PSHBU40 SERIES 40 Watts AC DC Open Frame Medical Power Supplies Single Output Rev A FEATURES • Wide Operating Voltage, 80 to 275VAC, 47 to 63Hz • Single Outputs • Supports Risk Management Process • Input to Output: 2MOPP APPLICATIONS Size: 4in x 1in x 1.28in (101.6mm x 25.4mm x 32.6mm) • Breathing Therapy Device • Blood Pressure System • Portable Medical Device • ECG ‘ EEG • Medical Tablet • • • • High ESD Immunity Ultra Low Earth Leakage Current Short Circuit and Over Load Protection IEC60601-1 Edition 3.1, ES606011:2005(R2012), CSA C22.2 No. 60601-1:14, EN60601-1:2006/A1:2013 Safety Approvals DESCRIPTION The PSHBU40 series of AC DC open frame medical power supplies offers up to 40 watts of output power in a 4” x 1” x 1.28” unit. This series consists of single output models and a wide operating voltage of 80 to 275VAC. Each model in this series has an ultra low earth leakage current and is protected against short circuit and over load conditions. This series has IEC60601-1 edition 3.1, ES60601-1:2005(R2012), CSA C22.2 No. 60601-1:14, EN60601-1:2006/A1:203 safety approvals. MODEL SELECTION TABLE Input Voltage Range Model Number PSHBU40-102 PSHBU40-103 PSHBU40-104 PSHBU40-105 PSHBU40-106 PSHBU40-107 PSHBU40-108 PSHBU40-109 PSHBU40-110 PSHBU40-111 PSHBU40-112 80 to 275VAC Output Voltage Range(1) 5~6VDC 6~8VDC 8~11VDC 11~13VDC 13~16VDC 16~21VDC 21~27VDC 27~33VDC 33~40VDC 40~50VDC 50~59VDC Output Current Min Load Max Load 4.66A 5.60A 4.00A 5.33A 3.28A 4.50A 3.07A 3.63A 2.50A 3.07A 1.90A 2.50A 1.48A 1.90A 1.21A 1.48A 1.00A 1.21A 0.80A 1.00A 0.68A 0.80A Ripple & Noise Min. Max. 50mVp-p 60mVp-p 60mVp-p 80mVp-p 80mVp-p 110mVp-p 110mVp-p 130mVp-p 130mVp-p 160mVp-p 160mVp-p 210mVp-p 210mVp-p 270mVp-p 270mVp-p 330mVp-p 330mVp-p 400mVp-p 400mVp-p 500mVp-p 500mVp-p 590mVp-p Output Power Total Regulation Efficiency 28W 32W 36W 40W 40W 40W 40W 40W 40W 40W 40W ±5% ±5% ±5% ±5% ±5% ±5% ±3% ±3% ±3% ±3% ±3% 79% 80% 85% 87.5% 88% 88% 88% 88% 88% 88% 88% SPECIFICATIONS All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted. We reserve the right to change specifications based on technological advances. SPECIFICATION TEST CONDITIONS INPUT SPECIFICATIONS Operate Input Voltage Range Safety Approval Input Voltage Range Input Frequency Input Current Inrush Current OUTPUT SPECIFICATIONS Output Voltage Line Regulation(4) Total Regulation(5) Output Power Output Current No Load Consumption Ripple & Noise (20MHz bandwidth)(6) Transient Response Time Start-Up Time Hold-Up Time(7) Temperature Coefficient 4/28/2016 Sine Wave Low Line High Line Low Line High Line Min Typ 80 100 47 Full Load, Vin=100VAC Full Load, Vin=240VAC Full Load, 25ºC, Cool Start, Vin=100VAC Full Load, 25ºC, Cool Start, Vin=240VAC Max Unit 275 240 63 1.0 0.6 30 60 VAC VAC Hz A A See Table Full Load, Vin=100~120VAC or 200~240VAC 1 See Table See Table See Table 0.1 See Table Io=Full Load to Half Load, Vin=100VAC Full Load, Vin=100~240VAC Full Load, Vin=110VAC All Conditions W 4 2 12 -0.04 Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] % +0.04 ms S ms %/ºC Page 1 of 4 Rev A PSHBU40 SERIES 40 Watts AC DC Open Frame Medical Power Supplies Single Output SPECIFICATIONS All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted. We reserve the right to change specifications based on technological advances. SPECIFICATION