DTHPU180A SERIES 180 Watts AC/DC Medical Desktop Power Supply Single Outputs Rev A Re Size: 7.40in x 3.52in x 1.94in (188mm x 89.5mm x 49.3mm) OPTIONS FEATURES • Output Connectors • ON/OFF Switch • Wide Operating Voltage of 90~260VAC, 47 to 63Hz • IEC-320-C14 Input Inlet • Single Output • Input to Output: 2MOPP • Class I APPLICATIONS • • • • • • • • • Over Voltage, Short Circuit, and Over Load Protection Optional ON/OFF Switch RoHS Compliant Medical Safety 3rd (IEC60601-1 3rd Edition) ANSI/AAMI ES 60601-1:2005(UL/cUL 3rd Edition), EN 60601-1:2006 (TUV/T-mark 3rd Edition) Safety Approvals DESCRIPTION Patient Monitor Blood Pressure System Portable Medical Devices ECG Machine The DTHPU180A series of AC DC medical desktop power supplies provides up to 180 watts of output power in a 7.40” x 3.52” x 1.94” package. This series consists of single output models ranging from 12VDC to 48VDC and a wide operating voltage of 90~260VAC. Each model is RoHS compliant and has over voltage, short circuit, and over load protection. This series has an IEC-320-C14 input inlet, medical safety (IEC 60601-1 3rd Edition) approvals, and ANSI/AAMI ES 60601-1:2005(UL/cUL 3rd Edition), EN 60601-1: 2006 (TUV/T-mark 3rd Edition) safety approvals. Please call factory for order details. MODEL SELECTION TABLE Model Number DTHPU180A-105 DTHPU180A-107 DTHPU180A-108 DTHPU180A-109 DTHPU180A-110 DTHPU180A-111 4/27/2016 Input Voltage Range Output Voltage Output Current 90~260VAC 12VDC 19VDC 24VDC 30VDC 33VDC 48VDC 14A 9.47A 7.50A 6.00A 5.455A 3.75A Ripple & Noise Min.-Max. Load No-Min. Load 100mVp-p 120mVp-p 100mVp-p 190mVp-p 100mVp-p 240mVp-p 100mVp-p 300mVp-p 100mVp-p 330mVp-p 100mVp-p 480mVp-p Total Regulation Output Power No Load Consumption Efficiency ±5% ±5% ±4% ±3% ±3% ±3% 168W 180W 180W 180W 180W 180W 0.5W 0.5W 0.5W 0.5W 0.5W 0.5W 89% 91% 91% 92% 93% 93% Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] Page 1 of 4 Rev A DTHPU180A SERIES 180 Watts AC/DC Medical Desktop Power Supply Single Outputs SPECIFICATIONS All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted. We reserve the right to change specifications based on technological advances. SPECIFICATION TEST CONDITIONS INPUT SPECIFICATIONS Operate Input Voltage Range Safety Approval Input Voltage Range Input Frequency Power Factor Correction Inrush Current OUTPUT SPECIFICATIONS Output Voltage Line Regulation(3) Total Regulation(4) Output Power Output Current Ripple & Noise (20MHz bandwidth)(5) Transient Response Time Hold-Up Time(6) Start-Up Time Temperature Coefficient PROTECTION Short Circuit Protection Over Load Protection Over Voltage Protection ENVIRONMENTAL SPECIFICATIONS Operating Case Temperature Storage Temperature Operating Humidity Storage Humidity Surge Voltage Operating Altitude Vibration Electro Static Discharge Cooling Flammability Rating MTBF GENERAL SPECIFICATIONS Efficiency(7) Insulation Resistance Dielectric Withstanding Voltage (P-S) Dielectric Withstanding Voltage (P-G) Safety Ground Leakage Current PHYSICAL SPECIFICATIONS Weight Typ Max Unit 260 240 63 1 2.2 0.9 60 120 VAC VAC Hz 90 100 47 0.95 Sine Wave Low Line High Line Low Line High Line Input Current Min Full Load, Vin=100VAC Full Load, Vin=240VAC Full Load, 25ºC, Cool Start, Vin=100VAC Full Load, 25ºC, Cool Start, Vin=240VAC A A See Table Full Load, Vin=100~120VAC or 200~240VAC 1 % 4 ms ms S %/ºC See Table See Table See Table See Table Full Load, Vin=110VAC Full Load, Vin=110VAC Full Load, Vin=100~240VAC All conditions 20 2 +0.04 -0.04 Recovers automatically after fault condition is removed 110 112 Derate linearly from 100% load at 40ºC to 50% load at 70ºC 10~95%RH Non-Condensing -10 -40 0 0 All Conditions 10~500Hz, 10min./1cycle, 60min. each along X, Y, Z axes Air Discharge, IEC61000-4-2 Contact Discharge, IEC61000-4-2 Operating Temperature at 25ºC, Calculated per MIL-HDBK-217F Full Load, Vin=230VAC Primary to Secondary, 500VDC, 25ºC/70% RH Primary to Secondary, limit current <10mA Primary to PE, limit current <10mA Vin=240VAC, 60Hz Automatic Recovery 150 132 70 85 95 95 2 3000 5 8 6 Free Air Convection UL94V-1 100,000 % % ºC ºC %RH %RH kV m G kV kV hours See Table 50 MΩ VAC VAC mA 4000 1500 0.25 31.53~33.58oz (894~952g) 7.40in x 3.52in x 1.94in (188mm x 89.5mm x 49.3mm) Dimensions (L x W x H) SAFETY & EMC CHARACTERISTICS Safety Approvals EMC Emission Protection Class ANSI/AAMI ES 60601-1:2005 (UL/cUL 3rd Edition EN 60601-1:2006 (TUV/T-mark 3rd Edition) Compliance to EN55011 (CISPR11), EN61000-3-2, 3 Class I B Class NOTES (1) (2) (3) (4) (5) (6) (7) 4/27/2016 Output can provide up to peak load when the power supply starts up. Staying in rated load continuously is not allowed. Each output is checked to be within voltage accuracy at factory in 60% rated load condition. Line regulation is defined by changing ±10% of input voltage range nominal line at rated load. Load regulation is defined by changing ±40% of measured output load from 60% rated load. Ripple & Noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and nominal line. Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated load and nominal line. Efficiency is measured at rated load and nominal line. Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] Page 2 of 4 Rev A DTHPU180A SERIES 180 Watts AC/DC Medical Desktop Power Supply Single Outputs DERATING CURVES Input Voltage Derating Curve Temperature Derating Curve MECHANICAL DRAWINGS *This series is required to use AWG#16/5C/4FT output cable. The regulation and efficiency will be changed by modified output cable. Optional output connectors are available. 4/27/2016 Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] Page 3 of 4 Rev A DTHPU180A SERIES 180 Watts AC/DC Medical Desktop Power Supply Single Outputs COMPANY INFORMATION Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one example of our commitment to producing a high quality, well-documented product for our customers. Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects. Contact Wall Industries for further information: Phone: Toll Free: Fax: E-mail: Web: Address: 4/27/2016 (603)778-2300 (888)597-9255 (603)778-9797 [email protected] www.wallindustries.com 37 Industrial Drive Exeter, NH 03833 Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] Page 4 of 4