PMP150 Series MEDICAL 1 3 2 - 1 5 0 W AT T S - A C / D C S I N G L E O U T P U T FEATURES • • • • • • • • Low safety ground leakage current Class I and Class II models are certified to medical and ITE safety standards Wide input range Optional output connectors Over voltage protection Over current protection Compliant with CEC and Energy Star Efficiency level IV RoHS Compliant SPECIFICATIONS INPUT OPERATING MTBF 150,000 hours minimum at full load at 25°C ambient, calculated per MIL-HDBK217F Average active 87% min. (except 85% min.for PMP150-12 and PMP150-13) 30-110 KHz Voltage range 90-264 VAC Frequency Inrush current Earth leakage current 47-63 Hz 60 A @ 115 VAC or 120 A @ 230 VAC, at 25°C cold start 2.0 A (rms) for 115 VAC 1.0 A (rms) for 230 VAC 220 µA max. @ 264 VAC, 63 Hz Touch current 100 µA max. @ 264 VAC, 63 Hz Operating temperature: 0°C to +60°C Power factor 0.98 Typical at 115 VAC Storage temperature: -40°C to +85°C Input – Output: 4000VAC Input – Case:1500VAC Output – Case: 500VAC Relative humidity: Derating: 5% to 95% non-condensing Derate from 100% at +40°C linearly to 50% at +60°C Output voltage See table Safety Output current See table UL 60601-1, CSA C22.2 No. 601.1, TÜV EN 60601-1, UL 60950-1, CSA C22.2 No. 60950-1, TÜV EN 60950-1 Ripple and noise 1% peak to peak maximum at full load C-Tick AS/NZS CISPR11 Group1, Class A Line regulation ±0.5% maximum at full load EMC standards Load regulation See table Minimum load current 0A all models (IEC60601-1-2) EN55011 / EN55022 and FCC, VCCI: Class B conducted . class B radiated Transient response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change Immunity standards EN61000-3-2, EN61000-3-3, EN61000-42, EN61000-4-3, EN61000-4-4, EN610004-5, EN61000-4-6, EN61000-4-8, EN61000-4-11 Overvoltage protection Provided and set at 112-140% of its nominal output voltage Protected to short circuit conditions Input current Isolation OUTPUT Overcurrent protection Switching frequency ENVIRONMENT STANDARDS AND APPROVALS Temperature coefficient ±0.04% /°C maximum Hold-up time 15 ms minimum at 115 VAC Doc. No. PBC9 Rev.C 28-4-11 Efficiency MECHANICAL Dimensions 207.6x82x49.7mm Weight 960g Powering Progress – www.powerbox.com.au 387 MEDICAL PMP150 Series 1 3 2 - 1 5 0 W AT T S - A C / D C S I N G L E O U T P U T OUTPUT VOLTAGE/CURRENT RATING CHART MODEL(1) CLASS I PMP150-12 PMP150-13 PMP150-13-2 PMP150-14 PMP150-15 PMP150-18 OUTPUT CLASS II PMP150F-12 PMP150F-13 PMP150F-13-2 PMP150F-14 PMP150F-15 PMP150F-18 V1 12.0 V 15.0 V 19.0 V 24.0 V 27.0 V 48.0 V MIN. CURRENT MAX. CURRENT 0A 0A 0A 0A 0A 0A 11.00 A 9.00 A 7.90 A 6.25 A 5.56 A 3.13 A TOL. RIPPLE & NOISE (2) MAX POWER AVERAGE ACTIVE EFFICIENCY (TYPICAL) @ 115 / 230 VAC ±5% ±5% ±5% ±5% ±5% ±5% 120 mV 150 mV 190 mV 240 mV 270 mV 480 mV 132 W 135 W 150 W 150 W 150 W 150 W 87 / 86 % 87 / 86 % 88 / 88 % 88 / 88 % 88 / 88 % 88 / 88 % NOTES: 1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet. 2. Ripple and noise is maximum peak-to-peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 µF tantalum capacitor in parallel with a 0.1 µF ceramic capacitor across the output. OUTPUT POWER DERATING CURVE MECHANICAL SPECIFICATIONS NOTES: 1. 2. 3. 4. Dimensions shown in inches [mm] Tolerance 0.02 [0.5] maximum Weight: 681 grams (1.505 lbs.) approx. Refer to Section titled “OPTIONAL OUTPUT CONNECTORS”. Add the suffix assigned for a selected connector to a wanted model number, e.g. PMP135-14-B1, for ordering. 5. The length of output cable for PMP135-12, PMP135-12-1, PMP135-13, PMP135F-12, PMP135F-12-1, and PMP135F-13 is 37.4 (950) PIN CHART MODEL PMP150-12 PMP150-13 PMP150-13-2 PMP150-14 PMP150-15 PMP150-18 388 PIN 1 2 3 4 PMP150F-12 PMP150F-13 PMP150F-13-2 PMP150F-14 PMP150F-15 PMP150F-18 V1 Return +V1 V1 Return +V1 Doc. No. PBC9 Rev.C 28-4-11