FUJI EDS2-21C

PLATINUM RESISTANCE BULB FOR GENERAL USE
DATA SHEET
FTF
These platinum resistance bulbs are convenient for measuring a wide range of temperatures in plants and the like.
The resistance element is protected by an internal tube
and moreover by an external tube in order to fully withstand high temperature, high pressure and corrosion.
A contact strip is inserted between the internal and
external protecting tubes, which enables a quick response
and ensures high resistance to vibration.
SPECIFICATIONS
Applicable standard:
JIS C 1604—1997
Resistance element:
Kind of element and nominal resistance value;
Pt100W
Class;
B class, A class
Prescribed current; 2mA
Temperature accuracy;
Class
Accuracy
A
± (0.15 + 0.002t) °C
B
± (0.3 + 0.005t) °C
Note:(1) t is measuring temperature indicated in °C and unrelated to the +
and – signs.
(2) Accuracy means maximum allowable limit when subtracting the
re-sistance value of resistance
element from the temperature in
terms of resistance calculated according to the standard resistance
table.
Protecting tube material:
S ; 304 stainless steel
W ; 316 stainless steel
*T ; 316L stainless steel
Mounting method and shape of protecting tube:
Protecting
tube shape
Mounting
method
Insertion type
A
B
C
F
G
H
J
L
—
M
N
P
Screw-in type
Screw-in type
Flange type
Mounting method:
Insertion type;
Either movable flange (FTR), movable airtight flange (FTT), or movable
airtight screw (FTS) is employed for
mounting
Screw-in type;
G3/4 (PT3/4) or R3/4 (PT3/4) external
thread (impossible with H type)
Flange type;
JIS 10K-25A • JIS 20K-25A
JIS 30K-25A or JIS 63K-25A
(impossible with N or P type)
Insertion length: See page 3
Structure of element:
1; Single element 2-wire method
(100W B class)
3; Single element 3-wire method
(100W B class)
4; Double element 2-wire method
(100W B class)
6; Double element 3-wire method
(100W B class)
7; Single element 3-wire method
(100W A class)
8; Double element 3-wire method
(100W A class)
EDS2-21c
Date
Jan. 31, 2007
FTF
Feasible combinations of mounting method and element structure:
A B C F G H J L M N P
1, 3
7
CODE SYMBOLS
1 2 3 4 5 6 7 8
F T F
9 10 11 12 13
Description
4 -
Element
Singleelement
2-wiremethod(100ΩBclass)
Singleelement
3-wiremethod(100ΩBclass)
Doubleelement(doublewinding)
2-wiremethod(100ΩBclass)
Doubleelement(doublewinding)
3-wiremethod(100ΩBclass)
Singleelement
3-wiremethod(100ΩAclass)
Doubleelement
3-wiremethod(100ΩAclass)
Note:100Ω,0.5classisstandard.
1
4
3
6, 8
Note: Manufacturing possible
Manufacturing possible
Flameproof-explosionproof-possible
Working conditions:
See page 4
Terminal box:
Material; Diecast aluminum alloy
Conduit connection;
G1/2 (PF1/2) (JIS 15C)
Two conduit connections
should be provided with
double element 3-wire
re-sistance bulb.
Coating; Baked melamine resin
Finish color;Silver
Nominal resistance table:
See page 3
Scope of delivery:Platinum resistance bulb
Note: *(1) Combination of protecting tube shape and mounting
dimension
Protecting
tube shape
Mounting
method
4
* 6
7
8
A
B
C
F
G
H
J
L
M
N
P
Mounting method *(1)
Insertiontype
Insertiontype
Insertiontype
Screw-intype(forlowtemperature)
Screw-intype(forlowtemperature)
Screw-intype(forlowtemperature)
Screw-intype(forhightemperature)
Screw-intype(forhightemperature)
Flangetype
Flangetype
Flangetype
S
W
* T
Protecting tube material
304stainlesssteel
316stainlesssteel
316Lstainlesssteel
Mounting dimension
FlangeJIS10K-25A
FlangeJIS20K-25A
FlangeJIS30K-25A
FlangeJIS63K-25A(impossiblewith
mountingmethodNorP)
MountingthreadG3/4(PF3/4)
MountingthreadR3/4(PT3/4)
(CombinationwithHtypemounting
methodisexcluded)
Insertiontype
1
2
3
4
7
8
Insertion type
A
B
C
0
Screw-in type
Insertion length [mm]
Fillininsertionlengthincmunit
(Accordingtotableofinsertion
lengths)
Ex.:Insertionlength500mm050
1000mm100
015
F
G
H
J
L
—
Screw-in type
295
Flange type
M
N
P
(a) When the element code No. 6 is used, the terminal box form
should be provided with two conduit connections.
