PSHBU58 SERIES
60 Watts
AC/DC Medical Open Frame Power Supplies
Single Output
Rev A
Size: 4in x 1.50in x 1.04in (101.6mm x 38.1mm x 26.5mm)
FEATURES
•
•
•
•
•
•
Wide Operating Input Voltage of 80~275VAC
Single Outputs
Over Voltage, Over Load, and Short Circuit Protection
Protection Class I
Cooling by Free Air Convection
High Efficiency up to 90%
APPLICATIONS
•
•
•
•
•
•
•
•
•
•
Output Power Up to 60 Watts
Support Risk Management Process
Input to Output: 2MOPP
High ESD Immunity
IEC60601-1 Edition 3.1, ES60601-1:2005(R2012), CSAC22.2
NO.60601-1:14, and EN60601-1:2006/A1:2013 Safety Approvals
DESCRIPTION
Breathing Therapy Devices
Blood Pressure Systems
Portable Medical Device
ECG ` EEG
Medical Tablets
The PSHBU58 series of AC/DC medical open frame power supplies provide up to 60 watts of continuous
output power. This series consists of single output models with a wide operating input voltage range of
80~275VAC and output voltages ranging from 5VDC to 55VDC. Each model is protected against short
circuit, over voltage and over load conditions and have IEC60601-1 Edition 3.1, ES60601-1:2005(R2012),
CSAC22.2 NO.6060-1-1:14, and EN60601-1:2006/A1:2013 safety approvals.
MODEL SELECTION TABLE
Model Number
PSHBU58-102
PSHBU58-103
PSHBU58-104
PSHBU58-105
PSHBU58-106
PSHBU58-107
PSHBU58-108
PSHBU58-109
PSHBU58-110
PSHBU58-111
PSHBU58-112
11/10/2015
Input Voltage
Range
Setting Voltage
Range(1)
80~275VAC
5~6VDC
6~8VDC
8~11VDC
11~13VDC
13~16VDC
16~21VDC
21~27VDC
27~33VDC
33~40VDC
40~50VDC
50~55VDC
Output Current
Min Load Max Load
5.50A
6.60A
5A
6.66A
4.54A
6.25A
4.61A
5.45A
3.75A
4.67A
2.85A
3.75A
2.22A
2.85A
1.81A
2.22A
1.50A
1.81A
1.20A
1.50A
1.09A
1.20A
Ripple &
Noise(6)
No Load
Consumption
Output
Power
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
100mVp-p
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
0.5W
33W
40W
50W
60W
60W
60W
60W
60W
60W
60W
60W
Total Regulation Efficiency(8)
±5%
±5%
±5%
±5%
±5%
±5%
±3%
±3%
±3%
±3%
±3%
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
80%
82%
86%
87%
87%
87%
88%
88%
89%
90%
90%
Page 1 of 4
PSHBU58 SERIES
60 Watts
AC/DC Medical Open Frame Power Supplies
Single Output
Rev A
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
INPUT SPECIFICATIONS
Safety Approval Input Voltage Range
Input Operate Voltage Range
Input Frequency
Input Current
Inrush Current
Safety Ground Leakage Current
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation(4)
Total Regulation
Output Power
Output Current
Ripple & Noise
Transient Response Time
Start-Up Time
Hold-Up Time(7)
Temperature Coefficient
PROTECTION
Short Circuit Protection
Over Load Protection
Over Voltage Protection
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Operating Humidity
Storage Humidity
Electro Static Discharge
Operating Altitude
Vibration
Cooling
Flammability Rating
Surge Voltage
MTBF
GENERAL SPECIFICATIONS
Efficiency
Insulation Resistance
Dielectric Withstanding Voltage
Min
Typ
Max
Unit
Safety Approval & Specification in Label
100
240
VAC
Derate linearly from 100% load at 90VAC to 80% load at 80VAC)
Sine Wave
Low Line
Full Load, Vin=100VAC
High Line
Full Load, Vin=240VAC
Low Line
Full Load, 25ºC, Cool Start, Vin=100VAC
High Line
Full Load, 25ºC, Cool Start, Vin=100VAC
Vin=240VAC, Fi=60Hz
80
47
275
63
1.2
0.6
30
60
0.25
VAC
Hz
A
A
mA
See Table
Full Load, Vin=100~120VAC or 200~240VAC
PSHBU58-102 – PSHBU58-107
PSHBU58-108 – PSHBU58-112
Full Load, Vin=110VAC
Full Load, Vin=100~240VAC
1
0.25
Full Load, Vin=100VAC
All Conditions
%
60
See Table
See Table
4
0.5
12
±0.04
Recovers automatically after fault condition is removed
Crowbar Mode
110
110
Derate linearly from 100% load at 40ºC to 50% load at 70ºC
10~95% RH
Non-Condensing
-10
-40
0
0
W
ms
S
ms
%/ºC
Automatic Recovery
150
275
70
85
95
95
Air Discharge, IEC61000-4-2
Contact Discharge, IEC61000-4-2
All Condition
10~500Hz, 10min./1cycle, 60min. each along X, Y, Z axes
Line-Neutral
Line-PE & Neutral PE
Operating Temperature at 25ºC, Calculated per MIL-HDBK-217F
%
±5
±3
%
%
ºC
ºC
% RH
% RH
3000
5
Free Air Convection
UL94V-1
1
2
M
G
kV
100k
Full Load, Vin=230VAC
h
See Table
50
MΩ
Primary to Secondary, limit current <10mA
Primary to PE, limit current <10mA
PHYSICAL SPECIFICATIONS
Weight
4000
1500
VAC
4.94oz (140g)
4.00in x 1.50in x 1.04in
(101.6mm x 38.1mm x 26.5mm)
Dimensions (L x W x H)
SAFETY
Safety Approvals
EMC Emission
Protection Classes
11/10/2015
IEC60601-1 Edition 3.1
ES60601-1: 2005 (R1012)
CSAC22.2 NO.60601-1:14
EN60601-1:2006/A1:2013
Compliance to EN55011 (CISPR11), EN60601-1-2
Class B
Class I
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 2 of 4
PSHBU58 SERIES
60 Watts
AC/DC Medical Open Frame Power Supplies
Single Output
Rev A
NOTES
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
Setting voltage range is a factory setting and cannot be adjusted.
Output can provide up to peak load when the power supply starts up. Continuous staying in more than rated load is not allowed.
In 60% rated load condition, each output is checked to be within voltage accuracy.
Line regulation is defined by changing ±10% of input voltage from nominal line at rated load.
Load regulation is defined by changing ±40% of measured output load from 60% rated load.
Ripple & Noise is measured by using 20MHz bandwidth limited oscilloscope and terminated each output with a 0.47µF capacitor at rated load and
nominal line.
Hold up time is measured from the end of the last charging pulse to the time which the main output drops down to low limit of main output at rated
load and nominal line.
Efficiency is measured at rated load and nominal line.
DERATING CURVES
Input Voltage Derating Curve
Temperature Derating Curve
MECHANICAL DRAWINGS
PIN
1
2
3
4
Notes:
1.
2.
3.
11/10/2015
PIN CHART
PSHBU58-1XX
RTN
RTN
OUT
OUT
Net Weight: 140g approx.
Input Connector mates with Molex housing
09-50-3031/35977-0300 and Moles
2478/35922 series crimp terminal.
Output connector mates with Molex housing
09-50-3061/35977-0600 and Molex
2478/35922 series crimp terminal.
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 4
Rev A
PSHBU58 SERIES
60 Watts
AC/DC Medical Open Frame Power Supplies
Single Output
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely
with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide
ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your
future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
11/10/2015
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 4