FCX – A SERIES DIFFERENTIAL PRESSURE TRANSMITTER FHC, FKC...2 DATA SHEET The FCX –A differential pressure transmitter accurately measures differential pressure, liquid level or gauge pressure and transmits a proportional 4 to 20mA signal. The transmitter utilizes a unique micromachined capacitive silicon sensor with state-of-the-art microprocessor technology to provide exceptional performance and functionality. FEATURES 1. High accuracy 0.1% accuracy for all calibrated spans is a standard feature for all DP models covering 0.1kPa{1m bar} draft range to 3000kPa{30 bar} high differential. Fuji's microcapacitance silicon sensor assures this accuracy for all elevated or suppressed calibration ranges without additional adjustment. 2. Minimum environmental influence The "Advanced Floating Cell" design which protects the pressure sensor against changes in temperature, static pressure, and overpressure substantially reduces total measurement error in actual field applications. 3. Smart / Traditional convertible Fuji micro-electronics manufacturing technology offers free selection of Smart / Traditional transmitters. A small plug-in communication module upgrades your model FHC to smart type model FKC, which has full remote communication capabilities. A Hand Held Communicator (HHC), model FXW can remotely display or reconfigure all transmitter parameters at any point on the loop without affecting the transmitter signal. 4. Fuji/HART bilingual communication module The communication module is "bilingual" to speak both Fuji proprietary protocol and HART. Any HART compatible devices can communicate with FCX-A/C series transmitters. 5. Application flexibility Example options that render the FCX –A suitable for almost any process applications includes. – Analog indicator at either the electronics side or terminal side – Full range of hazardous area approvals – Built-in RFI filter and lightning arrestor – 4 12 digits LCD meter – Stainless steel electronics housing – Wide selection of materials SPECIFICATIONS Functional specifications Type: Model FHC: 4 to 20mA, Traditional type Model FKC: 4 to 20mA with digital signal, Smart type Service: Liquid, gas, or vapour Static pressure, span, and range limit: Span limit [kPa] {m bar} Type Static pressure [MPa] {bar} Min. FHC F C 11 –0.1 to + 3.2 {–1 to + 32} F C 22 –0.1 to + 10 {–1 to + 100 } F C 23 –0.1 to + 10 {–1 to + 100 } F C 24 –0.1 to + 10 {–1 to + 100 } F C 25 –0.1 to + 10 {–1 to + 100 } F C 26 –0.1 to + 10 {–1 to + 100 } F C 33 –0.1 to + 16 {–1 to + 160 } F C 34 –0.1 to + 16 {–1 to + 160 } F C 35 –0.1 to + 16 {–1 to + 160 } F C 36 –0.1 to + 16 {–1 to + 160 } F C 38 –0.1 to + 16 {–1 to + 160 } F C 43 –0.1 to + 42 {–1 to + 420 } F C 44 –0.1 to + 42 {–1 to + 420 } F C 45 –0.1 to + 42 {–1 to + 420 } F C 46 –0.1 to + 42 {–1 to + 420 } F C 47 –0.1 to + 42 {–1 to + 420 } { { { { { { { { { { { { { { { { 0.1 1} 0.6 6} 3.2 32 } 6.4 64 } 13 130 } 50 500} 3.2 32} 6.4 64 } 13 130 } 50 500 } 300 3000 } 3.2 32 } 6.4 64 } 13 130 } 50 500 } 200 2000 } FKC { { { { { { { { { { { { { { { { 0.1 1} 0.1 1} 0.32 3.2 } 0.64 6.4 } 1.3 13 } 5 50 } 0.32 3.2 } 0.64 6.4 } 1.3 13 } 5 50 } 30 300 } 0.32 3.2 } 0.64 6.4 } 1.3 13 } 5 50 } 20 200 } Range limit [kPa] {m bar} Max. FHC/FKC { { { { { { { { { { { { { { { { +/– 1 { +/– 10 } +/– 6 { +/– 60 } +/– 32 { +/– 320 } +/– 64 { +/– 640 } +/– 130 { +/– 1300 } +/– 500 { +/– 5000 } +/– 32 { +/– 320 } +/– 64 { +/– 640 } +/– 130 { +/– 1300 } +/– 500 { +/– 5000 } +/– 3000 { +/– 30000 } +/– 32 { +/– 320 } +/– 64 { +/– 640 } +/– 130 { +/– 1300 } +/– 500 { +/– 5000 } +/– 2000 { +/– 20000 } 1 10 } 6 60 } 32 320 } 64 640 } 130 1300 } 500 5000 } 32 320 } 64 640 } 130 1300 } 500 5000 } 3000 30000 } 32 320 } 64 640 } 130 1300 } 500 5000 } 2000 20000 } Remark : To minimize environmental influence, span should be greater than 1/40 of the max. span in most applications. EDS6-85e Date May. 31, 2000 FHC, FKC...2 – Lower limit of static pressure (vacuum limit) ; Silicone fill sensor: See Fig. 1 F C 38 and F C 47: –0.5kgf/cm2 Fluorinated fill sensor: 66kPa abs (500mmHg abs) at temperature below 80°C – The maximum span of each sensor can be converted to different units using below factors. 1MPa= 103KPa=10bar=10.19716kgf/cm2= 145.0377psi 1kpa=10mbar=101.9716mmH2O=4.01463inH2O Over range limit: To maximum static pressure limit Output signal: Model FHC: 4 to 20mA DC 2-wire, linear signal Model FKC: 4 to 20mA DC (linear or square root) with digital signal superimposed on the 4 to 20mA signal Power supply: Transmitter operates on 10.5V to 45V DC at transmitter terminals. 10.5V to 32V DC for the units with optional arrester. Load limitations: see figure below [Ω] Load resistance 2000 1533Ω 1500 1000 600Ω Operating area 500 24V 250 10 16.1 10.5V Power voltage 45V 50 [V] Note: For communication with FXW, min. of 250 Ω required. Hazardous locations: Authorities BASEEFA Factory Mutual CSA RIIS SAA Flameproof Ex ds IIC T5, T6 Class I II III Div. 1 Groups B thru. G Class I II III Div. 1 Groups C thru. G Ex ds IIB+H2 T4 Ex d II C T5, T6 IP66/67 Intrinsic safety EEx ia IIC T4, T5 Class I II III Div. 1 Groups A thru. G Class I II III Div. 1 Groups A thru. G — Ex ia IIC T5, T6 IP66/67 Type N Nonincendive Ex N II T5 Class I II III Div. 2 Groups A thru. G Class I II III Div. 2 Groups A thru. G — Ex n IIC T5, T6 IP66/67 Model FKC: Zero and span are adjustable from the HHC. Zero is also adjustable externally from the adjustment screw. Damping: Adjustable electrical damping. Model FHC: The time constant is adjustable to 0, 0.3, 1.2, 4.8, or 19.2 seconds. Model FKC: The time constant is adjustable between 0 to 38.4 seconds. (4 steps) Zero elevation/suppression: –100% to +100% of URL Normal/reverse action: Model FHC: Selectable by moving a jumper pin located on the electronics unit. Model FKC: Selectable from HHC Indication: Analog indicator or 4 12 digit LCD meter, as specified. Burnout direction:Output hold Output 21.6mA selectable Output 3.8mA Model FHC: Unless otherwise specified, the burnout is in hold position. Model FKC: Selectable from HHC Loop-check output: Model FHC: Transmitter can output constant signal of 4mA, 12mA, or 20mA if MODE SWITCH is set to the loop check mode. Model FKC: Transmitter can be configured to provide constant signal 3.8mA through 21.6mA by HHC. Temperature limit: Ambient: – 40 to +85°C (–20 to +80°C for LCD indicator) (– 40 to +60°C for arrester option) (–10 to +60°C for fluorinated oil filled transmitters) For explosionproof units (flameproof or intrinsic safety), ambient temperature must be within the limits specified in each standard. Process: – 40 to +120°C for silicone fill sensor –20 to +80°C for fluorinated oil fill sensor Storage: – 40 to +90°C Humidity limit: 0 to 100% RH Communication: (Model FKC only) With HHC (Model FXW, consult Data Sheet No. EDS8-47), following information can be remotely displayed or reconfigured. } Items Zero/span adjustment: Model FHC: Zero is adjustable from the external adjustment screw. The adjustment screw can also function to adjust span when MODE SWITCH (located on the electronics unit) is in the span mode. INHIBIT mode to disable the adjustment screw is also available. Tag No. Model No. Serial No. Engineering unit Range limit Measuring range Damping Output mode Burnout direction Adjustment Output adjust Data Self diagnoses Printer External switch lock Transmitter display(*) Display Set v v v v v v v v v v — v v — v v v v — v — v v v v v v — — — v v Note: (*) HHC's version must be more than 5.0 (or FXW 1– 2), to use this function. 