X3GM-3,X3GM-6 Data sheet

X3GM-3,X3GM-6 Data sheet
Pre-amplified/3V Excitation/Gauge
Features
.
.
.
.
.
Applications
. Industrial instrumentation
. Pressure switch, Pneumatic device
. Medical device
3V excitation, Battery operation available
Volt level output
On-chip amplification and temperature compensations
Pre-calibration of offset voltage and span
Non-corrosive liquid measurable (X3GMC only)
Part number for ordering
X3GM C - 3 025KP G S R
Terminal leads direction (See Outline Diagram)
R:
Rated Pressure (Pa)
Model
X3GM
Length of the pressure port
Terminal leads configuration
3 : 3mm
S : Surface mount package
Pressure type
6 : 6mm
Available pressure media
G : Gauge (Positive pressure)
No mark : Non-corrosive gas only
GV : Gauge (Negative pressure)
C
: Non-corrosive gas and liquid
GW : Gauge (Bipolar pressure)
Pressure type
Gauge pressure
X3GM-3
X3GM-6
Surface mount package, 3mm pressure port
Surface mount package, 6mm pressure port
Model
Package configuration
Measurable pressure range (kPa)
100 100
0
100
0 25
0 50
0 100
0 200
0 1000
X3GM-3100KPGWSR
X3GM-3100KPGVSR
X3GM-3025KPGSR
X3GM-3050KPGSR
X3GM-3100KPGSR
X3GM-3200KPGSR
X3GM-3001MPGSR
X3GM-3100KPGWSRH
X3GM-3100KPGVSRH
X3GM-3025KPGSRH
X3GM-3050KPGSRH
X3GM-3100KPGSRH
X3GM-3200KPGSRH
X3GM-3001MPGSRH
X3GMC-3100KPGWSR
X3GMC-3100KPGVSR
X3GMC-3025KPGSR
X3GMC-3050KPGSR
X3GMC-3100KPGSR
X3GMC-3200KPGSR
X3GMC-3001MPGSR
Part number for ordering
X3GMC-3100KPGWSRH X3GM-6100KPGWSR
X3GMC-3100KPGVSRH X3GM-6100KPGVSR
X3GMC-3025KPGSRH
X3GM-6025KPGSR
X3GMC-3050KPGSRH
X3GM-6050KPGSR
X3GMC-3100KPGSRH
X3GM-6100KPGSR
X3GMC-3200KPGSRH
X3GM-6200KPGSR
X3GMC-3001MPGSRH
X3GM-6001MPGSR
X3GM-6100KPGWSRH
X3GM-6100KPGVSRH
X3GM-6025KPGSRH
X3GM-6050KPGSRH
X3GM-6100KPGSRH
X3GM-6200KPGSRH
X3GM-6001MPGSRH
X3GMC-6100KPGWSR
X3GMC-6100KPGVSR
X3GMC-6025KPGSR
X3GMC-6050KPGSR
X3GMC-6100KPGSR
X3GMC-6200KPGSR
X3GMC-6001MPGSR
X3GMC-6100KPGWSRH
X3GMC-6100KPGVSRH
X3GMC-6025KPGSRH
X3GMC-6050KPGSRH
X3GMC-6100KPGSRH
X3GMC-6200KPGSRH
X3GMC-6001MPGSRH
Specifications
Model/Rated pressure
100KPGW
100KPGV
025KPG
050KPG
100KPG
200KPG
Recommended operating conditions
Pressure type
Gauge pressure
100
100
25
50
100
200
Rated pressure
1.020
1.020
0.255
0.510
1.020
2.040
Measurable pressure range
100 100
0 100
0 25
0 50
0 100
0 200
Pressure media
1
XFGM : Non-corrosive gas only, XFGMC : Non-corrosive gas and liquid
Excitation voltage
3.0 0.3
Absolute maximum rating
Maximum load pressure
Twice of rated pressure
Maximum excitation voltage
6
Operating temperature
10 80
Storage temperature
20 100
Operating humidity
30 80 (No dew condensation)
Electric performances/characteristics (Excitation voltage Vcc=3.0V constant, Ambient temperature Ta=25 )
Current consumption
less than 6
Output impedance
less than 10
Source current
less than 0.2
Sink current
less than 2
Mechanical response time
2 (For the reference)
Full scale span voltage
1.5
Offset voltage 2
0.5 0.0375
Full scale span voltage 2
2.0 0.0375
Accuracy 2
2.5
Note ; 1) Please consult us available liquid pressure media when you choose the C models.
2) Excluding input voltage error.
1
001MPG
Unit
1000
10.20
0 1000
kPa
kg/cm2
kPa
VDC
1.5times of rated pressure
VDC
RH
mA
mA
mA
msec
V
V
V
FS/0 50
X3GM-3,X3GM-6 Data sheet
Pre-amplified/3V Excitation/Gauge
Outline dimensions
Unit : mm
X3GM-3
2.54 2.54
0.8
Pressure port
2
6
5
3
7
1
4
7
3
Vent hole
0.25
X.
6.5
MA
15
0.8
3.5
0.15
3
Index
Recommended footprint for PCB
10.3
5
4
Labelled faceplate
1.7
1
2.54 2.54
1.4
6
1
2
9.4
Land-pads
3
(Unit : mm)
Unit : mm
X3GM-6
2.54 2.54
0.8
Pressure port
2
6
5
3
7
1
4
7
Vent hole
3
10.3
6
5
4
2-R0.5
3.5
Recommended footprint for PCB
2- 1(Holes for projections)
1.4
1.7
1
Projections
1
2
3
Labelled faceplate
Land-pads
6.2
Projections
2.54 2.54
0.8
MA
0.5
X.
15
0.25
9.5
6
Index
6.2
9.4
(Unit : mm)
2
Pre-amplified/3V Excitation/Gauge
X3GM-3,X3GM-6 Data sheet
Connection diagram
X3GM-3, X3GM-6
Vcc(DC3V)
Vcc
Vout
Vout
GND
C2
C1
Remarks :
1)C1 : Connect 680pF capacitor within 2cm distance from leads.
2) C2 : Connect 0.01 F or more capacitor within 2cm distance
from leads.
GND 3)4 and 5 pins must be No Connection. Open them as floating
ones completely, and do NOT connect to other line or each
other.
Note ; Please read instruction “Notes” before using the sensor.
Fujikura reserves the right to change specifications without notice.
If you have any questions regarding technical issues or specifications, please contact us.
Sensor Engineering Department 5-1 Kiba 1-chome, Koto-ku, Tokyo 135-8512, Japan
Phone +81-(0)3-5606-1072 Fax. +81-(0)3-5606-1538
3
E-mail : [email protected]