Download Data Sheet

PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
Rev D
O Type: Open Frame
U Type: U-Chassis
C Type: Enclosed Case
DN Type: Din Rail
Size: 3in x 2in x 1.16in
Size: 3.6in x 2.44in x 1.54in
Size: 3.6in x 2.44in x 1.54in
Size: ~3.60in x 2.45in x 1.54in
OPTIONS
• Mechanical Type
• Output Voltage
• Protection Type
FEATURES
•
•
•
•
•
•
•
•
•
•
Protection Type Class I and Class II
Active Power Factor Correction
2 x 3 Inch Footprint
Low Leakage Current Under 75µA
High Efficiency up to 92%
Adjustable Output Voltage
Built-In EMI Filter
5000m Operating Altitude
100 Watts Maximum Output Power
4000VAC Input to Output MOPP Insulation
•
•
•
•
•
•
•
85~264VAC (120~370VDC) Input Voltage Range
-25ºC to 80ºC Operating Temperature Range
Over Voltage, Over Load, and Short Circuit Protection
Low Standby Power Consumption under 0.3W
Compliant to RoHS II and REACH, CE Marked
Designed to Meet Efficiency Level VI
ANSI/AAMI ES60601-1, EN60601-1, & IEC60601-1 3rd
Edition Safety Approvals
• Open Frame, U-Chassis, Enclosed Case, and Din Rail
Mechanical Options Available
DESCRIPTION
The PSMAD100 series of AC/DC medical power supplies provides 100 watts of output power in a compact 2 x 3 inch footprint. These supplies
feature a universal 85-264VAC (120~370 VDC) input, enabling them to be used anywhere in the world. The off load power draw is less than 0.3
watts, which complies with many energy-saving initiatives. 12V~48VDC single output voltages are available for this series, all of which have a
±10% adjustment range. These supplies also feature a low leakage current of less than 75μA at 264VAC and are designed to withstand 4000VAC,
input to output. The PSMAD100 series has an operating temperature range of -25°C to +80°C, active power factor correction, and a high efficiency
up to 92%. These supplies are also protected against short circuit, over voltage, and over current conditions. The PSMAD100 series has
ANSI/AAMI ES60601-1, EN60601-1, and IEC60601-1 3rd edition medical safety approvals, is CE marked, designed to meet Efficiency Level VI,
and meets the conducted and radiated EMI requirements of EN55011, EN55022 and FCC Part 18. Open frame, U-chassis, enclosed case, and
DIN rail mechanical options are available. Class I and Class II protection types are also available.
MODEL SELECTION TABLE
(1)
Model Number
PSMAD100-12S-O
PSMAD100-15S-O
PSMAD100-24S-O
PSMAD100-28S-O
PSMAD100-36S-O
PSMAD100-48S-O
PSMAD100-12S-U
PSMAD100-15S-U
PSMAD100-24S-U
PSMAD100-28S-U
PSMAD100-36S-U
PSMAD100-48S-U
PSMAD100-12S-C
PSMAD100-15S-C
PSMAD100-24S-C
PSMAD100-28S-C
PSMAD100-36S-C
PSMAD100-48S-C
PSMAD100-12S-DN
PSMAD100-15S-DN
PSMAD100-24S-DN
PSMAD100-28S-DN
PSMAD100-36S-DN
PSMAD100-48S-DN
2/23/2016
Input Voltage Range
85 - 264VAC
(120 – 370VDC)
85 – 264VAC
(120 – 370VDC)
85 – 264VAC
(120 – 370VDC)
85 – 264VAC
(120 – 370VDC)
Output Voltage
Output Current
12 VDC
15 VDC
24 VDC
28 VDC
36 VDC
48 VDC
12 VDC
15 VDC
24 VDC
28 VDC
36 VDC
