RACM100 - recom

Features
•
•
•
•
•
•
•
•
Regulated
Converters
Long 5 Year Warranty
2MOPP/250VAC
Suitable for built in Class II Applications
Wide Input Voltage Range (85-264VAC)
Low Leakage Current (<75µA)
5000m Operation
Active Power Factor Correction
Connector Set available
AC/DC Converter
RACM100
Description
100 Watt
5
Enclosed
Case Style
Single Output
5Y
The RACM100 is a compact 3.6”x2.44” high efficiency AC/DC power supply with 2xMOPP safety
approval for medical applications. These space saving enclosed power supplies have an universal input
voltage range (85-264VAC), 4kVAC isolation, require no minimum load and can be used at ambient
temperatures of between -25°C and +85°C. The 12V, 15V, 24V or 48V output voltages are
fully protected and have tolerances of less than ±0.2% over the entire input voltage range and less
than ±0.5% over the entire load range. The RACM100 series is certified to medical safety standard
IEC/ES/EN-60601-1 3rd Edition and feature BF rated outputs with less than 75µA leakage current. It
has a built-in Class B EMI filter and comes with a five year warranty.
wa
EAR
rranty
Selection Guide
Part
Number
Input
Voltage Range
(VAC)
Output
Output
Input Power
Voltage Current
@ No Load (VDC)(A)
(W)
Efficiency
typ.
(%)
RACM100-12S
85-264
128.34
0.3
91
RACM100-15S
85-264
156.67
0.3
92
RACM100-24S
85-264
244.17
0.3
92
RACM100-48S
85-264
482.09
0.3
91
Specifications (measured at TA= 25°C, 250VAC, full load and after warm-up)
2MOPP
250VAC
BASIC CHARACTERISTICS
Parameter
Condition
Min.
Typ.
85VAC
120VDC
Input Voltage
Max.
264VAC
370VDC
Input Current
115VAC, fulll load
230VAC, full load
1.15A
0.55A
Absolute max. Input Current
cold start, 230VAC
60A
Input Frequency Range
AC Input
47Hz
63Hz
Start-up Time
1 Second
Rise Time
20ms
Hold up Time
115VAC, full load
0%
Power Factor
0.95
Operating Frequency
(measured @ 20MHz BW)
EN-55011 Certified
EN-55022 Certified
ES-60601 Certified
IEC/EN-60601 Certified
16ms
Minimum Load
Output Ripple and Noise
E314885
60kHz
12VDC, with 10µF/25V MLCC
15VDC, with 10µF/25V MLCC
24VDC, with 1µF/50V MLCC
48VDC, with 0.1µF/100V MLCC
120mVp-p
150mVp-p
160mVp-p
340mVp-p
continued on next page
Refer to Applications Notes
www.recom-power.com
REV.: 2/2015
MED-1
RACM100
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 250VAC, full load and after warm-up)
Efficiency vs. Input Voltage
Efficiency vs. Load
94
230VAC
92
Efficiency (%)
Efficiency (%)
94
90
115VAC
88
86
84
10
20
30
40
50
60
70
80
100% Load
92
90
50% Load
88
86
84
85 90 110 130 150 170 190 230 240 250 264
90 100
Load (%)
Input Voltage (V)
Load vs. Input Voltage
Power Factor vs. Load
1
125
80
60
40
20
85
230VAC
Power Factor [ ]
% of Full Load
100
100
115
130
145
160
175
115VAC
0
264
30
40
Input Voltage (V)
50
60
70
80
90
100
Load (%)
REGULATIONS
Parameter
Condition
Value
Output Voltage Accuracy
230VAC, full load
±1%
Line Voltage Regulation
low line to high line, full load
Load Voltage Regulation
0% to 100% load
10% to 100% load
±0.5% max.
±0.4% max.
Transient Peak Deviation
load step from 50% - 75% change at 2.5A/µs
3% Vout max.
Transient Recovery Time
load step from 50% - 75% change at 2.5A/µs
500µs typ.
Output Voltage Trim
±0.2%
±10%
PROTECTIONS
Parameter
Input Fuse
Condition
internal line and neutral
Short Circuit Protection (SCP)
Over Load Protection (OLP)
Over Voltage Protection (OVP)
Isolation Voltage
continuous, auto-recovery
% of Iout rated
Hiccup Mode, 115% min. / 150% max.
% of Vout nominal
Latch Mode, 115% min. / 135% max.
I/P to O/P
I/P to Chassis
O/P to Chassis
working voltage
Means of Protection
Leakage Current
Value
T3.15A / 250VAC, slow blow type
4kVAC / 1 minute
1.5kVAC / 1 minute
1.5kVAC / 1 minute
250VAC / continuous
2MOPP
264VAC
Medical Device Classification
75µA max.