2 Performance specifications Accuracy rating: (including linearity, hysteresis, and repeatability) For spans greater than 1/10 of URL: ± 0.1% of span For spans below 1/10 of URL (Model FKC only): 0.1 × URL ± 0.05+0.05 % of span Span Option 0.075% Linearity: 0.05% of calibrated span Stability: ± 0.1% of upper range limit (URL) for 24 months Temperature effect: Effects per 55°C change between the limits of – 40°C and +85°C ) ( Range code (6th digit in Code Zero shift symbols) “1”/1kPa {10mbar} max. span “2”/6kPa {60mbar} max. span Total effect ± 0.25+0.2 URL %/55°C Span ( ) “3”/32kPa {320mbar} max. span “4”/64kPa {640mbar} max. span “5”/130kPa {1300mbar} max. ± 0.2+0.05 URL %/55°C span Span “6”/500kPa {5000mbar} max. span “7”/2000kPa {20000mbar} max. span “8”/3000kPa {30000mbar} max. span ( ) ± 0.3+0.2 URL %/55°C Span ( ) ± 0.25+0.05 URL %/55°C Span ( ) “1” /1kPa {10m bar} sensor “2” /6kPa {60 m bar} sensor “2” “3” “4” Zero shift (% of URL) Range code Time constant “1” 0.8 s “2” 0.5 s “3” 0.3 s “4” through “8” 0.2 s Dead time approx. 0.3 s Mounting position effect: Zero shift, less than 0.12kPa {1.2m bar} for a 10° tilt in any plane. No effect on span. This error can be corrected by adjusting Zero. (Double the effect for fluorinated fill sensors) Dielectric strength: 500V AC, 50/60Hz 1 min., between circuit and earth. Insulation resistance: More than 100MΩ at 500V DC. Turn-on time: 4 sec. Internal resistance for external field indicator: 12Ω or less. Performance specifications for square root output: (Model FKC only) Accuracy rating: Span Output Double the effects for material code (7th digit in Code symbols) “H”, “M”, “T”, “B”, “L” and “U”. Static pressure effect: Static pressure code (5th digit in Code symbols) Supply voltage effect: Less than 0.05% of calibrated span per 10V RFI effect: Less than 0.2% of URL for the frequencies of 20 to 1000MHz and field strength 30 V/m when electronics covers on. (Classification: 2-abc: 0.2% span per SAMA PMC 33.1) Step response: (without electrical damping) Span shift (% of calibrated span) ±0.2% / 1MPa{10bar} –0.2% /3.2MPa{32bar} ±0.2% / 3.2MPa{32bar} –0.2% /3.2MPa{32bar} ±0.1%/10MPa{100bar} –0.2%/10MPa{100bar} Double the Zero shift for material code (7th digit in Code symbols) “H”, “M”, “T”, “B”, “L” and “U”. Overrange effect: over 0.1 × URL below 0.1 × URL 50 to 100% ± 0.1% ± (0.05+0.05 × 0.1 × URL/Span)% 20 to 50% ± 0.25% ± 2.5 × (0.05+0.05 × 0.1 × URL/Span)% 10 to 20% ± 0.5% ± 5 × (0.05+0.05 × 0.1 × URL/Span)% Temperature effect: effect per 55°C change between the limits of – 40°C and +85°C Range code “1” and “2” “3” through “8” Shift at 20% output point URL ± 0.6+0.5 Span %/55°C URL ± 0.5+0.125 Span %/55°C ( ( ) ) Low flow cut-off: Customer configurable for any point between 7 to 20% of output Physical specifications Static pressure code (5th digit in Code symbols) Zero shift (% of URL) “1” / 1kPa {10m bar} sensor “2” / 6kPa {60m bar} sensor “2” “3” “4” ± 0.3% / 1MPa ± 0.3% / 3.2MPa ± 0.3% /10MPa ± 0.3% /16MPa ± 0.5% /42MPa {10bar } {32bar } {100bar } {160bar } {420 bar} Double the effects for material code (7th digit in Code symbols)“H”, “M”, “T”, “B”, “L” and “U”. Electrical connections: G1/2, 1/2-14 NPT, Pg13.5, or M20 × 1.5 conduit, as specified. Process connections: 1/4-18 NPT or Rc1/4 on 54mm centers, as specified. Meets DIN 19213. 