 x 49.2mm 116.8 x 62.0 x 39.2mm Package Weight with Fan without Fan 270g 255g Dimension Drawing (mm) 62.0 23.1 FC 41.3 Top View 33.0 FC 109.6 116.8 continued on next page MED-4 REV.: 2/2015 40.0 10.0 mm minimum www.recom-power.com 23.1 RACM150(/F) AC/DC Converter 33.0 FC 109.6 Series Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up) 116.8 Dimension Drawing (mm) Side View 10.0 mm minimum 10.0 40.0 1.2 39.2 CON1 - Input Pin1Line Pin3Neutral Mates with JST housing: VHR-3N JST crimp terminals: SVH-21T-P1.1 CON2 - Output 10.0 10.4 Bottom View Pin1,2,3-Vout Pin45,6+Vout Mates with JST housing: VHR-6N JST crimp terminals: SVH-21T-P1.1 CON2 - Output 3.5 Pin1-Fan Pin2+Fan Mates with Molex housing: 22-01-1022 Molex crimp terminals: 2759 Tolerance: ±0.5mm FC: fixing center FAN Rated Voltage: 12V (7-13.8) Input Power: 0.96W typ. 1.8W max. Speed:6000RPM Air Flow: 7CFM/Min.; 30dBA max. expected Lifetime (40°C): >70khours continuous Cable length: 55mm inlcuding connector www.recom-power.com REV.: 2/2015 MED-5 RACM150(/F) AC/DC Converter Series Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up) PACKAGING INFORMATION Parameter Packaging Dimension (LxWxH) Type Cardboard Box Packaging Quantity Storage Temperature Range Value 418 x 308 x 105mm 10pcs with fan without fan -40°C to +80°C -40°C to +80°C The product information and specifications are subject to change without prior notice. RECOM products are not authorized for use in safety-critical applications (such as life support) without RECOM’s explicit written consent. A safety-critical application is defined as an application where a failure of a RECOM product may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The buyer shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications. MED-6 REV.: 2/2015 www.recom-power.com