Type BF applied device
continued on next page
MED-2
REV.: 2/2015
www.recom-power.com
RACM100
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 250VAC, full load and after warm-up)
Internal Clearance
Creepage
I/P to O/P
I/P to O/P
8mm min.
8mmmin.
Isolation Resistance
500VDC
100MW min.
Insulation Grade
Reinforced Insulation
ENVIRONMENTAL
Parameter
Condition
Relative Humidity
Value
non-condensing
5% to 95% RH
with derating
-25°C to +85°C
Temperature Coefficient
±0.02% / °C
Operating Temperature Range
Operating Altitude
5000m max.
MTBF (+25°C)
according to MIL-HDBK-217F, full load
790.3 x 103 hours
Derating Graph
Output Current (%)
100
230VAC
80
115VAC
60
40
30
20
0
-25
0
20
40
60
55
80
100
85
Ambient Temperature (°C)
SAFETY AND CERTIFICATIONS
Certificate Type (Safety)
Report / File Number
T141119W01-E-1
1408016005
IEC/EN Medical Safety
Standard
IEC/EN-60601-1
Medical Report + ISO14971 Risk Assessment
ANSI/AAMI Medical Safety
E314885
ES60601-1
CAN/CSA Medical Safety
E314885
C22.2 No. 60601-1
Certificate Type (Others)
Conditions
EMI Standard
ESD
Standard / Criterion
Conducted
CISPR 11, EN-55011, Class B
Radiated
CISPR 11, EN-55011, Class A
Conducted and Radiated
FCC18, Class B
Air ±8kV; Contact ±6kV
EN61000-4-2, Criteria A
Radiated Immunity
20V/m
EN61000-4-3, Criteria A
Fast Transient
±2kV
EN61000-4-4, Criteria A
L-N ±1kV and L-PE/N-PE ±2kV
EN61000-4-5, Criteria A
20Vr.m.s
EN61000-4-6, Criteria A
Surge
Conducted Immunity
Power Frequency Magnetic Field
10A/m
Voltage Dip and Interruptions
100/230VAC, 50Hz
EN61000-4-8, Criteria A
30%
500ms
EN61000-4-11, EN60601-1-2, Criteria A
>95%
10ms
EN61000-4-11, EN60601-1-2, Criteria A
>95% 5000ms
EN61000-4-11, EN60601-1-2, Criteria B
continued on next page
www.recom-power.com
REV.: 2/2015
MED-3
RACM100
AC/DC Converter
Series
Specifications (measured at TA= 25°C, 250VAC, full load and after warm-up)
Harmonic Current
full load
EN61000-3-2; Class D
Voltage Flicker
EN61000-3-3, PASS
Thermal Shock
MIL-STD-810F
Shock
IEC68-2-27
Vibration
IEC68-2-6
DIMENSION and PHYSICAL CHARACTERISTICS
Parameter
Type
Value
Case Material
Aluminum
Package Dimension (LxWxH)
91.4 x 62.0 x 39.2mm
Package Weight
210g
Dimension Drawing (mm)
24.5
Trim Adj.
62.0
FC 41.3
CON2
CON1
Top View
CON1 - Input
Pin1Line
Pin3Neutral
22.1
Mates with
JST housing: VHR-3N
JST crimp terminals: SVH-21T-P1.1
76.2
FC 84.2
91.4
CON2 - Output
Pin1,2-Vout
Pin3,4+Vout
Side View
Tolerance: ±0.5mm
FC:
fixing center
1.20
39.2
Mates with
JST housing: VHR-4N
JST crimp terminals: SVH-21T-P1.1
continued on next page
REV.: 2/2015
www.recom-power.com
0.4
MED-4
39.2
RACM100
AC/DC Converter
1.20
Series
Specifications (measured at TA= 25°C, 250VAC, full load and after warm-up)
Dimension Drawing (mm)
10.0
10.4
Bottom View
3.5
PACKAGING INFORMATION
Parameter
Packaging Dimension (LxWxH)
Type
Cardboard Box
Packaging Quantity
Value
418 x 258 x 105mm
10pcs
Storage Temperature Range
-40°C to +85°C
The product information and specifications are subject to change without prior notice. RECOM products are not authorized for use in safety-critical applications (such as life support) without RECOM’s explicit written
consent. A safety-critical application is defined as an application where a failure of a RECOM product may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The buyer
shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications.
www.recom-power.com
REV.: 2/2015
MED-5