3 FHC, FKC...2 Process-wetted parts material: V H M T 316 stainless steel(*1) 316 stainless steel(*1) Diaphragm 316L stainless 316 stainless steel steel Hastelloy-C Hastelloy-C lining 316 stainless steel(*1) 316 stainless steel(*1) B L U Hastelloy-C lining Monel lining Tantalum lining Wetted sensor body Monel Monel lining Tantalum Tantalum lining Hastelloy-C Hastelloy-C lining Monel lining Tantalum lining Monel Tantalum Vent/drain 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel Hastelloy-C Monel Tantalum Notes: (*1) SCS14 per JIS G 5121 Remark: Chlorine service: The fill fluid is fluorinated oil. Degreasing: Process-wetted parts are cleaned, but the fill fluid is standard silicone oil. Not for use on oxygen or chlorine measurement. NACE specification: Metallic materials for all pressure boundary parts comply with NACE MR-01-75. ASTM B7M or L7M bolts and 2HM nuts (Class II) are available. Static pressure rating for code “3” (16 MPa) is degraded to 10MPa. Vacuum service: Special silicone oil and filling procedure are applied. See figure × 1. Sensor O-rings: Viton and teflon selectable. Availability of above material design depends on ranges and static pressure. Refer to “Code symbols”. Non-wetted parts material: Electronics housing: Low copper die-cast aluminum alloy (standard), finished with epoxy/polyurethane double coating, or 316 stainless steel, as specified. Bolts and nuts: Cr-Mo alloy (standard), 304 stainless steel (for static pressure code “1”, “2”, and “3” only), or 630 stainless steel (for static pressure code “3” and “4” only). Static pressure rating for code “3” with 304 stainless steel bolts is degraded to 10MPa. Fill fluid: Silicone oil (standard) or fluorinated oil (Daifloil) Mounting bracket: Carbon steel with epoxy coating or 304 stainless steel, as specified Environmental protection: IEC IP67 and NEMA 4X Mounting: On 60.5mm(JIS 50A) pipe using mounting bracket, direct wall mounting, or direct process mounting. Mass{weight}: Transmitter approximately 4.4kg without options. Add; 0.5kg for mounting bracket 0.8kg for indicator option 4.5kg for stainless steel housing option Silicone oil (Code:Y,G,N) Silicone oil (Code:R) [kPa abs] {mbar abs} 101 {1010} Operating pressure Material code (7th digit Process cover in Code symbols) Operating area 20 {200} 4 {40} 2.7 {27} None-operating area –40 A plug-in analog indicator (1.5% accuracy) can be housed in the electronics compartment or in the terminal box of the housing. 1 An optional 4 2 digits LCD meter is also available. Arrester: A built-in arrester protects the electronics from lightning surges. Lightning surge immunity: 4kV (1.2 × 50µs) Oxygen service: Special cleaning procedures are followed throughout the process to maintain all process wetted parts oil-free. The fill fluid is fluorinated oil. 4 60 85 120 Fig. 1 Relation between process temperature and operating pressure Customer tag: Coating of cell: A stainless steel tag with customer tag data is wired to the transmitter. Cell's surface is finished with epoxy/polyurethane double coating. Specify if environment is extremely corrosive. ACCESSORIES Oval flanges: Optional features Indicator: –15 Process temperature[°C ] (Model FFP, refer to Data Sheet No. EDS6-10) Converts process connection to 1/2-14 NPT or to Rc1/2; in carbon steel or in 316 stainless steel. Equalizing