Download Data Sheet

PSHBU40 SERIES
40 Watts
AC DC Open Frame Medical Power Supplies
Single Output
Rev A
FEATURES
• Wide Operating Voltage, 80 to
275VAC, 47 to 63Hz
• Single Outputs
• Supports Risk Management
Process
• Input to Output: 2MOPP
APPLICATIONS
Size: 4in x 1in x 1.28in (101.6mm x 25.4mm x 32.6mm)
• Breathing Therapy
Device
• Blood Pressure
System
• Portable Medical
Device
• ECG ‘ EEG
• Medical Tablet
•
•
•
•
High ESD Immunity
Ultra Low Earth Leakage Current
Short Circuit and Over Load Protection
IEC60601-1 Edition 3.1, ES606011:2005(R2012), CSA C22.2 No. 60601-1:14,
EN60601-1:2006/A1:2013 Safety Approvals
DESCRIPTION
The PSHBU40 series of AC DC open frame medical power
supplies offers up to 40 watts of output power in a 4” x 1” x
1.28” unit. This series consists of single output models and a
wide operating voltage of 80 to 275VAC. Each model in this
series has an ultra low earth leakage current and is protected
against short circuit and over load conditions. This series has
IEC60601-1 edition 3.1, ES60601-1:2005(R2012), CSA C22.2
No. 60601-1:14, EN60601-1:2006/A1:203 safety approvals.
MODEL SELECTION TABLE
Input Voltage
Range
Model Number
PSHBU40-102
PSHBU40-103
PSHBU40-104
PSHBU40-105
PSHBU40-106
PSHBU40-107
PSHBU40-108
PSHBU40-109
PSHBU40-110
PSHBU40-111
PSHBU40-112
80 to 275VAC
Output Voltage
Range(1)
5~6VDC
6~8VDC
8~11VDC
11~13VDC
13~16VDC
16~21VDC
21~27VDC
27~33VDC
33~40VDC
40~50VDC
50~59VDC
Output Current
Min Load Max Load
4.66A
5.60A
4.00A
5.33A
3.28A
4.50A
3.07A
3.63A
2.50A
3.07A
1.90A
2.50A
1.48A
1.90A
1.21A
1.48A
1.00A
1.21A
0.80A
1.00A
0.68A
0.80A
Ripple & Noise
Min.
Max.
50mVp-p
60mVp-p
60mVp-p
80mVp-p
80mVp-p
110mVp-p
110mVp-p 130mVp-p
130mVp-p 160mVp-p
160mVp-p 210mVp-p
210mVp-p 270mVp-p
270mVp-p 330mVp-p
330mVp-p 400mVp-p
400mVp-p 500mVp-p
500mVp-p 590mVp-p
Output Power Total Regulation Efficiency
28W
32W
36W
40W
40W
40W
40W
40W
40W
40W
40W
±5%
±5%
±5%
±5%
±5%
±5%
±3%
±3%
±3%
±3%
±3%
79%
80%
85%
87.5%
88%
88%
88%
88%
88%
88%
88%
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
INPUT SPECIFICATIONS
Operate Input Voltage Range
Safety Approval Input Voltage Range
Input Frequency
Input Current
Inrush Current
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation(4)
Total Regulation(5)
Output Power
Output Current
No Load Consumption
Ripple & Noise (20MHz bandwidth)(6)
Transient Response Time
Start-Up Time
Hold-Up Time(7)
Temperature Coefficient
4/28/2016
Sine Wave
Low Line
High Line
Low Line
High Line
Min
Typ
80
100
47
Full Load, Vin=100VAC
Full Load, Vin=240VAC
Full Load, 25ºC, Cool Start, Vin=100VAC
Full Load, 25ºC, Cool Start, Vin=240VAC
Max
Unit
275
240
63
1.0
0.6
30
60
VAC
VAC
Hz
A
A
See Table
Full Load, Vin=100~120VAC or 200~240VAC
1
See Table
See Table
See Table
0.1
See Table
Io=Full Load to Half Load, Vin=100VAC
Full Load, Vin=100~240VAC
Full Load, Vin=110VAC
All Conditions
W
4
2
12
-0.04
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
%
+0.04
ms
S
ms
%/ºC
Page 1 of 4
Rev A
PSHBU40 SERIES
40 Watts
AC DC Open Frame Medical Power Supplies
Single Output
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
PROTECTION
Short Circuit Protection
Over Load Protection
ENVIRONMENTAL SPECIFICATIONS
Operating Case Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Operating Altitude
Surge Voltage
Vibration
Electro Static Discharge
Cooling
MTBF
GENERAL SPECIFICATIONS
Efficiency(8)
Dielectric Withstanding Voltage (P-S)
PHYSICAL SPECIFICATIONS
Weight
Min
Recovers automatically after fault condition is removed
110
Derate linearly from 100% load at 40ºC to 50% load 70ºC
10~95% RH
Non-Condensing
-10
-40
0
0
All Conditions
Line-Neutral
Line-PE & Neutral-PE
10~500Hz, 10min./1cycle, 60min. each along X, Y, Z axes
Air Discharge, IEC61000-4-2
Contact Discharge, IEC61000-4-2
Operating Temperature at 25ºC, Calculated per MIL-HDBK-217F
Full Load, Vin=230VAC
Primary to Secondary, limit current <10mA
Typ
Max
Automatic Recovery
150
70
85
95
95
3000
1
2
5
15
8
Free Air Convection
100,000
Unit
%
ºC
ºC
%RH
%RH
m
kV
G
kV
Hours
See Table
4000
VAC
2.65oz (75g)
4in x 1in x1.28in
(4.00mm x 25.4mm x 32.6mm)
Mates with JST housing VHR-3N and JST
SVH series crimp terminal
Mates with JST housing VHR-6N and JST
SVH series crimp terminal
Dimensions (L x W x H)
Input Connector
Output Connector
SAFETY & EMC CHARACTERISTICS
Safety Approvals
EMC
Flammability Rating
Protection Classes
IEC60601-1 Edition 3.1
ES60601-1:2005(R2012)
CSAC22.2 No 60601-1:14
EN60601-1:2006/A1:2013
Compliance to EN55011 (CISPR11), EN60601-1-2
Class B
UL94V-1
Double Insulated, Class II
NOTES
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
4/28/2016
Output is at factory setting and cannot be adjusted.
Output can provide up to peak load when the power supply starts up. Staying in more than the rated load continuously is not allowed.
Each output is checked to be within voltage accuracy at factory in 60% rated load condition.
Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
Load regulation is defined by changing ±40% of measured output load from 60% rated load.
Ripple & Noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and
nominal line.
Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated
load and nominal line.
Efficiency is measured at rated load and nominal line.
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 4
Rev A
PSHBU40 SERIES
40 Watts
AC DC Open Frame Medical Power Supplies
Single Output
DERATING CURVES
Input Voltage Derating Curve
Temperature Derating Curve
MECHANICAL DRAWINGS
1
OUT
4/28/2016
PIN Chart PSHBU40-1XX
2
3
OUT
RTN
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
4
RTN
Page 3 of 4
Rev A
PSHBU40 SERIES
40 Watts
AC DC Open Frame Medical Power Supplies
Single Output
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely
with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide
ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your
future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
4/28/2016
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]