ABSOLUTE PRESSURE TRANSMITTER (DIRECT MOUNT TYPE) FKH...5 DATA SHEET The FCX–AIII absolute pressure transmitter (Direct mount type) accurately measures absolute pressure and transmits proportional 4 to 20mA signal. The transmitter utilizes the unique micromachined capacitive silicon sensor with state-of-the-art microprocessor technology to provide exceptional performance and functionality. FEATURES 1. High accuracy 0.2% accuracy for all calibrated spans is the standard feature for all AP models covering 8.125 to 3000kPa {0.13 to 30bar} high pressure range. Fuji's micro-capacitance silicon sensor assures this feature for all suppressed calibration ranges without additional adjustment. 2. Minimum inventory Electronics unit, communication module, local indicators and electronics housing are interchangeable among all FCX– AIII models. 3. Fuji/HART bilingual communication module The communication module is "bilingual" to speak both Fuji proprietary protocol and HART. Any HART compatible devices can communicate with FCX– AIII series transmitters. 4. Application flexibility Example features that render the FCX– AIII suitable for almost any process applications includes: — Full range of hazardous location approvals — Built-in RFI filter and lightning arrester — 5-digits LCD meter — The maximum span of each sensor can be converted to in different units using below factors. 5. Burnout current flexibility (Under Scale: 3.2 to 4.0mA, Over Scale: 20.0 to 22.5mA) Burnout signal level is adjustable using Model FXW Hand Held Communicator (HHC) to comply with NAMUR NE43. SPECIFICATIONS Functional specifications Type: 4 to 20mA with digital signal Service: Liquid, gas, or vapour Span, range, and overrange limit: Type Span limit [kPa abs] {bar abs} Min. FKH 02 FKH 03 FKH 04 Max. 8.125 {0.08125} 31.25 {0.3125} 187.5 {1.875} Output signal: Power supply: 130 {1.3} 500 {5} 3000 {30} Range limit [kPa abs] {bar abs} Overrange limit [MPa] {bar} 0.5 {5} 1.5 {15} 9 {90} 0 to 130 {0 to 1.3} 0 to 500 {0 to 5} 0 to 3000 {0 to 30} 4 to 20mA DC with digital signal superimposed on the 4 to 20mA signal. Transmitter operates on 10.5V to 45V DC at transmitter terminals. 10.5V to 32V DC for the units with optional arrester EDSX5-97h Date Aug. 10, 2011 FKH...5 Load limitations: see figure below [Ω] Load resistance 2000 1500 1474Ω R= E-10.5 (Imax+0.9)x10-3 1000 577Ω Operating area 500 24V 250 0 10 16.35 10.5V Power voltage (DC) 45V 50 E[ V ] Note: For communication with FXW, min. of 250 required. Hazardous locations: SEE TABLE 3 Zero/span adjustment: Zero and span are adjustable either from the HHC(1). Zero is also adjustable externally from the adjustable screw. Damping: Adjustable electrical damping The time constant is adjustable between 0.06 to 32.0 seconds. Zero elevation/suppression: Zero may be elevated within the specified range limit of each sensor model. Normal/reverse action: Configurable from HHC(1). Indication: Analog indicator or 5-digit LCD meter, as specified. Burnout direction: Selectable from HHC(1) If self-diagnostic detect transmitter failure, the analog signal will be driven to either “Output Hold”, “Output Overscale” or “Output Underscale” modes. “Output Hold”: Output signal is hold as the value just before failure happens. “Output Overscale”: Adjustable within the range 20.0mA to 22.5mA from HHC(1) “Output Underscale”: Adjustable within the range 3.2mA to 4.0mA from HHC(1) 3.2 4 20 22.5 [mA] Over scale Burnout Normal operating range Under scale Probable under range Burnout Temperature limit: Ambient: −40 to +85°C (−20 to +80°C for LCD indicator) (−40 to +60°C for arrester option) For explosionproof units (flameproof or intrinsic