RACM150 - recom

Features
•
•
•
•
•
•
•
•
Regulated
Converters
Long 5 Year Warranty
2MOPP/250VAC
Suitable for built in Class II Applications
Wide Input Voltage Range (85-264VAC)
Low Leakage Current (<100µA)
5000m Operation
Active Power Factor Correction
Connector Set available
AC/DC Converter
RACM150
Description
150 Watt
5
Enclosed
Case Style
Single Output
5Y
The RACM150-S(/F) is a compact 4.6” x 2.44” high efficiency AC/DC power supply with 2xMOPP safety
approval for medical applications. These space saving enclosed power supplies have a universal input
voltage range (85-264VAC), 4kVac isolation, require no minimum load and can be used at
ambient temperatures of between -25°C and +80°C. The 12V, 15V, 24V or 48V output voltages
are fully protected and have tolerances of less than ±0.2% over the entire input voltage range and
less than ±0.5% over the entire load range. The RACM150-S(/F) series is certified to medical safety
standard IEC/ES/EN-60601-1 3rd Edition and feature BF rated outputs with less than 100µA
leakage current. It has a built-in Class B EMI filter and comes with a five year warranty.
wa
EAR
rranty
Selection Guide
Part
Number
Input
Output
Output
Efficiency
Peak Output
Voltage Range
Voltage
Current (A)
typ.
Power (W)
(VAC) (VDC)115/230VAC(%)115/230VAC
RACM150-12S
85-264
12
10.0 / 10.84
91
120 / 130
RACM150-15S
85-264
15
8.33 / 9.0
92
125 / 135
RACM150-24S
85-264
24
5.2 / 5.63
92
125 / 135
RACM150-48S
85-264
48
2.5 / 2.71
91
120 / 130
RACM150-12S/F 85-264
1212.591 150
(1)
RACM150-15S/F (1)
85-264
1510.092 150
RACM150-24S/F (1)
85-264
246.2592 150
RACM150-48S/F (1)
85-264
483.1391 150
Notes:
Note1:
Please note that removing the fan from the /F version will not give the same performance as the equivalent fanless type. The two versions are not identical.
Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up)
2MOPP
250VAC
BASIC CHARACTERISTICS
Parameter
Condition
Min.
Typ.
85VAC
120VDC
Input Voltage
Max.
264VAC
370VDC
Input Current
115VAC, fulll load
230VAC, full load
1.7A
0.8A
Inrush Current
cold start, 115VAC
cold start, 230VAC
30A
60A
230VAC, with fan
230VAC, without fan
No Load Input Power
Input Frequency Range
AC Input
0.6W
0.25W
47Hz
0.7s
Rise Time
20ms
Hold up Time
30ms
Minimum Load
1s
0%
Power Factor
0.95
Operating Frequency
(measured @ 20MHz BW)
1W
0.3W
EN-55011 Certified
EN-55022 Certified
ES-60601 Certified
IEC/EN-60601 Certified
63Hz
Start-up Time
Output Ripple and Noise
E314885
60kHz
12VDC, with 1µF/25V MLCC
15VDC, with 1µF/25V MLCC
24VDC, with 1µF/50V MLCC
48VDC, with 0.1µF/100V MLCC
120mVp-p
150mVp-p
220mVp-p
250mVp-p
continued on next page
www.recom-power.com
Refer to Applications Notes
REV.: 2/2015
MED-1
RACM150(/F)
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up)
Efficiency vs. Load
230VAC
90
80
115VAC
70
60
50
10
20
30
40
50
60
Efficiency vs. Input Voltage
90
70
80
89
Efficiency (%)
Efficiency (%)
100
88
87
86
85
84
85 90 110 130 150 170 190 230 240 250 264
90 100
Load (%)
Load vs. Input Voltage
125
100
80
60
40
20
85
100
115
130
145
160
Deviation vs. Load
1
Deviation (%)
% of Full Load
Input Voltage (V)
175
0.5
0
-0.5
264
-1
0
10 20 30 20 50 60 70 80 90 100
Input Voltage (V)
Load (%)
REGULATIONS
Parameter
Condition
Value
Output Voltage Accuracy
230VAC, full load
Line Voltage Regulation
low line to high line, full load
±0.1% typ. / ±0.2% max.
Load Voltage Regulation
0% to 100% load
±0.1% typ. / ±0.5% max.
Output Voltage Trim
±0.1% typ. / ±1% max.
±10%
Transient Peak Deviation
load step from 50% - 75% change at 2.5A/µs
3% Vout max.
Transient Recovery Time
load step from 50% - 75% change at 2.5A/µs
500µs typ.
PROTECTIONS
Parameter
Input Fuse
Condition
internal line and neutral
Short Circuit Protection (SCP)
Over Load Protection (OLP)
Over Voltage Protection (OVP)
Isolation Voltage
Value
T3.15A / 250VAC, slow blow type
continuous, auto-recovery
% of Iout rated
Hiccup Mode, 115% min. / 150% max.
% of Vout nominal
Latch Mode, 115% min. / 135% max.
I/P to O/P
I/P to Chassis
O/P to Chassis
working voltage
4kVAC / 1 minute
1.5kVAC / 1 minute
1.5kVAC / 1 minute
250VAC / continuous
continued on next page
MED-2
REV.: 2/2015
www.recom-power.com
RACM150(/F)
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up)
Means of Protection
2MOPP
Leakage Current
264VAC
100µA max.
