Download Data Sheet

DTHPU180A SERIES
180 Watts
AC/DC Medical Desktop Power Supply
Single Outputs
Rev A
Re
Size: 7.40in x 3.52in x 1.94in (188mm x 89.5mm x 49.3mm)
OPTIONS
FEATURES
• Output Connectors
• ON/OFF Switch
• Wide Operating Voltage of
90~260VAC, 47 to 63Hz
• IEC-320-C14 Input Inlet
• Single Output
• Input to Output: 2MOPP
• Class I
APPLICATIONS
•
•
•
•
•
•
•
•
•
Over Voltage, Short Circuit, and Over Load Protection
Optional ON/OFF Switch
RoHS Compliant
Medical Safety 3rd (IEC60601-1 3rd Edition)
ANSI/AAMI ES 60601-1:2005(UL/cUL 3rd Edition), EN 60601-1:2006
(TUV/T-mark 3rd Edition) Safety Approvals
DESCRIPTION
Patient Monitor
Blood Pressure System
Portable Medical Devices
ECG Machine
The DTHPU180A series of AC DC medical desktop power supplies provides up to 180 watts of output
power in a 7.40” x 3.52” x 1.94” package. This series consists of single output models ranging from 12VDC
to 48VDC and a wide operating voltage of 90~260VAC. Each model is RoHS compliant and has over
voltage, short circuit, and over load protection. This series has an IEC-320-C14 input inlet, medical safety
(IEC 60601-1 3rd Edition) approvals, and ANSI/AAMI ES 60601-1:2005(UL/cUL 3rd Edition), EN 60601-1:
2006 (TUV/T-mark 3rd Edition) safety approvals. Please call factory for order details.
MODEL SELECTION TABLE
Model Number
DTHPU180A-105
DTHPU180A-107
DTHPU180A-108
DTHPU180A-109
DTHPU180A-110
DTHPU180A-111
4/27/2016
Input Voltage
Range
Output
Voltage
Output
Current
90~260VAC
12VDC
19VDC
24VDC
30VDC
33VDC
48VDC
14A
9.47A
7.50A
6.00A
5.455A
3.75A
Ripple & Noise
Min.-Max. Load No-Min. Load
100mVp-p
120mVp-p
100mVp-p
190mVp-p
100mVp-p
240mVp-p
100mVp-p
300mVp-p
100mVp-p
330mVp-p
100mVp-p
480mVp-p
Total
Regulation
Output
Power
No Load
Consumption
Efficiency
±5%
±5%
±4%
±3%
±3%
±3%
168W
180W
180W
180W
180W
180W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
89%
91%
91%
92%
93%
93%
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 1 of 4
Rev A
DTHPU180A SERIES
180 Watts
AC/DC Medical Desktop Power Supply
Single Outputs
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
INPUT SPECIFICATIONS
Operate Input Voltage Range
Safety Approval Input Voltage Range
Input Frequency
Power Factor Correction
Inrush Current
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation(3)
Total Regulation(4)
Output Power
Output Current
Ripple & Noise (20MHz bandwidth)(5)
Transient Response Time
Hold-Up Time(6)
Start-Up Time
Temperature Coefficient
PROTECTION
Short Circuit Protection
Over Load Protection
Over Voltage Protection
ENVIRONMENTAL SPECIFICATIONS
Operating Case Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Surge Voltage
Operating Altitude
Vibration
Electro Static Discharge
Cooling
Flammability Rating
MTBF
GENERAL SPECIFICATIONS
Efficiency(7)
Insulation Resistance
Dielectric Withstanding Voltage (P-S)
Dielectric Withstanding Voltage (P-G)
Safety Ground Leakage Current
PHYSICAL SPECIFICATIONS
Weight
Typ
Max
Unit
260
240
63
1
2.2
0.9
60
120
VAC
VAC
Hz
90
100
47
0.95
Sine Wave
Low Line
High Line
Low Line
High Line
Input Current
Min
Full Load, Vin=100VAC
Full Load, Vin=240VAC
Full Load, 25ºC, Cool Start, Vin=100VAC
Full Load, 25ºC, Cool Start, Vin=240VAC
A
A
See Table
Full Load, Vin=100~120VAC or 200~240VAC
1
%
4
ms
ms
S
%/ºC
See Table
See Table
See Table
See Table
Full Load, Vin=110VAC
Full Load, Vin=110VAC
Full Load, Vin=100~240VAC
All conditions
20
2
+0.04
-0.04
Recovers automatically after fault condition is removed
110
112
Derate linearly from 100% load at 40ºC to 50% load at 70ºC
10~95%RH
Non-Condensing
-10
-40
0
0
All Conditions
10~500Hz, 10min./1cycle, 60min. each along X, Y, Z axes
Air Discharge, IEC61000-4-2
Contact Discharge, IEC61000-4-2
Operating Temperature at 25ºC, Calculated per MIL-HDBK-217F
Full Load, Vin=230VAC
Primary to Secondary, 500VDC, 25ºC/70% RH
Primary to Secondary, limit current <10mA
Primary to PE, limit current <10mA
Vin=240VAC, 60Hz
Automatic Recovery
150
132
70
85
95
95
2
3000
5
8
6
Free Air Convection
UL94V-1
100,000
%
%
ºC
ºC
%RH
%RH
kV
m
G
kV
kV
hours
See Table
50
MΩ
VAC
VAC
mA
4000
1500
0.25
31.53~33.58oz (894~952g)
7.40in x 3.52in x 1.94in
(188mm x 89.5mm x 49.3mm)
Dimensions (L x W x H)
SAFETY & EMC CHARACTERISTICS
Safety Approvals
EMC Emission
Protection Class
ANSI/AAMI ES 60601-1:2005 (UL/cUL 3rd Edition
EN 60601-1:2006 (TUV/T-mark 3rd Edition)
Compliance to EN55011 (CISPR11), EN61000-3-2, 3
Class I
B Class
NOTES
(1)
(2)
(3)
(4)
(5)
(6)
(7)
4/27/2016
Output can provide up to peak load when the power supply starts up. Staying in rated load continuously is not allowed.
Each output is checked to be within voltage accuracy at factory in 60% rated load condition.
Line regulation is defined by changing ±10% of input voltage range nominal line at rated load.
Load regulation is defined by changing ±40% of measured output load from 60% rated load.
Ripple & Noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and
nominal line.
Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated
load and nominal line.
Efficiency is measured at rated load and nominal line.
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 4
Rev A
DTHPU180A SERIES
180 Watts
AC/DC Medical Desktop Power Supply
Single Outputs
DERATING CURVES
Input Voltage Derating Curve
Temperature Derating Curve
MECHANICAL DRAWINGS
*This series is required to use AWG#16/5C/4FT output cable. The regulation and efficiency will be changed by modified output cable.
Optional output connectors are available.
4/27/2016
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 4
Rev A
DTHPU180A SERIES
180 Watts
AC/DC Medical Desktop Power Supply
Single Outputs
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely
with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide
ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your
future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
4/27/2016
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 4