48 VDC
12 VDC
15 VDC
24 VDC
28 VDC
36 VDC
48 VDC
12 VDC
15 VDC
24 VDC
28 VDC
36 VDC
48 VDC
8.34 A
6.67 A
4.17 A
3.58 A
2.78 A
2.09 A
8.34 A
6.67 A
4.17 A
3.58 A
2.78 A
2.09 A
8.34 A
6.67 A
4.17 A
3.58 A
2.78 A
2.09 A
8.34 A
6.67 A
4.17 A
3.58 A
2.78 A
2.09 A
Ripple & Noise Output Power Efficiency
120mVp-p
150mVp-p
160mVp-p
180mVp-p
190mVp-p
340mVp-p
120mVp-p
150mVp-p
160mVp-p
180mVp-p
190mVp-p
340mVp-p
120mVp-p
150mVp-p
160mVp-p
180mVp-p
190mVp-p
340mVp-p
120mVp-p
150mVp-p
160mVp-p
180mVp-p
190mVp-p
340mVp-p
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
100W
91%
92%
92%
92%
91%
91%
91%
92%
92%
92%
91%
91%
91%
92%
92%
92%
91%
91%
91%
92%
92%
92%
91%
91%
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Package Type
Open Frame
U-Chassis
Enclosed Case
Din Rail
Page 1 of 7
PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
Rev D
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
264
370
63
1.15
0.55
0.3
VAC
VDC
Hz
INPUT SPECIFICATIONS
Operating Input Voltage Range
Input Frequency
Input Current
No Load Input Power
Power Factor Correction
Input Inrush Current
Input Protection
OUTPUT SPECIFICATIONS
Output Voltage
Initial Set Voltage Accuracy
Line Regulation
Load Regulation
AC Input
DC Input
AC Input
115VAC and Full Load
230VAC and Full Load
230VAC
85
120
47
230VAC
Internal Fuse in line and Neutral
60
T3.15A/250VAC
230VAC and Full Load
Low Line to High Line at Full Load
No Load to Full Load
10% Load to 90% Load
-1.0
-0.2
-0.5
-0.4
-10
Ripple & Noise (20MHz bandwidth)
With 10µF/25V 1206 X7R MLCC capacitor
With 10µF/25V 1206 X7R MLCC capacitor
With 1µF/50V 1206 X7R MLCC capacitor
With 1µF/50V 1206 X7R MLCC capacitor
With 1µF/50V 1206 X7R MLCC capacitor
With 0.1µF/100V 1206 X7R MLCC capacitor
Transient Response
Load step from 50~75% change at 2.5A/µs
See Table
+1.0
+0.2
+0.5
+0.4
+10
See Table
See Table
0
120
150
160
180
190
340
12V output model
15V output model
24V output model
28V output model
36V output model
48V output model
Peak Deviation
Recovery Time
1000
20
115VAC and Full Load
16
-0.02
%
%
%
%
mVp-p
3
Rise Time
A
%
500
Start-Up Time
% Vout
µs
ms
ms
+0.02
ms
%/ºC
Continuous, Automatic Recovery
115
150
%
115
135
%
% if Iout rated; Hiccup Mode
% of Vout (nom); Latch Mode
Natural Convection with Derating
-25
-40
+85
+85
5000
Non-Condensing
5
MIL-HDBK-217F Ta=25ºC, Full Load
95
MIL-STD-810F
IEC68-2-27
IEC68-2-6
790,300
ºC
ºC
M
% RH
hours
See Table
60
Isolation Voltage
1 minute (2MOPP Insulation)
Isolation Resistance
Leakage Current
500VDC
264VAC
2/23/2016
W
0.95
Voltage Adjustability
Output Power
Output Current
Minimum Load
Hold Up Time
Temperature Coefficient
PROTECTION
Short Circuit Protection
Over Load Protection
Over Voltage Protection
ENVIRONMENTAL SPECIFICATIONS
Operating Ambient Temperature
Storage Temperature
Operating Altitude
Relative Humidity
Thermal Shock
Shock
Vibration
MTBF
GENERAL SPECIFICATIONS
Efficiency
Switching Frequency
A
Input to Output
Input to F.G.
Output to F.G.