TEST CONDITIONS PROTECTION Short Circuit Protection Over Load Protection ENVIRONMENTAL SPECIFICATIONS Operating Case Temperature Storage Temperature Operating Humidity Storage Humidity Operating Altitude Surge Voltage Vibration Electro Static Discharge Cooling MTBF GENERAL SPECIFICATIONS Efficiency(8) Dielectric Withstanding Voltage (P-S) PHYSICAL SPECIFICATIONS Weight Min Recovers automatically after fault condition is removed 110 Derate linearly from 100% load at 40ºC to 50% load 70ºC 10~95% RH Non-Condensing -10 -40 0 0 All Conditions Line-Neutral Line-PE & Neutral-PE 10~500Hz, 10min./1cycle, 60min. each along X, Y, Z axes Air Discharge, IEC61000-4-2 Contact Discharge, IEC61000-4-2 Operating Temperature at 25ºC, Calculated per MIL-HDBK-217F Full Load, Vin=230VAC Primary to Secondary, limit current <10mA Typ Max Automatic Recovery 150 70 85 95 95 3000 1 2 5 15 8 Free Air Convection 100,000 Unit % ºC ºC %RH %RH m kV G kV Hours See Table 4000 VAC 2.65oz (75g) 4in x 1in x1.28in (4.00mm x 25.4mm x 32.6mm) Mates with JST housing VHR-3N and JST SVH series crimp terminal Mates with JST housing VHR-6N and JST SVH series crimp terminal Dimensions (L x W x H) Input Connector Output Connector SAFETY & EMC CHARACTERISTICS Safety Approvals EMC Flammability Rating Protection Classes IEC60601-1 Edition 3.1 ES60601-1:2005(R2012) CSAC22.2 No 60601-1:14 EN60601-1:2006/A1:2013 Compliance to EN55011 (CISPR11), EN60601-1-2 Class B UL94V-1 Double Insulated, Class II NOTES (1) (2) (3) (4) (5) (6) (7) (8) 4/28/2016 Output is at factory setting and cannot be adjusted. Output can provide up to peak load when the power supply starts up. Staying in more than the rated load continuously is not allowed. Each output is checked to be within voltage accuracy at factory in 60% rated load condition. Line regulation is defined by changing ±10% of input voltage from nominal line at rated load. Load regulation is defined by changing ±40% of measured output load from 60% rated load. Ripple & Noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47uF capacitor at rated load and nominal line. Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated load and nominal line. Efficiency is measured at rated load and nominal line. Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] Page 2 of 4 Rev A PSHBU40 SERIES 40 Watts AC DC Open Frame Medical Power Supplies Single Output DERATING CURVES Input Voltage Derating Curve Temperature Derating Curve MECHANICAL DRAWINGS 1 OUT 4/28/2016 PIN Chart PSHBU40-1XX 2 3 OUT RTN Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] 4 RTN Page 3 of 4 Rev A PSHBU40 SERIES 40 Watts AC DC Open Frame Medical Power Supplies Single Output COMPANY INFORMATION Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008 certification is just one example of our commitment to producing a high quality, well-documented product for our customers. Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your future projects. Contact Wall Industries for further information: Phone: Toll Free: Fax: E-mail: Web: Address: 4/28/2016 (603)778-2300 (888)597-9255 (603)778-9797 [email protected] www.wallindustries.com 37 Industrial Drive Exeter, NH 03833 Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 • Fax 603-778-9797 website: www.wallindustries.com • e-mail: [email protected] Page 4 of 4