valves: (Model FFN, refer to Data Sheet No. EDS6-10) Available in CS or in 316 stainless steel and in pressure rating 16MPa or 42MPa. Hand-held communicator: (Model FXW, refer to Data Sheet No. EDS 8-47) Communication module: (standard for model FKC) When using this module for model FHC, remote setting function becomes available. Remark: When the communication module is connected, the operation mode of external zero/span adjustable screw is changed to zero adjustment. CODE SYMBOLS 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 Description 2 Type 4 to 20mA, Traditional type 4 to 20mA with digital signal, Smart type FHC FKC Connections S T V W X Process connection Oval flange screw Conduit connection Rc1/4 1/4-18NPT 1/4-18NPT 1/4-18NPT 1/4-18NPT 7/16-20UNF 7/16-20UNF M10 (or M12)(*1) M10 (or M12)(*1) 7/16-20UNF G 1/2 1/2-14NPT Pg 13.5 M20×1.5 Pg 13.5 Span and materials 11V 11H 22V Static pressure [MPa] {bar} –0.1 to +3.2 {–1 to +32} [kPa] (m bar) 0.1/0.1…1/1 {1/1…10/10} –0.1 to +10 {–1 to 100} 0.6/0.1…6/6 {6/1…60/60} Span limit (*2) FHC/FKC 22H 33V –0.1 to +16 3.2/0.32…32/32 {–1 to +160} {32/3.2…320/320} 33H 33M 33T 34V 6.4/0.64…64/64 {64/6.4…640/640} 34H 34M 34T 35V 13/1.3…130/130 {130/13…1300/1300} 35H 35M 35T 36V 50/5…500/500 {500/50…5000/5000} 36H 36M 36T 38V 43V 300/30…3000/3000 {3000/300…30000/30000} –0.1 to +42 3.2/0.32…32/32 {–1 to +420} {32/3.2…320/320} 43H 43M 44V 6.4/0.64…64/64 {64/6.4…640/640} 44H 44M 45V 13/1.3…130/130 {130/13…1300/1300} 45H 45M 46V 50/5…500/500 {500/50…5000/5000} 46H 46M 47V 23B 23L 23U 24B 24L 24U 25B 25L 25U 26B 26L 26U 200/20…2000/2000 {2000/200…20000/20000} –0.1 to +10 {–1 to+100} 3.2/0.32…32/32 {32/3.2…320/320} 6.4/0.64…64/64 {64/6.4…640/640} 13/1.3…130/130 {130/13…1300/1300} 50/5…500/500 {500/50…5000/5000} Process cover Diaphragm Wetted cell body 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316L stainless steel Hast. C 316 stainless steel Hast. C lining 316L stainless steel Hast. C 316 stainless steel Hast. C lining 316L stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining Tantalum Tantalum lining 316L stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining Tantalum Tantalum lining 316L stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining Tantalum Tantalum lining 316 stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining Tantalum Tantalum lining 316 stainless steel 316L stainless steel Hast. C 316 stainless steel 316 stainless steel Hast. C lining Monel Monel lining 316L stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining 316L stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining 316 stainless steel Hast. C 316 stainless steel Hast. C lining Monel Monel lining 316 stainless steel 316 stainless steel Hast. C lining Monel lining Tantalum lining Hast. C lining Monel lining Tantalum lining Hast. C lining Monel lining Tantalum lining Hast. C lining Monel lining Tantalum lining Hast. C Monel Tantalum Hast. C Monel Tantalum Hast. C Monel Tantalum Hast. C Monel Tantalum Hast. C lining Monel lining Tantalum lining Hast. C lining Monel lining Tantalum lining Hast. C lining Monel lining Tantalum lining Hast. C lining Monel lining Tantalum lining Notes: (*1) The thread is M12, if 42MPa {420bar} static pressure is specified. (*2) 100: 1 turn down is possible for model FKC, but should be used at the span greater than 1/40 of the maximum span for better performance. 