(b) Feasible combinations of element and mounting method.
A B C F G H J L M N P
1, 3
7
4
6, 8
Manufacturing possible
Explosionproof structure possible
*(2) Cabtyre cable of 1m lg. is provided when the filling compound is used.
Structure/Explosionproofing
Standardtype
Plasticterminaltype
Flameprooftype(d2G4)
Systemintrinsicallysafeexplosionproofing
(i3nG5)forconnectionwithzener
barrier
A
B
C
F
*
*
*
*
*
*
0
1
2
3
4
5
6
7
8
Lining and filling compound *(2)
Lining
Fillingcompound
None
None
None
Paraffin
None
Silicongrease
Leadhomogen
None
Leadhomogen
Paraffin
Leadhomogen
Silicongrease
Kel-F
None
Kel-F
Paraffin
Kel-F
Silicongrease
Asterisked(*)items:Non-standard
Insertion length (Unit: mm) [Dimension "L" of "Outline Diagram"]
Mounting
method
Insertion
length
Insertion length
A
B
C
F
G
H
J
L
M
N
P
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
950
1,050
1,150
1,250
1,350
1,450
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
950
1,050
1,150
1,250
1,350
1,450
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
950
1,050
1,150
1,250
1,350
1,450
1,550
1,650
1,750
1,850
1,950
2,450
2,950
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
1,000
1,100
1,200
1,300
1,400
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
1,000
1,100
1,200
1,300
1,400
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
1,000
1,100
1,200
1,300
1,400
1,500
1,600
1,700
1,800
1,900
2,400
2,900
150
200
250
300
350
400
450
500
550
600
650
700
750
800
900
1,000
1,100
1,200
1,300
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
950
1,050
1,150
1,250
1,350
150
200
250
300
350
400
450
500
550
600
650
700
750
800
900
1,000
1,100
1,200
1,300
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
950
1,050
1,150
1,250
1,350
200
250
300
350
400
450
500
550
600
650
700
750
800
850
950
1,050
1,150
1,250
1,350
1,450
1,550
1,650
1,750
1,850
2,350
2,850
Nominal resistance value of resistance bulb
JIS C 1604-1997
Temp.
[°C]
0
Temp.
[°C]
−100
0
100
200
300
400
500
600
Temp.
[°C]
−0
60.26
100.00
0
100.00
138.51
175.86
212.05
247.09
280.98
313.71
0
−10
56.19
96.09
10
103.90
142.29
179.53
215.61
250.53
284.30
316.92
10
−20
52.11
92.16
20
107.79
146.07
183.19
219.15
253.96
287.62
320.12
20
−30
48.00
88.22
30
111.67
149.83
186.84
222.68
257.38
290.92
323.30
30
−40
43.88
84.27
40
115.54
153.58
190.47
226.21
260.78
294.21
326.48
40
−50
39.72
80.31
50
119.40
157.33
194.10
229.72
264.18
297.49
329.64
50
−60
35.54
76.33
60
123.24
161.05
197.71
233.21
267.56
300.75
60
−70
31.34
72.33
70
127.08
164.77
201.31
236.70
270.93
304.01
70
−80
27.10
68.33
80
130.90
168.48
204.90
240.18
274.29
307.25
80
−90
22.83
64.30
90
134.71
172.17
208.48
243.64
277.64
310.49
90
−100
18.52
60.26
100
138.51
175.86
212.05
247.09
280.98
313.71
100
Remarks: The resistance values given in the above table are those when nominal resistance value "R 0" is 100W.