kHz
4000
1500
1500
0.1
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
VAC
75
GΩ
µA
Page 2 of 7
PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
Rev D
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
PHYSICAL SPECIFICATIONS
O Type
U Type
C Type
DN Type
Weight
5.50oz (156g)
6.84oz (194g)
7.41oz (210g)
8.18oz (232g)
3in x 2in x 1.16in
(76.2mm x 50.8mm x 1.16mm)
3.6in x 2.44in x 1.54in
(91.4mm x 62.0mm x 1.54mm)
~3.60in x 2.45in x 1.54in
(~76.3mm x 62.23mm x 39.2mm)
O Type
Dimensions (L x W x H)
U & C Types
DN Type
SAFETY & EMC CHARACTERISTICS
ANSI/AAMI ES60601-1
EN60601-1
IEC60601-1
Safety Approvals
EMI
EN55011, EN55022 and FCC Part 18
Harmonic Currents
Voltage Flicker
ESD
Radiated Immunity
Fast Transient
Surge
Conducted Immunity
Power Frequency Magnetic Field
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
Full Load
EN60601-1-2
230VAC 50Hz
EN60601-1-2
100VAC 50Hz
Dip and Interruptions
Conducted
Radiated
30%
60%
>95%
>95%
30%
60%
>95%
>95%
500mS
100mS
10mS
5000mS
500mS
100mS
10mS
5000mS
Class B
Class A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria A
Perf. Criteria B
Perf. Criteria A
Perf. Criteria B
Perf. Criteria A
Perf. Criteria B
NOTES
(1)
(2)
Protection types Class I and Class II are available for this series. Class I comes standard and for Class II add the suffix “B” to the model number.
See page 7 for model number setup for model number setup.
Din Rail option is only available for enclosed case type models.
DERATING CURVES
Derating Curve vs. Ambient Temperature
C Type and DN Type
2/23/2016
Derating Curve vs. Ambient Temperature
O Type and U Type
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 3 of 7
Rev D
PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
Derating Curve vs. Input Voltage
EFFICIENCY GRAPHS
Efficiency vs. Output Load
PSMAD100-24S-x
2/23/2016
Efficiency vs. Input Voltage
PSMAD100-24S-x
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 4 of 7
Rev D
PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
MECHANICAL DRAWINGS
O Type: Open frame
2/23/2016
C Type: Enclosed Case
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 5 of 7
Rev D
DN Type: Din Rail
Notes:
1.
2.
3.
2/23/2016
PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
U Type: U-Chassis
All dimensions in inch (mm)
Tolerance: x.xx±0.02 (x.x±0.5) x.xxx±0.01 (x.xx±0.25)
M3x0.5 screw locked torque MAX 5Kgf.xm/0.49N.m
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 6 of 7
PSMAD100 SERIES
100 Watts
Medical AC/DC Power Supplies
Single Output
Rev D
CONNECTORS
CON1-Input Connector
Pin 1
Pin 3
CON2-Output Connector
Pin 1,2
Pin 3,4
Line
Neutral
-Vout
+Vout
Mates with
JST housing: VHR-4N
JST crimp terminals: SVH-21T-P1.1
Mates with
Mounting holes marked with
must be
JST housing: VHR-3N
connected
to
safety
earth
for
CLASS
I
application
JST crimp terminals: SVH-21T-P1.1
MODEL NUMBER SETUP
PSMAD
100
Series Name
Output Power
100: 100 Watts
-
12
S
Output Voltage
Output Quantity
12: 12VDC
S: Single
-
O
B
Package Type
Protection Type
O: Open Frame
15: 15VDC
U: U-Chassis
24: 24VDC
C: Enclosed Case
None: Class I
B: ClassII
DN: DIN Rail (1)
28: 28VDC
36: 36VDC
48: 48VDC
NOTES
1. DIN Rail Option is only available for enclosed case models.
COMPANY INFORMATION
Wall Industries, Inc. has created custom and modified units for over 50 years. Our in-house research and development
engineers will provide a solution that exceeds your performance requirements on-time and on budget. Our ISO9001-2008
certification is just one example of our commitment to producing a high quality, well-documented product for our customers.
Our past projects demonstrate our commitment to you, our customer. Wall Industries, Inc. has a reputation for working closely
with its customers to ensure each solution meets or exceeds form, fit and function requirements. We will continue to provide
ongoing support for your project above and beyond the design and production phases. Give us a call today to discuss your
future projects.
Contact Wall Industries for further information:
Phone:
Toll Free:
Fax:
E-mail:
Web:
Address:
2/23/2016
(603)778-2300
(888)597-9255
(603)778-9797
[email protected]
www.wallindustries.com
37 Industrial Drive
Exeter, NH 03833
Wall Industries, Inc. • 37 Industrial Drive, Exeter, NH 03833 • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
Fax 603-778-9797
website: www.wallindustries.com • e-mail: [email protected]
Page 7 of 7