safety), ambient temperature must be within the limits specified by each standard. Process: −40 to +85°C for silicone fill sensor Storage: −40 to +90°C Humidity limit: 0 to 100% RH Communication: With HHC(1) (Model FXW, consult Data Sheet No. EDS8-47), following items can be remotely displayed or configured. Note: HHC’s version must be higher than 7.0 (or FXW 1– 4), for FCX AIII. Local configurator with LCD display (option): Local confi gurator with 3 push button and LCD display can support following items. Items By communication By local configurator with FXW (with 3 push button) Display Set Display Set Tag No. v v v v Model No. v v v v Serial No. & Software Version v — v — Engineering unit v v v v Range limit v — v — Measuring range v v v v Damping v v v v Output mode v — v — Burnout direction v v v v Calibration v v v v Output adjust — v — v Data v — v — Self diagnoses v — v — Printer (In case of FXW with printer option) v — — — External switch lock v v v v Transmitter display v v v v Linearize v v — — Rerange v v v v Saturate current v v v v Write protect v v v v History – Calibration history – Ambient temperature history v v v — v v v — Probable over range EMC Conformity: EN61326-1: 2006 Output Limits comforming the NAMUR NE43 by order. (Note) 2 (1) HHC: Hand Held Communicator Performance specifications Accuracy rating: (including linearity, hysteresis, and repeatability). For spans greater than 1/10 of URL: ±0.2% of span For spans below 1/10 of URL: 0.1 x URL ± (0.1 + 0.1 span ) % of span Stability: ±0.2% of upper range limit (URL) for 10 years (In case of 6th digit code "3", "4") Temperature effect: Effect per 28°C change between the limits of −40°C and +85°C URL Zero shift: ± (0.4 + 0.2 span )%/28°C URL Total effect: ± (0.475 + 0.2 span )%/28°C Overrange effect: Zero shift, 0.3% of URL for any overrange to maximum limit Update rate: 60 msec Step response: Time constant. 0.08 s (at 23°C) Dead time: about 0.12 s (without electrical damping) Mounting position effect: Zero shift, less than 0.1kPa for a 10° tilt in any plane. No effect on span. This error can be corrected by adjusting zero. Dielectric strength: 500V AC, 50/60Hz 1 min., between circuit and earth. Insulation resistance: More than 100M at 500V DC. Internal resistance for external field indicator: 12 or less Physical specifications Electrical connections: G1/2, 1/2-14NPT, Pg13.5, or M20 x 1.5 conduit, as specified. Process connections: 1/2-14 NPT, 1/4-18NPT, Rc1/2 or Rc1/4 as specified. Process-wetted parts material: Material code (7th digit Process cover in "Code symbols") V 316 stainless steel Diaphragm Wetted sensor body 316L stainless 316 stainless steel steel Vent/drain 316 stainless steel Non-wetted parts material: Electronics housing: Low copper die-cast aluminum alloy (standard), fi nished with polyester coating, as specified. Fill fluid: Silicone oil Mounting bracket: 304 stainless steel Environmental protection: IEC IP67 and NEMA 4X Mounting: On 60.5mm (JIS 50A or 2B) pipe using mounting bracket, direct wall mounting, or direct process mounting. Mass{weight}: Transmitter approximately 2.2kg without options. Add; 0.5kg for mounting bracket 3 FKH...5 Optional features Indicator: A plug-in turnable analog indicator (2.5% accuracy) An optional 5 digits LCD meter is also available. Local configurator with LCD display: An optional 5 digits LCD meter with 3 push buttons can support items as using communication with FXW. Arrester: A built-in arrester protects the electronics from lightning surges. Lightning surge immunity: 4KV (1.250µs) Degreasing: Process-wetted parts are cleaned, but the fill fluid is standard silicone oil. Not for use for oxygen or chlorine measurement. NACE specification: Metallic materials for all pressure boundary parts comply with NACE MR-01-75. Customer tag: A stainless steel tag for customer tag data is wired to the transmitter. 