Medical Device Classification
Type BF applied device
Internal Clearance
Creepage
I/P to O/P
I/P to O/P
Isolation Resistance
500VDC
8mm min.
8mm min.
100MW min.
Insulation Grade
Reinforced Insulation
ENVIRONMENTAL
Parameter
Condition
Relative Humidity
Value
non-condensing
5% to 95% RH
Temperature Coefficient
±0.02% / °C
Operating Temperature Range
(refer to derating graph)
without fan and with derating
with fan and with derating
-25°C to +80°C
-25°C to +80°C
Operating Altitude
5000m max.
MTBF (+25°C)
according to MIL-HDBK-217F, full load
786.1 x 103 hours
Derating Graph
Vin= 230VAC
150
150
140
140
Output Current (%)
Output Current (%)
Vin= 115VAC
120
100
80
60
RACM150-xxS/F
RACM150-xxS (xx=15, 24)
RACM-xxS (xx=12 48)
40
20
0
-25
0
120
100
80
60
RACM150-xxS/F
RACM150-xxS (xx=15, 24)
RACM-xxS (xx=12 48)
40
20
0
-25
10 20 30 40 50 60 70 80 90 100
Ambient Temperature (°C)
0
10 20 30 40 50 60 70 80 90 100
Ambient Temperature (°C)
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
IEC/EN Medical Safety
Report / File Number
T141119W01-E-1
1408016005
Standard
IEC/EN-60601-1
Medical Report + ISO14971 Risk Assessment
ANSI/AAMI Medical Safety
E314885
ES60601-1
CAN/CSA Medical Safety
E314885
C22.2 No. 60601-1
Certificate Type (Others)
EMI Standard
Conditions
Standard / Criterion
Conducted
CISPR 11, EN-55011, Class B
Radiated
CISPR 11, EN-55011, Class A
Conducted and Radiated
FCC18, Class B
Air ±8kV; Contact ±6kV
EN61000-4-2, Criteria A
Radiated Immunity
20V/m
EN61000-4-3, Criteria A
Fast Transient
±2kV
EN61000-4-4, Criteria A
ESD
continued on next page
www.recom-power.com
REV.: 2/2015
MED-3
RACM150(/F)
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up)
Surge
Conducted Immunity
Power Frequency Magnetic Field
Voltage Dip and Interruptions
100/230VAC, 50Hz
Harmonic Current
L-N ±1kV and L-PE/N-PE ±2kV
EN61000-4-5, Criteria A
20Vr.m.s
EN61000-4-6, Criteria A
10A/m
EN61000-4-8, Criteria A
30%
500ms
EN61000-4-11, EN60601-1-2, Criteria A
>95%
10ms
EN61000-4-11, EN60601-1-2, Criteria A
>95% 5000ms
EN61000-4-11, EN60601-1-2, Criteria B
full load
EN61000-3-2; Class D
Voltage Flicker
EN61000-3-3, PASS
Thermal Shock
MIL-STD-810F
Shock
IEC68-2-27
Vibration
IEC68-2-6
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Type
Value
Case Material
Aluminum
Package Dimension (LxWxH)
with fan
without Fan
116.8 x 62.0 x 49.2mm
116.8 x 62.0 x 39.2mm
Package Weight
with Fan
without Fan
270g
255g
Dimension Drawing (mm)
62.0
23.1
FC 41.3
Top View
33.0
FC 109.6
116.8
continued on next page
MED-4
REV.: 2/2015
40.0
10.0 mm minimum
www.recom-power.com
23.1
RACM150(/F)
AC/DC Converter
33.0
FC 109.6
Series
Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up)
116.8
Dimension Drawing (mm)
Side View
10.0 mm minimum
10.0
40.0
1.2
39.2
CON1 - Input
Pin1Line
Pin3Neutral
Mates with
JST housing: VHR-3N
JST crimp terminals: SVH-21T-P1.1
CON2 - Output
10.0
10.4
Bottom View
Pin1,2,3-Vout
Pin45,6+Vout
Mates with
JST housing: VHR-6N
JST crimp terminals: SVH-21T-P1.1
CON2 - Output
3.5
Pin1-Fan
Pin2+Fan
Mates with
Molex housing: 22-01-1022
Molex crimp terminals: 2759
Tolerance: ±0.5mm
FC: fixing center
FAN
Rated Voltage:
12V (7-13.8)
Input Power:
0.96W typ. 1.8W max.
Speed:6000RPM
Air Flow:
7CFM/Min.; 30dBA max.
expected Lifetime (40°C): >70khours continuous
Cable length:
55mm inlcuding connector
www.recom-power.com
REV.: 2/2015
MED-5
RACM150(/F)
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 230VAC, full load and after warm-up)
PACKAGING INFORMATION
Parameter
Packaging Dimension (LxWxH)
Type
Cardboard Box
Packaging Quantity
Storage Temperature Range
Value
418 x 308 x 105mm
10pcs
with fan
without fan
-40°C to +80°C
-40°C to +80°C
The product information and specifications are subject to change without prior notice. RECOM products are not authorized for use in safety-critical applications (such as life support) without RECOM’s explicit written
consent. A safety-critical application is defined as an application where a failure of a RECOM product may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The buyer
shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications.
MED-6
REV.: 2/2015