(Unit: W)
FTF
Working conditions
Model
1
3
S
4
FTF
A W
6
T
7
8
Working temperature, working pressure
Application
Mass{weight}
Outline diagrams
Working temp.; –200 to +650 °C
Working pressure; Atmospheric pressure
When both movable airtight screw and
flange are used, working pressure can
be up to 0.1MPa {1.02kgf/cm2}
Used for corrosive gas or liquid,
and when the insertion length is
desired to be adjustable
0.5 to 2 kg
Fig. 1
Fig. 2
(
)
S
1
FTF 3 B W
T
7
Ditto
Ditto
Used when especially short
time lag is required
0.5 to 1.5 kg
Fig. 3
S
1
FTF 3 C W
T
7
Ditto
Ditto
Used when greater insertion
length is required
0.7 to 4.4 kg
Fig. 4
Used for corrosive gas or liquid.
Used for comparatively high
temp., high pressure.
J type is used when the temp.
in the terminal box is expected
to rise or when there is heatretaining material
0.5 to 1.8 kg
Fig.
Fig.
Fig.
Fig.
Used for corrosive gas or liquid.
L type is used when the temp.
in the terminal box is expected
to rise or when there is heatretaining material, or used when
especially short time lag is
required
0.5 to 1.5 kg
Fig. 9
Fig. 10
Used for water, air, corrosive
gas etc., and especially in case
greater insertion length is required
1 to 3 kg
Fig. 11
S
1
G
FTF 3 L W
T
7
5
MPa
4
0
-200
500°C
3.5
3
0
-200
100
500°C
2
1.5
MPa
S
1
FTF 3 H W
T
7
100
MPa
1
3
S
4 F
FTF
W
J
6
T
7
8
0
-200
100
500°C
5
6
7
8
1
3
S
4
FTF
H W
6
T
7
8
Working temp.; –200 to +650 °C
Working pressure;
Less than flange (JIS) nominal pressure
JIS 10K-25 • JIS 20K-25
JIS 30K-25 • JIS 63K-25
Material; 304, 316,
316L stainless steel
Used for corrosive gas or liquid
etc., and especially in case airtightness is required
1.4 to 3.8 kg
Fig.
Fig.
Fig.
Fig.
Fig.
Fig.
Fig.
S
1
FTF 3 N W
T
7
Working temp.; –200 to +650 °C
Working pressure;
Less than flange (JIS) nominal pressure
JIS 10K-25 • JIS 20K-25
JIS 30K-25
Material; 304, 316,
316L stainless steel
Ditto
Used when especially short
time lag is required
1.4 to 3.8 kg
Fig. 19
Fig. 20
Fig. 21
Ditto
Used when especially long
insertion length is required
1.2 to 6.3 kg
Fig. 22
Fig. 23
Fig. 24
S
1
FTF 3 P W
T
7
Ditto
12
13
14
15
16
17
18
Note:When the resistance bulb is used at a place of low temperature (below 0 °C) or of high temperature, use a bulb which is filled internally with paraffin or silicon compound after connecting the cable.
Paraffin can be used in a range of 0 to 50 °C and silicon in a range of – 60 to + 150 °C.
OUTLINE DIAGRAM
[Unit:mm]
(Refer to working conditions of page 4)
Determine insertion length "L" by collating the 5th digit of
code symbols with the table shown on page 3.
For example, when the 5th digit of code symbol is A, select "L" from 22 lengths, 150mm to 1450mm, by the table
shown on page 3.
Screw-in type
Dimension "A" on the neck part of screw-in type is as follows.