4 ACCESSORIES Hand held communicator: (Model FXW, refer to Data Sheet No.EDS 8-47) CODE SYMBOLS 1 2 3 4 5 6 7 8 F KH 0 9 10 11 12 13 5 14 15 21 Description 0 Connections 5 6 7 8 Conduit connection Case type Process connection G1/2 1/2-14NPT Pg 13.5 M20 x 1.5 T type T type T type T type 1/2-14NPT 1/2-14NPT 1/2-14NPT 1/2-14NPT Span limit [kPa abs] {bar abs} 8.125…130 {0.08125…1.3} 31.25…500 {0.3125…5} 187.5…3000 {1.875…30} 2 3 4 Material Process cover 316 stainless steel V Diaphragm 316L stainless steel Wetted cell body 316 stainless steel Indicator and arrester Indicator None Analog, 0 to 100% linear scale Analog, custom scale None Analog, 0 to 100% linear scale Analog, custom scale Digital, 0 to 100% linear scale Digital, custom scale Digital, 0 to 100% linear scale Digital, custom scale Digital, 0 to 100% linear scale (Local configurator unit with LCD displey) Digital, Custom scale (Local configurator unit with LCD displey) Digital, 0 to 100% linear scale (Local configurator unit with LCD displey) Digital, Custom scale (Local configurator unit with LCD displey) A B D E F H L P Q S 1 2 4 5 Arrester None None None Yes Yes Yes None None Yes Yes None None Yes Yes Approvals for hazardous locations None (for ordinary locations) TIIS, Flameproof (Cable gland seal) (*1) TIIS, Intrinsic safety FM, Flameproof (or explosionproof) (*4) FM, Intrinsic safety and nonincentive FM Combined of flameproof and intrinsic safety (*4) ATEX Flameproof (*3) ATEX Intrinsic safety ATEX Type n ATEX Combined of flameproof and intrinsic safety (*3) IECEx Scheme, Flameproof (*3) IECEx Scheme, Intrinsic safety CSA, Flameproof (or explosionproof) (*4) CSA, Intrinsic safety and nonincentive NEPSI, Flameproof (or exprosionproof) NEPSI, Intrinsic safety NEPSI, Combined of flameproof and intrinsic safety A C G D H V X K P M R T E J F S U (Note 1) (Note 4) (Note 4) (Note 3) (Note 3) (Note 3) (Note 4) Mounting bracket None Yes (stainless steel) A C Optional specification Stainless tag None (*5) Yes Y B (Note 5) Special applications and fill fluid Y G N L (*1) (*3) (*4) (*5) (*6) Filled liquid None (standard) Degreasing NACE specification Silicon oil Silicon oil Silicon oil Process adaptor None (1/2 -14NPT) Rc1/4 Rc1/2 1/4-18NPT Y A B C Note1: Note3: Note4: Note5: Note6: Treatment <Other options> (*6) Instruction manual unattached (Note 6) Available for 4th digit code "5". Available for 4th digit code "6", "8". Available for 4th digit code "6". Customer tag number can be engraved on standard stainless steel name plate. If extra tag plate is required select "Yes". If other option is not necessary, 21st digit code is blank. In case of 21st digit code is blank, instruction manual attached. 5 FKH...5 OUTLINE DIAGRAM (Unit:mm) IN CASE OF 4TH DIGIT CODE "5, 7, 8" IN CASE OF 4TH DIGIT CODE "6" 2 HOLES 9 DIA. MTG. PIPE JIS 50A (2B) ZERO ADJ. SCREW 10 EARTH TERMINAL 8 17 100 ZERO 212 "B" CONDUIT CONN. WITH ANALOG IND. (21) (11) WITH DIGITAL IND. 127 20 DETAIL "B" DETAIL "A" DETAIL "D" φ81 φ18 F φ15 K E 16 φ6 70 D "A" PRESS. CONN. MTG. BLACKET SEE TABLE2 85 4th digit of the code symbols "D" ADAPTER (OPTION) SEE TABLE2 U-BOLT M8 J SEE TABLE2 SEE TABLE1 Conduit conn. D E F 5 G 1/2 18 2 6 1/2-14NPT 16 4 7 Pg13.5 10.5 4.5 8 M20×1.5 16 4 ADAPTER CONN.(L) Rc 1/2 NPT 1/4 Rc 1/4 G 38 H 31 J 27 K 16 18 25 22 8 TABLE 2 TABLE 1 H 108 L G 27 1/2-14NPT 35 170 OPTION PARTS FOR FLAMEPROOF OF TIIS (JAPAN) CONNECTION DIAGRAM <SS TAG PLATE> 10th digit code C (Cable gland type) NOTE1) G1/2 φ13 φ33.5 −/CK− CK+ 75 + DETAIL C 106 17 19 20 "C" CONDUIT CONN. G1/2 CABLE GLAND NOTE1) IN CASE OF 10TH CODE“C” , φ11 CABLE IS SUITBLE. 