5 FHC, FKC...2 1 2 3 4 5 6 7 8 FHC 2 F KC 2 9 10 11 12 13 14 15 Description Indicator and arrester Indicator None Analog, 0 to 100% linear scale Analog, 0 to 100% sq. root scale Analog, custom scale Analog, double scale None Analog, 0 to 100% linear scale Analog, 0 to 100% sq. root scale Analog, custom scale Analog, double scale Digital, 0 to 100% Digital, custom scale Digital 0 to 100% square root Digital, 0 to 100% Digital, custom scale Digital 0 to 100% square root A B C D J E F G H K L P M Q S N A B C D E M N H J K P R T Q A B C D E F Approvals for hazardous locations None (for ordinary locations) JIS, Flameproof (Conduit seal) JIS, Flameproof (Cable gland seal) FM, Flameproof (or explosionproof) CSA, Flameproof (or explosionproof) BASEEFA, Flameproof (Conduit seal) BASEEFA, Flameproof (Cable gland seal) (Conduit connection G 1/2 only) FM, Intrinsic safety and Nonincendive CSA, Intrinsic safety and Nonincendive CENELEC, Intrinsic safety CENELEC, Intrinsic safety and BASEEFA, Type N SAA, Flameproof (Conduit seal) (*5) SAA, Intrinsic safety (*5) SAA, Type N (Non-sparking) (*5) Side vent/ drain and mounting bracket Side vent/drain Mounting bracket None None Specify "A", "B", or "C" for the None Yes, carbon steel None Yes, stainless steel 7th digit code "B", "L", or "U" Yes None Yes Yes, carbon steel Yes Yes, stainless steel Stainless steel parts Stainless steel tag plate Stainless steel elec, housing Coating of cell None None None Yes None None None Yes None Yes Yes None None None Yes Yes None Yes None Yes Yes Yes Yes Yes Special applications and fill fluid Treatment Fill fluid None (standard) Silicone oil None (standard) Fluorinated oil Degreasing Silicone oil Oxygen service Fluorinated oil (7th digit code, "V" only) Chlorine service Fluorinated oil (7th digit code "H", "T", "B", "U") NACE specification Silicone oil (Not available for 7th digit code "T", "U" and 15th digit code "A", "B") Vacuum service Silicone oil for vacuum use Sensor O-ring Viton Teflon Bolt/nut Cr-Mo alloy hexagon socket head cap screw/carbon steel nut Cr-Mo alloy hexagon bolt/nut NACE bolt/nut (ASTM A193 B7M/A194 2HM) (*6) NACE bolt/nut (ASTM A320 L7M/A194 2HM) 304 stainless steel/304 stainless steel } (*7) (*9) 630 stainless steel/304 stainless steel } (*8) } Y B C E M N P Q Y W G A D N R A B A B C D E F } } Notes: 6 Arrester None None None None None (*3) Yes Yes Yes Yes Yes (*3) None None (Model FKC only) (*4) None Yes Yes (Model FKC only) (*4) Yes (*3) The scale is selectable "JIS and SI unit" or "Linear and sq. root" or "Linear and sq. root by 10". (*4) Incase of FKC, specified the output mode linear or sq. root. Unless specified, the output mode is linear. In case of 9th digit code "P", "S" with FKC, specified the output indication. Unless specified, the indication is output mode. (*5) Available for 4th digit code "S", "T", "W". (*6) Static pressure should be –0.1 to +10MPa{–1 to +100bar}. (*7) Available for 5th digit code "1", "2", "3". In case of stainless steel bolt with 5th digit code "3", static pressure should be –0.1 to +10MPa {–1 to + 100bar}. (*8) Available for 5th digit code "3", "4". (*9) In case of tropical use, select a stainless bolts and nuts. OUTLINE DIAGRAM (Unit:mm) 4th of Code symbols "S, V, W, X" type is not attached U-bolt M8 Zero/span adj. screw (10) "A" Press conn. 52.5 88 Field ind. (option) *Cable gland (option) (27) "B" Conduit conn. Vent/drain plug 154 With field ind. Details of "A" Details of "B" φ13 φ15 2.5 17 8 E F φ7 43 59 Earth terminal (both side) See table 1 Details of "C" "C" Conduit conn. Oval flange screw See table 1 D Side vent (option) See table 1 Conduit conn. D E F 41.3 G1/2 See table 1 Press. conn. G Oval flange screw Earth