FUJIKURA XFPM-050KPG-P1

XFPM(NIBP) Data sheet
Pre-amplified/5V Excitation/Gauge
Features
.
.
.
.
Applications
. Non-invasive Blood Pressure(NIBP)monitors
. Medical device
Accuracy, 1.0 FS
Volt level output
On-chip amplification and temperature compensations
Pre-calibration of offset voltage and span
Part number for ordering
XFPM - 050KP G R - P1
Prefix (NIBP)
P1 : Accuracy 1.0 FS
Model
XFPM
Rated pressure(Pa)
Terminal leads direction (See Outline Diagram)
No mark :
Pressure type
G : Gauge
Pressure type
R
:
Gauge pressure
XFPM
Model
Package configuration
Measurable pressure range (kPa)
0 50
Dual-In-line-Package (DIP)
Part number for ordering
XFPM-050KPG-P1
XFPM-050KPGR-P1
Specifications
Model/Rated pressure
050KPG(R)-P1
Recommended operating conditions
Pressure type
Gauge pressure
Rated pressure
300(40)
Measurable pressure range
0 300(0 40)
Pressure media
Non-corrosive gas only
Excitation voltage
5 0.25
Absolute maximum rating
Maximum load pressure
Twice of rated pressure
Maximum excitation voltage
8
Operating temperature
40 125
Storage temperature
40 125
Operating humidity
30 80 (No dew condensation)
Electric performances/characteristics (Excitation voltage Vcc=5.0V constant, Ambient temperature Ta=25 )
Current consumption
less than 10
Output impedance
less than 10
Source current
less than 0.2
Sink current
less than 2
Mechanical response time
2 (For the reference)
Full scale span voltage
3.5
Offset voltage 1
0.5 0.1
4.0 0.1
Full scale span voltage 1
Temperature sensitivity of offset(TSO)
0.2
Accuracy 1,2
1.0
Note ; 1) Excluding input voltage error.
2) Excludes offset calibration error and temperature error of offset.
1
Unit
mmHg(kPa)
mmHg(kPa)
VDC
VDC
RH
mA
mA
mA
msec
V
V
V
V/0 50
FS/0 50
XFPM(NIBP) Data sheet
Pre-amplified/5V Excitation/Gauge
Outline dimensions
Unit (mm)
XFPM
4
5
6
3
2
1
C0.5
3
2
4
3
14
1
4.5
0.6
Vent hole
R1.5
15.24(600mil)
5
Pressure port
0.8
max 15
2.2
0.5
6
8
5
0.2
4
12
Terminal leads
Labelled faceplate
6.8
2.8
Recommended footprint for PCB
6- 1.8(Diameter of lands)
2.54 2.54
0.5
60
1.2
2.54 2.54
6- 0.9(Diameter of through holes)
15.2
unit(mm)
Unit (mm)
XFPM-R
4
0.2
8
0.8
Pressure port
2.2
0.5
3
14
R1.5
4
5
6
3 2
4.5
1
15.24(600mil)
5
max 15
6
4
5
12
C0.5
1
2
0.6
Vent hole
3
Terminal leads
Recommended footprint for PCB
2.54 2.54
6- 0.9(Diameter of through holes)
6- 1.8(Diameter of lands)
2.54 2.54
60
1.2
0.5
Labelled faceplate
2.8
6.8
15.2
2
unit(mm)
XFPM(NIBP) Data sheet
Pre-amplified/5V Excitation/Gauge
Connection diagram
XFPM
Vcc(DC5V)
Vcc
Vout
Vout
GND
C2
C1
Remarks :
1)C1 : Connect 680pF capacitor within 2cm distance from leads.
2) C2 : Connect 0.01 F or more capacitor within 2cm distance
from leads.
GND 3) 4 and 5 pins must be No Connection. Open them as floating
ones completely, and do NOT connect to other line or each
other.
Transfer Function
Vout=Vs (P
+ ) (Pressure Error Temperature Error Multiplier
Vs)
Vs=5.0V
Notes ; Output, Vout, is not perfectly ratiometric power supply voltage Vs.
Model
XFPM-050KPGR-P1
pressure range
0 +300mmHg
0.00233
Temperature Error Multiplier=1
Temperature Error Multiplier
3
2
1
0
0
5
10
15
20
25
30
35
40
45
50
Temperature
Note ; Please read instruction “Notes” before using the sensor.
Fujikura reserves the right to change specifications without notice.
If you have any questions regarding technical issues or specifications, please contact us.
Sensor Engineering Department 5-1 Kiba 1-chome, Koto-ku, Tokyo 135-8512, Japan
Phone +81-(0)3-5606-1072 Fax. +81-(0)3-5606-1538
E-mail : [email protected]
3
0.1
Pressure Error 1)(mmHg)
8.57