6 TABLE 3 Authorities ATEX Authorities Intrinsic safety Ex II 1 G Ex ia IIC T5 Tamb = –40°C to +50°C Ex ia IIC T4 Tamb = –40°C to +70°C Entity Parameters: Ui=28V, Ii=94.3mA, Pi=0.66W, Ci=26nF (Without Arrester), Li=0.6mH (Without analog indicator), Ci=36nF (With Arrester), Li=0.7mH (With analog indicator) Factory Mutual Tamb A,B,D Y,G,N −40°C to +85°C L,P,1,2 Y,G,N −20°C to +80°C Q,S,4,5 Y,G,N −20°C to +60°C E,F,H Y,G,N −40°C to +60°C Entity Parameters: Vmax=28V, Imax=94.3mA, Pi=0.66W, Ci=35.98nF, Li=0.694mH CSA Ex II 2 GD Ex d IIC T6 IP66/67 T85°C Tamb = −40°C to +65°C Ex d IIC T5 IP66/67 T100°C Tamb = −40°C to +85°C Factory Mutual Class I Div.1 Groups B, C, D T6 Type 4X Class II III Div.1 Groups E, F, G T6 Type 4X Tamb max = +60°C CSA Class I Div.1 Groups C, D Class II Div.1 Groups E, F, G Class III Div.1 Note) “Seal Not Required” enclosure is allowed. Class I Div.1 Groups A, B, C, D Class II Div.1 Groups E, F, G Class III Div.1 Temp Code T5 Tamb max = +50°C Temp Code T4 Tamb max = +70°C Entity Parameters: Vmax=28V, Imax=94.3mA, Ci=25nF (Without Arrester), Ci=36nF (With Arrester),Li=0.6mH (Without analog meter), Li=0.7mH (With analog meter) Ex ia IIC T4 Tamb max = +60°C Entity Parameters: Ui=28V, Ii=94.3mA, Pi=0.66W, Ci=38.4nF, Li=0.694mH IECEx Scheme Ex ia IIC T4 Tamb = –40°C to +70°C Ex ia IIC T5 Tamb = –40°C to +50°C Entity Parameters: Ui=28V, Ii=94.3mA, Pi=0.66W, Ci=26nF (Without Arrester), Li=0.6mH (Without analog indicator), Ci=36nF (With Arrester), Li=0.7mH (With analog indicator) Ex ia IIC T4 Ex d IIB+H2 T6 / Ex ia IIC T4 Model code 9th digit 13th digit IECEx Scheme Ex d IIC T5 IP66/67 Tamb = −40°C to +85°C Ex d IIC T6 IP66/67 Tamb = −40°C to +65°C TIIS Ex do IIB+H2 T4 Tamb max = +60°C Maximum process temp. = +120°C NEPSI Ex d IIB+H2 T6 Tamb = −40°C to +60°C Y,G,N −40°C to +85°C L,P,1,2 Y,G,N −20°C to +80°C Q,S,4,5 Y,G,N −20°C to +60°C E,F,H Y,G,N −40°C to +60°C Ex II 3 GD EEx nL IIC T5 Tamb = −40°C to +50°C EEx nL IIC T4 Tamb = −40°C to +70°C Specific Parameters: Model without arrester: Ui=42.4V, Ii=113mA, Pi=1W, Ci=25.18nF, Li=0.694mH Model with arrester: Ui=32V, Ii=113mA, Pi=1W, Ci=35.98nF, Li=0.694mH EEx nAL IIC T5 Tamb = −40°C to +50°C EEx nAL IIC T4 Tamb = −40°C to +70°C Specific Parameters: Model without arrester: Umax=42.4V, Imax=113mA, Pmax=1W, Model with arrester: Umax=32V, Imax=113mA, Pmax=1W Tamb A,B,D Type n Nonincendive Authorities ATEX TIIS NEPSI ATEX Class I II III Div.1 Groups A, B, C, D, E, F, G T4 Entity Type 4X Model code 9th digit 13th digit Flameproof Factory Mutual (pending) Class I II III Div.2 Groups A, B, C, D, F, G T4 Entity Type 4X Model code 9th digit 13th digit Entity Parameters: Ui=42.4V, Ii=113mA, Pi=1W, Ci=35.98nF, Li=0.694mH CSA Tamb A,B,D Y,G,N −40°C to +85°C L,P,1,2 Y,G,N −20°C to +80°C Q,S,4,5 Y,G,N −20°C to +60°C E,F,H Y,G,N −40°C to +60°C Class I Div.2 Groups A, B, C, D Class II Div.2 Groups E, F, G Class III Div.2 Temp Code T5 Tamb max = +50°C Temp Code T4 Tamb max = +70°C Entity Parameters: Vmax=28V, Ci=25.18nF (Without Arrester), Ci=35.98nF (With Arrester), Li=0.694mH 7 FKH...5 Caution on Safety *Before using this product, be sure to read its instruction manual in advance. International Sales Div Sales Group Gate City Ohsaki, East Tower, 11-2, Osaki 1-chome, Shinagawa-ku, Tokyo 141-0032, Japan http://www.fujielectric.com Phone: 81-3-5435-7280, 7281 Fax: 81-3-5435-7425 http://www.fjielectric.com/products/instruments/ Information in this catalog is subject to change without notice. Printed in Japan