terminal H.P. L.P. S G1/2 17 8 Rc1/4 7/16-20UNF screw depth 13 M4 T 1/2-14NPT 16 5 1/4-18NPT 7/16-20UNF screw depth 13 No. 8-32UNC V Pg13.5 1/4-18NPT M10 or M12 screw depth 13 M4 54 W M20x1.5 1/4-18NPT M10 or M12 screw depth 13 M4 70 104 X Pg13.5 1/4-18NPT 7/16-20UNF screw depth 13 M4 8 4.5 16 5 8 4.5 Table 1 231 4 holes 9 dia. 4th of Code symbols Down 37 210 φ80 105 φ80 Up 70 38 58 G φ18.5 Fastener (flameproof model only) 130 120 43.5 Mtg. pipe JIS 50A (2B) CONNECTION DIAGRAM Note *: Cable gland is supplied in case of flameproof packing type. ø11 cable is suitable. S T H F K CV W X V H M 2 T + – CK+ CK– + – ORDERING INFORMATION When ordering this instrument, specify: 1. CODE SYMBOLS 2. Measuring range 3. Output orientation (burnout direction) when abnormality is occurred in the transmitter. (Unless otherwise specified, output hold function is supplied). 4. Output mode in case of FKC (linear or square root output). (Unless otherwise specified, output mode is linear). 5. Indication method (indicated value and unit) in case of the digital indicator/actual scale (codes P and S on 9th digit) when FKC is in the square root output mode. (Unless otherwise specified, the indication is square root 0 to 100%). 6. TAG No. (up to 26 alphanumerical characters), if required. 7 FHC, FKC...2 4th of Code symbols "S, V, W, X" type is not attached "A" or "B" Press. conn. U-bolt M8 Zero/span adj. screw Oval flange screw See table 1 58 Fastener (flameproof model only) (10) Mtg. pipe JIS 50A (2B) φ80 120 105 52.5 70 φ80 Up 88 Down 38 58 *Cable gland 40 4 holes 9 dia. (27) 210 Details of "C" Earth terminal (both side) See table 1 G φ4 234 F 17 2 1 0 –0.1 +0.2 43 91 φ13 φ15 φ16 186 Details of "D" 4th of code symbols "S" type φ12+0.1 "D" Conduit conn. "C" Conduit conn. With field ind. Vent/drain plug Details of "A" Field ind. (option) G1/2 E See table 1 Details of "B" 4th of Code symbols "T, V, W, X" type φ16 10 1/4–18NPT φ4 Conduit conn. 4th of Code symbols E F S G1/2 T 1/2-14NPT 16 V Pg13.5 W M20x1.5 X Pg13.5 17 G Oval flange screw 8 7/16-20UNF screw depth 13 Earth terminal L.P. H.P. M4 7/16-20UNF screw depth 13 8 4.5 M10 screw depth 13 M10 16 5 screw depth 13 7/16-20UNF 8 4.5 screw depth 13 4–M8x12 No. 8-32UNC 5 M4 54 M4 70 M4 104 Table 1 The applicable standards used to demonstrate compliance are :- Note *: Cable gland is supplied in case of flameproof packing type. ø11 cable is suitable. EMI (Emission) EN 50081-1 :-1992 CONNECTION DIAGRAM S H T F K CV W X + B L 2 U – Test item Applicable Electromagnetic Radiation Disturbance CK + CK – + – Frequency ranme 30-1000MHz Basic standard EN55022 Class B EMS (Immunity) EN 50082-1 :-1992 The product conforms to the requirements of the Electromagnetic compatibility Directive 89/336/EEC as detailed within the technical construction file number TN510412. No. Test item 1 Electrostatic discharge Radio-frequency electromagnetic field. Fast transients common mode 2 3 Test specification Basic standard Performance criteria 8kV (Air) IEC 801-2 : 1984 27-500MHz 3V/m IEC 801-3 : 1984 (Unmodulated) 0.5kV, 5/50(Tr/Th)ns IEC 801-4 : 1988 5kHz Rep. B A B "LVD - The transmitter is not covered by the requirements of the LVD standard." Head office 11-2 Osaki 1-chome, Shinagawa-ku, Tokyo, 141-0032 Japan Phone: 81-3-5435-7111 http://www.fujielectric.co.jp/eng/sg/KEISOKU/welcome.htm Sales Div. International Sales Dept. No.1, Fuji-machi, Hino-city, Tokyo, 191-8502 Japan Phone: 81-42-585-6201, 6202 Fax: 81-42-585-6187 Information in this catalog is subject to change without notice. Printed in Japan