5th digit of code symbols
A
F, G
50mm
J, L
150mm
Insertion type
Fig. 1
Fig. 2
Fig. 5
1
S
3 F
A
W 74 - •••
FTF 4
F
J
T
7
Fig. 6
1
S
3 F
A
W 84 - •••
FTF 4
F
J
T
7
Fig. 7
S
6 F
A
W 74 - •••
FTF 8
F
J
T
Fig. 8
S
6 F
A
W 84 - •••
FTF 8
F
J
T
1
S
A
3
FTF 4 A W 04 - ••• F
T
7
S
6
A
FTF 8 A W 04 - ••• F
T
Fig. 3
1
S
A
FTF 3 B W 04 - ••• F
7
T
Fig. 4
1
S
A
FTF 3 C W 04 - ••• F
7
T
FTF
Flange type
Fig. 9
Fig. 10
Fig. 11
S
1
A
G
W 74 - •••
FTF 3
F
L
T
7
Fig. 12
1
S
A
3
FTF 4 M W 14 - ••• F
T
7
Fig. 13
1
S
A
3
FTF M W 24 - ••• F
4
T
7
Fig. 14
1
S
A
3
FTF M W 34 - ••• F
4
T
7
Fig. 15
1
S
A
3
FTF 4 M W 44 - ••• F
T
7
S
1
A
G
W 84 - •••
FTF 3
F
L
T
7
1
S
A
FTF 3 H W 74 - ••• F
7
T
Fig. 16
S
6
A
FTF 8 M W 14 - ••• F
T
Fig. 20
1
S
A
FTF 3 N W 24 - ••• F
7
T
Fig. 17
S
6
A
FTF 8 M W 24 - ••• F
T
Fig. 21
1
S
A
FTF 3 N W 34 - ••• F
7
T
Fig. 18
S
6
A
FTF 8 M W 34 - ••• F
T
Fig. 22
1
S
A
FTF 3 P W 14 - ••• F
7
T
Fig. 19
1
S
A
FTF 3 N W 14 - ••• F
7
T
Fig. 23
1
S
A
FTF 3 P W 24 - ••• F
7
T
FTF
Terminal box for flameproof type
100
125
G1/2
(PF1/2)
Fig. 24
1
S
A
FTF 3 P W 34 - ••• F
7
T
Cap
Movable airtight screw (type FTS)
2, 4, 6 in 4th digit of code
Type
φd1
φd2
a
b
Protecting tube
Approx. ℓ outer dia. (φD)
Mass
{weight}
[kg]
FTS6E 31
R1 1/4
58
(PT1 1/4)
30
100
30
1.2
FTS6D 28
R1 1/4
58
(PT1 1/4)
30
100
25 to 27
1.2
FTS4C 24
R1 (PT1) 50
20
80
21.3 to 22
0.8
FTS2A 13
R3/4
(PT3/4)
20
80
12
0.8
38
Note: Material is SS400
1, 3, 5 in 4th digit of code
Type
φd1
φd2
φa
c
Protecting Mass
Approx. ℓ tube outer {weight}
dia. (φD)
[kg]
FTS5E 31
G1 1/4
69
(PF1 1/4)
58
30
100
30
1.2
FTS5D 28
G1 1/4
69
(PF1 1/4)
58
30
100
25 to 27
1.2
FTS3C 24
G1 (PF1) 59
50
20
80
21.3 to 22
0.8
FTS1A 13
G3/4
(PF3/4)
38
20
80
12
0.8
48
Note: Material is SS400
b
Movable airtight flange (type FTT)
Movable flange (type FTR)
JIS 10K-50 flange
31
Protecting tube
outer dia. (φD)
30
28
26.8• 27
φd
24
JIS 10K-50 flange
Material
SS400
21.3 • 21.7 • 22
JIS 10K-25 flange
50
Protecting tube
outer dia. (φD)
49
31
30
28
25 to 27
24
21.3 to 22
13
12
φd
Type
φd
FTT2D
28
Protecting tube
outer dia. (φD)
26.8 • 27
FTT2C
24
21.3 • 21.7 • 22
FTT2A
13
12
FC200
Type
φd
FTT1D
28
FTT1C
24
21.3 • 21.7 • 22
FTT1A
13
12
SS400
JIS 10K-25 flange
Material
JIS 5K-25 flange
Protecting tube
outer dia. (φD)
26.8 • 27
Material
Material
FC200
Type
φd
Protecting tube
outer dia. (φD)
FTR2D
28
25 to 27
FTR2C
24
21.3 to 22
FTR2A
13
12
Material
FC200
JIS 5K-25 flange
Type
φd
Protecting tube
outer dia. (φD)
FTR1D
28
25 to 27
FTR1C
24
21.3 to 22
FTR1A
13
12
Material
FC200
FTF
CautiononSafety
*Beforeusingthisproduct,besuretoreaditsinstructionmanualinadvance.
Head Office
GateCityOhsaki,EastTower,11-2,Osaki1-chome,
Shinagawa-ku,Tokyo141-0032,Japan
http://www.fesys.co.jp/eng
Instrumentation Div.
International Sales Dept.
No.1,Fuji-machi,Hino-city,Tokyo,191-8502Japan
Phone:81-42-585-6201,6202Fax:81-42-585-6187
http://www.fic-net.jp/eng
Information in this catalog is subject to change without notice.
PrintedinJapan